vilnason 2180 Ultrasound Physiotherapy Device

User Instruction Manual

1. Product Introduction

The vilnason 2180 Ultrasound Physiotherapy Instrument is a new hand-held ultrasonic device designed to alleviate various kinds of chronic or acute pain. It features a fashionable appearance, easy operation, and a comfortable handle, making it suitable for personal and household use.

The instrument is primarily composed of a main unit, a 4.0cm² therapeutic probe, and an adapter.

Intended Use: This ultrasonic instrument is suitable for relieving pain, muscle spasms, soft tissue contracture, and softening scars.

vilnason 2180 Ultrasound Physiotherapy Device
Figure 1: vilnason 2180 Ultrasound Physiotherapy Device

2. Safety Precautions and Contraindications

2.1 General Safety Information

  • Do not connect this device with other electronic instruments.
  • Ensure the treatment site and treatment head are cleaned with alcohol or mild soapy water before use.
  • Always use ultrasound couplant during treatment. Apply enough couplant to the skin surface and ensure the treatment head is in full contact with the skin for effective ultrasound transmission.
  • The ultrasonic couplant is a consumable; use only couplant with domestic medical device registration and within its validity period.
  • Monitor the equipment and patient status during treatment to ensure safety.
  • If the treatment head's temperature exceeds 42°C, the time indicator will flash twice as a warning.
  • A continuously lit time indicator signifies good contact between the treatment head and skin; a flashing indicator means poor contact.
  • If any abnormality occurs with the instrument or patient, stop treatment immediately and take appropriate measures.
  • If discomfort occurs after treatment, stop using it immediately and consult a doctor.

2.2 Contraindications

Do not use the device for the following patients or with the consent of a doctor:

  • Patients with implantable electronic devices such as cardiac pacemakers.
  • Patients using oxygen generators or artificial hearts.
  • Patients with malignant tumors.
  • Individuals with hemophilia or hemorrhagic diseases.
  • Patients with heart disease (especially those with implanted pacemakers or other implantable instruments).
  • Patients with acute infectious diseases or purulent wounds.
  • Those who have had deep X-ray therapy or radioisotope therapy.
  • Pregnant women.
  • Without proper examination by a physician.

3. Setup

The vilnason 2180 is a hand-held device that does not require complex installation. Simply connect the power adapter to begin use.

3.1 Connecting the Power Supply

  1. Place the equipment on a flat and stable surface to prevent tilting, vibration, or falling.
  2. Ensure the power switch on the device is in the "OFF" position before connecting the power cord.
  3. Insert the adapter interface into the device's power connection socket.
  4. Insert the adapter plug into a correctly rated power supply network (e.g., 110V for US plug).
  5. Use only the medical adapter provided by the manufacturer. Using an unsuitable or damaged adapter can affect or damage the instrument.
Device with US plug adapter
Figure 2: Connecting the US Plug Adapter

4. Operating Instructions

4.1 Device Components

Labeled diagram of device controls
Figure 3: Device Controls and Indicators
  1. Time indicator: Displays remaining treatment time.
  2. Time selection key: Adjusts treatment duration (5, 10, 15 minutes).
  3. Power indicator: Shows device power status.
  4. Output mode indicator: Displays the selected intensity mode (L, M, H).
  5. Output mode selection key: Cycles through Low (L), Medium (M), and High (H) intensity modes.
  6. Power switch: Turns the device ON/OFF.
  7. Ultrasonic probe: The therapeutic head that delivers ultrasound waves.
  8. Power Supply: Connection point for the adapter.

4.2 Usage Steps

  1. After connecting the power supply (as described in Section 3.1), turn the power switch to the "ON" position.
  2. Apply a generous amount of ultrasound couplant to the skin area you wish to treat.
  3. Press the "Time selection key" to choose your desired treatment duration (5, 10, or 15 minutes).
  4. Press the "Output mode selection key" to select the desired intensity level: Low (L), Medium (M), or High (H).
  5. Place the ultrasonic probe firmly and flatly on the skin surface where the couplant has been applied. Ensure full contact. The time indicator will light up steadily when good contact is made.
  6. Move the probe slowly and continuously over the treatment area. Do not hold the probe stationary for extended periods.
  7. For optimal penetration, stay for approximately 30 seconds on each spot, allowing the ultrasound wave to enter the skin surface.
  8. The device will automatically shut off after the selected time has elapsed.
Over-temperature protection and timing adjustment features
Figure 4: Over-temperature Protection and Smart Timing
Diagram showing 1MHz ultrasound depth and effective radiation area
Figure 5: Ultrasound Penetration Depth and Area

4.3 Application Examples

The device can be used on various body parts for pain relief:

Person using device on knee
On the knee for joint pain
Person using device on elbow
On the elbow for muscle relief
Person using device on ankle
On the ankle for sprains or stiffness
Person using device on shoulder
On the shoulder for muscle tension

4.4 Product Demonstration Video

Video: Demonstration of the vilnason 2180 device in use for pain relief on various body parts.

