1. Important Safety Information
Please read this user manual carefully before operating the Diode Laser Hair Removal Machine. Failure to follow instructions may result in injury or damage to the device.
General Cautions
- Do not use the machine in environments with electromagnetic interference or vibration that could affect other equipment.
- Avoid using the machine in environments containing flammable anesthetics or oxidizing gases (e.g., N2O, O2).
- Ensure combustible materials like cotton or wool textiles are not present in oxygen-rich environments during operation.
- Always ensure cleaning and disinfection solutions have evaporated completely before operation.
- Dispose of abandoned or replaced spare parts according to local regulations or return to the manufacturer to prevent environmental pollution.
- The content of this manual is copyrighted. Unauthorized copying or storage is prohibited.
Laser Safety
The GL090 Diode Laser Hair Removal Machine emits high-intensity infrared laser beams. To protect eyes, wear suitable protection glasses (NOHD standard) for both operator and patient.
WARNING: Laser Radiation. Avoid direct eye or skin exposure.
- Do not point the laser directly at eyes or skin.
- Do not power down the machine directly during operation.
- Familiarize yourself with the machine's structure and this manual before operation.
- Only trained and certified personnel should operate this device.
- The cooling function must be turned off when the machine is in standby to prevent probe damage.
- This is a Class IV laser product. Do not look directly at the laser or reflected laser.
- People not wearing protective glasses are forbidden near the laser during operation.
- Do not attempt to repair anything not described in this manual. Maintenance should only be performed by manufacturer-authorized personnel.
- Ensure the treatment room has high-density shading material on all doors and windows. Display a warning sign outside the room.
- Keep the machine away from flammable substances, anesthetics, and oxygen. Ensure flammable liquids used for cleaning have evaporated before use.
- Do not direct the laser at areas not intended for treatment.
2. Product Overview
The MYOSLIM Diode Laser Hair Removal Machine is a professional device designed for permanent hair removal and skin rejuvenation. It utilizes advanced 3-wavelength technology (755nm, 808nm, 1064nm) combined with a freezing point cooling system for effective and comfortable treatments across various skin and hair types.
Figure 2.1: MYOSLIM Diode Laser Hair Removal Machine highlighting painless, HD touch screen, high power handle, and 3 wavelengths.
Key Features:
- Painless Permanent Hair Removal: Achieved through freezing point technology.
- HD Touch Screen: User-friendly interface for easy operation.
- High Power Handle: Efficient and effective treatment.
- Triple Wavelength Technology: 755nm, 808nm, and 1064nm for comprehensive treatment across all skin and hair types.
- Advanced Cooling System: Air cooling, water cooling, sapphire contact cooling, and semi-conductor cooling ensure continuous 24-hour operation.
- Durable: 40 million shots life for long-term use.
- Multi-language Support: 8 kinds of languages available.
Figure 2.2: Device highlighting 808nm diode laser, freezing point hair removal, skin rejuvenation, and 755nm, 808nm, 1064nm wavelengths.
3. Setup
Unpacking and Inspection
Upon receiving the machine, open the package and check all spare parts against the packing list. If any parts are missing or damaged, contact customer support immediately.
Installation Procedure
- Place the device on a stable, dedicated area.
- Connect the handpiece to the fiber optic connector. Ensure it is securely plugged in. (Note: The handpiece must remain connected during operation to prevent errors.)
- Verify that the remote interlock switch is inserted into the designated "Remote interlock" port.
- Connect the foot switch to the "Foot Switch" port.
- Plug the three-core power cord into a compatible AC220V power outlet.
- Locate the KCD (Key Control Device) on the back of the machine. Turn the key from "OFF" to "ON" in a clockwise direction.
- Ensure the emergency stop switch is in the released state (turn right to release if pressed).
- Allow a few minutes for the machine to complete its automatic detection procedure. The machine will enter standby mode, indicated by the standby screen. If the cooling system has issues, the machine will not enter standby mode and will remain on the initial splash screen.
