1. Introduction
This device is an intellectual precise measurement apparatus. It is designed and manufactured with imported components and sensors, offering high sensitivity, accuracy, and stable performance across various temperatures. Its compact size allows for easy storage and portability. The meter is capable of measuring dissolved oxygen values and the solvent's temperature. The SMART SENSOR AR8210 Dissolved Oxygen Meter is widely applied in industrial, electrical, agriculture, medicine, food industries, and for assessing drinking water quality. Please read this instruction manual thoroughly to ensure correct usage and accurate readings.
2. Package Contents
The product package includes the following items:
- Main unit: 1 pc
- USB cable: 1 pc
- Probe: 1 pc
- Electrode filling bottle: 1 pc
- Spare parts for protection: 1 pc
- Instruction Manual: 1 pc
3. Specifications
| Parameter | Value |
|---|---|
| Model Number | AR8210 |
| Dissolved Oxygen Measuring Range | 0.00-20.00 mg/L |
| Basic Measurement Error | ±0.4 mg/L |
| Measurement Response Time | Less Than 30 Seconds |
| Solvent Temperature Compensation Range | 0°C-40°C |
| Temperature Measuring Range | 0°C-40°C |
| Power Supply | 3.7V Li-polymer Rechargeable Battery |
| Display | Large Segment Type LCD Display |
| Working Temperature Range | 0°C-40°C |
| Working Humidity Range | Less Than 85%RH (Non-Condensing) |
| Product Size | 141*44*43mm |
| Package Size | 241*190*50mm |
| Certification | CE |
4. Setup
4.1 Initial Probe Preparation
The probe must be prepared and calibrated before its first use, and again after prolonged periods of inactivity. Ensure the probe is dry before starting.
- Carefully remove the new probe reservoir.
- Fill the reservoir approximately 3/4 full with fresh electrolyte. Ensure there are no air bubbles in the electrolyte. If bubbles are present, gently tap the probe with your finger to dislodge them.
- Screw the probe reservoir back onto the probe handle until the membrane makes contact with the gold cathode. It is crucial to keep the reservoir as bubble-free as possible.
- Note: If electrolyte overflows, gently clean it with a soft paper.
- Connect the probe connector to the main unit.
- Turn on the device and allow it to warm up for at least 30 minutes before use.
- If, after immersion in hydrogen peroxide water or saturated dissolved oxygen water, the measured value changes very little, it indicates the need to replace the probe reservoir and add fresh electrolytes.
4.2 Charging the Device
When the LCD display indicates a low battery or is empty, the device needs to be charged.
- Open the battery door located at the bottom of the device.
- Plug the provided USB cable into the USB port on the device and connect the other end to a suitable USB power source (e.g., computer, wall adapter).
- A full charge typically takes about 4 hours.
5. Operation
5.1 Button Functions
- Power ON/OFF / Data Hold button (labeled 'HOLD'):
- Short press: Turns the unit ON. After the unit is on, short press to hold the measured value.
- Long press (>3 seconds): Turns the unit OFF.
- Calibrate / Increase button (labeled 'CAL/▲'):
- In normal measurement mode, press for more than 3 seconds to enter Calibration mode.
- In compensation mode, short press to increase the value. Long press for fast increase.
- Backlight / Reduce button (labeled 'SET/▼'):
- Short press: Toggles the backlight ON or OFF.
- In compensation mode, short press to reduce the value. Long press for fast reduction.
5.2 LCD Display Screen
The large segment type LCD display provides clear readings and indicators:
- Top Row: Timer icon, CAL icon, Data Hold icon (HOLD), Battery power indication.
- Middle Section: Dissolved Oxygen measured value area (e.g., 0.00 mg/L).
- Lower Section: Measured temperature value of solvent under test (e.g., 30.3 °C).
- Units: Displays units for Dissolved Oxygen (mg/L), Salinity (ppt), Air Pressure (mmHg), and Temperature (°C/°F).
5.3 Auto Power Off Setup
To configure the auto power off function:
- Before turning on the unit, simultaneously press the 'HOLD' and 'CAL/▲' keys. This will enter the auto power on/off setup mode.
- The LCD display will show either "APO ON" or "APO OFF".
- Use the 'HOLD' key to toggle between "APO ON" (auto power off enabled) and "APO OFF" (auto power off disabled).
- Long press the 'HOLD' key to save your selection and return to normal measurement mode.
- APO ON mode: The unit will automatically turn off after 5 minutes of no key input.
- APO OFF mode: The auto power off function is disabled. The user must manually turn off the unit by long pressing the 'HOLD' key for more than 3 seconds.
5.4 Temperature Unit Selection
To change the temperature unit (°C or °F):
- Before turning on the unit, simultaneously press the 'HOLD' and 'SET/▼' keys. This will enter the temperature selection mode.
- The LCD display will show either "°C" or "°F".
- Use the 'HOLD' key to cycle through the available units.
- Long press the 'HOLD' key to save your selected unit and return to normal measurement mode.
6. Calibration
The instrument uses an air calibration method. It is crucial to calibrate the instrument after connecting the probe and allowing it to warm up for 30 minutes.
6.1 Air Calibration Procedure
- In normal measurement mode, press and hold the 'CAL/▲' key for 3 seconds to enter calibration mode.
- Place the probe in the air and wait for 2 to 3 minutes until the reading value stabilizes on the LCD.
- Once the reading is stable, press the 'CAL/▲' key until the LCD screen displays "PAS". The instrument will store the setting value and return to normal measurement mode.