5. Maintenance

  1. After each treatment, turn the power switch to the "OFF" position.
  2. Remove the adapter from the power outlet and disconnect it from the device.
  3. Clean the treatment site and the ultrasonic probe with alcohol or a mild soapy water solution. Ensure no liquid enters the device's internal components.
  4. Store the device in a clean, dry place, away from direct sunlight and extreme temperatures.
  5. Do not use abrasive cleaners or solvents on the device.
  6. If the adapter is damaged, contact the manufacturer or dealer for a replacement. Do not attempt to repair it yourself.

6. Troubleshooting

6.1 Device Not Turning On

  • Check if the power adapter is securely connected to both the device and the power outlet.
  • Ensure the power switch is in the "ON" position.
  • Verify that the power outlet is functional.

6.2 No Sensation or Weak Sensation During Treatment

  • Ensure sufficient ultrasound couplant is applied to the skin.
  • Verify that the ultrasonic probe is in full and firm contact with the skin. The time indicator should be steadily lit.
  • Check if the selected intensity mode (L, M, H) is appropriate for your needs. Increase the intensity if necessary.
  • Move the probe slowly and continuously over the treatment area.

6.3 Treatment Head Overheating

  • If the time indicator flashes twice, the treatment head temperature has exceeded 42°C. Stop treatment immediately.
  • Allow the device to cool down before resuming use.

7. Specifications

Parameter Specification
Security classification Class II BF application equipment
Operation mode Continuous operation
Network power input 100-240V, 50/60Hz, 600mA
Adapter output 15V = 1.2A
Acoustic working frequency 1MHz ±10%
Rated output power 4.8W ±20% (When duty cycle is 50%)
Maximum output power of time 9.6W
Actual output power Low gear: 2.88W ±20%
Medium gear: 3.84W ±20%
High grade: 4.80W ±20%
Maximum sound intensity of beam 9.6W/cm²
Pulse repetition rate 100Hz ±10%
Duty cycle Low: 30%, Medium: 40%, High gear: 50%
Pulse duration 3ms, 4ms, 5ms
Pulse repetition period 10ms
Ratio of time maximum output power to output power 2, 2.5, 3.33
Usage time 5 minutes, 10 minutes, 15 minutes, three gears adjustable
Effective radiation area (AER) 4.0cm² ±20%
Effective sound intensity (le) Low grade: 0.72W/cm² ±20%
Medium: 0.96W/cm² ±20%
High-grade: 1.20W/cm² ±20%

8. User Tips

  • Use with Gel: Always apply a specialized ultrasound gel to the treatment area. This ensures proper transmission of the ultrasonic waves into the skin and helps you feel the effects.
  • Slow Movements: Move the device slowly and continuously over the skin. Rapid movements may reduce the effectiveness of the treatment.
  • Heat Sensation: The device may produce a warming sensation, especially on higher settings. This is normal and can contribute to pain relief.
  • Consistency is Key: For chronic conditions, consistent daily use over several sessions (e.g., 10 daily sessions) is often recommended to observe significant improvement.
  • Plug Compatibility: If the included power cord is not compatible with your region's outlets (e.g., US plug vs. EU plug), you may need a converter.
  • No Beep for Timer: Be aware that the timer may not have an audible alert when it expires, so keep track of your session duration.

9. Frequently Asked Questions (FAQ)

9.1 Can this device be used in water?

No, this is an electronic device and should not be used in water or submerged in any liquids. Ensure the device and your hands are dry during operation.

9.2 Is this device useful for plantar fasciitis?

Yes, the vilnason 2180 can be used for pain relief associated with conditions like plantar fasciitis, as it helps with muscle and joint pain relief.

9.3 How long should I use the device for each session?

The device has a smart timer with options for 5, 10, or 15 minutes. The appropriate duration depends on the treated area and individual needs. Consult the operating instructions for guidance.

10. Warranty and Support

For warranty information, technical support, or replacement parts, please contact your original point of purchase or the manufacturer directly. Keep your purchase receipt as proof of purchase.

A detailed user manual is also available in PDF format for download: User Manual (PDF).