MYAMi Programmable DC Power Supply

Contents

User Manual for MYAMi Programmable DC Power Supply

Adjustable Laboratory Digital Tube Programmable 15V-80V, 10A-60A DC Switching Power Supply

1. Security Overview

This chapter covers the operation of the power supply and important safety instructions to be followed during storage. Users should carefully read the following instructions before operation to ensure safety and maintain the machine in optimal condition.

1.1 Security Symbols

The following safety symbols may appear in this manual or on this product:

  • ⚠ Warning: A warning statement that identifies conditions and actions that may endanger the life of the operator.
  • ⚠ Attention: The disclaimer states that damage may occur to this product or other devices connected to it.
  • ⚠ Danger: Warning High Voltage.
  • ⊤ Grounding Terminal: Indicates a grounding terminal.
  • ⊤ Grounding (Earth) Terminal: Indicates an earth grounding terminal.
  • ⊤ Shell Grounding Terminal: Indicates a shell grounding terminal.

1.2 Safety Guidelines

  • Do not place heavy objects on the housing.
  • Avoid serious impact or improper disposal that may cause machine damage.
  • Preventive measures should be taken to release static electricity when connecting instruments.
  • Do not block the ventilation openings of the side panels and rear panel.
  • Do not disassemble the machine unless you are a professional.
  • Power Supply: AC input voltage: 200-240VAC, 50/60Hz. Ensure the grounding wire of the power cable is connected to the grounding terminal to prevent electric shock.
  • Fuse: Ensure the correct fuse model is used before turning on the device. To prevent fires, only fuses with the product's specified specifications are permitted. Before replacing the fuse, first cut off the power supply and identify the cause of the fuse failure.
  • Cleaning: Cut off the power supply before cleaning. Wet a soft cloth with mild detergent and water, and do not spray the detergent directly. Do not use products containing chemicals or detergents with abrasive properties such as benzene, toluene, xylene, and acetone.
  • Operational Environment: Place of use: Indoors, away from direct sunlight, dust, and strong magnetic fields. Relative humidity: <80%. Altitude: <2000m. Temperature: 0°C~40°C.
  • Storage Environment: Indoor. Relative humidity: <70%. Temperature: 10°C~70°C.

2. Product Description

2.1 Introduction

This series of products features a high-precision single-output programmable DC switch-mode power supply, designed for portability, compactness, and adjustable operation modes. The system is fully microprocessor (MPU)-controlled and can be easily connected to computers via RS-232 communication interfaces, meeting users' needs for automated testing and control. Its software commands strictly adhere to SCPI standards, enabling users to develop custom automation applications. The fully digital system allows precise data input through keyboard and knob controls, ensuring speed and accuracy. It is widely applicable in electronics development, device testing, telecommunications, laboratories, and research institutions.

MYAMi Programmable DC Power Supply front view

Figure 1: MYAMi Programmable DC Power Supply

2.2 Characteristics

  • Compact and lightweight, combining the features of a desktop and a system.
  • Dual five-digit digital display for voltage and current, clear and intuitive.
  • High resolution and high precision.
  • Constant voltage/constant current automatic switching.
  • Equipped with an output control switch for more flexible operation.
  • Convenient and fast operation and setting interface, high-speed adjustment knobs and digital key inputs.
  • Built-in buzzer for alerts or warnings.
  • The system supports 100 sets of storage and retrieval functions, allowing you to store or retrieve 100 sets of voltage, current, and other data.
  • Remote Sensing to Achieve Compensation for Load Line Voltage Drop.
  • Overload, reverse polarity protection, overvoltage protection, overcurrent protection, overtemperature protection.
  • One-click lock feature prevents accidental operations.
  • The temperature-controlled fan speed ensures low noise levels for the instrument and prolongs the fan's service life.
  • Equipped with RS232 interface.
  • Supports SCPI and MODBUS programming instruction sets.

2.3 Front Panel and Rear Panel Description

The front panel features a digital display for voltage and current, along with various control buttons and an adjustment knob. The rear panel provides connectivity options and power input.

