ZIBOO UT276A+ UT278A+

Contents

ZIBOO UT276A+ UT278A+ Clamp Ground Resistance Tester

User Manual

1. Introduction

Thank you for purchasing the ZIBOO Clamp Ground Resistance Tester. This instrument, also known as a loop resistance tester, is designed for precise measurement of grounding resistance. It features a modern black luxury screen, simultaneous display of resistance and current, data storage, data access, alarm functions, and automatic shutdown. Its robust design offers shockproof, dustproof, and moisture-proof protection, making it an indispensable tool for telecommunications, electricity, meteorology, computer rooms, oil fields, electromechanical installation and maintenance, and various industrial applications.

The instrument is controlled by a microprocessor and utilizes a fast filtering technique to minimize interference, ensuring accurate ground resistance detection. It is suitable for measuring various grounding resistances, including those in telecommunications, electricity, meteorology, and computer systems.

Safety Rules and Precautions

Before using the instrument, please read and strictly observe the following safety rules and precautions to prevent electric shock or personal injury:

  • The instrument is designed, produced, and inspected according to IEC61010 safety specifications.
  • Always pay special attention to safety when using this instrument.
  • Avoid using high-frequency signal generators (e.g., mobile phones) near the meter during measurement to prevent errors.
  • Familiarize yourself with the text and symbols on the instrument's body.
  • Ensure the instrument and accessories are in good condition before use.
  • Before powering on, press the trigger once or twice to ensure the jaws are properly closed.
  • Do not measure in flammable environments, as sparks may cause an explosion.
  • Do not withhold the trigger or clamp any wires when turning on the meter.
  • After power-on, the "OL Ω" symbol should be displayed. The measured object can then be clamped.
  • Avoid storing the instrument for long periods in high temperature, humidity, condensing conditions, or direct sunlight.
  • Ensure the meter is turned off when replacing batteries.
  • If the low battery symbol (Battery Level Indicator) appears, replace the batteries promptly to maintain measurement accuracy.
  • Keep the jaw contact surfaces clean; do not wipe with corrosive or rough objects.
  • When opening the trigger, avoid impacting the clamp, especially the jaw joint.
  • A slight noise from the clamp head during resistance measurement is normal. The "deep-beep-beep--" sound indicates an alarm.
  • Adhere to the specified measuring range and operating environment.
  • Do not measure lead current exceeding the clamp meter's upper limit.
  • All use, disassembly, calibration, and maintenance must be performed by authorized personnel.
  • If the instrument is deemed unsafe for continued use, stop operation immediately, seal it, and have it handled by a qualified organization.
  • Strictly follow all instructions related to the safety warning sign (Alarm Indicator Symbol) in this manual.

2. Product Overview

ZIBOO UT276A+ Clamp Ground Resistance Tester

Figure 1: ZIBOO UT276A+ Clamp Ground Resistance Tester

Key Features

Key features of the tester including resistance measurement, maximum range, alarm function, LCD, automatic range, and data storage.

Figure 2: Key Features

  • Resistance Measurement: Measures ground resistance.
  • Maximum Range: Capable of measuring up to 1200Ω.
  • Alarm Function: Provides audible and visual alerts for out-of-range measurements.
  • LCD (Liquid Crystal Display): Clear digital display for readings.
  • Automatic Range: Automatically adjusts to the appropriate measurement range.
  • Data Storage: Stores up to 300 groups of measurement data.
  • High-Precision Measurement: Ensures accurate readings.
  • Insulation Protection: Designed for user safety.
  • Seamless High Precision Clamp Head Design: Features a 32mm jaw opening for accurate detection.

Product Components and Dimensions

Diagram showing product dimensions and labeled parts: Jaw, Jaw Opening Trigger, LCD Display, AL Alarm Button, MODE Mode Key, Power Button, HOLD Hold Key, MEM Storage Key, SET setting key.

