1. Safety Information
1.1 General Introduction
Thank you very much for using our Nerve Muscle Stimulator. For your safe use of this device, please read this user manual carefully before your first use. Please keep this user manual with your device for future reference.
This device is a portable electrotherapy device, featuring three therapeutic modes: Transcutaneous Electrical Nerve Stimulator (TENS), Electrical Muscle Stimulation (EMS), and Massage. These modes are used for pain relief and electrical muscle stimulation. The stimulator sends gentle electrical current to underlying nerves and muscle groups via electrodes applied to the skin. The parameters of the device are controlled by the buttons on the touch screen panel. The intensity level is adjustable according to the needs of the individual user.
1.2 Medical Background
EXPLANATION OF TENS
Transcutaneous Electrical Nerve Stimulation (TENS) is a non-invasive, drug-free method of controlling pain. TENS uses tiny electrical impulses sent through the skin to nerves to modify your pain perception. TENS does not work for everyone. However, in most patients, it is effective in reducing or eliminating pain, allowing for a return to normal activity.
HOW TENS WORKS
The TENS unit sends comfortable impulses through the skin that stimulate the nerve (or nerves) in the treatment area. In many cases, this stimulation will greatly reduce or eliminate the pain sensation the patient feels. Pain relief varies by individual patient, mode selected for therapy, and the type of pain. In many patients, the reduction or elimination of pain lasts longer than the actual period of stimulation (sometimes as much as three to four times longer). In others, pain is only modified while stimulation actually occurs.
EXPLANATION OF EMS
Electrical Muscle Stimulation (EMS) is an internationally accepted and proven way of treating muscular injuries. It works by sending electronic pulses to the muscle needing treatment; this causes the muscle to exercise passively. This device has low frequency and in conjunction with the square wave pattern allows direct work on muscle groupings. This is widely used in hospitals and sports clinics for the treatment of muscular injuries and for the re-education of paralyzed muscles, to prevent atrophy in affected muscles and improving muscle tone and blood circulation.
HOW EMS WORKS
The EMS units send comfortable impulses through the skin that stimulate the nerves in the treatment area. When the muscle receives this signal, it contracts as if the brain has sent the signal itself. As the signal strength increases, the muscle flexes as in physical exercise. When the pulse ceases, the muscle relaxes and then this cycle is repeated until therapy is completed.
The goal of electrical muscle stimulation is to achieve contractions or vibrations in the muscles. Normal muscular activity is controlled by the central and peripheral nervous systems, which transmit electrical signals to the muscles. EMS works similarly but uses an external source (the stimulator) with electrodes attached to the skin for transmitting electrical impulses into the body. The impulses stimulate the nerves to send signals to a specifically targeted muscle, which reacts by contracting, just as it does with normal muscular activity.
1.3 Indication for Use (IFU)
TENS:
The device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), lower extremities (leg), abdomen and bottom due to strain from exercise or normal household work activities.
EMS:
The device is designed to be used to stimulate healthy muscles in order to improve or facilitate muscle performance.
Important Safety Information
Read the instruction manual before operating. Be sure to comply with all "Contraindications", "Warnings", "Cautions", and "Adverse reactions" in the manual. Failure to follow instructions may cause harm to the user or device.
1.4 Contraindications
- This device should not be used for symptomatic local pain relief unless etiology is established or a pain syndrome has been diagnosed.
- This device should not be used when cancerous lesions are present in the treatment area.
- Stimulation should not be applied over swollen, infected, inflamed areas or skin eruptions (e.g., phlebitis, thrombophlebitis, varicose veins, etc.).
- Electrodes must not be applied to sites that might cause current/stimulation to flow through the carotid sinus region (anterior neck) or trans-cerebrally (through the head).
- Patients with an implanted electronic device, such as a cardiac pacemaker, implanted defibrillator, or any other metallic or electronic device should not use this device without first consulting a doctor.
- This device should not be used over poorly innervated areas.
- This device should not be used on patients with serious arterial circulatory problems in the lower limbs.
- This device should not be used on patients with abdominal or inguinal hernia.
- Patients with heart disease, epilepsy, cancer, or any other health condition should not use this device without first consulting a physician.
1.5 Warnings, Cautions, and Adverse Reactions
Warnings
- Apply stimulation only to normal, intact, clean, healthy skin.
- Do not apply stimulation when in the bath or shower.
- Do not apply stimulation while sleeping.
- Do not use this device during pregnancy unless directed by your physician.
- Stimulation should not be applied over the neck, mouth, or eyes. Severe spasm of the laryngeal and pharyngeal muscles may occur, and the contractions may be strong enough to close the airway or cause difficulty in breathing.
- Stimulation should not be applied across the chest in that the introduction of electrical current into the heart may cause cardiac arrhythmias.
