1. Introduction
The Hantek DSO4000C series digital oscilloscopes are advanced 4-channel benchtop instruments designed for precise waveform analysis and signal generation. Featuring a 7-inch 64K color TFT display, a 1GS/s sample rate, and bandwidths ranging from 80MHz to 250MHz, these oscilloscopes are suitable for a wide array of electrical and electronic testing applications. They also integrate a 1-channel arbitrary waveform generator, enhancing their versatility for simulating various signals.
Video: Overview of the Hantek DSO4000C series digital oscilloscope.
2. Product Overview
The Hantek DSO4000C series combines a 4-channel oscilloscope with an external DVM and an arbitrary/function waveform generator. Its advanced digital trigger system ensures high sensitivity and low jitter, making it ideal for capturing complex signals. The intuitive interface and independent controls for oscilloscope and waveform generator functions allow for simultaneous and easy operation.
Key Features:
- 4-channel oscilloscope with EXT input and DVM function.
- Bandwidth options: 80MHz, 100MHz, 200MHz, 250MHz.
- 1GS/s real-time sample rate.
- Minimum vertical range of 500μV/div.
- Over 32 types of automatic measurement functions.
- Advanced digital trigger system with high sensitivity and low jitter.
- Over 14 types of trigger functions including edge, overtime, pulse, pattern, and interval.
- Integrated 1-channel arbitrary/function waveform generator (25MHz, 12-bit resolution, 200MHz DDS) capable of generating sine, square, triangular, ramp, trapezoidal, impulse, and DC waves.
- USB Host/Device for PC communication and U-stick storage/system updates.
- Support for SCPI remote control commands.
- Optional RS232, LAN, HDMI ports, touch screen, WIFI communication, and SD card storage.

Figure 2.1: Front view of the Hantek DSO4000C series oscilloscope.

Figure 2.2: Angled view of the Hantek DSO4000C series oscilloscope.

Figure 2.3: Right side view of the Hantek DSO4000C series oscilloscope.

Figure 2.4: Left side view of the Hantek DSO4000C series oscilloscope.

Figure 2.5: Top view of the Hantek DSO4000C series oscilloscope.

Figure 2.6: Rear view of the Hantek DSO4000C series oscilloscope.
3. Package Contents
Upon unboxing your Hantek DSO4000C series oscilloscope, please verify that all the following items are included and in good condition:
- 1 x Hantek DSO4000C Series Oscilloscope
- 4 x Oscilloscope Probes
- 1 x BNC Cable
- 1 x USB Cable
- 1 x Power Cable
- 1 x English User Manual
- 1 x Calibration Certificate

