1. General Information
The MYOSLIM EM-01 EMS Body Sculpt series utilizes advanced HI-EMT technology, employing high-energy focused electromagnetic waves (HI-EMT) designed to induce powerful muscle contractions, thereby promoting muscle growth and reshaping the internal structure of muscles. Additionally, it claims to trigger significant fat breakdown through intense muscle contractions, causing fatty acids to be released from triglycerides and metabolized by the body over several weeks.
Advanced HI-EMT Technology
The EMS Body Sculpt series employs advanced HI-EMT technology, utilizing high-energy focused electromagnetic waves to achieve intense muscle contractions, promoting muscle growth, and deep internal structural reshaping.
Touch-Controlled Display
The touch-controlled display significantly simplifies device operation, allowing clear visibility of information from various positions. The brightness of the display is adjustable to accommodate different office lighting conditions. User-friendly, the screen guides you throughout the entire treatment process.
Real-time Monitoring and Guidance
Throughout the treatment process, the display shows remaining treatment time, utilized channels and treatment methods, applied treatment types, attachments used, and other necessary information. This aids users in real-time monitoring of treatment progress and provides necessary guidance.
Upgradeability
As your therapeutic needs evolve, you may wish to upgrade your equipment later on. The EMS Body Sculpt series offers flexible upgrade options to meet your future treatment requirements.
1.1 Indications for Use
The EMS Body Sculpt is intended for:
- Improvement of abdominal tone, strengthening of the abdominal muscles, and development of a fitter abdomen.
- Strengthening, toning, and firming of buttocks, thighs, and calves.
- Improvement of muscle tone and firmness, for strengthening muscles in arms.
1.2 Operating Environment
The operating environment of the EMS Body Sculpt instrument series is crucial to ensuring the effectiveness and safety of the device.
- Clean and Hygienic Environment: The operating environment should be clean and hygienic to prevent any external contamination or bacterial infections. Regular cleaning and disinfection of the equipment and its accessories are essential.
- Appropriate Temperature and Humidity: Ensure that the operating environment's temperature and humidity are within the appropriate range specified by the device manufacturer. Extreme temperatures or excessive dryness/humidity may impact the device's performance.
- Adequate Lighting: Provide sufficient lighting to ensure that operators can clearly see and operate the equipment correctly. This is especially important for conducting precise beauty procedures.
- Good Ventilation: The area where the equipment is operated should have adequate ventilation to ensure air circulation and prevent the accumulation of any harmful gases.
- Safety Measures: The operating environment should have appropriate safety measures, including fire extinguishers and first aid equipment, to address emergencies. Operators should be familiar with the equipment's safety procedures and know how to handle emergency situations.
- Avoidance of Interference Sources: The operating environment for beauty equipment should be kept away from any electromagnetic fields or other devices that could potentially cause interference, ensuring the normal operation of the equipment.
- Operator Training: All operators using beauty equipment should undergo proper training to understand the correct usage and safety procedures of the equipment.
2. Safety Precautions and Warnings
Precautions Before Use
- Do not use when feeling full; wait at least one hour after meals.
- Before using the instrument, remove any metal jewelry you are wearing.
- Before starting the operation, touch the probe vertically to the skin, then turn on the switch of the main unit and adjust the intensity of the operation mode.
- Start at the lowest intensity level, and then increase the intensity sequentially according to the user's skin tolerance.
- It is forbidden to eat or drink during the operation, and try to eat after one hour after the operation.
Contraindications
Do not use the device if any of the following conditions apply:
- Wounds on the skin (or avoid using on wounded areas).
- Patients with heart disease (such as those with arrhythmias), especially those with pacemakers.
- Cancer patients.
- Patients with infectious diseases.
- Pregnant women, lactating women, or during menstrual period.
- Major surgery patients, diabetic patients.
- Having any metal in the body.
- People with cerebrovascular disease (stroke patients).
Possible Side Effects
These effects may include but are not limited to:
- Muscle soreness.
- Reduction in muscle fat.
- Temporary muscle spasms.
- Transient joint or muscle tendon pain.
- Localized redness or skin flushing.
3. Device Components and Accessories
3.1 Front and Rear Panel of the Main Unit

The front panel features a touch-controlled display screen for operation and an EMS handle connection. The main power switch is also located on the front.

The rear panel includes the handle connection port, the main power input socket, and the main power switch.