Front Panel Controls:

  1. Main Switch
  2. Function Key
  3. Numbering Key
  4. Direction Key
  5. Output Negative Terminal
  6. Remote Sensing Terminal Negative
  7. Remote Sensing Terminal Positive
  8. Output Positive Terminal
  9. Adjustment Knob
  10. Digital Tube Display
  11. Function Combination Key
  12. One-click Lock Key

Rear Panel Connections:

  1. M1 Communication Interface (Standard RS232, optional RS485, USB, LAN)
  2. M4 Airborne Control Interface (Customizable)
  3. M3 BNC External Trigger Interface
  4. M2 BNC Interface (Customizable functionality)
  5. Power Fan
  6. Power Socket (including fuse holder)
  7. Shell Grounding Terminal

2.4 Keyboard Description

The front panel contains 24 keys (excluding the POWER key). The text above each key indicates its primary function, which is activated by pressing directly. The text below the key shows its function key-first press [Shift], then press the corresponding key to activate its function. A "beep" sound is emitted when a key is pressed (buzzer enabled).

Key NameFunction of Tonic ChordUpward ScrollKey NameFunction of Tonic ChordUpward Scroll
0Enter 0Not customized7Enter the number 7Not customized
1Enter number 1Not customized8Enter the number 8Not customized
2Enter number 2Not customized9Enter the number 9Not customized
3Enter number 3Not customized.Enter decimal pointNot customized
4Enter number 4Not customized←, →Move the cursor left and rightNot have
5Enter the number 5Not customized↑, ↓Flip pagesNot have
6Enter the number 6Not customizedEnterConfirm keyNot have
USet voltageNot customizedISet currentNot customized
SavePreserveNot customizedCallRetrieveNot customized
OutputTurn Output On or OffNot haveLockButton LockMenu
ShiftUpper - DANGUE FUNCTIONAL SELECTIVITY KeyNot haveEscCancel keyNot have

3. Setup and Initial Configuration

3.1 Power Connection

  • Ensure the power supply is switched off.
  • Connect the power cord to the power socket on the rear panel.
  • Connect the other end of the power cord to a grounded AC outlet (200-240VAC, 50/60Hz).
  • Verify that the shell grounding terminal on the rear panel is properly connected to earth ground for safety.

3.2 Remote Control Interface Settings

The power supply supports remote control via RS232, RS485, USB, or LAN interfaces. These interfaces are located on the rear panel.

Communication Settings:

To establish communication with a PC, configure the COM port with the following settings:

  • Baud rate: 9600
  • Check digit: None
  • Data bits: 8
  • Stop position: 1
  • Data outflow control: None

Note:

  • If the power does not respond during remote control, check:
    • If the interface cable is disconnected.
    • If the interface line and power supply are connected to the corresponding pins of the PC.
    • If the interface cable is securely fastened.
    • If the communication parameters of the remote control program are consistent with those specified above.
    • If the command end of the remote control program is a line break (hex 0x0A).

3.3 Remote Sensing Connection

When the load consumes a large current, a voltage drop occurs on the connection line between the power supply and the load terminals. Under constant voltage conditions, remote sensing can automatically compensate for the voltage drop in the load line.

To use the remote sensing function, you must first set the power supply to remote sensing mode and turn it on only when the output is off. Use twisted pair cables for the Sense lead and avoid twisting them with the load line.

To enable/disable remote sensing:

  1. Press [Shift], then [Lock], and finally [Up] (↑).
  2. Navigate to the menu and select "SENSE".
  3. Use the key or knob to turn it ON or OFF.

Wiring Diagram for Remote Power Measurement:

Connect the SENSE+ terminal on the power supply to the positive terminal of the load.

Connect the SENSE- terminal (often implied or shared with output negative) to the negative terminal of the load.

Connect the OUTPUT+ terminal to the positive terminal of the load.

Connect the OUTPUT- terminal to the negative terminal of the load.

(Note: The diagram in the original manual shows SENSE+ and OUTPUT+ connected to the positive side of the load, and SENSE- and OUTPUT- connected to the negative side of the load, with the SENSE lines bypassing the main load current path to measure voltage directly at the load.)

4. Operation

The voltage and current units in this instrument and its manual are all expressed in volts (V) and amperes (A). The instrument is factory-set to front-panel operation mode. Once powered on, you can perform all required settings directly on the panel.

4.1 Input Modes

Digital Keypad Input:

Use the 0-9 numeric keys to enter the value, then press [Enter] to confirm. If you enter incorrect data, press [Esc] to clear the current data and re-enter it.

Knob Input:

For continuous signal adjustment, use the digital adjustment knob. Press the [Shift] key to enable cursor movement. Rotate the knob right to increment the cursor position and left to decrement. Data changes immediately upon adjustment.

4.2 Output Voltage Setting

Press the [U] key to make the voltage display window flash.

