1. Introduction
1.1 Safety Brief
This manual contains important safety instructions that must be followed in the operation and storage environment of the SPPS/SPPS-S series. To ensure your personal safety, and ensures that this product works in the best environment, please read this manual carefully before using.
When you get a brand-new power supply, you need to do the necessary checks to make sure the instrument is working properly:
- To check whether there are damages caused during transportation.
- To check whether all the accessories are complete.
- To check whether the output voltage and output current are normal after turning on the device.
If finding out any problems, please contact the merchant immediately.
1.2 Safety Symbols
The safety symbols below will appear in this manual or on the DC power supply:
- Attention: Indicates important information or warnings.
- High Voltage: Indicates presence of dangerous voltage.
- Grounding: Indicates a grounding terminal.
1.3 Product Brief
SPPS/SPPS-S Series adjustable DC regulated power supply is specially designed for laboratories, schools and students. The output voltage and current can be between 0 and nominal value, continuously adjustable. The stability and ripple factor of the power supply are very good and have a perfect protection circuit. Can work at full load for a long time. This power supply can be used as both a regulated power supply and a regulated current supply.
2. Setup
2.1 Unboxing and Initial Inspection
Upon receiving your KUAIQU DC Power Supply, carefully unpack all components. Verify that all accessories are present and that the unit has no visible damage from transit. Refer to the 'Packing List' section to ensure all items are included.

Figure 2.1: KUAIQU DC Power Supply and included accessories (U-shaped output cable, power cable).
2.2 Power Connection
Ensure the power supply is switched off before connecting it to the mains. Locate the AC power input and power switch on the rear panel of the device.

Figure 2.2: Rear panel of the power supply with power input and switch.
Connect the provided power cord to the AC power input. Ensure the input voltage matches the requirements of the device (e.g., AC 220V or AC 110V). Incorrect AC voltage input will cause serious damage to the device.
2.3 Power On/Off
Press the power switch on the rear panel to turn on the device. The display screen will initialize and show the parameter settings from the last shutdown. Press the power switch again to turn off the power supply.
Video 2.1: Demonstration of powering on the device, adjusting settings, and using the USB charging function.
2.4 Connecting the Load
To connect a load to the power supply:
- Rotate the terminal knob by turning it counterclockwise.
- Insert the load terminal (e.g., banana plug or U-shaped cable).
- Turn the terminal knob clockwise to secure the connection.
Warning: Improper connection may result in damage to the power supply unit or the load connected to the power supply. When connecting a battery load, do not reverse the polarity as this may damage the power supply.
3. Operating Instructions
3.1 Panel Overview

Figure 3.1: Front and back functional components of the KUAIQU DC Power Supply.
3.2 Status Indicators
- Constant Voltage (CV) Output Indicator: The indicator is on when the power supply is working in constant voltage mode.
- Constant Current (CC) Indicator: This indicator is on when the power supply is working in constant current mode.
3.3 Output Parameter Setting
The device features a 4-digit color LCD display showing voltage, current, and power. Voltage is accurate to 0.01V and current to 0.001A.

Figure 3.2: Accurate 4-digit display for voltage, current, and power.
To set the output voltage and current:
- Press the V/A button to switch between voltage adjustment and current adjustment modes.
- Use the ADJUST rotary encoder knob to change the values.
- Normally, the voltage and current work in fine-tuning mode. To switch to coarse-tuning mode, press the ADJUST knob. Each press moves the cursor one place to the left, allowing for faster adjustments.

Figure 3.3: Precision Encoder Adjustment Knob for accurate control.
3.4 Save/Recall Settings (Memory Function)
The front panel allows saving and recalling up to 4 groups of internal memories (M1-M4).

Figure 3.4: Memory function for one-button recall of stored settings.
- Save Settings: Press the corresponding memory key (M1-M4) to set the desired voltage and current values. Press the same memory key again to automatically save these settings.
- Recall Settings: Press the corresponding memory key (M1-M4) to recall the parameter values saved in that memory slot. The corresponding key light will illuminate, indicating the active memory group.
Note: When recalling other settings from memory, the output will be automatically turned off for safety.
3.5 Output On/Off
Use the On/Off button to control the output. Press it to turn on the output (the key light will be on). Press it again to turn off the power output (the key light will be off).

