1. Introduction
This manual provides detailed instructions for the safe and efficient operation of your MYAMi Laboratory Adjustable Programmable DC Power Supply. Please read this manual thoroughly before use and keep it for future reference.
1.1 Product Overview
This series of products features a high-precision single-output programmable DC switch-mode power supply, designed for portability, compactness, and adjustable operation modes. The system is fully microprocessor (MPU)-controlled and can be easily connected to computers via RS-232 communication interfaces, meeting users' needs for automated testing and control. Its software commands strictly adhere to SCPI standards, enabling users to develop custom automation applications. The fully digital system allows precise data input through keyboard and knob controls, ensuring speed and accuracy. It is widely applicable in electronics development, device testing, telecommunications, laboratories, and research institutions.
1.2 Key Features
- Compact and lightweight design.
- Dual five-digit digital display for voltage and current.
- High resolution and precision.
- Constant voltage/constant current automatic switching.
- Equipped with an output control switch.
- Fast operation with high-speed adjustment knobs and digital key inputs.
- Built-in buzzer for alerts.
- Supports 100 sets of storage and retrieval functions.
- Remote Sensing for load line voltage drop compensation.
- Comprehensive protection: Overload, reverse polarity, overvoltage, overcurrent, overtemperature.
- One-click lock feature.
- Temperature-controlled fan for low noise and extended service life.
- Equipped with RS232 interface.
- Supports SCPI and MODBUS programming instruction sets.
2. Safety Overview
This chapter covers important safety instructions. Users must read these instructions carefully before operation to ensure safety and maintain optimal machine condition.
2.1 Safety Symbols
| Symbol | Description |
|---|---|
| ⚠ Warning | Identifies conditions and actions that may endanger the life of the operator. |
| ⚠ Attention | Indicates potential damage to the product or connected devices. |
| ⚡ Danger High Voltage | Indicates presence of high voltage. |
| ★ (Circle with horizontal line) | Grounding terminal. |
| ★ (Three horizontal lines) | Grounding (earth) terminal. |
| ★ (Square with vertical line) | Shell grounding terminal. |
2.2 General Safety Guidelines
- Do not place heavy objects on the housing.
- Avoid serious impact or improper disposal that may cause machine damage.
- Take preventive measures to release static electricity when connecting instruments.
- Do not block the ventilation openings of the side panels and rear panel.
- Do not disassemble the machine unless you are a professional.
2.3 Power Supply Safety
- AC input voltage: 200-240VAC, 50/60Hz.
- Ensure the grounding wire of the power cable is connected to the grounding terminal to prevent electric shock.
2.4 Fuse Safety
- Ensure the correct fuse model is used before turning on the device.
- To prevent fires, only fuses with the product's specified specifications are permitted.
- Before replacing the fuse, first cut off the power supply and identify the cause of the fuse failure.
2.5 Cleaning Safety
- Cut off the power supply before cleaning.
- Wet a soft cloth with mild detergent and water, and do not spray the detergent directly.
- Do not use products containing chemicals or detergents with abrasive properties (e.g., benzene, toluene, xylene, acetone).
2.6 Operational Environment
- Place of use: Indoors, away from direct sunlight, dust, and strong magnetic fields.
- Relative humidity: <80%
- Altitude: <2000m
- Temperature: 0°C ~ 40°C
2.7 Storage Environment
- Indoor
- Relative humidity: <70%
- Temperature: 10°C ~ 70°C
3. Setup
3.1 Unpacking and Inspection
Carefully remove the power supply from its packaging. Inspect the unit for any signs of physical damage. Verify that all standard accessories, including the power cord, user manual, and any interface cables (if applicable), are present.
3.2 Front Panel Description
The front panel provides all controls and displays for operating the power supply.

| No. | Component | Description |
|---|---|---|
| 1 | Mains Switch | Turns the unit on/off. |
| 2 | Function Key | Accesses various functions. |
| 3 | Numbering Keypad | For direct numerical input. |
| 4 | Direction Key | Navigation keys. |
| 5 | Output Negative Terminal | Negative output terminal. |
| 6 | Remote Sensing Terminal Negative | Negative terminal for remote sensing. |
| 7 | Remote Sensing Terminal Positive | Positive terminal for remote sensing. |
| 8 | Output Positive Terminal | Positive output terminal. |
| 9 | Adjustment Knob | For fine adjustment of parameters. |
| 10 | Digital Display Tube | Displays voltage and current. |
| 11 | Function Combination Key | Used in combination with other keys for advanced functions. |
| 12 | One-click Lock Key | Locks the keypad to prevent accidental changes. |
3.3 Rear Panel Description
The rear panel houses power input, cooling, and communication interfaces.

