1. Introduction
The Myoslim15T 6500W RF EMS Body Sculpt Machine utilizes advanced HI-EMT technology, employing high-energy focused electromagnetic waves designed to induce powerful muscle contractions. This process promotes muscle growth and reshapes the internal structure of muscles. Additionally, it claims to trigger significant fat breakdown through intense muscle contractions, causing fatty acids to be released from triglycerides and metabolized by the body over several weeks. This device is intended for professional salon use to provide body sculpting and fat reduction treatments.
The touch-controlled display significantly simplifies operation, offering clear visibility and adjustable brightness. Real-time monitoring and guidance are provided throughout the treatment process, showing remaining time, channels, methods, and attachments used. The EMS Body Sculpt series also offers flexible upgrade options to meet evolving therapeutic needs.
1.1 Indications for Use
- Improvement of abdominal tone, strengthening of abdominal muscles, and development of a fitter abdomen.
- Strengthening, toning, and firming of buttocks, thighs, and calves.
- Improvement of muscle tone and firmness for strengthening muscles in arms.
2. Important Safety Instructions
DANGER
Risk of electric shock. Do not open the device casing. No user-serviceable parts inside. Refer servicing to qualified personnel only.
CAUTION
- This device is for adult use only. Do not use on minors or individuals without a health clearance certificate.
- Do not use if you have metal implants, heart disease (especially with pacemakers), cancer, diabetes, or are pregnant/lactating/menstruating.
- Remove all metal jewelry before operation. Avoid contact with open wounds.
- Do not use on a full stomach. Allow at least 1 hour after meals and refrain from eating/drinking during and 1 hour post-treatment.
- Never replace the treatment handle while the device is powered on. Always turn off the power and calibrate the handle before replacement.
- Keep away from strong electromagnetic sources. Do not daisy-chain multiple devices on a single power outlet.
- This product is intended to be used in dry areas only.
SAVE THESE INSTRUCTIONS
CE Mark:
This symbol indicates "European Conformity", declaring conformity with all relevant EU health, safety, and environmental protection regulations.
3. Setup
3.1 Operating Environment
- Clean and Hygienic Environment: Ensure the operating environment is clean and hygienic to prevent contamination. Regular cleaning and disinfection of equipment and accessories are essential.
- Appropriate Temperature and Humidity: Maintain the operating environment's temperature and humidity within the manufacturer's specified range to avoid impacting device performance.
- Adequate Lighting: Provide sufficient lighting for clear operation, especially for precise beauty procedures.
- Good Ventilation: Ensure adequate air circulation to prevent accumulation of harmful gases.
- Avoidance of Interference Sources: Keep the equipment away from electromagnetic fields or other devices that could cause interference.
3.2 Device Start-Up
- Connect one end of the power cord to the machine and the other end into the main socket. Do not connect multiple machines to the same socket.
- Move the manual switch on the rear panel to the "o" position.
- Press the power button on the front panel.
3.3 Handle Exchange
The handle must be calibrated every time it is installed or replaced. When replacing the handle, you must turn off the power of the machine. The handle cannot be replaced when the machine is powered on.

4. Operating the Device
4.1 Preparing for Therapy
- Communicate with patients to understand their needs and explain the procedure.
- Ensure the patient is in good health and has no contraindications.
- Advise the patient to stay hydrated.
- Ensure the patient wears appropriate clothing that allows access to the treatment area.
- Advise the patient to avoid eating large quantities before the treatment.
- Familiarize yourself and the patient with the treatment process.
- Advise the patient to avoid drinking alcohol before treatment.
4.2 Placing the Handle
The treatment handles can be placed on various body parts. Ensure the handle is securely attached to the body using the provided belt.





