1. Introduction
This manual provides detailed instructions for the operation and maintenance of the HeltecBMS HT-RT01 High Precision Intelligent Lithium Battery Pack Internal Resistance Test Analyzer. This instrument is designed for testing and high-speed sorting of low-resistance and large-scale lithium battery packs. It measures internal resistance (mΩ) and voltage (V) to assess battery performance and facilitate matching for battery pack assembly.
The instrument utilizes a high-performance single-crystal microcomputer chip from ST Microelectronics and a high-resolution A/D conversion chip for precise measurement. It employs a 1.000KHZ AC positive current synthesized by a phase-locked loop as the measurement signal source.

2. Product Features
- **High Precision Measurement**: Achieved through a Microchip Technology high-resolution 18-bit AD conversion chip.
- **Dual Display**: Double 5-digit display with a highest resolution of 0.1µΩ/0.1mV.
- **Wide Range & Automatic Switching**: Covers a broad range of measurement needs with automatic multi-unit switching.
- **Automatic Polarity Discrimination**: No need to manually distinguish battery polarity; the instrument automatically identifies and displays.
- **Balanced Input Kelvin Four-wire Measuring Probe**: Ensures high anti-interference and more accurate results by avoiding contact resistance and wire resistance interference.
- **1KHZ AC Current Measurement**: Provides high accuracy for internal resistance measurements.
- **Versatile Compatibility**: Suitable for various battery/pack measurements below 100V, including ternary lithium, lithium iron phosphate, lead acid, lithium ion, lithium polymer, alkaline, dry battery, nickel-metal hydride, nickel-cadmium, and button batteries.
- **Computer Connectivity**: Equipped with a serial connection terminal for expanded instrument measurement and analysis functions via PC software.
3. Packaging List
Please check the contents of your package against the list below:

- HT-RT01 Tester x1
- 4-wire resistance test pen x1
- Test fixture x1
- USB data cable x1
- Power supply cord x1
- Manual x1
4. Setup
- **Power Connection**: Connect the power supply cord to the instrument and then to a 110V/220V AC power outlet. Turn on the main power switch located on the rear panel.
- **Test Probe/Fixture Connection**: Correspondingly connect the precision test fixture or the 4-wire resistance test pen to the red (+) and black (-) sockets on the instrument's front panel.
- **Initial Power On**: Press the 'O/Enter' button to turn on the device. The display will show the internal resistance and voltage values.
5. Operation
5.1. Measuring Internal Resistance and Voltage
- **Connect Battery**: Place the battery into the test fixture or connect the test pen to the battery terminals. Ensure correct polarity (red to positive, black to negative).
- **Read Measurement**: The screen will automatically display the internal resistance (in mΩ) and voltage (in V) of the battery under test. The instrument defaults to automatic gear selection.

5.2. Battery Screening
The instrument can be used to quickly screen and match batteries based on their internal resistance and voltage. The smaller the internal resistance, the longer the lifespan and better rate performance of lithium ions.
When assembling battery packs, it is crucial to screen and match cells based on capacity, internal resistance, and voltage. The performance of a battery pack is limited by the worst cell (the "barrel" principle).

5.3. Language Switch
To switch the display language between Chinese and English:
- Press the "Set" key.
- Select the "Language" menu option.
- Press the "Sel" key to toggle between Chinese and English.
- Press the "O/Enter" key to confirm your selection.
5.4. Calibration
5.4.1. Calibrating Resistance
To calibrate the resistance:
- Connect a standard resistance (e.g., 20mΩ) to the tester.
- Press the "Set" key.
- Select the "Cal R" (Calibration Resistance) menu option.
- Press the "O/Enter" key to enter the calibration menu.
- Use the "Vol-/Res+" and "Sel" keys to adjust the displayed value to match the standard resistance.
- Press "O/Enter" to save.
5.4.2. Calibrating Voltage
To calibrate the voltage:
- Connect a battery with a known, accurate voltage to the tester.
- Press the "Set" key.
- Select the "Cal V" (Calibration Voltage) menu option.
- Press the "O/Enter" key to enter the calibration menu.
- Use the "Vol-/Res+" and "Sel" keys to adjust the displayed value to match the known voltage.
- Press "O/Enter" to save.
5.5. Connecting to Computer for Data Analysis
The HT-RT01 can connect to a computer via USB for advanced data analysis and battery screening.
- **Install Software**: Contact the merchant to download and install the necessary driver software and the "Battery Tester" application.
- **Enable Serial Port on Tester**:
- Press the "Set" key to enter the menu.
- Navigate to the "Serial Port Setting" menu (usually item 5 or 6).
- Select "ON" using the "Sel" key.
- Press "O/Enter" to confirm. A USB icon will appear on the tester's screen.
- **Connect USB**: Connect the tester to your computer using the provided USB data cable.
- **Open Software**: Launch the "Battery Tester" software on your computer.
- **Select COM Port**: In the software, select the correct COM port (e.g., COM1) for the tester.
- **Start Monitoring**: Open the serial port in the software to view real-time resistance and voltage data. You can set values for voltage and internal resistance to quickly screen batteries (PASS/NG).