| No. | Component | Description |
|---|---|---|
| 1 | M1 Communication Interface | Standard RS232, optional RS485, USB, LAN. |
| 2 | M4 Airborne Control Interface | Customizable capabilities. |
| 3 | M3 BNC External Trigger Interface | For external trigger signals. |
| 4 | M2 BNC Interface | Customizable functionality. |
| 5 | Power Fan | For cooling the unit. |
| 6 | Power Socket | Includes fuse holder. |
| 7 | Shell Grounding Terminal | For safety grounding. |
3.4 Power Connection
- Ensure the power supply is switched off.
- Connect the provided power cord to the power socket (6) on the rear panel.
- Plug the other end of the power cord into a grounded AC outlet (200-240VAC, 50/60Hz).
- Ensure the shell grounding terminal (7) is properly connected to earth ground for safety.
3.5 Load Connection
- Ensure the power supply output is off before connecting any load.
- Connect the positive (+) terminal of your load to the Output Positive Terminal (8) on the front panel.
- Connect the negative (-) terminal of your load to the Output Negative Terminal (5) on the front panel.
- For remote sensing, connect the sense lines from the load to the Remote Sensing Terminals (6 and 7).
3.6 Communication Interface Setup
The power supply supports RS232, RS485, USB, and LAN interfaces for remote control. For detailed setup, refer to the "Telecontrol" section.
4. Operating Instructions
The instrument is factory-set to front-panel operation mode. Once powered on, you can perform all required settings directly on the panel.
4.1 Keyboard Operation
The front panel contains 24 keys (excluding the POWER key). The text above each key indicates its primary function, activated by pressing directly. The text below the key shows its function key—first press [Shift], then press the corresponding key to activate its function. A "beep" sound is emitted when a key is pressed (buzzer enabled).
| Key Name | Primary Function | Shift Function |
|---|---|---|
| 0-9 | Enter Number | Not customized |
| . | Enter Decimal Point | Not customized |
| ◀ ▶ | Move Cursor Left/Right | Not applicable |
| ▲ ▼ | Flip Pages | Not applicable |
| Enter | Confirm Key | Not applicable |
| U | Set Voltage | Not customized |
| I | Set Current | Not customized |
| Save | Preserve | Not customized |
| Call | Retrieve | Not customized |
| Output | Turn Output On/Off | Not applicable |
| Lock | Button Lock | Menu |
| Esc | Cancel Key | Not applicable |
4.2 Input Modes
4.2.1 Digital Keypad Input
Use the 0-9 numeric keys to enter the value, then press [Enter] to confirm. If you enter incorrect data, press [Esc] to clear the current data and re-enter it.
4.2.2 Knob Input
A digital adjustment knob is used for continuous signal adjustment. Press the [Shift] key (Function Combination Key) to allow the cursor position to shift left or right. Rotating the knob increments or decrements the cursor position, changing the input data immediately.
4.3 Setting Output Voltage
- Press the [U] key to make the voltage display window flash.
- Method 1 (Keypad): Press the 0-9 numeric keys, then press [Enter] to set the output voltage.
- Method 2 (Knob): Use the direction keys to move the cursor to the desired position and adjust the voltage setting by rotating the knob.
Example: To set output voltage to 32.000V, press [U] [3] [2] [.] [0] [0] [0] [Enter].
4.4 Setting Output Current
- Press the [I] key to make the current display window flash.
- Method 1 (Keypad): Press the 0-9 numeric keys, then press [Enter] to set the output current.
- Method 2 (Knob): Use the direction keys to move the cursor to the desired position and adjust the current setting by rotating the knob.
Example: To set output current to 10.000A, press [I] [1] [0] [.] [0] [0] [0] [Enter].
4.5 Overvoltage Protection (OVP) Setting
OVP safeguards the power supply and connected devices against exceeding preset voltage levels. Enable OVP before operation.
- Press [Shift] then [Direction Up] to navigate to the operating parameters menu.
- Select [OVP-S.] to adjust the protection threshold.
- Method 1 (Keypad): Press 0-9 numeric keys followed by [Enter] to set the OVP value.
- Method 2 (Knob): Move the cursor to the desired position and adjust the OVP setting by rotating the knob.
- To enable/disable OVP, navigate to [OVP-E.] in the operating parameters menu and toggle ON/OFF.
Note: If output voltage exceeds the OVP limit, output shuts off, displaying "OVP". Restart with ON/OFF button after resolving the issue.
4.6 Overcurrent Protection (OCP) Setting
OCP safeguards the power supply by limiting load current to preset thresholds. Enable OCP and set the threshold value before operation.
- Press [Shift] then [Direction Up] to navigate to the operating parameters menu.
- Select [OCP-S.] to adjust the protection threshold.
- Method 1 (Keypad): Press 0-9 numeric keys followed by [Enter] to set the OCP value.
