1. Safety Tips
- This product features a unique anti-interference design and filtering algorithm. Ensure proper grounding and separation from the AC ground wire.
- Use in an environment that complies with the technical specifications of this product.
- Avoid direct sunlight.
- Installation, wiring, and maintenance should only be performed by personnel with professional knowledge and safety training.
- It is strictly forbidden to open the chassis with electricity.
2. Product Overview
The Erilles S-Type Load Cell and Indicator system is designed for high-precision measurement of tension and compression forces in various industrial applications, such as crane hoppers. It offers multiple measuring ranges, good stability, and high precision, with customizable options.

Figure 1: S-Type Load Cell and Indicator highlighting key features.

Figure 2: Different S-type load cell styles and their corresponding weighing ranges.
Video 1: Product overview, unboxing, and demonstration of the indicator's features and robust design.
3. Specifications
3.1. Technical Parameters (Load Cell)
| Parameter | Value |
|---|---|
| Rated load | 5kg~5T |
| Output sensitivity | 2.0±0.05 mV/V |
| Non-linearity | ±0.03~±0.05%F.S |
| Lag error | ±0.03%F.S |
| Repeatability | ±0.03%F.S |
| Creep | ±0.03%F.S |
| Zero temperature coefficient | ±0.05%F.S/10℃ |
| Sensitivity temperature coefficient | ±0.05%F.S/10℃ |
| Input resistance | 380±30Ω |
| Output resistance | 350+5Ω |
| Operating temperature | -20℃~+70℃ |
| Safe overload | 150 %F.S |
| Insulation resistance | ≥2000MΩ/100VDC |
| Incentive voltage | 5V~12V |
| Protection class | IP65, IP66 |
3.2. Indicator Parameters
| Specification | Details |
|---|---|
| Display | 5-digit LED + TFT LCD, dual display (gross, net, peak, unit) |
| Power Supply | DC10-28V / AC 85-265V (100-240V, 50/60HZ) |
| Load Cell Capacity | Up to 8 x 350Ω load cells |
| Signal Range | 0.5~3.0 mV/V |
| Excitation Voltage | 5V |
| Sampling Rate | Adjustable 10 / 40 / 640 / 1280 Hz |
| Internal Resolution | 24-bit Σ-Δ ADC, internal resolution 1/1,000,000 |
| Display Resolution | 1/100,000 |
| Nonlinearity | 0.005% F.S |
| Max Power Consumption | 5W |
| Operating Temperature | -25℃ to +45℃ |
| Relative Humidity | <85% RH |
| Weight | Approx. 300g |
| Dimensions | 96 × 48 × 110 mm |
| Panel Cutout | 92 × 46 mm |
3.3. Model Options
| Model | Description |
|---|---|
| Base Model | Dual five-digit display, 3 sets of relays, and one set of switching inputs |
| Analog Output Version | Base Model + Analog Output (0-5V, 0-10V, 0-5-10V, 0-20mA, 4-20mA, 4-12-20mA) |
| RS485 Version | Base Model + RS485 output |
| Full-function Version | Base Model + Analog Output + RS485 Output |
4. Setup and Installation
4.1. Load Cell Installation
The S-type load cell can be used for both tension and compression measurements. Ensure proper mounting to avoid side loads or twisting, which can affect accuracy.

Figure 3: Load cell installation styles and compatible accessories.
4.2. Indicator Panel Cutout and Dimensions
The indicator requires a panel cutout of 92mm x 46mm for flush mounting.

Figure 4: Indicator dimensions and panel cutout specifications.
4.3. Wiring Diagram
Ensure all connections are secure and follow the diagram carefully. Incorrect wiring can damage the device.

Figure 5: Indicator wiring diagram for power, sensor, and outputs.

Figure 6: Load cell wiring diagram. Note: Green and White lines may need to be swapped for tension measurement.
5. Operating Instructions
5.1. Display and Buttons
The indicator features a dual five-digit display for real-time gross, net, and peak monitoring. The front panel includes buttons for navigation and function control.

Figure 7: Indicator front panel with display and control buttons.
- Fn (Function) Button: Enter the menu or go back to the previous level. Long press to enter the decimal point selection interface.
- Left Arrow Button: Enter the calibration interface, move the cursor. Move the main menu interface and parameter selection screen down. Long press to enter the zeroing option.
- Up Arrow / Tare Button: Change the menu, increase the value. Long press to enter the peeling option.
- ENT (Enter) Button: The menu interface is the confirmation menu. The display interface is the second window display (unit, total weight, peak switching). Long press to enter the clear peak interface.
- The second window of the PEAK light is always on, and the peak light flashes to wait for the peak to be triggered.
- The lights of OUT1/OUT2/OUT3 are triggered by relays 1, 2, and 3, respectively, and are turned off by untriggered states.
5.2. Software Interface and Settings
The indicator's software allows for detailed configuration and data monitoring. Access settings via the menu button.