6. Specifications
| Parameter | Value |
|---|---|
| Measurement Parameters | AC resistance, DC resistance |
| Precision (IR) | ±0.5% |
| Precision (V) | ±0.5% |
| Measuring Range (IR) | 0.001mΩ ~ 200Ω |
| Measuring Range (V) | 0.001V ~ ±100VDC |
| Frequency | AC 1KHz |
| Signal Source Current (2mΩ/20mΩ gear) | 50mA |
| Signal Source Current (200mΩ/2Ω gear) | 5mA |
| Signal Source Current (20Ω/200Ω gear) | 0.5mA |
| Measurement Range (Resistance) | 6 gear adjustment |
| Measurement Range (Voltage) | 3 gear adjustment |
| Test Pace | 5 times /S |
| Calibration (Resistance) | Manual calibration |
| Calibration (Voltage) | Manual calibration |
| Power Supply | AC110V/AC220V |
| Supply Current | 50mA~100mA |
| Range | Automatic Switching |
| Measuring Input | Balance System |
| Measuring Probes | LCR Kelvin 4-wire clamp |
| Language | Switch language (Chinese/English) |
| USB Port | Connecting to computer |
| Size | 190*180*80mm |
| Weight | 1.1kg |
7. Maintenance
To ensure the longevity and accuracy of your HeltecBMS HT-RT01, please follow these maintenance guidelines:
- **Cleaning**: Regularly wipe the instrument with a soft, dry cloth. Do not use abrasive cleaners or solvents.
- **Storage**: Store the instrument in a dry environment, away from direct sunlight, extreme temperatures, and high humidity.
- **Probe Care**: Keep the test probes clean and free from debris. Ensure the connections are secure before each use.
- **Software Updates**: Periodically check with the merchant for any available software or firmware updates to ensure optimal performance and compatibility.
8. Troubleshooting
If you encounter issues with your HT-RT01, please refer to the following common problems and solutions:
8.1. Measurement Inaccuracies or Errors
- **Interference**: Do not use a mobile phone near the instrument during measurement, as radio frequency signals can interfere and cause errors. Keep the instrument away from high-frequency switching power supplies.
- **Environmental Conditions**: The instrument cannot enter mist and must be used in a dry environment. Moisture can lead to inaccurate measurements.
- **Poor Contact**: When using the test fixture or test clamp, ensure the contact is good. Otherwise, the measured internal resistance value may be too large or incorrect.
- **Battery Surface Condition**: Pay attention to whether the positive and negative electrodes of the battery have dirt or oxide layers. Rub the electrodes with fine sandpaper to expose the metallic luster for accurate readings.
- **Test Clip Parallelism**: During the test, keep the test clips parallel to reduce the influence of eddy current and improve test accuracy.
8.2. Negative Resistance/Voltage Display
- If the display shows a negative value for resistance or voltage, it indicates that the test probes are connected in reverse polarity to the battery. Correctly connect the positive (red) and negative (black) electrodes.
8.3. No Reading or Incorrect Reading
- **Power Supply**: Ensure the instrument is correctly connected to a 110V/220V power supply and the main power switch is ON.
- **Input Voltage Limit**: The maximum input voltage for this product is 100V. Do NOT test battery packs exceeding 100V, as this can damage the instrument.
- **Software Connection**: If connecting to a computer, ensure the instrument's serial port settings are open *before* connecting the USB cable and opening the software.
9. User Tips
- For consistent and accurate results, always ensure the battery terminals are clean and free of corrosion before testing.
- When using the test fixture, make sure the battery is seated firmly to establish good contact with the probes.
- Regularly calibrate your device, especially if you notice any discrepancies in readings or after prolonged use.
10. Safety and Compliance
10.1. Disclaimer
⚠️ WARNING: This product can expose you to chemicals including lead, which is known to the State of California to cause cancer and birth defects or other reproductive harm. For more information go to www.P65Warnings.ca.gov.
10.2. General Safety Precautions
- Always handle batteries with care.
- Do not attempt to open or modify the instrument.
- Ensure the instrument is used within its specified voltage limits (max 100V).
- Keep the instrument out of reach of children.
11. Warranty and Support
The HeltecBMS HT-RT01 comes with a one-year warranty. For any technical support, software downloads, or warranty claims, please contact the merchant directly.