Note: The handle can only be used by the original machine and cannot be used interchangeably.
3.2 Handle

3.3 List of Accessories
| Serial | Name | Quantity |
|---|---|---|
| 1 | Host | 1 set |
| 2 | Handle | 1 |
| 3 | 7*100CM Belt | 1 |
| 4 | Power cable | 1 |
| 5 | Manual | 1 copy |
4. Device Installation
4.1 Handle Film Application
Apply the protective film to the working part of the handle as shown in the diagram on page 8 of the original manual.
4.2 Installing the Handle
Align the handle with the connection port on the main unit and fasten the clip. Lock the key ring to secure the handle in place. Refer to the diagram on page 8 of the original manual for visual guidance.
4.3 Connecting Power
Connect one end of the power cord to the machine and the other end to a power socket. Ensure the power cord is plugged directly into the main socket and avoid connecting multiple machines to a single socket. Refer to the diagram on page 8 of the original manual for visual guidance.
4.4 Switching On/Off
- Move the manual switch on the rear panel to the 'I' position to turn on.
- Press the button on the front panel to activate the device.
- To turn off, press and hold the power button on the front panel.
- Move the manual switch on the rear panel to the 'O' position to fully power off.
5. Basic Device Operation
5.1 Handle Exchange
When replacing the handle, you must turn off the power of the machine. The handle cannot be replaced when the machine is powered on.
5.2 Device Start-up/Shut-down
- Connect one end of the power cord to the machine and plug the other end into the socket. Plug the power cord directly into the main socket. Do not connect multiple machines to the socket.
- Move the manual switch on the rear panel to the "I" position.
- Press the button on the front panel.
- When the treatment is finished, press and hold the power button to turn off the machine.
5.3 Navigation Controls
The pre-treatment regimen can be modified via on-screen controls. The treatment interface displays the following items and functions:
- Return: Navigates back to the previous screen.
- Home: Returns to the main menu.
- Intensity: Adjusts the treatment intensity (0-100%).
- Frequency: Three different periods of frequency (F1: 15 Hz, F2: 25 Hz, F3: 5 Hz) with free adjustment.
- Time: Sets the treatment duration (e.g., 30:00 minutes).
- Start/Pause Work Button: Initiates or pauses the treatment.
- Stop Working Button: Terminates the treatment.
Note: Adjust the corresponding intensity/stage frequency/time according to different customer needs. Click the start/pause work button to enter the operation. If the customer requests to pause, they can click the start/pause work button to pause the work, and after the work is completed, they can directly click the return key to return.
5.4 Unit Settings
From the main screen, you can access various settings:
- Auto Mode: Allows selection of gender (Male/Female), age range (<30, 30-39, 40-49, 50-59, 60+), and treatment area (ABS, Shoulder, Arm, Glutes, Thighs, Hamstring). Intensity and time can then be set.
- Manual Mode: Allows direct adjustment of Intensity, F1, F2, F3 frequencies, and Time.
- Language Settings: Options include Chinese and English.
- Password Protected Settings: This section allows advanced adjustments for Intensity, F1, F2, F3 frequencies, Time, Current (mA), and Aging Time. Access requires a password.
6. Therapy Application
6.1 Preparing the Therapy
- Communicate with patients to understand their needs and comfort levels.
- Ensure the patient is in good health and has no contraindications.
- Advise the patient to stay hydrated before and after the session.
- Ensure the patient wears appropriate, comfortable clothing.
- Advise the patient to avoid eating large quantities before the session.
- Familiarize yourself with the treatment process and device operation.
- Advise the patient to avoid drinking alcohol before the session.
6.2 Placing the Handle
The handle should be placed directly on the target treatment area. Ensure proper contact with the skin. Refer to the diagrams below for correct placement on various body parts:

The image illustrates correct handle placement for:
- Belly: Centered on the abdominal area.
- Buttock: Centered on each buttock.
- Arm: Placed on the bicep or tricep area.
Additional areas for handle placement include Thigh, Calf, and 'Arse' (gluteal region), as indicated in the manual's diagrams. Please operate according to the controller model you purchased and follow the manufacturer's official documentation to ensure the best experience.
7. Maintenance and Repair
7.1 Maintenance
7.1.1 Daily Inspection
Managers must regularly inspect instruments and components. Check whether the power socket has a reliable connection to the ground. Please check whether the power cord of the instrument is damaged before use, and stop using the instrument if it is damaged.
7.1.2 Cleaning
Instrument cleaning is performed on a regular basis to ensure sanitation and proper functioning of the equipment. The device housing can be wiped with a damp cloth, soapy water, or a neutral detergent while the instrument is powered off. The logo of the machine is silk screen printed, and the rest of the parts are spray painted. Do not use alcohol-based detergents.
7.1.3 Replacing the Fuse
When replacing the fuse, to prevent electric shock, please turn off the power and unplug the power plug. The fuse is located below the handle junction on the back of the machine. An additional spare fuse is included.
7.2 Repair
To repair the machine, you must contact the company's business department or dealer, and do not repair it without authorization. Any unpacking without consent will not be guaranteed by our company.
8. Troubleshooting
8.1 Common Faults and Troubleshooting Methods
| Fault Phenomenon | Causes | Exclusions and Workarounds |
|---|---|---|
| Device not working | The power plug is not inserted into the socket Power switch not on Power fuse failure | Check the power Check the power switch Replace the fuse |
| Handle not working | Check the handle for loose connection | Reseat the handle |
9. Specifications
9.1 Technical Parameters
| Parameter | Value |
|---|---|
| Name | EMS Body Sculpt |
| Input power supply | AC110V-220V 50/60Hz |
| Output maximum power | 0-6500W |
| Frequency | 5-200HZ |
| Device size | 47*35*23(cm) |
| Device weight | 10.5(kg) |
| Packaging size | 47*33*42(cm) |
| Package weight | 13.7(kg) |
9.2 Electromagnetic Compatibility (EMC)
Radiated Electromagnetic Field Immunity Test
| Parameter | Details |
|---|---|
| Port | Enclosure |
| Basic Standard | IEC 61000-4-3:2020 |
| Test Level | 30V/m with 80% AM. 1kHz Modulation. |
| Standard require | A |
| Tester | Sam Liu |
| Temperature | 25°C |
| Humidity | 54% |
Test Equipment Used:
| Description | Manufacturer | Model | Identifier | Cal. Date | Cal. Due |
|---|---|---|---|---|---|
| Signal Generator | IFA | 2023B | N/A | 15/03/2022 | 14/03/2023 |
| Power Amplifier | AR | 150W1000 | N/A | 15/03/2022 | 14/03/2023 |
| Power Antenna | AR | 25S1G4A | N/A | 15/03/2022 | 14/03/2023 |
Block Diagram of Test Setup: The test setup involves a 7x4x3 Chamber with the Equipment Under Test (EUT) and support positioned. A Power Amplifier, Signal Generator, and PC Controller are connected to manage the forward power. An EUT Monitoring camera is used in the Contro Room. The diagram on page 16 of the original manual provides a visual representation of this setup.
10. Warranty Information
The main machine is guaranteed for one year from the date of purchase (with the receipt), except for fragile parts, non-warranty scope, or expired warranty period. Accessories have a warranty period of 3 months. Man-made damage is excluded from the warranty.
Maintenance and updates can be provided, with appropriate fees charged. If the instrument breaks and needs to be sent back for repair, the buyer shall bear the return shipping cost within six months from the date of purchase, and the seller shall bear the shipping-in cost. For periods longer than six months, the buyer bears all shipping costs.
Warranty Exclusions
Free warranty service is not provided for failures caused by personal reasons, including:
- Faults caused by unauthorized disassembly and modification of this product.
- Faults caused by accidental fall.
- Failure caused by lack of reasonable maintenance.
- Failure to operate according to the correct instructions in this instruction manual.
Maintenance Record Sheet
A maintenance record sheet is provided in the original manual (page 17) to document product name, purchase date, dealer information, customer details, and maintenance history including cause of issue and content of maintenance.
11. User Tips
- Consistency is Key: For best results in muscle toning and firmness, consistent use of the MYOSLIM EM-01 over several weeks is recommended.
- Easy to Use: The device is designed to be user-friendly, making it easy to integrate into your routine.
- Efficient Workouts: The HI-EMT technology provides an efficient workout, stimulating significant muscle contractions in a short period, comparable to weeks of conventional training.
12. Frequently Asked Questions (FAQ)
- Q: What are the warranty terms for the MYOSLIM EM-01?
- A: The main machine has a one-year warranty from the date of purchase, excluding fragile parts and man-made damage. Accessories are covered for three months. For detailed terms, please refer to the 'Warranty Information' section of this manual.