1. Product Overview
The HoldPeak VC4060B is a versatile 2-in-1 electrical installation tester, combining the functionalities of a high-precision digital multimeter and an insulation resistance tester. This device is designed for comprehensive electrical diagnostics in various applications, offering reliable and accurate measurements.
Key Features:
- 2-in-1 Functionality: Combines insulation testing (0.01 MΩ–60 GΩ) with standard digital multimeter measurements (AC/DC voltage, current, resistance, capacitance, inductance, frequency, temperature).
- High Voltage Safety: Certified for 1000V insulation tests (IEC1010) with dedicated L/E ports, automatic discharge, and overload protection (1000V DC/750V AC).
- Precision Measurements: Features True RMS for AC measurements and a 6000-count LCD display for laboratory-level accuracy.
- Efficiency Functions: Includes Non-Contact Voltage (NCV) detection, data hold, backlight, and auto power-off to enhance usability and battery life.
- Comprehensive Kit: Supplied with essential accessories for immediate use.
Video: Overview of the HoldPeak VC4060B Digital Multimeter and Insulation Tester.
2. Safety Information
WARNING: Read and understand all safety information and operating instructions before using this instrument. Failure to follow these instructions may result in electric shock, fire, or serious injury.
- Always ensure the meter is in good working condition and free from damage before use.
- Do not apply voltage or current that exceeds the maximum rated values for the meter.
- Use caution when working with voltages above 30V AC RMS, 42V peak, or 60V DC, as these pose a shock hazard.
- Always disconnect power to the circuit and discharge all high-voltage capacitors before making resistance, continuity, diode, or capacitance measurements.
- Ensure the test leads are properly connected to the correct input jacks for the desired measurement.
- Do not operate the meter in explosive gas, vapor, or dust environments.
- Replace batteries immediately when the low battery indicator appears to ensure accurate readings.
- Do not use the meter if the battery cover is not properly closed.
3. Package Contents
Carefully unpack the box and ensure all items listed below are present and undamaged. If any item is missing or damaged, please contact your vendor.
- 1 x HoldPeak VC4060B Digital Multimeter and 1000V Insulation Resistance Tester
- 1 x Portable Carrying Case
- 1 x Pair of Test Leads (Red and Black)
- 1 x Pair of Alligator Clips (Red and Black)
- 1 x Temperature Probe (K-type thermocouple)
- 1 x Digital Multimeter Manual (Please note: A detailed PDF manual can be obtained by contacting customer service via email.)

Figure 3.1: Contents of the HoldPeak VC4060B package.
4. Device Layout
The HoldPeak VC4060B features a clear 3-inch backlit LCD display and an intuitive control panel for easy operation.

Figure 4.1: Front panel of the HoldPeak VC4060B.
4.1. Display
The large LCD shows measurement values, units, function indicators, and battery status.
4.2. Rotary Function Dial
Used to select the desired measurement function (e.g., V~, V-, Ω, F, Hz, °C/°F, A~, A-, NCV, L, mA, Insulation Test).
4.3. Buttons
- RANGE: Manually select measurement range.
- ΔREL (Relative Mode): Measures the difference between a stored value and the current reading.
- HZ/DUTY: Toggles between frequency and duty cycle display.
- MAX/MIN: Records maximum and minimum values during a measurement session.
- SELECT: Toggles between different functions within a single dial position (e.g., AC/DC, Resistance/Continuity/Diode).
- TEST: Initiates insulation resistance testing.
- HOLD: Freezes the current display reading.
4.4. Input Jacks
- VΩHz+ (Red): Positive input for Voltage, Resistance, Frequency, Capacitance, Diode, Continuity, and Temperature.
- COM (Black): Common (negative) input for all measurements.
- mA (Red): Positive input for milliampere current measurements.
- 20A (Red): Positive input for up to 20A current measurements.
- L (Red): Live terminal for Insulation Resistance Testing.
- E (Black): Earth terminal for Insulation Resistance Testing.
5. Setup
5.1. Battery Installation
The HoldPeak VC4060B requires batteries (not included) for operation. Please ensure you use the correct type and polarity.
- Turn the multimeter OFF and disconnect all test leads from the input jacks.
- Locate the battery compartment on the rear of the device.
- Use a screwdriver to loosen the screw(s) securing the battery cover.
- Remove the battery cover.
- Insert the required batteries (typically AA or AAA, refer to the battery compartment label for exact type and quantity) observing the correct polarity (+ and -).
- Replace the battery cover and secure it with the screw(s).
6. Operating Instructions
6.1. General Operation
- Power On/Off: Rotate the function dial from OFF to any desired measurement function to power on. Rotate back to OFF to power off.
- Backlight: Press and hold the 'HOLD' button briefly to toggle the backlight on/off.
- Auto Power-Off: The meter will automatically power off after approximately 15 minutes of inactivity to conserve battery life. Press any button or rotate the dial to resume operation.
6.2. Insulation Resistance Testing
This meter can perform insulation resistance tests at 500V and 1000V.

