Erilles S-Type Load Cell & Indicator

Contents

S-Type Load Cell & Indicator Instruction Manual

Model: S-Type Load Cell & Indicator Series

Brand: Erilles

1. Safety Tips

  • This product features a unique anti-interference design and filtering algorithm. Ensure proper grounding and separation from the AC ground wire.
  • Use in an environment that complies with the technical specifications of this product.
  • Avoid direct sunlight.
  • Installation, wiring, and maintenance should only be performed by personnel with professional knowledge and safety training.
  • It is strictly forbidden to open the chassis with electricity.

2. Product Overview

The Erilles S-Type Load Cell and Indicator system is designed for high-precision measurement of tension and compression forces in various industrial applications, such as crane hoppers. It offers multiple measuring ranges, good stability, and high precision, with customizable options.

S-Type Load Cell and Indicator with features listed: Multiple measuring ranges, Good stability, High precision, Customizable.

Figure 1: S-Type Load Cell and Indicator highlighting key features.

Diagram showing two S-type load cell styles with different weighing ranges: 5KG-100KG and 200KG-5T.

Figure 2: Different S-type load cell styles and their corresponding weighing ranges.

Video 1: Product overview, unboxing, and demonstration of the indicator's features and robust design.

3. Specifications

3.1. Technical Parameters (Load Cell)

ParameterValue
Rated load5kg~5T
Output sensitivity2.0±0.05 mV/V
Non-linearity±0.03~±0.05%F.S
Lag error±0.03%F.S
Repeatability±0.03%F.S
Creep±0.03%F.S
Zero temperature coefficient±0.05%F.S/10℃
Sensitivity temperature coefficient±0.05%F.S/10℃
Input resistance380±30Ω
Output resistance350+5Ω
Operating temperature-20℃~+70℃
Safe overload150 %F.S
Insulation resistance≥2000MΩ/100VDC
Incentive voltage5V~12V
Protection classIP65, IP66

3.2. Indicator Parameters

SpecificationDetails
Display5-digit LED + TFT LCD, dual display (gross, net, peak, unit)
Power SupplyDC10-28V / AC 85-265V (100-240V, 50/60HZ)
Load Cell CapacityUp to 8 x 350Ω load cells
Signal Range0.5~3.0 mV/V
Excitation Voltage5V
Sampling RateAdjustable 10 / 40 / 640 / 1280 Hz
Internal Resolution24-bit Σ-Δ ADC, internal resolution 1/1,000,000
Display Resolution1/100,000
Nonlinearity0.005% F.S
Max Power Consumption5W
Operating Temperature-25℃ to +45℃
Relative Humidity<85% RH
WeightApprox. 300g
Dimensions96 × 48 × 110 mm
Panel Cutout92 × 46 mm

3.3. Model Options

ModelDescription
Base ModelDual five-digit display, 3 sets of relays, and one set of switching inputs
Analog Output VersionBase Model + Analog Output (0-5V, 0-10V, 0-5-10V, 0-20mA, 4-20mA, 4-12-20mA)
RS485 VersionBase Model + RS485 output
Full-function VersionBase Model + Analog Output + RS485 Output

4. Setup and Installation

4.1. Load Cell Installation

The S-type load cell can be used for both tension and compression measurements. Ensure proper mounting to avoid side loads or twisting, which can affect accuracy.

Diagram showing S-type load cell installation style with arrows indicating tension and compression, and examples of lifting ring, joint bearing, and hook accessories.

Figure 3: Load cell installation styles and compatible accessories.

4.2. Indicator Panel Cutout and Dimensions

The indicator requires a panel cutout of 92mm x 46mm for flush mounting.

Diagram showing the dimensions of the indicator (96mm x 48mm x 110mm) and the required panel cutout (92mm x 46mm).

Figure 4: Indicator dimensions and panel cutout specifications.

4.3. Wiring Diagram

Ensure all connections are secure and follow the diagram carefully. Incorrect wiring can damage the device.

Detailed wiring diagram for the indicator, showing connections for AC100-240V power, DC10-28V power, sensor inputs (AO+, AO-, S+, S-, E+, E-), digital inputs (DI+, DI-), RS485 (B-, A+), and relay outputs (OUT1, OUT2, OUT3).

Figure 5: Indicator wiring diagram for power, sensor, and outputs.

Load cell wiring diagram showing Red (Motivation+), Green (Signal+), Black (Motivation-), and White (Signal-). Note: Green and White lines may need to be swapped for tension measurement.

Figure 6: Load cell wiring diagram. Note: Green and White lines may need to be swapped for tension measurement.

5. Operating Instructions

5.1. Display and Buttons

The indicator features a dual five-digit display for real-time gross, net, and peak monitoring. The front panel includes buttons for navigation and function control.

Front view of the indicator showing the dual digital displays and control buttons: Fn (Function), left arrow, up arrow/Tare, and ENT (Enter/Peak).

Figure 7: Indicator front panel with display and control buttons.

  • Fn (Function) Button: Enter the menu or go back to the previous level. Long press to enter the decimal point selection interface.
  • Left Arrow Button: Enter the calibration interface, move the cursor. Move the main menu interface and parameter selection screen down. Long press to enter the zeroing option.
  • Up Arrow / Tare Button: Change the menu, increase the value. Long press to enter the peeling option.
  • ENT (Enter) Button: The menu interface is the confirmation menu. The display interface is the second window display (unit, total weight, peak switching). Long press to enter the clear peak interface.
  • The second window of the PEAK light is always on, and the peak light flashes to wait for the peak to be triggered.
  • The lights of OUT1/OUT2/OUT3 are triggered by relays 1, 2, and 3, respectively, and are turned off by untriggered states.