- Stimulation should not be applied across the head.
- Stimulation should not be applied over the carotid sinus nerve, particularly in patients with a known sensitivity to the carotid sinus reflex.
- Replacement of a component could result in an unacceptable risk. Please contact us if you have any problem with the device.
- Keep this device out of the reach of infants, toddlers, and children.
- Do not recharge, use, or leave the battery and device in any high-temperature environment such as a location near fire or in direct sunlight which may cause the battery to overheat, ignite, or rupture.
- There is potential hazard from simultaneous connection of a patient to high-frequency surgical equipment and the stimulator that may result in burns and possible damage to the stimulator.
- Operation proximity (e.g., 1 m) to shortwave or microwave therapy equipment may produce instability in the stimulator output.
- When the device is transferred to another patient, replace used pads with new pads.
- Please avoid the use of the equipment as much as possible in the following occasions or conditions, as it may result in incorrect operation: talking on a mobile phone; near the signal emission base station for broadcasting, television, communication, radar, navigation, etc.; near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic resonance imaging.
- Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of the equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
Cautions
- This device is for single patient use only.
- Keep yourself informed of the contraindications.
- This device is not intended for use on an unattended patient who is noncompliant, emotionally disturbed, has dementia, or a low IQ.
- This device should not be used while driving, operating machinery, close to water, or during any activity in which involuntary muscle contractions may put the user at undue risk for injury.
- Never use the device in rooms where aerosols (sprays) are used or pure oxygen is being administered.
- Do not use this device at the same time as other equipment which sends electrical pulses to your body.
- Do not confuse the electrode cables and contacts with your headphones or other devices, and do not connect the electrodes to other devices.
- Do not use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the control panel.
- Check the wires and associated connectors before each use.
- Turn the device off before applying or removing the electrodes.
- Electrical stimulators should be used only with the leads and electrodes recommended for use by the manufacturer.
- Do not use it in the presence of explosive atmosphere or flammable mixture.
Adverse Reactions
- Potential allergic reactions due to the electrical stimulation or gel including skin irritation, redness, burning, or hypersensitivity. If skin irritation occurs, discontinue use and consult your physician.
- Discomfort occurs due to improper stimulation levels; reduce the stimulation intensity or turn the device off directly.
- Headache and other painful sensations during or following the application of electrical stimulation near your eyes, to your head, and face. You should stop using the device and consult with your physician if you experience any adverse reactions from the device.
Notes
- Please note that only original recognized accessories, detachable parts, and materials approved by standard should be used.
- Please note that the adapter you choose for battery charging must comply with ANSI/AAMI/ES60601-1 or IEC 60950 standards.
2. Device Overview
2.1 Illustration of the Device

- 1. A/B output connector: Connects electrode wires.
- 2. USB port: For charging the device.
- 3. LCD screen: Displays modes, intensity, and timer.
- 4. On/off switch (II/O button): Long press to turn on/off. Short press for Pause/Start.
- 5. Lock Control (Clock/Lock button): Long press to lock/unlock buttons. Short press for Timer adjustment.
- 6. Channel A intensity adjustment (A+/A- buttons): Adjusts intensity for Channel A.
- 7. Channel B intensity adjustment (B+/B- buttons): Adjusts intensity for Channel B.
- 8. Function selector (M/F button): Long press to switch between TENS and EMS functions. Short press to select different modes within TENS/EMS.
- 9. Electrode wires: Connect the device to the electrode pads.
2.2 Features of the Device
- LCD display
- A/B dual output
- Adjustable timer up to 90 minutes
- 28 modes (TENS & EMS)
- Rechargeable built-in lithium battery
- 40 level intensity settings
- Allows 2 or 4 pads to be applied to different targeted areas at the same time
- Free premium accessories: self-adhesive electrode pads
2.3 Package Contents

- TENS Device x 1pc
- Oval Electrode pads x 2pcs
- Palm Electrode pads x 6pcs
- Rectangular Electrode pads x 2pcs
- Electrode wires x 2pcs
- USB Cable x 1pc
- Manual x 1pc
- Pouch x 1pc
- Packaging Box x 1pc
3. Setup
Connecting Pads and Wires
- Connect the pads to the wire by snapping them on, then remove the protective film. Note: Adding a few drops of water on the pads will increase their longevity.
- Place pads where the massage sensations are desired. Plug into one of the ports on the device. For additional relief, use the second port and repeat steps 1 & 2 for two additional pads.
Important: Make sure both pads are on the skin and not overlapping. Ensure your skin is free from any dirt, oil, or lotion. When using two or four pads simultaneously, ensure all pads are placed on your body prior to energizing the unit, otherwise, you won't be able to increase intensity.