  • Method 1 (Digital Keypad): Press the 0~9 numeric keys, then press the [Enter] key to set the output voltage.
  • Method 2 (Knob): Use the cursor keys (←, →) to move the cursor to the specified position and adjust the voltage setting by rotating the knob.

Example: To set the output voltage to 32.000V, press [U] [3] [2] [.] [0] [0] [0] [Enter].

4.3 Output Current Setting

Press the [I] key to make the current display window flash.

  • Method 1 (Digital Keypad): Press the 0~9 numeric keys, then press the [Enter] key to set the output current.
  • Method 2 (Knob): Use the cursor keys (←, →) to move the cursor to the specified position and adjust the current setting by rotating the knob.

Example: To set the output current to 10.000A, press [I] [1] [0] [.] [0] [0] [0] [Enter].

4.4 Overvoltage Protection (OVP) Setting

Overvoltage protection safeguards power supplies and Devices Under Test (DUTs) against exceeding preset voltage levels. Before operation, enable overvoltage protection.

Setting OVP Threshold:

  1. Press [Shift], then [Up] (↑).
  2. Navigate to the operating parameters menu, then select [OVP-S.].
  3. Method 1 (Digital Keypad): Press the 0-9 numeric keys followed by [Enter] to set the overvoltage protection value.
  4. Method 2 (Knob): Use the cursor keys (←, →) to move the cursor and adjust the OVP setting by rotating the knob.

Example: To set OVP to 33.0V, press [Shift], then [Up] (↑), navigate to [OVP-S], then [3] [3] [.] [0] [Enter].

Enable/Disable OVP:

  1. Press [Shift], then [Up] (↑).
  2. Go to the operation parameters menu, then select [OVP-E.].
  3. Toggle the protection function between ON/OFF.

Important:

  • The external voltage applied to the output must not exceed 120% of the rated voltage, otherwise internal components may be damaged.
  • When the power supply is under overvoltage protection, check for external causes. Once resolved, use the ON/OFF button to restart it.

4.5 Overcurrent Protection (OCP) Setting

Overcurrent protection safeguards the power supply by limiting load current to preset thresholds. Before operation, enable the overcurrent protection and set the threshold value. When the load current exceeds this limit, the output immediately shuts off, with the digital display on the panel showing "OCP". The current adjustment range is restricted by this protection threshold.

Setting OCP Threshold:

  1. Press [Shift], then [Up] (↑).
  2. Navigate to the operating parameters menu, then select [OCP-S.].
  3. Method 1 (Digital Keypad): Press the 0-9 numeric keys followed by [Enter] to set the overcurrent protection value.
  4. Method 2 (Knob): Use the cursor keys (←, →) to move the cursor and adjust the OCP setting by rotating the knob.

Example: To set OCP to 11.0A, press [Shift], then [Up] (↑), navigate to [OCP-S], then [1] [1] [.] [0] [Enter].

Enable/Disable OCP:

  1. Press [Shift], then [Up] (↑).
  2. Go to the operation parameters menu, then select [OCP-E.].
  3. Toggle the protection function status between ON/OFF.

4.6 Output Delay Time Setting

This setting is disabled on standard models in this series.

For models that support it:

  1. Press [Shift], then [Lock], and finally [Up] (↑).
  2. Go to the Run Parameters menu, then select [DELAY] to adjust the delay setting.

Note: This Delay setting applies only during automatic loop execution. Therefore, when memory is stored, this Delay setting is also stored in the memory location.

4.7 Feature Settings

To access feature settings, press [Shift], then [Lock], and finally [Up] (↑). Use the page flip keys (↑, ↓) to select the corresponding menu item.

  • INIT: Restore factory settings.
  • OUT: Output status.
  • V-M: Maximum voltage input.
  • I-M: Maximum current input.
  • BAUD (RS232 Baud Rate Setting): Set the communication baud rate.
    Code1.2K2.4K4.8K9.6K19.2K38.4K57.6K11.52K
    Baud rate1200240048009600192003840057600115200
  • BEEP: Prompt sound (Buzzer settings).
  • H.K.: Shortcut key (Quick Access Keys).
  • V.SELF: Automatic voltage fine-tuning (Voltage Self-Check Function Settings).
  • ADDR: Communication address.
  • SENSE: Compensation function (Remote Sensing).
  • COM-P: Communication protocol.
  • VER: Version number (Check Software Version).

4.8 Store and Retrieve Operations

The power supply can save up to 100 groups of common parameters in non-volatile memory for convenient and quick recall.

