1. Safety Warnings and Precautions
Always strictly follow the safety rules and precautions listed in this manual to prevent electric shock, personal injury, or equipment damage.
- Pay close attention to the label text and symbols on the instrument panel and back panel.
- Do not place or store the instrument for extended periods under high temperature and humidity, in locations with condensation or direct sunlight.
- If the battery voltage is low, replace the battery promptly.
- If the instrument will not be used for a long time, remove the battery to prevent leakage.
- When replacing the battery, observe correct polarity.
- Use, disassembly, and repair of this instrument must be performed by authorized personnel only.
- If the instrument is deemed dangerous to continue using, stop operation immediately, seal it, and have it handled by an authorized institution.
Electrical Symbols
| Symbol | Description |
|---|---|
| ⚡ | Extremely dangerous! The operator must strictly abide by the safety rules, otherwise there is a danger of electric shock, resulting in personal injury or casualty accident. |
| ⚠ | Danger! The operator must strictly abide by the safety rules, otherwise there is a danger of electric shock, resulting in personal injury or casualty accident. |
| ⚠ (Triangle with exclamation mark) | Warning! Safety rules must be strictly observed, otherwise there is personal injury or equipment damage. |
| □□ | Double insulation |
| ∼ | AC (Alternating Current) |
| ——— | DC (Direct Current) |
2. Product Introduction
The SWEVY Flexible Current Clamp Meter is designed for field testing of AC high current and leakage current. It features high precision, good linearity, and strong anti-interference ability, making it an excellent alternative to conventional current meters and large diameter clamp meters for measuring high current leakage. It is widely used across various sectors including electric power, communications, meteorology, railways, oil fields, construction, measurement, scientific research, teaching, industrial, and mining enterprises.
This instrument is particularly suitable for current testing in industrial environments where signals are severely distorted, such as dense wiring, transformer wall-to-wall core current tests, and thick wire cable tests. It also supports relay protection, thyristor rectification, variable frequency speed regulation, semiconductor switch, power electronic conversion equipment, and arc welding applications.
Utilizing advanced Rogowski Coil technology, the flexible current sensor is a toroidal coil evenly wound on non-ferromagnetic materials. This design ensures no hysteresis effect, no magnetic saturation, high linearity, and strong anti-interference. The output signal is the current-to-time differential, which, when integrated, accurately restores the input current. The measured current range spans from milliamperes to tens of thousands of amps.
The flexible current clamp coil part has no exposed metal conductor, ensuring non-contact measurement for safety and speed. Its compact size, light weight, aesthetic appearance, and soft, flexible nature make it ideal for use in narrow environments and dense wiring spaces. It offers high accuracy, reliability, bandwidth, and frequency response. Users can customize the coil length as needed.
The meter includes alarm value setting and alarm indication functions, along with capabilities for reading, saving historical data, and data retention. It is an indispensable tool for electrical safety testing.
Figure 1: SWEVY SW-1012A Flexible Current Clamp Meter highlighting 10000A current, 150mm coil inner diameter, alarm setting, AC current/leakage current measurement, and AC/DC voltage test capabilities.
Figure 2: SWEVY SW-1015A Flexible Current Clamp Meter emphasizing 150mm coil inner diameter, 1mA AC current resolution, AC/DC voltage measurement, and 500 data storage capacity.
3. Instrument Structure and Components
Figure 3: Diagram illustrating the components and dimensions of the SW-1015A model.
- Rogowski sensing coil
- Open coil knob
- LCD Monitor
- Button area (MODE, HOLD, SET, Power button)
- Voltage input jack
- Battery cover
4. Setup and Initial Use
4.1. Battery Installation
The instrument requires DC4.5V power, supplied by three 1.5V AAA alkaline batteries. Ensure the battery voltage is sufficient for proper operation.
- Ensure the instrument is powered off.
- Locate the battery cover (Component 6 in Figure 3) on the back of the device.
- Open the battery cover.
- Insert three 1.5V AAA alkaline batteries, paying close attention to the correct polarity (+/-).
