1. Overview
This product is a mini digital LCR tweezers bridge. It is powered by a rechargeable lithium polymer battery for extended use and standby time. It features a TFT color display screen with a clear and intuitive interface. It intelligently identifies resistors, capacitors, and inductors, accurately measuring their values with a maximum measurement frequency of 100kHz and supports sweep frequency measurement function. In addition to traditional LCR measurement, it also features diode test, continuity test, ±32VDC voltage measurement, and a signal generator output function.
The body is small, elegant, and portable. The elastic tweezer arms paired with removable gold-plated measuring tips offer flexible and reliable operation. It is suitable for various usage scenarios such as factory SMD work, repair, teaching, and R&D.

Figure 1: ZOYI Tweezer LCR Meter ZT-MD2 Features
2. Safety Precautions
To avoid possible electric shock, fire, and personal injury, please read the safety precautions before use.
- Do not measure voltages higher than the rated range to prevent damage to the meter and personal injury.
- Do not perform LCR measurements on live circuits. Ensure capacitors are fully discharged before measurement.
- Do not use or store this product for extended periods in high temperature, dusty environments, or under direct sunlight.
- Do not use this product in the presence of explosive gases, vapors, or in damp environments.
3. Setup
3.1. Package Contents
Verify that all items are present in the package. If any items are missing or damaged, contact your vendor.

Figure 2: ZT-MD2 Package Contents

Figure 3: ZT-MD1 Package Contents
3.2. Charging the Device
The ZOYI LCR Meter supports rechargeable power and can be used while charging. It can be charged with a standard phone charger via its USB-C port.
Video 1: ZT-MD2 Unboxing and Charging Demonstration
3.3. Interface Overview
Familiarize yourself with the device's display and controls.
| No. | Function | Description |
|---|---|---|
| 1 | Frequency Measurement | Press KEY4 (Frequency) to switch between different measurement frequencies. |
| 2 | Level Measurement | Press KEY3 (Level) to switch between different measurement levels. |
| 3 | Range/Function | Press KEY2 (Function) to switch between different measurement functions. |
| 4 | Battery Status | Yellow: low battery; Lightning symbol: charging; Green: fully charged. |
| 5 | Data Hold | Short press KEY1 (Power) to hold current measurement value and save to record file. |
| 6 | Unit | Displays the unit of the current measurement value. |
| 7 | Measurement Speed | Long press KEY3 to adjust measurement speed: S=Slow, M=Medium, F=Fast. |
| 8 | Measurement Mode | A: Auto mode, P: Parallel equivalent mode, S: Series equivalent mode. |
| 9 | Type Symbol | Displays the type symbol corresponding to the current measurement function. |
| 10 | Main Value | Displays the main measurement parameter value. |
| 11 | Secondary Value | Displays the secondary measurement parameter value. |
| 12 | Secondary Value | Displays the secondary measurement parameter value. |
4. Operating Instructions
4.1. Key Operation Reference
| Key Name | KEY4 (Frequency) | KEY3 (Level) | KEY2 (Function) | KEY1 (Power) |
|---|---|---|---|---|
| Measurement Interface | Short: Freq Set Long: Freq Set | Short: Speed Long: Level | Short: Tare Long: Function | Short: Hold Long: Power Off |
| Setting Interface | Short: Exit Long: Cancel | Short: Up Long: Up | Short: Down Long: Down | Short: Confirm Long: Power Off |
4.2. LCR Measurement
- Press KEY2 to select the desired LCR measurement mode.
- Press KEY3 to choose the measurement level (typically 0.6 V).
- Press KEY4 to select the measurement frequency (typically 1 kHz).
- Place the tweezer tips across the component under test. Once the reading stabilizes, read the value displayed on the screen.
Note:
- When measuring small resistances, ensure good surface contact. Oxidation on the surface may affect accuracy.
- Discharge capacitors before measurement to prevent residual charge from damaging the instrument.
- Use high frequency for small capacitors and low frequency for large capacitors; use high frequency for small inductors and low frequency for large inductors; for resistors, use 1 kHz.
- Since the tweezer uses a two-wire measurement method, contact resistance cannot be ignored. It is recommended to use the REL (tare) function.
- For impedances greater than 10 Ω or capacitances greater than 1 nF, calibration with REL is not necessary.

Figure 4: LCR Measurement Examples
4.3. Diode and Continuity Measurement
Switch manually to the diode mode, then place the tweezer tips across the diode terminals. The screen will display the diode's polarity and forward voltage.
Note:
- The diode symbol shown on the screen corresponds to the physical orientation of the diode under test, allowing polarity identification.
- The forward voltage reading of the diode is for reference only. The measurement range is Vf 0.1 V-0.9 V. If the reading exceeds this range, the screen will display "OL." For higher precision requirements, use a professional diode tester.
4.4. DC Voltage Measurement
- Press KEY2 to switch to the voltage measurement function. The screen will display a measurement range of -32 V to +32 V.
- Place the tweezer tips across the voltage to be measured.
- Read the voltage value shown on the screen. The device will automatically determine the polarity based on the sign.
Note:
- Do not exceed the rated input voltage range at the tweezer tips; otherwise, the device may be damaged.
- The device cannot measure AC voltage. Do not apply AC input.
- Do not apply voltage while in other measurement modes.

Figure 5: DC Voltage Measurement
4.5. Sweep Frequency Mode Measurement
In sweep frequency mode, the parameters of a component can be observed at different frequencies, making it convenient for analyzing the frequency characteristics of the component.
- Press KEY2 to switch to the sweep frequency function.
- Press KEY4 to select capacitance, resistance, or inductance.
- Place the tweezer tips across the component under test and read the parameter values displayed on the screen.
4.6. Signal Generator
The device is equipped with a signal generator output function. The available output signals include sine wave, square wave, triangle wave, staircase wave, half wave, full wave, and noise wave. The maximum frequency for sine wave is 100 kHz, while the maximum frequency for other signals is 10 kHz.
- Press KEY2 to switch to the signal function, then press KEY1 to confirm.
- Press KEY1 to select the waveform, then use KEY2 or KEY3 to choose the waveform type.
- Use the same method to adjust the signal frequency, amplitude, and duty cycle.
- Once settings are completed, the configured waveform will be output synchronously from both tweezer tips.

Figure 6: Seven Selectable Waveforms
4.7. Sorting Mode
When measuring resistance, capacitance, or inductance, press and hold KEY2 and KEY3 simultaneously to enter the sorting measurement mode.
In sorting mode, press and hold KEY4 to set the nominal value and tolerance.
Based on the preset nominal value and tolerance, the instrument will automatically determine PASS or FAIL during measurement and record the count.
5. Calibration
When calibration is required, users can perform calibration according to the following instructions. Calibration includes the following ranges: 0 Ω, 10 Ω, 100 Ω, 1 KΩ, 10 KΩ, 100 KΩ, 1 MΩ, 10 MΩ, OPEN, and Voltage. Each range can be calibrated individually, or all ranges can be calibrated sequentially.
- Enter the settings menu, select the Calibration Settings option, and use KEY2 or KEY3 to choose the calibration range.
- Place the tweezer tips firmly across the standard source, then press KEY1 to confirm and start calibration.
- The calibration process takes approximately 20 seconds. Upon completion, the buzzer will sound a single "beep" and the screen will display "Success." If the calibration value is incorrect, the buzzer will sound two short "beeps" and the screen will display "Failure."
Note:
- Use SMD carbon film resistors as calibration standards. Do not use wire-wound resistors.
- During short-circuit calibration, ensure the tweezer tips are reliably shorted and kept as far as possible from interference sources.
- For OPEN calibration, do not connect any resistor—keep the tweezer tips open.
- Do not connect the charging cable or Bluetooth adapter during calibration.
- To restore factory calibration data, simply perform a factory reset.
6. Maintenance
Do not attempt to open or repair this product, or modify the circuitry.
Clean the device with a damp cloth and mild detergent. Do not use corrosive agents or solvents.
Dust or moisture on the test ports may affect measurement accuracy.
If the tweezer tips become worn, oxidized, or damaged, they can be replaced as follows:
- Remove the 4 screws from the tweezer tips (screw type: 3.5PM2*3). Be sure to use matching tools.
- Install the new tweezer tips. Tighten the screws securely.
7. Specifications
7.1. Measurement Parameters
| Main Parameter | Secondary Parameter |
|---|---|
| L (Inductance) | D (Loss angle tangent) |
| C (Capacitance) | Q (Quality factor) |
| R (Resistance) | Z (Impedance) |
| V (Voltage) | X (Reactance) |
| - | P (Phase angle) |
7.2. Specifications Accuracy
| Function | Range | 100 Hz | 1 kHz | 10 kHz | 100 kHz | Auto Range |
|---|---|---|---|---|---|---|
| Resistance | 1MΩ-10MΩ | 5%±5 | 5%±5 | - | - | 0 Ω - 10 MΩ |
| 1KΩ-1MΩ | 1%±5 | 0.5%±5 | 1%±5 | 5%±5 | ||
| 1Ω-1KΩ | 1%±5 | 0.5%±5 | 1%±5 | 5%±5 | ||
| 0Ω-1Ω | 5%±5 | 5%±5 | - | - | ||
| Capacitance | 1mF-20mF | 5%±5 | 5%±5 | - | - | 50 pF - 5 mF |
| 1uF-1mF | 2%±5 | 2%±5 | 2%±5 | 5%±5 | ||
| 1nF-1uF | 1%±5 | 0.5%±5 | 2%±5 | 5%±5 | ||
| 1pF-1nF | 2%±5 | 2%±5 | 2%±5 | 5%±5 | ||
| Inductance | 1H-60H | 5%±5 | 5%±5 | - | - | 50 μH - 60 H |
| 1mH-1H | 5%±5 | 1%±5 | 2%±5 | 5%±5 | ||
| 10uH-1mH | 0.5%±5 | 0.7%±5 | 3%±5 | 5%±5 | ||
| 1uH-10uH | 3%±5 | 3%±5 | 5%±5 | 5%±5 |
Voltage: Measurement Range: ±0.1 V-±32 V. Accuracy: 0.5% ± 2
Parameters: L, C, R, D, Q, Rs, Z, X, P
Functions: Auto mode, Resistance mode, Capacitance mode, Inductance mode, Diode/Continuity mode, Voltage measurement, Sweep frequency measurement, Signal output
Frequencies: 100Hz, 120Hz, 1KHz, 10KHz, 100KHz
Signal Levels: 0.1V, 0.3V, 0.6V
Signal Types: Sine wave, Square wave, Triangle wave, Staircase wave, Half wave, Full wave, Noise wave
7.3. General Technical Specifications
| Parameter | Specification |
|---|---|
| Display | 1.47-inch TFT |
| Range | Auto |
| Sampling Rate | 1 time/sec |
| Languages | Chinese, English |
| Screen Orientation | Auto, Left, Right |
| Backlight Brightness | 10%~100% |
| Volume Settings | 0%~100% |
| Auto Power Off | Off, 5 min, 30 min, 60 min, 120 min |
| Calibration Settings | 0Ω-10MΩ OPEN |
| Low Battery Warning | Yes |
7.4. Mechanical Specifications
| Parameter | Specification |
|---|---|
| Dimensions | 145*29*14mm |
| Weight | 50g |
| Battery | Lithium polymer rechargeable 3.7 V, 300 mAh |
| Housing Material | CNC aluminum alloy |
| Warranty | 1 year |
7.5. Environmental Conditions
| Condition | Temperature | Humidity |
|---|---|---|
| Operating Conditions | 0~40℃ | <75% |
| Storage Conditions | -20~60℃ | <80% |
8. User Tips
- Always ensure components are fully discharged before testing, especially capacitors, to prevent damage to the meter.
- For precise measurements of small components, ensure clean contact points on both the component and the meter's tips.
- Utilize the auto-ranging feature for general measurements, but switch to specific frequencies and levels for more detailed analysis or when encountering unstable readings.
9. Warranty and Support
9.1. Limited Warranty and Liability
This product is covered by a one-year warranty starting from the date of purchase. However, the warranty does not cover damages caused by accidents, negligence, misuse, modifications, contamination, or abnormal operating environments.
This manual is subject to change without prior notice. The information contained herein is considered accurate; if the user discovers any errors or omissions, please contact the manufacturer.
The company shall not be held responsible for any accidents or hazards resulting from improper operation by the user. The functions described in this manual are not intended as justification for using the product for special purposes.
9.2. Firmware Upgrade
With the device powered off, press and hold the KEY4 (Frequency key), then simultaneously press and hold the KEY1 (Power key) until "Bootloader" appears on the screen.
- Connect the device to a computer using a Type-C data cable. A USB drive will appear on the computer.
- Copy the prepared upgrade file into the USB drive. The upgrade will start automatically.
- Once completed, the screen will return to the measurement interface, indicating a successful upgrade.