Figure 6.1: Insulation Resistance Testing, NCV, and Continuity Testing.
- Turn the rotary dial to the desired insulation test voltage (500V/500MΩ or 1000V/2GΩ).
- Connect the red test lead to the 'L' input jack and the black test lead to the 'E' input jack.
- Connect the test leads to the circuit or component under test.
- Press the 'TEST' button to initiate the insulation test. The display will show the resistance value.
- Release the 'TEST' button to stop the test.
6.3. Non-Contact Voltage (NCV) Detection
The NCV function allows for quick detection of live AC voltage without direct contact.
- Turn the rotary dial to the 'NCV' position.
- Move the top part of the meter near the conductor or outlet.
- If AC voltage is detected (typically 50-600V AC), the meter will emit an audible beep and a flashing red light.
6.4. Continuity Testing
Use this function to check for open or short circuits.
- Turn the rotary dial to the 'Ω' position. Press 'SELECT' until the continuity symbol (speaker icon) appears on the display.
- Connect the red test lead to the 'VΩHz+' jack and the black test lead to the 'COM' jack.
- Touch the test probes to the two points of the circuit.
- If the resistance is below approximately 50Ω, the buzzer will sound, indicating continuity.
6.5. TRMS Multimeter Functions
The meter offers a wide range of standard multimeter measurements with True RMS accuracy.

Figure 6.2: Examples of various TRMS Multimeter measurements.
For all measurements below, connect the red test lead to 'VΩHz+' (or 'mA'/'20A' for current) and the black test lead to 'COM'.
- AC/DC Voltage (V~ / V-): Select 'V~' for AC voltage or 'V-' for DC voltage. Press 'SELECT' if needed to toggle.
- AC/DC Current (A~ / A- / mA): Select 'A~' or 'A-' for high current, or 'mA' for milliampere current. Ensure correct input jack is used.
- Resistance (Ω): Select 'Ω'. Press 'SELECT' to cycle through Resistance, Continuity, and Diode Test.
- Capacitance (F): Select 'F'.
- Inductance (L): Select 'L'.
- Frequency (Hz) / Duty Cycle: Select 'Hz'. Press 'HZ/DUTY' to switch between frequency and duty cycle.
- Diode Test: Select 'Ω'. Press 'SELECT' until the diode symbol appears.
- Temperature (°C/°F): Select '°C/°F'. Connect the temperature probe to the 'VΩHz+' and 'COM' jacks.
7. Maintenance
7.1. Cleaning
Wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents. Keep the input terminals free from dirt and moisture.
7.2. Battery Replacement
When the low battery indicator appears on the display, replace the batteries as described in Section 5.1. Remove batteries if the meter is not used for extended periods to prevent leakage.
7.3. Fuse Replacement
If the current measurement function stops working, the fuse may need replacement. Refer to the detailed PDF manual for specific fuse types and replacement procedures, or contact qualified service personnel.
8. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| No display or dim display | Low batteries; Meter off; Damaged display | Replace batteries; Turn dial to function; Contact support |
| Incorrect readings | Wrong function selected; Poor test lead contact; Out of range; Damaged test leads | Select correct function; Ensure good contact; Select higher range; Replace test leads |
| Current measurement not working | Blown fuse; Wrong input jack | Replace fuse (refer to manual); Use correct input jack (mA or 20A) |
| Insulation test not working | Incorrect lead connection; Test button not pressed | Connect leads to L and E; Press and hold TEST button |
9. Specifications

Figure 9.1: Technical Specifications of the HoldPeak VC4060B.
| Parameter | Range / Value | Accuracy |
|---|---|---|
| Display Type | Digital Only (6000 Counts) | - |
| DC Voltage (DCV) | 600mV ~ 1000V | ±(0.5% + 5) |
| AC Voltage (ACV True RMS) | 6V ~ 750V | ±(0.8% + 10) / ±(1.0% + 10) |
| DC Current (DCA) | 6mA ~ 20A | ±(1.0% + 5) / ±(2.0% + 5) |
| AC Current (ACA) | 6mA ~ 20A | ±(1.5% + 10) / ±(2.0% + 10) |
| Resistance | 600Ω ~ 60MΩ | ±(0.8% + 5) / ±(1.2% + 5) |
| Capacitance | 60nF ~ 60mF | ±(2.5% + 5) / ±(5.0% + 8) |
| Inductance (L) | 60mH ~ 100H | ±(1.5% + 5) |
| Frequency (Hz) | 10Hz ~ 50kHz | ±(1.0% + 3) |
| Temperature (°C/°F) | -40°C ~ 1000°C / -40°F ~ 1832°F | ±(1.0% + 5) / ±(1.5% + 15) |
| Insulation Test Voltage | 500V, 1000V | - |
| Insulation Resistance Range | 500MΩ, 2GΩ | ±(4% + 2) |
| Dimensions (approx.) | 200mm (H) x 90mm (W) x 50mm (D) | - |
| Weight (approx.) | 480g | - |
| Battery Included | No | - |
10. Warranty and Support
This HoldPeak VC4060B Digital Multimeter and Insulation Tester comes with a 365-day warranty from the date of purchase, covering manufacturing defects. Please retain your proof of purchase for warranty claims.
For technical support, warranty service, or to obtain a detailed PDF version of the user manual, please contact the seller or manufacturer directly. Provide your product model number (VC4060B) and a clear description of the issue when seeking assistance.