- Method 2 (Knob): Move the cursor to the desired position and adjust the OCP setting by rotating the knob.
- To enable/disable OCP, navigate to [OCP-E.] in the operating parameters menu and toggle ON/OFF.
Note: If load current exceeds the OCP limit, output shuts off, displaying "OCP".
4.7 Output Delay Time Setting
This feature is disabled on standard models in this series. If available, press [Shift], then [Lock], then [Direction Up] to go to Run Parameters menu and select [DELAY] to adjust. This setting applies only during automatic loop execution.
4.8 Feature Settings
Press [Shift], then [Lock], then [Direction Up] to access the feature settings menu. Use direction keys to navigate.
- RS232 Baud Rate Setting: Select 'BAUD' and set the baud rate code (e.g., 1200, 2400, 9600, etc.) to match remote computer control requirements.
- Buzzer Settings: Select 'BEEP' to enable or disable the buzzer.
- Quick Access Keys: Select 'H.K.' to enable/disable. When enabled, pressing 0-9 numeric keys retrieves settings from memory groups 0-9.
- Voltage Self-Check Function: Select 'V.SELF' to enable/disable. When enabled, it monitors and adjusts output voltage to minimize deviation.
- Store Operation: Press [Save], enter a group number (0-99), then [Enter] to save current parameters.
- Retrieve Operation: Press [Call], enter a group number (0-99), then [Enter] to retrieve stored parameters.
- Check Software Version: Select 'VER' to display the software version.
- Remote Sensing: Select 'SENSE' to enable/disable. Compensates for voltage drop in load lines under high current. Use twisted pair cables for sense leads.
4.9 Output Switch
The [Output] button on the front panel controls the activation and deactivation of the power supply output.
4.10 Keyboard Lock
The [Lock] button activates or deactivates the keyboard lock. When locked, the numeric keypad displays '-L-L-L-L', and all keys/knobs are locked except [Lock] and [Output].
4.11 Adjustment Knob
Turn the adjustment knob to change the cursor position or adjust parameter values.
4.12 Defensive Functions
The instrument is equipped with overvoltage, overcurrent, and overtemperature protection features. Overtemperature (OTP) protection automatically shuts down the output if abnormal high temperatures occur.
4.13 Operation Modes
4.13.1 Constant Voltage (CV) Operation Mode
- Connect the load to the output terminals (positive to positive, negative to negative) while the output is off.
- Set the current limit value using the [I] key and keypad/knob.
- Set the desired voltage value using the [U] key and keypad/knob.
- Press [Output] to enable output. The display will show the actual measured output.
- Confirm CV mode: Check if the CV indicator light is on. If the CC light is on, increase the current limit.
4.13.2 Constant Current (CC) Operation Mode
- Connect the load to the output terminals while the output is off.
- Set the voltage limit value using the [U] key and keypad/knob.
- Set the desired current value using the [I] key and keypad/knob.
- Press [Output] to enable output. The display will show the actual measured output.
- Confirm CC mode: Check if the CC indicator light is on. If the CV light is on, increase the voltage limit.
4.13.3 Constant Voltage/Constant Current State Switching
In output mode, the power supply operates in CV mode (green indicator) when output current is below the set value. When the current reaches the set value, it switches to CC mode (red indicator). The power supply automatically switches between CV and CC based on the load.
4.14 Telecontrol (Remote Control)
All power supplies in this series support remote control via RS232 or USB interfaces for computer communication, enabling all panel functions through host computer software.
4.14.1 Interface Settings
The instrument's rear panel features RS232, RS485, USB, and LAN interfaces.
4.14.2 Communication Settings
Set up the COM port on your PC with the following configuration:
- Baud rate: 9600
- Check digit: None
- Data bits: 8
- Stop position: 1
- Data outflow control: None
Troubleshooting Remote Control: If the power supply does not respond, check:
- Interface cable connection.
- Interface line and power supply pin connections.
- Secure fastening of the interface cable.
- Consistency of communication parameters between remote control program and device.
- If the command end of the remote control program is a line break (hex 0x0A).
4.14.3 SCPI Programming Instruction Set
For details on the SCPI programming instruction set, refer to the programming manual (PDF).
4.15 Analog Quantity Control Module
This module allows for analog control of voltage and current parameters. It samples analog input and sends corresponding parameter settings to the power supply via an RS232 link, occupying the communication interface.

Control Connection Description:
- V (Positive Power Supply): Recommended 9V—24V (limit 7V—26V).
- K (Power Supply Output State Control): Output when grounding, no output when suspended or high level.
- U (Voltage Control Analog Input): 0---5V or 0---10V (or customized).
- I (Current Control Analog Input): 0---5V or 0---10V (or customized).
- G (Module Power Negative): Ground connection.