- Close the battery cover securely.
Warning: The instrument cannot be tested if the battery cover is not securely closed, as this poses a danger. Incorrect battery polarity will damage the meter. If the meter will not be used for a long time, remove the batteries.
4.2. Power On/Off
- To power on: Press the POWER button. The LCD will display readings.
- To power off: Press the POWER button again.
- The instrument features an automatic shutdown function after approximately 15 minutes of inactivity to conserve battery. Any button press will reset this 15-minute countdown.
- If the LCD display appears dim or dark after powering on, the battery voltage may be too low. Replace the batteries.
5. Operating Instructions
5.1. Current Measurement (AC High Current & Leakage Current)
High voltage, extremely dangerous! Current measurements must be performed by trained and authorized personnel who strictly adhere to safety rules. Do not measure currents above the instrument's upper limit to avoid electric shock, personal injury, or equipment damage.
- Power on the instrument.
- Rotate the open coil knob (Component 2 in Figure 3) to open the flexible clamp head.
- Position the flexible coil around the wire(s) to be measured. Ensure the clamp head is fully closed and locked after positioning.
- Observe the reading on the LCD monitor (Component 3 in Figure 3).
-
Specific Measurement Scenarios:
- Leakage current of electrical equipment: Clamp both the live and neutral wires together. (Note: This requires clamping two roots).
- Leakage current of grounding wire: Clamp only the grounding wire. (Note: This requires clamping a single root).
- Total current of the main line: Clamp only the main line. (Note: This requires clamping a single root).
- If the LCD displays "OL" (Overload), the measured current exceeds the meter's maximum upper limit. Use a meter with a higher current rating for testing.
Video 1: Demonstration of current and voltage measurement, alarm setting, and data storage using the SWEVY flexible current clamp meter.
Figure 4: The flexible coil being used to measure 10000A AC high current in a densely wired electrical distribution cabinet, demonstrating its suitability for narrow spaces.
Figure 5: Close-up of the flexible CT inductive clamp head, designed for accurate measurement even at low currents with a high resolution of 0.1A.
5.2. Voltage Measurement (AC/DC Voltage)
- Power on the instrument.
- Connect the test leads to the voltage input jack (Component 5 in Figure 3) and to the circuit under test.
- Press the MODE button to switch between AC V (AC voltage mode) and DC V (DC voltage mode).
- Read the voltage value displayed on the LCD.
Figure 6: The meter performing an AC/DC voltage test, showing the test leads connected to a power outlet and the display indicating 220V AC.
5.3. Alarm Value Setting
The meter allows setting an alarm value. When the measured value exceeds this preset alarm value, the instrument will emit a prompt.
- While in a measurement mode (e.g., voltage or current), press the SET button to enter alarm setting mode.
- Use the MODE and HOLD buttons to adjust the alarm threshold.
- Press SET again to confirm and save the alarm value.
- When the measured value surpasses the set alarm, the instrument will provide an audible or visual alert.
5.4. Data Retention, Storage, Review, and Erasure
The instrument supports storing up to 500 groups of measurement data.
- Data Retention (HOLD function):
- During a measurement, short press the HOLD button (less than 3 seconds) to display "HOLD" and freeze the current reading. The data is automatically numbered and stored.
- To cancel hold, short press the HOLD button again. The meter will resume measuring.
- If the stored data reaches 500 groups, the "FULL" symbol will flash. Press HOLD to clear the "FULL" flash and return to measurement mode.
- Data Review:
- Press and hold the HOLD button for approximately 3 seconds to enter data review mode. The first stored data group will be displayed.
- Press the SET button to view the next data group.
- Press the MODE button to view the previous data group.
- If no data is stored, "null" will be displayed.
- Long press the HOLD button to exit data review mode.
- Data Erasure:
- While in data review mode, press and hold the POWER button for more than 3 seconds.
- Select "Yes" to confirm clearing all stored data.
- The instrument will automatically return to the test state after data erasure is complete.