4. Operating Instructions
Basic Operation
- Start up: Turn the device on by short pressing the power button (II/O) in the middle of the panel unit.
- Select Channel A: Press the "A+" or "A-" buttons to choose Channel A. If "A" is flashing on the LCD screen, you can select the intensity and modes for Channel A.
- Select Mode: Short press the "M/F" button to select different modes in TENS & EMS functions. Long press "M/F" to switch between TENS & EMS functions. NOTE: Intensity level defaults to zero when starting the unit or when changing modes for your safety.
- Adjust Intensity: Press the "A+" or "A-" button to increase/decrease the strength for Channel A. Long press "A+" or "A-" button for 2 seconds to enter quick increase/decrease intensity, and release the button quickly when it comes to your desired intensity.
- Time Setting: Press the "clock" button (O/) at the bottom of the LCD panel to set the timer from 10-90 minutes for Channel A. The timer can be increased by 10-minute intervals, and the unit will automatically turn off after the set time has finished.
- Select Channel B: Press the "B+" or "B-" buttons to choose Channel B. If "B" is flashing on the LCD screen, you can then select the intensity and modes for Channel B.
- Select Mode (Channel B): Short press the "M/F" button to select different modes in TENS & EMS functions. Long press "M/F" to switch from TENS & EMS functions. NOTE: Intensity level defaults to zero when starting the unit or when changing modes for your safety.
- Adjust Intensity (Channel B): Press the "B+" or "B-" button to increase/decrease the strength for Channel B. Long press "B+" or "B-" button for 2 seconds to enter quick increase/decrease intensity, and release the button quickly when it comes to your desired intensity.
- Pause the device: Short press the ON/OFF switch button (II/O) during operation to pause. Repeat short press again to resume the device.
- Lock buttons: Long press the clock/lock button (O/) to lock the buttons (cannot set up). Long press again to unlock the buttons (can set up).
- Turn off the device: Long press the on/off switch (II/O) in the middle of the unit panel.
TENS/EMS Modes and Intended Use
| Mode Type | Mode # | Human Parts | Intended Use |
|---|---|---|---|
| TENS | 01 | neck | Used for symptomatic relief of chronic intractable pain and temporary relief of pain associated with sore and aching muscles due to strain from exercise or normal household and work activities. |
| 02 | neck | ||
| 03 | shoulder | ||
| 04 | back | ||
| 05 | waist | ||
| 06 | waist | ||
| 07 | abdomen | ||
| 08 | abdomen | ||
| 09 | Arm | ||
| 10 | hand | ||
| 11 | thigh | ||
| 12 | calf | ||
| 13 | knee | ||
| 14 | Joint | ||
| 15 | leg | ||
| 16 | leg | ||
| EMS | 17 | Up Arm | Used for relaxation of muscle spasm, increase of blood flow circulation, prevention or delay in muscle atrophy, muscle atrophy re-education, maintaining or increasing range of motion, and immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. |
| 18 | Up Arm | ||
| 19 | Down Arm | ||
| 20 | back | ||
| 21 | waist | ||
| 22 | abdomen | ||
| 23 | thigh | ||
| 24 | calf | ||
| 25 | leg | ||
| 26 | neck | ||
| 27 | back | ||
| 28 | waist | Used for relaxation of muscle spasm. |
5. Maintenance and Care
5.1 Tips for Skin Care
Follow these suggestions to avoid skin irritation, especially if you have sensitive skin:
- Wash the area of skin where you will be placing the electrodes with soap. Rinse thoroughly and dry the area completely before and after placing electrodes.
- Excess hair may be clipped with scissors; do not shave the stimulation area.
- Many skin problems arise from the "pulling stress" from adhesive patches that are excessively stretched across the skin during application. To prevent this, apply electrodes from the center outward; avoid stretching over the skin.
- To minimize "pulling stress", tape extra lengths of lead wires to the skin in a loop to prevent tugging on electrodes.
- When removing electrodes, always remove by pulling in the direction of hair growth.
- It may be helpful to rub skin lotion on the electrode placement area during treatment downtime when you are not wearing electrodes.
- Never apply electrodes over irritated or broken skin.
5.2 Cleaning the Device
- Clean the device after use with a soft, slightly moistened cloth.
- Do not use any chemical cleaners or abrasive agents for cleaning.
5.3 Electrodes
- Use this device only with the leads and electrodes provided by the manufacturer.
- Check your electrode pads before every use. Replace them if needed. Reusable electrode pads can cause slight skin irritation, lose adhesion properties, and deliver less stimulation if overused.
- The electrode pads are a kind of hydrogel and medical silicone.
5.4 Cleaning the Electrode's Cords
Clean the electrode cords by wiping them with a damp cloth. Coating them lightly with talcum powder will reduce tangles and prolong their life.