Store Operation:

  1. Press [Save].
  2. Set the knob group number (0-99).
  3. Press [Enter] to save the current running parameters.

Example: To store the setting at memory address 5, press [Save] [5] [Enter].

Retrieve Operation:

  1. Press the [Call] key.
  2. Set the group number (0-99).
  3. Press [Enter] to retrieve the group's operational parameters.

Example: To call the function with memory address 5, press [Call] [5] [Enter].

4.9 Output Control and Keyboard Lock

Out Switch:

The [On/off] button on the front panel controls the output's activation and deactivation.

Keyboard Lock:

The [Lock] button on the front panel activates or deactivates the keyboard lock. Pressing [Lock] displays the numeric keypad as '-L-L-L-L'. All other keys and knobs are locked except [Lock] and [On/off].

Adjustment Knob:

Turn the knob to change the cursor position or adjust values when in setting mode.

4.10 Operation Mode Description

Constant Voltage (CV) Operation Mode:

  1. Connect the load to the output terminals (+ and -) while the output is turned off.
  2. Enter the current limit value: Press [I] to enter current setting input mode and set the desired current value using the numeric keypad or knob.
  3. Enter the voltage value to output: Press [U] to enter voltage setting input mode and set the desired voltage value using the numeric keypad or rotary knob.
  4. Start output: Press [On/Off] to enable output. The display will show the actual measured output value.
  5. Confirm CV mode: Check if the CV light is on. If the CC light is on, increase the current limit to ensure the output remains in constant voltage mode.

Constant Current (CC) Operation Mode:

  1. Connect the load to the output terminals (+ and -) while the output is turned off.
  2. Enter the voltage limit value: Press [U] to enter voltage setting mode and set the desired voltage value using the numeric keypad or rotary knob.
  3. Enter the current value to output: Press [I] to enter current setting mode and set the desired current value using the numeric keypad or knob.
  4. Start output: Press [On/Off] to activate the output. The display will show the actual measured output value.
  5. Confirm CC mode: Check if the CC light is on. If the CV light is on, increase the voltage limit to ensure the output remains in constant current mode.

Constant Voltage/Constant Current State Switching:

In output mode, when the output current is below the set value, the power supply operates in constant voltage mode, with the constant voltage indicator light turning green. The output voltage remains stable at the set value until the current reaches the set value, at which point it switches to constant current mode, and the constant current indicator light turns red. The power supply automatically switches between constant voltage and constant current modes based on the load.

5. Maintenance

5.1 Periodic Inspection

  • To ensure optimal performance, perform regular inspections during product use.
  • Check whether the power input socket of the power supply is burned out.

5.2 Fuse Replacement

If the fuse needs replacement:

  1. Remove the power cord from the power supply.
  2. Use a small screwdriver to open and remove the fuse box, typically located near the power socket on the rear panel.
  3. Replace the old fuse with a new one of the correct specification.
  4. Close the fuse box and re-insert the power cord.

Fuse Specifications:

  • T4AL/250V (for models MY-K3010, 2030, 3020, 6010, 8008)
  • T8AL/250V (for MY-K1560, 2045, 3030, 6015, 8010)

6. Troubleshooting

  • No Output / OVP Display: If the power supply shows "OVP" and has no output, check for external causes of overvoltage. Resolve the issue and then use the ON/OFF button to restart the power supply. Ensure the external voltage applied to the output does not exceed 120% of the rated voltage.
  • No Output / OCP Display: If the power supply shows "OCP" and has no output, check if the load current exceeds the preset overcurrent protection threshold. Reduce the load or increase the OCP threshold if appropriate.
  • No Output / OTP Display: The instrument is equipped with overtemperature protection (OTP). If abnormal high temperatures occur, the output will automatically shut down. Allow the unit to cool down and ensure proper ventilation.
  • Remote Control Issues: Refer to the "Remote Control Interface Settings" section for troubleshooting communication problems (cable connection, pin alignment, baud rate, command format).
  • Keyboard Locked: If the keyboard is unresponsive, check if the [Lock] button has been pressed. Press [Lock] again to unlock. Only [Lock] and [On/off] buttons function when locked.

7. Specifications

When using this specification, ensure the device is powered on for at least 30 minutes.

7.1 General Specifications

FeatureDescription
Rated DC Output Temperature0°C ~ 40°C
ProtectOverload protection, polarity reversal protection, overvoltage protection, overcurrent protection, overtemperature protection
Lock KeyboardYes
Programming InterfaceStandard RS232 interface with SCPI and MODBUS command sets, optional RS485/LAN interfaces
Store/Retrieve100 groups
InsulativityBetween the output terminal and the housing or other terminals; ≤60VDC
Operational EnvironmentIndoor use, Altitude: ≤2000m, Ambient temperature: 0°C~40°C, Relative humidity: ≤80%, Installation level: II, Pollution level: 2
Storage EnvironmentEnvironmental temperature: -10 to 70°C, Relative humidity: ≤70%
Power InputAC220V ±10%, 50/60Hz
Size (unprotected)400(D)×215(W)×88(H)mm
WeightApproximately 5.5 kg

7.2 Performance Specifications (Common Models)

Model30V10A30V20A60V10A80V8A
Voltage Output
Voltage Range0~30V0~30V0~60V0~80V
Current Range0~10A0~20A0~10A0~8A
Overvoltage Crowbar0.1~34V0.1~34V0.1~64V0.1~88V
Overcurrent Protection0.1~12A0.1~24A0.1~12A0.1~8.8A
Power Effect≤0.01%+4mV
Load Effect≤0.01%+5mV
Recovery Time≤1.5ms (50% load change)
Ripple and Noise5mVrms, 30mVpp
Temperature Coefficient≤300ppm/°C
Set Precision±(0.03% of reading + 10mV) (25±5°C)
Set Resolution1 mV (5 digits) / 0.1 mV (6 digits)
Current Output
Power Effect≤0.2%+3mA
Load Effect≤0.2%+5mA
Ripple and Noise≤10mArms
Set Precision±(0.03% of reading + 10mA) (25±5°C)
Set Resolution1 mA (5 bits) / 0.1 mA (6 bits)
Display
Voltage Display5 digits display / 6 digits display
Current Display5 digits display / 6 digits display
Voltage Resolution1 mV (5 digits) / 0.1 mV (6 digits)
Current Resolution1 mA (5 bits) / 0.1 mA (6 bits)
Accuracy of Reading±(0.02% of reading + 5mV) (25±5°C); ±(0.05% of reading + 10mA) (25±5°C)

Note: If the technical indicators change, no further statement will be made.

7.3 Performance Specifications (High Power Models)

Model15V60A60V15A80V11A
Voltage Range0~15V0~60V0~80V
Current Range0~60A0~15A0~11A
Overvoltage Crowbar0.1~18V0.1~64V0.1~88V
Overcurrent Protection0.1~62A0.1~18A0.1~12A

8. User Tips

  • For precise voltage and current control, utilize the digital keypad for initial coarse adjustments and then fine-tune with the adjustment knob.
  • Always connect your load to the output terminals before turning on the power supply output to prevent sudden current surges.
  • Regularly check the ventilation openings to ensure they are not blocked, especially during prolonged operation at high power, to prevent overtemperature shutdowns.
  • When performing sensitive measurements, consider using the remote sensing function to compensate for voltage drops across the load wires, ensuring the voltage at the DUT is accurate.
  • Utilize the store/retrieve functions to save frequently used voltage and current settings for quick recall in different experimental setups.

9. Frequently Asked Questions (FAQ)

Q: What is the purpose of overvoltage and overcurrent protection?

A: OVP (Overvoltage Protection) and OCP (Overcurrent Protection) are safety features designed to protect both the power supply and the connected device (DUT) from damage due to excessive voltage or current. They automatically shut off the output if set limits are exceeded.

Q: How do I know if the power supply is in Constant Voltage (CV) or Constant Current (CC) mode?

A: The power supply has indicator lights (CV and CC) on the front panel. A green CV light indicates Constant Voltage mode, while a red CC light indicates Constant Current mode. The unit automatically switches between these modes based on the load and set limits.

Q: Can I control this power supply remotely?

A: Yes, this power supply supports remote control via RS232, and optionally RS485, USB, or LAN interfaces. You will need to use host computer software and configure the communication settings as described in the manual.

Q: What should I do if the unit displays "OVP" or "OCP"?

A: If "OVP" is displayed, check for any external conditions causing overvoltage and restart the unit. If "OCP" is displayed, verify your load and current settings, and adjust them if necessary. Always ensure your settings are appropriate for your application.

10. Warranty and Support

For warranty information, please refer to the terms and conditions provided by your seller at the time of purchase. For technical support or service inquiries, please contact the MYAMi Official Store or your authorized distributor.

A user manual in PDF format is also available for download: View User Manual (PDF)