Figure 3.1: Hantek DSO4000C series oscilloscope and accessories.
4. Setup
4.1 Power Connection
- Ensure the oscilloscope is placed on a stable, level surface with adequate ventilation.
- Connect the provided power cable to the power input on the rear panel of the oscilloscope.
- Plug the other end of the power cable into a suitable AC power outlet.
4.2 Probe Connection
- Connect the oscilloscope probes to the BNC input connectors (CH1, CH2, CH3, CH4) on the front panel.
- Ensure a secure connection by twisting the BNC connector until it locks.
- Attach the ground clip of each probe to a common ground point in your circuit.
4.3 Initial Power-On
- Press the power button on the front panel to turn on the oscilloscope.
- The device will perform a self-test and display the boot screen.
- Once the main waveform display appears, the oscilloscope is ready for use.
5. Operating Instructions
5.1 Oscilloscope Functions
The DSO4000C series offers comprehensive oscilloscope functionalities:
- Channel Selection: Use the dedicated channel buttons (CH1-CH4) to enable or disable channels and access their settings.
- Vertical Controls: Adjust the VOLTS/DIV knob to change the vertical scale (amplitude) for each channel. Use the position knob to shift the waveform vertically.
- Horizontal Controls: Adjust the SEC/DIV knob to change the horizontal time base. The position knob shifts the waveform horizontally.
- Acquisition Modes: Select from Normal, Peak Detect, Average, and High Resolution (HR) modes to optimize waveform capture for different signal types.
- Automatic Measurements: The oscilloscope can automatically measure over 32 parameters including Frequency, Period, Peak-to-Peak (Pk-Pk), RMS, Min, Max, Rise Time, Fall Time, Pulse Width, Duty Cycle, and more.
5.2 Trigger System
The advanced digital trigger system allows for stable capture of various signals:
- Edge Trigger: Triggers on the rising, falling, or both edges of a signal.
- Pulse Trigger: Triggers on pulses of specific width, polarity, and condition (e.g., <, >, !=, =).
- Video Trigger: Supports NTSC and PAL video signals, triggering on scan lines, line numbers, odd/even fields, or all fields.
- Slope Trigger: Triggers on rising or falling slopes with specified time ranges and conditions.
- Overtime Trigger: Triggers when a signal remains high or low for a specified duration.
- Pattern Trigger: Triggers on a defined logic pattern (0: Lower level; 1: High level) across channels.
- Interval Trigger: Triggers based on the time interval between two events.
- Serial Bus Triggers: Includes dedicated triggers for UART, LIN, CAN, SPI, and IIC protocols, allowing for detailed analysis of serial data communication.
5.3 Arbitrary Waveform Generator (AWG)
The integrated AWG provides a single channel output for generating various waveforms:
- Waveform Types: Sine, Square, Ramp, Exponential (EXP), Triangular, Trapezoidal, Impulse, and DC.
- Frequency Range: Sine (0.1Hz-25MHz), Square (0.1Hz-10MHz), Ramp (0.1Hz-1MHz), EXP (0.1Hz-5MHz).
- Amplitude: 5mV~3.5Vp-p (50Ω) or 10mV~7Vp-p (High impedance).
- Resolution: 12-bit vertical resolution, 4KSa waveform depth.
5.4 Connectivity
- USB Host/Device: Connect to a PC for data transfer and remote control, or use a USB stick for waveform storage and system updates.
- SCPI Commands: Supports a variety of SCPI commands for remote control and automation.
- Optional Ports: Depending on your model, optional RS232, LAN, HDMI ports, touch screen, WIFI communication, and SD card storage may be available for enhanced connectivity and control.
6. Maintenance
6.1 Cleaning
- Always power off and unplug the oscilloscope before cleaning.
- Use a soft, damp cloth with mild detergent to clean the exterior surfaces.
- Do not use abrasive cleaners or solvents that could damage the plastic or screen.
- For the display, use a soft, lint-free cloth specifically designed for electronics screens.
- Ensure no liquid enters the device through ventilation openings or ports.
6.2 Storage
- Store the oscilloscope in a cool, dry environment, away from direct sunlight, extreme temperatures, and high humidity.
- If storing for extended periods, disconnect all cables and place the unit in its original packaging or a protective case.
6.3 Calibration
Your Hantek oscilloscope comes with a calibration certificate. For continued accuracy, periodic recalibration by a qualified service center is recommended, especially after significant temperature changes or impacts.
7. Troubleshooting
This section provides solutions to common issues you might encounter with your Hantek oscilloscope.
7.1 No Power
- Check if the power cable is securely connected to both the oscilloscope and the power outlet.
- Verify that the power outlet is functional by plugging in another device.
- Ensure the power switch on the oscilloscope is in the 'ON' position.
7.2 No Waveform Display
- Check if the input probes are correctly connected to the channels and the circuit under test.
- Ensure the channels are enabled (usually indicated by a light next to the channel button).
- Adjust the vertical (VOLTS/DIV) and horizontal (SEC/DIV) scales to bring the waveform into view.
- Check the trigger settings. If the trigger level is set incorrectly, the waveform may not be stable or visible. Try setting the trigger mode to 'Auto'.
- Verify that the signal source is active and producing a signal within the oscilloscope's measurement range.
7.3 Unstable Waveform
- Adjust the trigger level to a point on the waveform that provides a stable display.
- Ensure the trigger source is set to the correct channel.
- Experiment with different trigger types (e.g., Edge, Pulse) to find the most suitable one for your signal.
- Check the probe compensation. Improperly compensated probes can distort waveforms.
7.4 Incorrect Measurements
- Ensure the probes are correctly calibrated for the attenuation factor (e.g., 1X, 10X).
- Verify that the measurement cursors are placed accurately on the waveform.
- Check the input coupling (AC/DC/GND) to ensure it matches your signal type.
8. Specifications
8.1 General Specifications
| Attribute | Value |
|---|---|
| Brand Name | Hantek |
| Model Number | DSO4084C, DSO4104C, DSO4204C, DSO4254C |
| DIY Supplies | Electrical |
| Certification | CE |
| Origin | Mainland China |
| Display Size | 7 Inches & Above |
| Display Resolution | 640*480 Pixels |
| Digital Channels | 4 |
| Battery Included | No |
| High-concerned chemical | None |
8.2 Oscilloscope Performance
| Parameter | DSO4254C | DSO4204C | DSO4104C | DSO4084C |
|---|---|---|---|---|
| Bandwidth | 250MHz | 200MHz | 100MHz | 80MHz |
| Real Time Sampling Rate | 1GSa/S | |||
| Max Waveform Capture Rate | 30,000wfm/s | |||
| Record Length | 64K (Max per single-channel), 32K (Max per dual-channel) | |||
| SEC/DIV Range | 2ns/div ~ 100s/div | |||
| Sample Rate and Delay Time Accuracy | ±50ppm | |||
| AD Converter | 8-bit resolution | |||
| VOLTS/DIV Range | 500μV/div to 10V/div | |||
| Rise Time (typical) | <1.4ns | ≤1.8ns | <3.5ns | ≤4.4ns |
| Input Impedance | 20pF±3 pF, 1MΩ±2% | |||
| Maximum Input Voltage | 300VRMS (10X) | |||