Figure 3: Product Parameters and Labeled Parts

The ZIBOO Earth Clamp Tester has dimensions of approximately 28.5cm (11.2in) in length and 5.8cm (2.28in) in width. Key components include:

  • 1. Clamp Jaw: The opening mechanism for clamping wires.
  • 2. Trigger: Used to open the clamp jaw.
  • 3. LCD: Liquid Crystal Display for showing measurements and status.
  • 4. POWER Key: Turns the device on/off.
  • 5. MODE Key: Switches between measurement modes.
  • 6. AL Alarm Function Key: Activates/deactivates and sets alarm thresholds.
  • 7. MEM Key: For data storage and retrieval.
  • 8. SET Key: Used to confirm settings.
  • 9. HOLD Key: Locks the current display data.
  • 10. CALIBRATION LOOP: 1Ω
  • 11. CALIBRATION LOOP: 10Ω

LCD Display Details

Detailed breakdown of the LCD display symbols: Clamp Head Open Symbol, Alarm Indicator Symbol, Greater-Than Sign, Data Lookup Symbol, Data Storage Symbol, Noise Indicator Symbol, Data Lock Symbol, Battery Level Indicator, Unit Instructions, Four Digit Resistance Data, Number of Storage Groups.

Figure 4: LCD Display Symbols

  1. Jaw Open Symbol
  2. Alarm Indicator Symbol
  3. Greater-Than Sign
  4. DC AC Symbol
  5. Data Lookup Symbol
  6. Data Storage Symbol
  7. Noise Indicator Symbol
  8. Data Lock Symbol
  9. Battery Level Indicator
  10. Unit Instructions
  11. Four Digit Resistance Data
  12. Number of Storage Groups

3. Setup

Battery Installation and Check

The device requires 4 AA alkaline dry batteries (5th alkaline battery 4PCS). To install, open the battery compartment on the back of the device. Insert the batteries according to the polarity markings. Close the compartment securely.

After turning on the device, observe the LCD. If the battery voltage low symbol (Battery Level Indicator) is displayed, the batteries are running low and should be replaced to ensure accurate measurements.

Boot Up Procedure

  1. Before powering on, press the trigger once or twice to ensure the jaws are well closed.
  2. Press the POWER button to switch on the instrument.
  3. The instrument will automatically calibrate. During boot-up, the clamp meter must be in a natural resting state; do not flip the clamp or apply external force to the jaw.
  4. Upon successful boot-up, "OL Ω" will be displayed, and the resistance measurement mode is automatically entered.
  5. If an "Er" symbol appears, it indicates a power-on error, likely due to improperly closed jaws or wires clamped during startup. Ensure the jaws are clear and closed before powering on again.

Alarm Settings

The alarm function alerts you when measured values exceed a set threshold. To configure:

  1. After power on, short press the "AL" key to turn the alarm function on or off.
  2. Long press the "SET" key to enter the alarm value setting mode.
  3. Use the "AL" and "MEM" keys to change the current number size (resistance, current, or voltage alarm values).
  4. Short press the "MODE" key to switch between different alarm modes (e.g., resistance, current, voltage).
  5. Long press the "SET" key again to save the settings and exit.

When the measured resistance exceeds the set alarm value and the alarm is active, the meter will flash the "Alarm Indicator Symbol" symbol and emit an intermittent "beep--beep beep--" sound. The maximum ground resistance alarm setting is 200 Ω.

4. Operating Instructions

Resistance Test

ZIBOO Earth Clamp Tester in use, clamping a wire in an electrical panel.

Figure 5: Tester in use

The basic principle of the clamp-type grounding resistance meter is to measure the loop resistance. The jaw section contains a voltage coil and a current coil. The voltage coil provides an excitation signal, inducing a potential V in the circuit. A current I is generated, and the meter measures V and I to calculate resistance R using the formula R=V/I.

To perform a resistance test:

  1. Ensure the device is booted up correctly and displays "OL Ω".
  2. Withhold the trigger to open the jaws.
  3. Clamp the circuit under test.
  4. Release the trigger to close the jaws around the conductor.
  5. Read the resistance value displayed on the LCD.