Control Accuracy: Internal sampling performed in 4,000 words.
Applicable CNC Models: B/C/D/V series, B6 / C6 / D6 / V6 series.
5. Maintenance
5.1 Periodic Inspection
- Perform regular inspections during product use to ensure optimal performance.
- Check whether the power input socket of the power supply is burned out.
5.2 Fuses Replacement
- Remove the power cord from the unit.
- Use a small screwdriver to carefully remove the fuse box located within the power socket on the rear panel.
- Replace the old fuse with a new one of the correct specification.
- Reinsert the fuse box and reconnect the power cord.
Fuse Specifications:
- T4AL/250V (for models MY-K3010, 2030, 3020, 6010, 8008)
- T8AL/250V (for models MY-K1560, 2045, 3030, 6015, 8010)
5.3 Cleaning
Refer to section 2.5 Cleaning Safety for detailed instructions.
6. Troubleshooting
This section provides solutions to common issues. For problems not listed here, please contact customer support.
- No Power: Check the power cord connection, the power outlet, and the fuse. Replace the fuse if blown.
- No Output: Ensure the output switch is ON. Check load connections. Verify OVP/OCP settings are not triggering protection.
- OVP/OCP Error Display: If "OVP" or "OCP" is displayed, the overvoltage or overcurrent protection has been triggered. Disconnect the load, check for faults in the load circuit, adjust OVP/OCP limits if necessary, and then restart the output.
- Remote Control Issues: Refer to section 4.14.2 Communication Settings for troubleshooting steps related to remote control connectivity.
- Unstable Output: Check load connections for loose wires. Ensure remote sensing is correctly configured if used. Verify the operational environment meets specifications.
7. Specifications
All specifications are subject to change without notice. Ensure the device is powered on for at least 30 minutes for accurate measurements.
| Parameter | Value |
|---|---|
| Rated DC Output (0°C~40°C) | |
| Voltage | 0~30V (Model dependent, up to 300V) |
| Current | 0~10A (Model dependent, up to 60A) |
| Overvoltage Crowbar | 0.1~34V (Model dependent, up to 88V) |
| Overcurrent Protection | 0.1~12A (Model dependent, up to 62A) |
| Voltage Output Characteristics | |
| Power Effect | ≤0.01%+4mV |
| Load Effect | ≤0.01%+5mV |
| Recovery Time | ≤1.5ms (50% load change) |
| Ripple and Noise | 5mVrms, 30mVpp |
| Temperature Coefficient | ≤300ppm/°C |
| Set Precision | ±(0.03% of reading + 10mV) (25±5°C) |
| Set Resolution | 1 mV (5 digits) / 0.1 mV (6 digits) |
| Current Output Characteristics | |
| Power Effect | ≤0.2%+3mA |
| Load Effect | ≤0.2%+5mA |
| Ripple and Noise | ≤10mArms |
| Set Precision | ±(0.03% of reading + 10mA) (25±5°C) |
| Set Resolution | 1 mA (5 bits) / 0.1 mA (6 bits) |
| Display | |
| Voltage Display | 5 digits display / 6 digits display |
| Current Display | 5 digits display / 6 digits display |
| Voltage Resolution | 1 mV (5 digits) / 0.1 mV (6 digits) |
| Current Resolution | 1 mA (5 bits) / 0.1 mA (6 bits) |
| Accuracy of Reading | ±(0.02% of reading + 5mV) (25±5°C); ±(0.05% of reading + 10mA) (25±5°C) |
| General | |
| Protection | Overload, polarity reversal, overvoltage, overcurrent, overtemperature |
| Lock Keyboard | Yes |
| Programming Interface | Standard RS232 (SCPI, MODBUS), optional RS485/LAN |
| Store/Retrieve | 100 groups |
| Insulation | Between output terminal and housing/other terminals; ≤60VDC |
| Operational Environment | Indoor use, Altitude: ≤2000m, Temp: 0°C~40°C, RH: ≤80%, Installation level: II, Pollution level: 2 |
| Storage Environment | Temp: -10°C~70°C, RH: ≤70% |
| Power Input | AC220V±10%, 50/60Hz |
| Size (unprotected) | 400(D)×215(W)×88(H)mm |
| Weight | Approximately 5.5 kg |
8. User Tips
- Users have noted the intuitive interface and accurate output, making it suitable for various lab tasks.
- The unit is generally reported to work properly and provide stable power.
- Utilize the memory storage functions (100 groups) for frequently used voltage/current settings to save time.
- For precise measurements, allow the unit to warm up for at least 30 minutes after powering on.
9. Warranty and Support
For warranty information, technical support, or service inquiries, please contact your seller or the manufacturer directly. Please have your model number (MY-L3003V-PC) and purchase details ready.