- Important: If the LCD screen shows "Err" during this step, it indicates that the calibration failed. This could be due to the value being outside the calibration range or a damaged probe.
6.2 Salinity and Altitude Compensation
After successful air calibration, if your sample liquid is saline and the test location is not at sea level, you need to perform compensation:
- In normal measurement mode, long press the 'SET/▼' key to enter compensation setup mode.
- The LCD screen will display "TDS" and the "Saltinity value" will be flashing.
- Use the 'CAL/▲' key to increase the salinity value and the 'SET/▼' key to reduce it (long press for faster adjustment).
- After setting the salinity, press the 'HOLD' key to save the setting and proceed to Altitude compensation mode (AP).
- The LCD screen will display "AP" and the "AP value" will be flashing.
- Use the 'CAL/▲' key to increase the altitude value and the 'SET/▼' key to reduce it (long press for faster adjustment).
- To switch between altitude units (mmHg and KPa), simultaneously press the 'CAL/▲' and 'SET/▼' keys.
- After setting the altitude, press the 'HOLD' key to save the setup value and return to normal measurement mode.
7. Maintenance
Regular maintenance of the probe and electrode is highly recommended every six months to ensure optimal performance and longevity. Always keep the probe clean and its head protected when not in use. Periodically clean the electrode and refill the electrolyte with fresh solution.
7.1 Probe and Electrode Cleaning Procedure
- Remove the probe from the instrument.
- Unscrew the old diaphragm set from the probe head.
- Clean the metal electrode and the old diaphragm thoroughly with distilled water.
- For Gold Cathode Electrode: If the gold cathode electrode appears dark, use polish paper with a drop of deionized water to polish the surface (5-10 turns). Afterward, clean the gold cathode with deionized water and dry it with clean paper. Ensure the surface of the gold cathode becomes bright again.
- For Silver Anode Electrode: The surface of the silver anode electrode should be silver-grey. If it appears dark grey, use polish paper to restore its brightness.
- Add fresh electrolyte to the new diaphragm, filling it approximately 3/4 of the way. Ensure there are no air bubbles in the diaphragm.
- Screw the new diaphragm onto the electrode until it is securely tightened.
- Remarks: If electrolyte overflows, clean it with soft paper.
8. Storage
8.1 Short-Term Storage (Overnight or Between Measurements)
For short periods, such as overnight or between measurements, keep the electrode connected to the instrument and place it in clean water.
8.2 Long-Term Storage
If the electrode will not be used for a long period, it should be stored in a dry area. First, clean the electrode with boiled distilled water and allow it to cool down. Ensure the unit is turned off during the cleaning process. The instrument should be stored in an environment with humidity less than 85% RH and a temperature not exceeding 40°C.
9. Troubleshooting
- "Err" displayed during calibration: This indicates that the calibration failed. Possible causes include the measured value being outside the calibration range or a damaged probe. Recheck probe preparation and try calibration again. If the issue persists, the probe may need replacement.
- Unstable or inaccurate readings:
- Ensure the probe is properly prepared with fresh electrolyte and no air bubbles.
- Check if the probe needs cleaning or if the gold cathode or silver anode are discolored. Refer to the maintenance section.
- Verify that the device has been properly calibrated, including any necessary salinity and altitude compensation.
- Allow sufficient warm-up time (30 minutes) after turning on the device.
- Battery empty / Device not turning on: The internal Li-polymer battery may be depleted. Connect the device to a USB power source using the provided cable and charge for approximately 4 hours.
- Backlight not working: Short press the 'SET/▼' key to toggle the backlight on or off.
10. User Tips
- Consistent Cleaning: Always rinse the probe with distilled water before and after each use to prevent contamination and ensure accurate readings.
- Regular Calibration: Calibrate the meter frequently, especially if you notice inconsistent readings or if the device has been stored for a long time. This ensures the highest accuracy.
- Bubble-Free Electrolyte: When refilling the probe reservoir, take extra care to eliminate all air bubbles, as they can significantly interfere with dissolved oxygen measurements.
- Proper Storage: Follow the short-term and long-term storage guidelines to protect the electrode and extend its lifespan.
- Warm-up Time: Always allow the device to warm up for at least 30 minutes after turning it on, particularly before taking critical measurements.
11. Warranty & Support
11.1 Warranty Period
The warranty period for the composite electrode is one year from the date of storage. During this period, if any malfunction of the instrument or electrode is found, the factory will repair or replace it with a functional unit, provided the issue is due to manufacturing defects.
11.2 Special Announcement
The manufacturer reserves the right to change the design and the user manual without prior notice to the end user.
Version: 6-82100-0116-03
Certification: CE
For a comprehensive guide and additional documentation, please refer to the official user manual PDF: User Manual (PDF)
12. Frequently Asked Questions (FAQ)
- Q: How often should I calibrate the meter?
- A: It is recommended to calibrate the meter before its first use, after prolonged periods of inactivity (weeks or months), and before making any precise measurements.
- Q: What does "Err" on the LCD during calibration mean?
- A: "Err" indicates that the calibration failed. This could be because the measured value is outside the acceptable calibration range or the probe itself is damaged. Ensure proper probe preparation and try calibrating again.
- Q: How do I know when to replace the electrolyte or probe reservoir?
- A: If, after immersing the probe in hydrogen peroxide water or saturated dissolved oxygen water, the measured value changes very little, it suggests that the electrolyte or probe reservoir needs to be replaced.
- Q: Can I use the meter immediately after turning it on?
- A: It is recommended to allow the device to warm up for at least 30 minutes after turning it on, especially before taking precise measurements, to ensure stable and accurate readings.