5.2. Software Interface and Settings

The indicator's software allows for detailed configuration and data monitoring. Access settings via the menu button.

Screenshot of the indicator's software interface showing Connection Settings Page with options like Connection address, Baud rate, Data bits, Parity, Stop bits, Communication port, Disconnect, Device Configuration, and Data Graph.

Figure 8: Software interface - Connection Settings Page.

Image showing multi-language software support (English, Spanish, German, French, Russian, Japanese, Korean, Chinese) and key features like High Precision Measurement, Dual LED Display, Multiple Output Options, Flexible Functions, and Industrial-Grade Design.

Figure 9: Multi-language software and key features of the indicator.

Screenshot of the software interface showing Data Display Page with functions like Current Weight, Peak Value, Valley Value, Remove Peel (Do Not Save), Remove Peel (Save), Clear Peel (Save), Set Zero (Do Not Save), Read once, Press and hold to read continuously, Weight Log, Import, Clear, Export, Export Excel, Delete.

Figure 10: Software interface - Data Display Page.

5.3. Calibration

The indicator supports various calibration methods:

  • Zero Calibration: Set unloaded sensor output to zero.
  • Multi-point Calibration: Enable calibration at multiple points to improve linearity.
  • Digital Calibration: Input sensor sensitivity and capacity directly (no weights required).
  • Measuring Range: Enter sensor full capacity.
  • Sensitivity: Input sensor sensitivity (mV/V).
  • Excitation Voltage: Default = 5.000 (5.000V).
  • Range Coefficient: Scaling coefficient, default 1.0000 = 1.0000.

Refer to the detailed parameter tables (C2.CAL, C3.rEL, C4.AdV, C5.COM, C6.aNa) in the full PDF manual for advanced settings and calibration procedures.

6. Maintenance

  • Keep the device clean and free from dust and moisture.
  • Regularly check wiring connections for tightness and signs of wear.
  • Avoid exposing the load cell or indicator to excessive shock or vibration.
  • If the device is not in use for an extended period, store it in a dry, stable environment.
  • Do not attempt to repair the device yourself. Contact qualified service personnel for assistance.

7. Troubleshooting

This section lists common issues and their descriptions. For detailed solutions, refer to the full PDF manual.

Fault Code/DescriptionMeaning
No effective peaks were capturedThe system did not detect valid peak values.
Err.roThe fault EEPROM failed to read and write. Press the OK key to release it.
Err.Ad / AL.AdADC failure. The sampling signal is too large, or the ADC reads maximum/minimum value (automatically canceled).
AL.OLSensor overload or exceed maximum range.
AL.AZRThe sensor fails to power on and clear, exceeding the set power-on zero value.
AL.PELManual peeling fails, does not meet the peeling conditions. (1) cannot be negative, (2) the relay cannot be suctioned, and (3) it needs to be in a stable state.
AL.MZrManual zeroing fails and does not meet the zeroing conditions.
Er.doMIf the value set by the user is too small, it will be automatically set to the minimum value that can be set.
Er.upIf the value set by the user is too large, it will be automatically set to the maximum value that can be set.
Er.iNVThe user input is invalid.
Er.pwEThe user entered the wrong password.
FA.roMThe configured parameter write failed, but the parameter has taken effect.
FA.AdCADC setup failed.
Er.buyThe system is busy and cannot perform operations, and generally the data is not collected and stabilized, and the zeroing or calibration operation is performed.

8. User Tips

  • The sensor and indicator offer fast response and stable readings, making them reliable for real-time monitoring.
  • Installation is generally straightforward and hassle-free.
  • For PLC integration, the provided register map is highly effective for instant setup and reliable data output.
  • The full-function model is particularly versatile, combining RS485, analog output, and relay control in a single unit, ideal for complex automation lines.
  • The indicator's display is bright and easy to read, even in varying light conditions.

9. Frequently Asked Questions (FAQ)

Q: What is the operating temperature range for the indicator?
A: The indicator operates reliably from -25℃ to +45℃.
Q: Can the load cell be used for both tension and compression?
A: Yes, the S-Type load cell is designed for measuring both tension and compression loads.
Q: Is the device waterproof?
A: The load cell and indicator have a protection class of IP65/IP66, making them resistant to dust and water ingress.
Q: How do I perform a zero calibration?
A: To perform a zero calibration, navigate to the calibration menu using the Fn button and select the zero calibration option. Ensure the sensor is unloaded before proceeding.
Q: What are the available output options for the indicator?
A: Depending on the model, output options include 3 relays, switching inputs, analog output (0-5V, 0-10V, 0-5-10V, 0-20mA, 4-20mA, 4-12-20mA), and RS485 communication.

10. Warranty and Support

For detailed warranty information, technical support, or service inquiries, please refer to the official user manual PDF or contact the manufacturer directly.

Official User Manual (PDF): Download PDF Manual

The manufacturer, Knoweigh Tech, specializes in smart weighing solutions and offers comprehensive after-sales support, including phone and video guidance.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.