Please operate according to the controller model you purchased and follow the manufacturer's official documentation to ensure the best experience. If you have any questions, please contact customer service for support.
4.3 Navigation Controls and Unit Settings
The device features a touch-screen interface for setting up and monitoring therapy. Pre-treatment regimens can be modified via on-screen controls.
4.3.1 Setting the Therapy by Controls
The treatment interface displays the following items and functions:
- Navigation controls
- Channels selector
- Stop therapy
- Start/pause therapy
- Setting up therapy parameters
4.3.2 Unit Settings (Mode 1)
The operation interface for Mode 1 is shown below. Before operation, take out the handle from the handle frame and use the belt to fix the handle to the desired operating part on the customer. The frequency of Mode 1 is 5-60Hz.
Enter the solution 1 interface to adjust the handle operation time according to customer needs, then click the start/stop button to proceed.
Note:
- Adjust the corresponding intensity according to different customer needs to start the operation. If the customer requests to pause, you can click start/pause.
- The work button suspends the work, and you can directly click the back button to return after the work is completed.
4.3.3 Unit Settings (Mode 2)
The operation interface for Mode 2 is shown below. Similar to Mode 1, fix the handle to the operating part. The frequency of Mode 2 is 30-100Hz.
Mode 2 allows for frequency adjustment across three different periods, offering free adjustment. Adjust the corresponding intensity/stage frequency/time according to customer needs, click the start/pause button to begin. If pausing is needed, click start/pause. After completion, click the return key.
4.3.4 General Settings
The device allows for various general settings adjustments:
- Language Settings: Adjust the display language.
- Sound Adjustment: Control the device's sound output.
- Brightness Adjustment: Modify screen brightness for optimal visibility.
- Set Date/Time: Configure the current date and time.
- System Adjustment: Access system-level settings.
- Technician Access: A password-protected section for technical adjustments.
5. Maintenance
5.1 Daily Inspection
Regularly inspect instruments and components. Check for a reliable power socket connection to the ground. Inspect the power cord for damage before use; discontinue use if damaged.
5.2 Cleaning
5.2.1 Cleaning the Surface of the Device
The machine's logo is silk-screen printed, while other parts are spray-painted. Use cotton paper towels or ordinary toilet paper moistened with water to wipe clean. Do not use alcohol-based detergents.
5.2.2 Cleaning of Accessories that Come into Contact with Patient
Parts that contact the skin should be cleaned and disinfected with 75% alcohol before and after each use. Do not soak directly in alcohol-based detergent; wipe with pure cotton material.
5.3 Replacing the Fuse
The fuse is located below the handle junction on the back of the machine. An additional spare fuse is included. To prevent electric shock, always turn off the power and unplug the power plug before replacing the fuse.
5.4 Transport and Storage
5.4.1 Storage
Store the instrument in its original packaging box in a well-ventilated room. The packaging box should be padded. The ambient temperature should be -10°C to 40°C, and relative humidity should not exceed 93%. Do not store near harmful gases, flammable, explosive substances, or corrosive gases.
5.4.2 Transportation
The instrument, in its packaging, is suitable for road, railway, air, and water transportation. During loading, unloading, and transportation, prevent severe vibration and shock. Ensure it is not affected by moisture and not mixed with flammable or corrosive substances. Specific requirements are detailed in the order contract.
6. Troubleshooting
6.1 Common Faults and Troubleshooting Methods
| Fault Phenomenon | Causes | Exclusions and Workarounds |
|---|---|---|
| Device not working |
|
|
| Handle not working | Check the handle for looseness | Reseat the handle |
7. Technical Specifications
| Attribute | Value |
|---|---|
| Brand Name | MYOSLIM |
| Item Type | Massage & Relaxation |
| Technology | EMS/HIEMT+RF 2 in 1 Slimming |
| Functions | Skin tight, butt lift, muscle build, muscle stimulation, Body Slimming, Muscle building, Skin Tightening, Cellulite Reduction, Body Muscle Sculpting, etc. |
| Treatment Area | Body (arm, waist, abdomen, buttocks, upper legs) |
| Power Supply | Alternating Current |
| Voltage | 50-1000V AC |
| Frequency | 200hz |
| Material | ABS |
| Certification | CE |
| Origin | Mainland China |
| Hand Instrument | Yes |
| Electronic | Yes |
| High-concerned chemical | None |
7.1 Electromagnetic Compatibility (EMC) Test
The device has undergone Radiated Electromagnetic Field Immunity Test to ensure compliance with safety standards.
| Parameter | Value |
|---|---|
| Port | Enclosure |
| Basic Standard | IEC 61000-4-3:2020 |
| Test Level | 30V/m with 80% AM. 1kHz Modulation. |
| Standard Require | A |
| Tester | Sam Liu |
| Temperature | 25°C |
| Humidity | 54% |
7.2 Test Equipment
| Description | Manufacturer | Model | Identifier | Cal. Date | Cal. Due |
|---|---|---|---|---|---|
| Signal Generator | IFA | 2023B | N/A | 15/03/2022 | 14/03/2023 |
| Power Amplifier | AR | 150W1000 | N/A | 15/03/2022 | 14/03/2023 |
| Power Antenna | AR | 25S1G4A | N/A | 15/03/2022 | 14/03/2023 |
7.3 Block Diagram of Test Setup

8. Warranty and Support
8.1 Term of Use
- The warranty period for the main instrument is 1 year (excluding man-made damage).
- The warranty period for accessories is 3 months (excluding man-made damage).
- Continued use beyond the service period requires approval from an engineer, who will conduct periodic performance tests.
- Instruments and accessories beyond their useful life should be disposed of according to local regulations.
8.2 Warranty Card Information
The main machine can be replaced with a new one within 7 days from the date of purchase (with receipt and intact appearance). The main machine is guaranteed for one year (excluding fragile parts, non-warranty scope, or expired period). Maintenance and updates are available, with appropriate fees charged. If the instrument needs repair, the buyer bears return shipping costs within six months of purchase, and the seller bears shipping-in costs. After six months, the buyer bears all shipping costs.
Free warranty service is NOT provided for failures caused by personal reasons, including:
- Faults caused by unauthorized disassembly and modification.
- Faults caused by accidental fall.
- Failure caused by lack of reasonable maintenance.
- Failure to operate according to the correct instructions in this manual.
8.3 After-sales Service Provided
Online support, video technical support, maintenance, and repair services are available. To repair the machine, you must contact the company's business department or dealer. Any unpacking without consent will void the warranty.
9. User Tips
While no specific user reviews or Q&A were provided, general tips for using professional body sculpting equipment include:
- Always perform a patch test on a small area of skin before full treatment to check for adverse reactions.
- Ensure proper hydration before and after sessions to support metabolic processes.
- Maintain a balanced diet and regular exercise routine to maximize and sustain treatment results.
- Follow the recommended treatment schedule and intensity levels for optimal safety and effectiveness.
- Regularly clean and inspect the handles and device for any wear or damage to ensure consistent performance and hygiene.