Figure 8.1: Oscilloscope Horizontal, Vertical, and Acquisition Specifications.

Figure 8.2: Oscilloscope Trigger, Inputs, and Measurement Specifications.
8.3 Arbitrary Waveform Generator Specifications
| Parameter | Value |
|---|---|
| Waveform Frequency (Sine) | 0.1Hz ~ 25MHz |
| Waveform Frequency (Square) | 0.1Hz ~ 10MHz |
| Waveform Frequency (Ramp) | 0.1Hz ~ 1MHz |
| Waveform Frequency (EXP) | 0.1Hz ~ 5MHz |
| Amplitude (50Ω) | 5mV ~ 3.5Vp-p |
| Amplitude (High impedance) | 10mV ~ 7Vp-p |
| DAC | 2K ~ 200MHz adjustable |
| Frequency Resolution | 0.001 |
| Channel | 1CH waveform output |
| Waveform Depth | 4KSa |
| Vertical Resolution | 12 bit |
| Frequency Stability | <30ppm |
| Output Impedance | 50 Ω |

Figure 8.3: Arbitrary Waveform Generator Specifications.
8.4 Trigger System (Detailed)

Figure 8.4: Detailed Trigger System Specifications (UART, LIN, CAN, SPI, IIC).
9. Warranty and Support
Hantek products are designed for reliability and performance. This product is covered by a manufacturer's warranty against defects in materials and workmanship. Please refer to the official Hantek website or contact your local distributor for specific warranty terms and conditions applicable to your region.
9.1 Technical Support
For technical assistance, troubleshooting, or service inquiries, please contact Hantek customer support or your authorized Hantek service provider. Have your product model and serial number ready when contacting support.
9.2 After-Sales Service Policy
- All delivered products are tested before shipping to ensure proper functionality.
- In case of missing or defective parts, notify the seller within 5 days of receiving the item for arrangement of re-shipment.
- After-sales service is offered within 1 year of purchase.
10. User Tips
To get the most out of your Hantek DSO4000C series oscilloscope, consider the following tips:
- Probe Compensation: Always compensate your probes before making critical measurements to ensure accurate readings. Most oscilloscopes have a dedicated probe compensation output.
- Grounding: Proper grounding is crucial for accurate and safe measurements. Always connect the probe's ground clip to the circuit's ground.
- Auto-Set Function: When first connecting a signal, use the 'Auto' button to quickly get a stable waveform display. You can then fine-tune the settings manually.
- Save Settings: If you frequently work with similar setups, save your oscilloscope settings to recall them quickly later.
- Firmware Updates: Periodically check the Hantek website for firmware updates to ensure your device has the latest features and bug fixes.
- Environmental Conditions: Operate the oscilloscope within its specified environmental conditions (temperature, humidity) to ensure optimal performance and longevity.