Figure 3.5: Steps to use the ON/OFF button for operation.
3.6 Locking the Front Panel
Press the LOCK button. When the button light is on, all buttons and knob operations on the front panel (except On/Off) will be locked. This prevents accidental changes to the output. To release the lock, press the LOCK button again; the light will turn off, and panel operation will resume.
3.7 USB Charging Function
The power supply includes a 5V/2A USB-A output port for charging various electronic devices.

Figure 3.6: USB-A 5V/2A charging port for fast charging.
3.8 Constant Voltage/Constant Current Characteristics
The power supplies operate as constant voltage/constant current automatic conversion types. They automatically switch between constant voltage and constant current states based on load changes. The intersection between these modes is called the conversion point.
- In constant voltage mode, the output voltage remains constant, and the output current increases with the load.
- When the current reaches the set limit, the power supply switches to constant current mode. The output current remains stable, and the output voltage decreases proportionally as the load increases further.
The CV indicator light is on during constant voltage, and the CC indicator is on when in constant current mode.
4. Specifications
The following specifications are measured at an ambient temperature of 25±5°C, relative humidity: <80%RH, and preheated for 30 minutes. Actual parameters may vary slightly.

Figure 4.1: Product dimensions and weight.
| Feature | Value |
|---|---|
| Output Power | 50-150W (up to 150W) |
| Type | Power Supply |
| EU LVD Voltage | Low voltage: Below 50Vac or 75Vdc |
| Rated Capacity | 40-1000VA (up to 40VA) |
| Output Current | 0-10A |
| Output Frequency | 50Hz or 60Hz |
| Output Type | SINGLE |
| Certification | CE |
| Working Temperature | 0-40℃, Relative Humidity <80%RH |
| Storage Temperature | -10-70℃, Relative Humidity <70%RH |
| Display Accuracy | 1% ±5 digits |
| Display Resolution | Voltage: 0.01V, Current: 0.001A |
| Product Weight | 1.2Kg - 1.4Kg (depending on model) |
| Product Dimensions | Length: 252mm, Width: 70mm, Height: 150mm (approximate, varies by model) |
| Parameter | SPPS305D | SPPS3010D | SPPS605D | SPPS1203D | SPPS2001D | SPPS3001D |
|---|---|---|---|---|---|---|
| Rated Voltage | 0-30V | 0-30V | 0-60V | 0-120V | 0-200V | 0-300V |
| Rated Current | 0-5A | 0-10A | 0-5A | 0-3A | 0-1A | 0-1A |
| Power | 150W | 300W | 300W | 360W | 200W | 300W |
| USB Output | 5V 2A | |||||
| Load Regulation (Voltage) | ≤ 0.3% + 10 mV | |||||
| Load Regulation (Current) | ≤ 0.1% + 2.5 mA | |||||
| Line Regulation (Voltage) | ≤ 0.01% + 2.5 mV | |||||
| Line Regulation (Current) | ≤ 0.1% + 2.5 mA | |||||
| Setting Resolution (Voltage) | 10mV | 10mV | 10mV | <100V: 10mV, ≥100V: 0.1V | 100mV | 100mV |
| Setting Resolution (Current) | 1mA | |||||
| Readback Resolution (Voltage) | 10mV | 10mV | 10mV | <100V: 10mV, ≥100V: 0.1V | 100mV | 100mV |
| Readback Resolution (Current) | 1mA | |||||
| Setting Accuracy (Voltage) | ≤0.05%+10mV | ≤0.05%+10mV | ≤0.01%+15mV | ≤0.01%+20mV | ≤0.01%+30mV | ≤0.01%+30mV |
| Setting Accuracy (Current) | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA |
| Readback Accuracy (Voltage) | ≤0.05%+10mV | ≤0.05%+10mV | ≤0.01%+15mV | ≤0.01%+20mV | ≤0.01%+30mV | ≤0.01%+30mV |
| Readback Accuracy (Current) | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA | ≤0.01%+10mA |
| Ripple (Voltage) | ≤ 60 mVp-p, ≤ 10 mVrms | ≤ 100 mVp-p, ≤ 15 mVrms | ≤ 200 mVp-p, ≤ 20 mVrms | ≤ 300 mVp-p, ≤ 30 mVrms | ≤ 1000 mVp-p, ≤ 140 mVrms | ≤ 1200 mVp-p, ≤ 140 mVrms |
| Ripple (Current) | ≤ 10 mArms | ≤ 15 mArms | ≤ 20 mArms | ≤ 30 mArms | ≤ 60 mArms | ≤ 80 mArms |
| Rise Time (No Load) | ≤200ms | |||||
| Rise Time (Full Load) | ≤300ms | |||||
| Fall Time (No Load) | ≤15S | ≤20S | ≤30S | ≤30S | ≤5S | ≤5S |
| Fall Time (Full Load) | ≤300ms | |||||
| Dynamic Response | ≤200ms | |||||
| Efficiency | ≥85% | |||||
| Input Voltage/Current | AC 220V (AC110V optional) | |||||
| Size/Net Weight | Size: about 24*8.5*16cm, weight: 1.3KG | |||||
| Package Size/Weight | Single unit: 33.5*13*22cm, weight: 1.7KG (8Pcs/Carton) | |||||
5. Maintenance
5.1 Product Maintenance
- Disconnect the power when the product is not in use.
- Unplug the power supply before cleaning.
- Do not use hydrocarbons, chlorides or similar solvents, or use abrasive cleaners.
5.2 Fuse Replacement
If the fuse blows, the power supply will stop working. To find and correct the cause of the blown fuse, then replace it with a fuse of the same specification.
Warning: For effective safety protection, it is only necessary to replace the fuse of a specific specification. Before replacing the fuse, the power must be turned off and the power cord must be unplugged from the power outlet.

Figure 5.1: Fuse box location on the rear panel.
5.3 Cooling System
The device features a built-in intelligent cooling fan. When the working temperature of the DC power supply exceeds 122°F/50°C, the cooling fan is automatically activated to prevent overheating.

Figure 5.2: Intelligent cooling fan operation.
6. Troubleshooting
If you encounter issues with your KUAIQU DC Power Supply, please refer to the following common problems and solutions:
- No Power: Check if the power cord is securely connected and the power switch on the rear panel is in the 'ON' position. Verify the mains power outlet is functional. If still no power, check the fuse (refer to Section 5.2).
- No Output Voltage/Current: Ensure the 'On/Off' output button is pressed and its indicator light is on. Check load connections for proper polarity and secure contact.
- Incorrect Output Readings: Verify that the multimeter or measuring device used for verification is calibrated and functioning correctly. Ensure the power supply is preheated for at least 30 minutes for stable readings.
- Overload/Short Circuit Protection Activated: The device has multiple security protections including over voltage, over heat, overload, short circuit, and leakage protection. If protection activates, disconnect the load, identify and resolve the cause (e.g., short circuit in the load), then restart the power supply.

Figure 6.1: Multiple Security Protection features.
7. User Tips
- Utilize Memory Presets: For frequently used voltage and current settings, save them to the M1-M4 memory buttons for quick recall and improved workflow efficiency.
- Lock Function for Stability: When performing sensitive experiments or long-duration tests, use the 'LOCK' button to prevent accidental adjustments to voltage and current, ensuring stable operation.
- Convenient USB Charging: Don't forget the built-in 5V/2A USB port on the front panel. It's handy for charging mobile devices or powering small USB-powered tools directly from your workbench.
- Monitor Display: The 4-digit color display provides real-time, accurate readings of voltage, current, and power. Keep an eye on these values to ensure your circuit is operating within expected parameters.
8. Warranty and Support
8.1 Product Warranty
This product is offered free maintenance service within one year from the date of purchase. This warranty does not cover:
- Lack of this product warranty card.
- Failures caused by improper use, such as improper handling and improper repair, modification or adjustment of the device.
- Consumable materials are not covered by the warranty.
- Naturally irresistible disasters such as floods, fires, earthquakes, etc.
Maintenance costs are charged for repairs that exceed the warranty period, and the costs incurred for maintenance are the responsibility of the user.
8.2 Contact Support
If you have any questions or require technical assistance, please contact us via email. We aim to respond as soon as possible.
Email: yzmdquantuoguan@163.com
9. Packing List
The standard package includes the following items:
- 1x Power Supply Unit
- 1x Power Cord
- 1x Output Load Cord (U-shaped output cable)
- 1x User's Manual
- 1x Warranty Card