Figure 8: Software interface - Connection Settings Page.

Figure 9: Multi-language software and key features of the indicator.

Figure 10: Software interface - Data Display Page.
5.3. Calibration
The indicator supports various calibration methods:
- Zero Calibration: Set unloaded sensor output to zero.
- Multi-point Calibration: Enable calibration at multiple points to improve linearity.
- Digital Calibration: Input sensor sensitivity and capacity directly (no weights required).
- Measuring Range: Enter sensor full capacity.
- Sensitivity: Input sensor sensitivity (mV/V).
- Excitation Voltage: Default = 5.000 (5.000V).
- Range Coefficient: Scaling coefficient, default 1.0000 = 1.0000.
Refer to the detailed parameter tables (C2.CAL, C3.rEL, C4.AdV, C5.COM, C6.aNa) in the full PDF manual for advanced settings and calibration procedures.
6. Maintenance
- Keep the device clean and free from dust and moisture.
- Regularly check wiring connections for tightness and signs of wear.
- Avoid exposing the load cell or indicator to excessive shock or vibration.
- If the device is not in use for an extended period, store it in a dry, stable environment.
- Do not attempt to repair the device yourself. Contact qualified service personnel for assistance.
7. Troubleshooting
This section lists common issues and their descriptions. For detailed solutions, refer to the full PDF manual.
| Fault Code/Description | Meaning |
|---|---|
| No effective peaks were captured | The system did not detect valid peak values. |
| Err.ro | The fault EEPROM failed to read and write. Press the OK key to release it. |
| Err.Ad / AL.Ad | ADC failure. The sampling signal is too large, or the ADC reads maximum/minimum value (automatically canceled). |
| AL.OL | Sensor overload or exceed maximum range. |
| AL.AZR | The sensor fails to power on and clear, exceeding the set power-on zero value. |
| AL.PEL | Manual peeling fails, does not meet the peeling conditions. (1) cannot be negative, (2) the relay cannot be suctioned, and (3) it needs to be in a stable state. |
| AL.MZr | Manual zeroing fails and does not meet the zeroing conditions. |
| Er.doM | If the value set by the user is too small, it will be automatically set to the minimum value that can be set. |
| Er.up | If the value set by the user is too large, it will be automatically set to the maximum value that can be set. |
| Er.iNV | The user input is invalid. |
| Er.pwE | The user entered the wrong password. |
| FA.roM | The configured parameter write failed, but the parameter has taken effect. |
| FA.AdC | ADC setup failed. |
| Er.buy | The system is busy and cannot perform operations, and generally the data is not collected and stabilized, and the zeroing or calibration operation is performed. |
8. User Tips
- The sensor and indicator offer fast response and stable readings, making them reliable for real-time monitoring.
- Installation is generally straightforward and hassle-free.
- For PLC integration, the provided register map is highly effective for instant setup and reliable data output.
- The full-function model is particularly versatile, combining RS485, analog output, and relay control in a single unit, ideal for complex automation lines.
- The indicator's display is bright and easy to read, even in varying light conditions.
9. Frequently Asked Questions (FAQ)
- Q: What is the operating temperature range for the indicator?
- A: The indicator operates reliably from -25℃ to +45℃.
- Q: Can the load cell be used for both tension and compression?
- A: Yes, the S-Type load cell is designed for measuring both tension and compression loads.
- Q: Is the device waterproof?
- A: The load cell and indicator have a protection class of IP65/IP66, making them resistant to dust and water ingress.
- Q: How do I perform a zero calibration?
- A: To perform a zero calibration, navigate to the calibration menu using the Fn button and select the zero calibration option. Ensure the sensor is unloaded before proceeding.
- Q: What are the available output options for the indicator?
- A: Depending on the model, output options include 3 relays, switching inputs, analog output (0-5V, 0-10V, 0-5-10V, 0-20mA, 4-20mA, 4-12-20mA), and RS485 communication.
10. Warranty and Support
For detailed warranty information, technical support, or service inquiries, please refer to the official user manual PDF or contact the manufacturer directly.
Official User Manual (PDF): Download PDF Manual