5.5 Maintenance
- Maintenance and all repairs should only be carried out by an authorized agency. The manufacturer will not be held responsible for the results of maintenance or repairs by unauthorized persons.
- The user must not attempt any repairs to the device or accessories. Please contact the seller if you have any problems.
- Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to warranty.
- Check the unit before each use for signs of wear and/or damage. Replace worn items as required.
5.6 Safekeeping
- Please do not put the main machine in a place exposed to sunlight, high temperature, humidity, lots of dust, or a place close to fire, or where it is easy to vibrate or shock.
- Please put where children cannot reach it.
5.7 Cleaning and Storage
Environmental conditions of transport and storage: Temperature: -20°C to +60°C, Humidity: Humidity does not exceed 85%.
Use a lightly moistened cloth to clean the device and wipe gently.
- Make sure the device is off when you clean it.
- Do not let water get into the device.
6. Troubleshooting
If your device does not seem to be operating correctly, refer to the chart below to determine what may be wrong. Should none of these measures correct the problem, the device should be serviced.
| Problem | Cause | Solution |
|---|---|---|
| One pad feels stronger than the other. | This is normal. Different areas of your body will react differently. | Nothing needs to be done. Make sure the pads are moist and are making good contact. |
| During massage, the skin feels a painful burning sensation or the stimulation becomes weakened. | Adhesive gel pads are not adhering firmly to the skin. The gel pads are too dry. Wires are damaged. | Apply a few drops of water to the adhesive surface of each pad and make sure the pads are pressed firmly to the skin during application. Use a new set of pads. |
| Unit is on, but no sensation is felt in the pads. | Both pads are not firmly on the skin. There is a loose connection. The intensity level needs to be increased. | Make sure both pads are firmly pressed to the skin (two pads minimum). One pad alone will not work. Make sure all connections are secure from the unit to the wires and the wires to the pads. |
| Adhesive gel pads do not stick to skin even after cleaning and moistening. | Adhesive material ran out. | Adhesive gel pads need to be replaced. |
| Unit does not turn on or the screen is dim. | Battery is low. | Recharge the battery. |
| Unit does not charge. | Wall adapter or USB cord. | Charge your unit using a USB cord only, using any USB power outlet. |
| Only feel stimulation one channel. | Your device features the A/B independent channel. | Try to set the mode and/or the intensity on each channel. Try to use a new set of wires. |
7. Specifications
General Specifications
- Material: ABS
- Shell Color: Black
- Screen Type: LCD Backlight Display
- Output Pulse Channel: 2 Channels
- Intensity Stage: 40 Levels
- Timer: 10-90 Minutes
- Power Supply: USB Charging
Product Properties
- Is Batteries Included: No (refers to external batteries, device has built-in rechargeable battery)
- High-concerned chemical: None
- Electronic: Yes
- Voltage: None (operates via USB charging)
- Application: BODY
- Brand Name: SANPKON
- Origin: Mainland China


8. Recommended Use Positions
The TENS & EMS Stimulator can be applied to various body parts for pain relief and muscle stimulation. Refer to the diagram below for recommended electrode placement.

Common application areas include:
- Neck
- Upper Arm
- Forearm
- Abdomen
- Back
- Waist
- Shoulder
- Thigh Leg
- Lower Leg
- Elbow
- Foot
- Knee
- Lumbar
- Cervical
9. User Tips
- Maximize Pad Longevity: To extend the life of your electrode pads, always clean the skin area thoroughly with soap and water before application. After use, gently remove the pads by lifting a corner and pulling in the direction of hair growth. Store them on their protective sleeve in a re-sealable bag in a cool, dry place. Adding a few drops of water to the adhesive side and letting it air dry can also help maintain stickiness.
- Proper Skin Preparation: Ensure your skin is clean, dry, and free from oils, lotions, or dirt before applying electrodes to prevent irritation and ensure optimal adhesion and conductivity.
- Intensity Adjustment: Start with a low intensity and gradually increase it to a comfortable level. If you experience any discomfort or tingling sensation, reduce the intensity or turn off the device.
10. Warranty and Support
For any issues, repairs, or warranty claims, please contact the seller directly. Unauthorized repairs will void the warranty.
Manufacturer Contact Information:
- Manufacturer Name: Dongguan Bohuikang Technology Co., Ltd
- Address: 5th Floor, No 5, second street, huaide shui xiang new strict, Humen town, Dongguan city, Guangdong, Mainland China
11. Medical Disclaimer
All information provided in this manual is intended for general knowledge only and is not a substitute for medical advice or treatment for specific medical conditions. The information contained herein is intended to provide broad consumer understanding and knowledge of the product offered.
This information should not be considered complete and should not be used in place of a visit, call, consultation, or advice from your health care provider.