If the large middle digit shows "OL Ω", the measured resistance exceeds the upper limit of the meter. If it shows "L0.01Ω", the measured resistance is below the lower limit.

Calibration Check: If you suspect the grounded value is not normal, use a random calibration ring (1Ω and 10Ω are available) to check the clamp meter's accuracy.

Data Lock (HOLD)

To lock and save the current displayed data:

  1. Once the measurement is stable, short press the "HOLD" key. The display will show "HOLD" and the data will be locked.
  2. Short press the "HOLD" key again to exit the lock mode.

Video 1: Demonstration of ZIBOO Earth Clamp Tester functions, including alarm setting, data hold, data access/storage, and practical application.

Data Storage / Review / Delete (MEM)

The instrument can store up to 300 groups of data. To store, review, or delete data:

  1. Storing Data: After a measurement is completed and stable, short press the "HOLD" key. The "MEM" symbol will flash, and the data will be automatically numbered and stored. If the memory is full, the instrument will flash an indicator.
  2. Reviewing Data: Short press the "MEM" key to display the "MR" symbol on the reading interface, indicating data lookup mode. Use the "AL" or "SET" keys to select the desired data by array number (step value of 1).
  3. Exiting Lookup Mode: Short press the "MEM" key again to exit the data lookup mode.
  4. Deleting Stored Data: While in data retrieval mode (after short pressing "MEM"), long press the "MEM" key, then press the "POWER" button to delete the stored data.

Field Application

The ZIBOO Earth Clamp Tester can be used in various grounding system configurations:

Multi-Point Grounding System

For systems like transmission tower grounding, communication systems, or buildings with multiple interconnected grounding bodies via an overhead ground wire, the meter measures the equivalent resistance. The measured resistance (R1) is the target grounding resistance, and R0 is the equivalent resistance of other parallel grounding resistances. Due to the physical separation of grounding points, R0 is often much smaller than R1, allowing R0 to be approximated as 0 for practical engineering purposes. This means the measured resistance is approximately R1.

Limited Point Grounding System

In scenarios where a small number (N, greater than 2) of grounding bodies are linked, such as five towers connected by an overhead ground wire, the assumption of R0=0 may introduce more error. For accurate results in such nonlinear systems, specialized software (Limited-Point Grounding System Solution software) is recommended to solve the N equations for N unknown grounding resistances. Users must measure the same number of testing values as the number of grounding bodies.

Single-Point Grounding System

The EARTH CLAMP TESTER series meters primarily measure loop resistance. For single-point grounding measurements, where a direct loop cannot be formed, two methods can be used:

(1) Two-Point Method:

  1. Near the measured grounding body (RA), find an independent grounding body (RB) with better grounding (e.g., a water pipe or building).
  2. Connect RA and RB with a single testing line.
  3. The meter measures the total series resistance (RT) which is RT = RA + RB + RL, where RL is the resistance of the testing line.
  4. By measuring the resistance of the testing line separately, you can deduce RA + RB. If this value is within acceptable limits, both grounding bodies are qualified.

(2) Three-Point Method:

  1. Near the measured grounding body (RA), find two independent grounding bodies (RB and RC) with better grounding.
  2. Step 1: Link RA and RB with a test line. Use the meter to get the first reading R1. (R1 = RA + RB)
  3. Step 2: Link RB and RC. Use the meter to get the second reading R2. (R2 = RB + RC)
  4. Step 3: Link RC and RA. Use the meter to get the third reading R3. (R3 = RC + RA)
  5. Calculation: From these three readings, you can calculate the individual grounding resistances:
    • RA = (R1 + R3 - R2) / 2
    • RB = R1 - RA
    • RC = R3 - RA

5. Maintenance

Battery Replacement

The instrument uses 4 AA alkaline dry batteries. When the battery voltage drops to 5.2V, the battery symbol (Battery Level Indicator) will be displayed on the LCD. Replace the batteries promptly to ensure continued accuracy of measurements. Always turn off the meter before replacing batteries.