6. Maintenance
6.1. Battery Replacement
When the battery power is insufficient, the meter will display a low battery symbol. Replace the batteries as follows:
- Ensure the instrument is powered off by pressing the POWER button.
- Open the battery cover.
- Replace the old batteries with new 1.5V AAA alkaline batteries, ensuring correct polarity.
- Close the battery cover securely.
Warning: Do not operate the instrument without the battery cover securely in place. Remove batteries if the instrument will not be used for extended periods.
7. Troubleshooting
| Problem | Possible Cause | Solution |
|---|---|---|
| LCD display is dim or dark. | Low battery voltage. | Replace the batteries. |
| "OL" symbol displayed during current measurement. | Measured current exceeds the instrument's maximum range. | Use a meter with a higher current rating. |
| "FULL" symbol displayed during data storage. | Data storage capacity (500 groups) is full. | Clear historical data (refer to Section 5.4). |
| Inaccurate leakage current measurement when clamping multiple conductors. | The instrument is designed for accurate leakage measurement when clamping live and neutral together, or a single grounding wire. Clamping multiple conductors (e.g., live, neutral, and ground) simultaneously for leakage can lead to inaccuracies. | Ensure correct clamping method for leakage current: live and neutral together (2 roots) or single grounding wire (1 root). Avoid clamping all conductors for leakage. |
| Instrument does not power on. | Batteries are dead, incorrectly installed, or battery cover is not closed. | Check battery charge, polarity, and ensure battery cover is securely closed. |
8. Technical Specifications
8.1. Model Specifications
| Feature | SW-1012A | SW-1015A |
|---|---|---|
| Coil length | 470mm | 470mm |
| Coil internal diameter | Φ 150mm | Φ 150mm |
| Current range | 0A-10000A | 0mA-10000A |
| Current resolution | 0.1A | 1mA |
8.2. General Technical Specifications
| Function | Specification |
|---|---|
| Measurement Modes | Current test, Leakage current test, AC/DC Voltage test |
| Current Accuracy | ±2% ±5dgt (Wire at center of flexible coil, 23℃±2℃) |
| AC Voltage Range | 0.0V~750V AC |
| AC Voltage Accuracy | ±1.5%rdg ±3dgt |
| AC Voltage Resolution | 0.1V |
| DC Voltage Range | 0.0V~1000V DC |
| DC Voltage Accuracy | ±1.5%rdg ±5dgt |
| DC Voltage Resolution | 0.1V |
| Sampling Speed | 2 times/second |
| Power Supply | DC4.5V (3 x 1.5V AAA alkaline batteries) |
| Backlight | Yes |
| LCD Size | 45mm × 28mm |
| Coil Thickness | Φ 8mm (Suitable for narrow environments and dense lines) |
| Instrument Total Weight | 220g (including 150mm caliber coil and battery) |
| Overflow Display | "OL" symbol |
| Data Storage | 500 groups |
| Data Full Indicator | "FULL" symbol |
| Low Battery Indication | Battery symbol empty when voltage is below 3.15V |
| Insulation Resistance | >100MΩ (1kV) |
| Operating Temperature & Humidity | -10℃~40℃, below 70% RH |
| Storage Temperature & Humidity | -10℃~60℃, below 80% RH |
| Safety Regulations | IEC1010-1, IEC1010-2-032, Pollution Level 2, CAT III (600V) |
| EMC Standard | IEC61326 |
| Instrument Dimensions (Host) | 151mm × 100mm × 35mm (without coil) |
9. User Tips
- For accurate leakage current measurements, ensure you are clamping only the specific conductors as instructed (e.g., live and neutral together, or a single grounding wire). Clamping multiple conductors (like live, neutral, and ground) simultaneously for leakage current can lead to unreliable readings.
- When measuring current, the accuracy can be affected by the position of the wire within the flexible coil. For best results, try to center the wire as much as possible.
- Regularly check battery levels, especially before critical measurements, to avoid interruptions or inaccurate readings due to low power.
10. Warranty and Support
For warranty information, technical support, or service inquiries, please contact your retailer or the manufacturer directly. Keep your purchase receipt as proof of purchase.