Figure 3.1: Front view of the machine, showing the touch screen and handpiece connection.
Figure 3.2: Back view of the machine, showing power input and cooling vents.
4. Operating Instructions
System Control
The machine's control system includes a touch screen, key switch, button switch, emergency stop switch, and indicator lights.
- Indicator Lights: Green (POWER) indicates power on. Yellow (READY) indicates ready for use. Blue (WORK) indicates laser output. Red light on handpiece flashes during laser output.
- Emergency Stop Switch: Red button with yellow circle. Press to immediately cut power to the control system and laser. Turn clockwise to release.
- KCD (Key Control Device): Located at the back. Turn to "OFF" and remove the key when not in use to prevent unauthorized operation.
Parameter Setting (Standby Mode Only)
Parameters can only be adjusted when the machine is in standby mode.
- Energy Density: Click "Energy" on the touch screen. Use the blue triangle buttons to adjust the energy by 1J/cm² increments. Range: 1 - 160 J/cm².
- Cooling System: The system defaults to cooler off. Click "COOLING" to turn it on (white "Snowflake" button turns blue, indicating automatic refrigeration). Click again to turn off cooling.
- Pulse Width: Click "Pulse width". Use the blue triangle buttons to adjust by 1ms increments.
- Frequency: Click "Rep Rate". Use the blue triangle buttons to adjust by 1Hz increments. Range: 0.5 - 10Hz.
Ready State
- The "ST / RY" button switches between "standby" and "ready" states.
- When "ST" is orange and "RY" is blue, the machine is in standby.
- Click "ST / RY" to connect to laser power. "RY" turns orange and "ST" turns blue, indicating the machine is ready. The yellow indicator light turns on.
- In ready state, press the foot pedal or button switch to initiate laser output.
- To adjust parameters, switch back to standby mode by clicking "ST / RY".
Expert Reserves
The machine allows saving and recalling up to four sets of customized treatment parameters.
- Registering Data: From the standby screen, click "USER PROG". Select a MODE (1-4) to modify. Use the digit keypad to change parameters. Click "SAVE" to store changes, then "BACK" to return to standby.
- Applying Parameters: From the standby screen, click "USER PROG" and select the desired MODE. The red button indicates successful selection. Click "BACK" to apply the parameters.
Intelligent Counter for Pulse
The system tracks the cumulative light pulses. The "Count" tab shows the current session's pulses, and "Total Count" records the machine's total optical pulses. The current pulse count resets upon reboot but can be manually cleared by clicking "RESET" during operation.
Treatment Method
The following procedure is a general guideline. Actual parameters and operation may vary based on clinical usage.
- Preparation:
- For patients with darker skin, protect from sun exposure (clothing, SPF>15 sunscreen) for 4-6 weeks before treatment. Hydroquinone medicine may be used for hyperpigmentation tendencies.
- Ensure the treatment area is clean and dry. Shave hair cleanly and carefully before treatment.
- Anesthesia may be used for sensitive areas or patients with low pain tolerance, but is generally not necessary.
- Application:
- Both operator and patient must wear protective glasses.
- Apply a thin layer of cooling gel to the treatment area to reduce epidermal temperature, lubricate, and ease pain.
- Perform a test treatment first to determine optimal energy density. Start with conservative settings and gradually increase until the desired effect is achieved.
- Press the laser handpiece firmly against the skin to ensure good contact. This helps reduce blood flow and hemoglobin absorption of laser energy.
- For effective skin cooling, ensure a 0.25-0.5 second pause between skin contact and laser operation. Move the handpiece to the next spot immediately after treatment.
- Maintain appropriate overlapping (under 10%) to avoid skin damage.
- Clean the sapphire treatment head frequently during treatment to prevent hair or other debris from absorbing laser energy, which can cause epidermal damage and pain.
- Post-Treatment:
- Apply chlorotetracycline eye ointment or erythromycin ointment to the treated area after each session to prevent local infection.
- Use sunblock (SPF>15: PA+++) to avoid direct sun exposure and prevent pigmentation.
Video 4.1: Product demonstration showing device operation, laser energy absorption by melanin, and treatment application with cooling gel.
Figure 4.2: Diagram illustrating how laser energy targets the hair follicle, damaging stem cells and leading to permanent hair removal.
Figure 4.3: Real experience: Before and after images demonstrating the effectiveness of laser hair removal.
5. Maintenance
Regular maintenance is crucial for ensuring the normal operation and longevity of your laser hair removal machine.
Mainframe Maintenance
- Keep the machine in a clean, dry, dust-free environment with an ambient temperature between 4°C and 40°C.
- If storing for a long period, shield the equipment to protect it from dust, water vapor, and other pollutants that could damage the laser output or the device.
- Store in a dry, ventilated area, free from corrosive gases and direct sunlight.
- If the machine is stored or transported in temperatures below 0°C for an extended time, the cooling water must be drained from the mainframe. This procedure should only be performed by trained professionals.
Laser Handpiece Maintenance
- Avoid dropping, hitting, or clashing the handpiece. Do not attempt to open it without authorization, as it contains precision optical components.
- Regularly check and clean the optical crystal face before use. Cleanliness significantly impacts output quality.
- Prevent damage to the handle's transmission system by avoiding extreme bending during use.
- Clean the laser head after each treatment using a soft, cleanser tissue, especially the sapphire surface, ensuring no spots or remnants remain.
Cooling System Maintenance
- Regularly check the cooling fan. A malfunctioning fan can lead to heat accumulation and damage the machine.
- Inspect the water storage tank, cooling capacity, and water quality regularly. Maintenance personnel should add or recharge cooling water as needed.
- The cooling water has specific requirements from the manufacturer. Contact them for refills.
Power Fuse Tube Replacement
- Turn off all power switches and unplug the power cord.
- Use a small flat screwdriver to open the fuse light jacket by rotating counter-clockwise.
- Remove the old fuse tube.
- Replace only with a manufacturer-authorized standard type fuse tube.
- Insert the new fuse tube into the jacket and screw it back in.
- Reconnect the power cord, turn on the power switch and key lock, then check if the machine runs smoothly.
6. Troubleshooting and Malfunction Alarms
The hair removal apparatus features multi-function protection and a malfunction alarm system. Alarms are triggered by issues such as safety interlock switches, flow switches, water temperature, fiber interlock switch, and current over-current. Protection instructions are displayed on the touch screen, with indicator lights turning red upon failure.
Alarm Resolution
- Safety Interlock Switch Alarm: Check if the shell is open, the safety interlock switch is loose, or the remote interlock is open.
- Flow Switches Alarm: Check for low water level in the tank or leaking water pipes/radiators.
- Water Temperature Alarm: Reduce or stop laser shooting for a while, then restart.
- Fiber Interlock Switch Alarm: Verify that the fibers are correctly connected to the optical output port.
- Current Over-current Alarm: Stop current operation, turn down the machine, restart, and re-operate.
If any alarm persists after checking and confirming the issue, contact after-sales maintenance staff. If the same alarm initiates more than three times, professional service is required.
Forced Shutdown
In emergency situations where quick cessation of laser output is necessary, press the emergency switch on the main screen. Note that forced shutdowns may result in the loss of current treatment parameters.
7. Specifications
Technical Parameters
| Parameter | Value |
|---|---|
| Laser Type | Semiconductor Laser |
| Laser Wavelength | 755nm, 808nm, 1064nm |
| Treatment Spot Size | 12*18mm |
| Power Density | 1 ~ 140 J/cm² (continuous adjustable) |
| Pulse Frequency | 1 ~ 10Hz |
| Cooling Method | Sealed inner-circled water cooling system, treatment head temperature -4 ~ 4°C |
| Power Supply | 110V±22V / 220V/110V, 50Hz/60Hz |
| Display Screen | 4.3-inch LCD display |
| Net Weight | 21kg |
| Dimensions (L x W x H) | 57 x 43 x 56 cm |
Working Conditions
- Ambient Temperature: 5°C ~ 40°C
- Relative Humidity: ≤80%
- Atmospheric Pressure Range: 860hPa ~ 1060hPa
Shipping and Reserve Conditions
- Temperature Range: -20°C ~ 60°C
- Relative Density Range: ≤100%
- Atmospheric Pressure Range: 500hPa ~ 1060hPa
8. Clinical Applications and Contraindications
Applications
The Diode Laser Hair Removal Machine is primarily used for hair removal.
Clinical Features
The machine emits 808nm laser, which is absorbed by melanin in the hair follicle without damaging surrounding epidermis. Light energy converts to heat, destroying the follicle for permanent hair removal. It offers stable performance, long life, and excellent clinical results.
- Safety: Stable performance, real-time controlled by intellectual CPU.
- Speed: Fast coverage with a large 10x18mm spot size, high efficiency.
- Efficacy: 808nm wavelength laser (0.75-1.50µm) is optimally absorbed by melanin, penetrating deep into the dermis to target hair at various depths.
- Pain-Free: Sapphire cooling tip cools the epidermis to 0-4°C, ensuring comfort.
- Convenience: User-friendly touch display interface.
- Permanence: Effective for all skin colors.
Contraindications
- History of keloid scarring or photosensitivity.
- Inflammatory skin conditions (e.g., eczema, active Herpes Simplex) at the treatment site.
- Recent depilation treatments (waxing, plucking, sugaring, threading) within the previous 6 weeks.
- Allergy to hydroquinone or other bleaching creams.
- Pregnancy, or until periods return and breastfeeding ends.
- Cancer or other cancers.
- Use of Isotretinoin within the previous 6 months.
- Tanned skin (active tan from sun exposure or tanning bed) within the previous 30 days.
Note: The hair removal machine may not be effective on white hair.
9. Possible Side Effects
- Superficial skin damage may occur after laser treatment.
- Temporary hyperpigmentation or depigmentation may occur, which usually recovers naturally within several months.
- Mild pain may occur during operation, but local anesthesia ointment can relieve it (anesthesia is not always necessary).
- Purpura (HSP) may appear at the treatment site within 1-3 days post-treatment.
- Minor swelling and redness may occur, typically subsiding within a few hours.
10. User Tips
- Always perform a test treatment on a small area to determine the optimal energy density for the patient's skin type and hair color.
- For darker skin tones, start with lower energy settings to minimize epidermal absorption and heat damage.
- Ensure the handpiece is in full contact with the skin during treatment to maximize effectiveness and cooling.
- Regularly clean the sapphire treatment head to maintain optimal performance and prevent skin damage.
- Treatment intervals vary by body area; typically 30 days for facial hair and 60 days for body/limb hair. The best time for treatment is when hair begins to regrow.
- Always wear protective eyewear during operation.
- Keep the machine and surrounding area clean and free of obstructions.
11. Shutdown Procedure
- Click the shutdown button on the touch screen. A "Shutdown Protection Pattern" will appear, and the machine will shut down after a period. Do not disconnect power during this time.
- Rotate the key switch to "OFF".
- Close the hull switch on the rear chassis and disconnect the external power supply.
- If a forced shutdown is necessary (e.g., by pressing the emergency switch or turning off the key switch), current treatment parameters may be lost. Use only in exceptional circumstances.
12. Warranty and Support
We provide comprehensive after-sales service, including installation, debugging, training, and maintenance. The machine comes with a one-year warranty, covering free repairs during this period. After the warranty, we continue to offer repair services and free technical advice.
For further details, please refer to the official user manual PDF: User Manual (PDF)