Cleaning and Storage

  • Keep the jaw contact surfaces clean. Do not use corrosive or rough objects to wipe them.
  • Store the instrument in a cool, dry place, away from high temperatures, humidity, condensation, and direct sunlight.
  • The device has an automatic shut-down feature after 5 minutes of inactivity to conserve battery. Before automatic shutdown, the device will flash for 30 seconds. Pressing the power button will delay the shutdown by 5 minutes.

Professional Servicing

Any disassembly, calibration, or maintenance of this instrument must be performed by authorized personnel only.

ZIBOO Earth Clamp Tester stored in its protective carrying case.

Figure 6: Tester in its protective carrying case

6. Troubleshooting

  • "Er" Symbol: Indicates a boot-strap error. This usually happens if the trigger is pressed or the jaw is open when the power is turned on. Ensure the jaws are closed and clear before powering on.
  • Low Battery Symbol (Battery Level Indicator): The battery voltage is below 5.0V. Replace the 4 AA alkaline batteries immediately to maintain measurement accuracy.
  • "OL Ω" Symbol: The measured resistance has exceeded the upper limit of the meter's range.
  • "L0.01Ω" Symbol: The measured resistance is below the lower limit of the meter's range.
  • "OL A" Symbol: The measured current has exceeded the upper limit of the meter's range.
  • Alarm Symbol (Alarm Indicator Symbol) Flashing with "beep--beep beep--" Sound: The measured value is greater than the critical alarm setting. Check your alarm settings and the measurement.
  • "NOISE" Symbol Flashing: Indicates a large interference current in the ground loop being tested. The instrument will also emit a "beep-beep-beep-" sound. The accuracy of the test cannot be guaranteed under these conditions.

7. Specifications

Model Differentiation

Feature UT276A+ UT278A+
Resistance Range 0.01 Ω - 500 Ω 0.01 Ω - 1200 Ω
Current Range 0.00mA - 20A 0.00mA - 20A
Product Type Standard Explosion-proof

General Technical Specifications

Parameter Value
Functions Ground resistance test, Loop resistance test
Ambient Temperature & Humidity 23°C ± 5°C, below 75%RH
Power Supply DC 6V (4 AA alkaline dry batteries)
Measure Method Mutual induction
Resistance Resolution 0.001 Ω
Current Resolution 0.01mA
Jaw Size 55mm × 32mm
Display Ω + A synchronous display, 4-digit LCD, black screen design
Ground Voltage No
Measure Time 2 times/second
Data Storage 300 groups (indicated by "MEM" symbol)
Data Review "MR" symbol indicates data lookup
Overflow display "OL" symbol indicates over-range overflow
Interference Test Automatically identifies interference signals, "NOISE" symbol indicates large interference current
Battery Voltage Real-time display, reminds timely charging
Automatic shut-down Turns off after 5 minutes
Power Consumption 50mA Max
Weight 1180g (including battery)
Working Temperature & Humidity -10°C ~ 40°C; below 80%RH
Storage Temperature & Humidity -20°C ~ 60°C; below 70%RH
Insulation Resistance Above 20MΩ (500V between circuit and case)
Pressure Resistance AC 3700V/rms (between circuit and case)
External Magnetic Field <40A/m
External Electric Field <1V/m
Safety Regulations IEC61010-1 (CAT III 300V, CAT IV 150V, Pollution degree 2); IEC61010-031; IEC61557-1 (earth resistance)

8. User Tips

  • Functionality: Be aware that this device primarily measures ground resistance. While it can detect leakage current, its core functionality is focused on earth resistance. If you require extensive leakage current measurement capabilities, consider specialized equipment.
  • Protective Case: The ZIBOO Earth Clamp Tester comes with a protective carrying case for safe storage and transport.
  • Initial Testing: It is recommended to thoroughly test the device upon receipt to ensure all functions are working as expected.

9. Accessories

Item Quantity
Earth Tester 1 PC
Battery 4 PCS (5th alkaline battery)
Test Loop 1 PC
User's Manual 1 SET
Carrying Case 1 PC

10. Support

For detailed information, including comprehensive diagrams and advanced troubleshooting, please refer to the official user manual PDF: