SWEVY SW-762 Conductivity Meter User Manual

Model: SW-762 | Version: 1.0.6

1. Introduction

The SWEVY SW-762 is an instrument designed for detecting the water quality of solutions. It accurately measures solution conductivity and temperature. This product is stable, reliable, easy to carry, and widely used in various fields including industry, laboratories, medicine, food, scientific research, and environmental protection.

2. Product Overview

2.1 Appearance and Components

SWEVY SW-762 Conductivity Meter
Figure 1: SWEVY SW-762 Conductivity Meter
Labeled diagram of the SW-762 Conductivity Meter showing its parts and dimensions
Figure 2: Labeled Diagram of SW-762

The meter features a dual display for conductivity and temperature values, making it easy to read at a glance. It utilizes a titanium electrode probe for rapid detection and high reaction sensitivity, ensuring accurate measurements. The probe also features a detachable design to extend its service life. The device has a waterproof and dustproof body, suitable for various environments.

2.2 LCD Display

Close-up of the LCD display showing conductivity and temperature values
Figure 3: LCD Display showing conductivity and temperature

The LCD provides the following indicators:

  • Battery level prompt
  • Calibration tips
  • Value lock prompt (HOLD)
  • Measurement value display area (µS, mS)
  • Temperature unit (°C, °F)
  • Temperature value display area

Product Demonstration Video

Video 1: Product demonstration showing usage, detachable probe, dual display, temperature compensation, calibration, and unit switching.

3. Battery Installation Instructions

  1. Insert the battery according to the positive and negative instructions inside the battery compartment.
  2. After installing the battery, please tighten the battery cover.
  3. When the instrument is not used for a long time, be sure to remove the battery and place it properly.

4. Turn On/Off

Short press the "Power key" to turn on the instrument. Long press the "Power key" to turn off the instrument.

5. Measurement Steps

  1. Install batteries and turn on the instrument.
  2. Before measurement, please clean the probe with distilled water and wipe it clean with medical cotton.
  3. Immerse the tip of the probe into the solution to be tested. Note that the position of the solution cannot exceed the water level line (The tip of the probe is 38mm from the water level line). After immersion, the value of the solution can be measured.

Notes for Measurement:

  • Please do not immerse the whole machine in the liquid; mainly immerse the front probe (electrode).
  • If there are bubbles in the water during measurement, please pick up the instrument repeatedly 3-4 times.
  • After the measurement, please clean the probe and cover the probe cover.
  • If the instrument is not used or the probe has been replaced for a long time, the instrument should be calibrated before the measurement, so as not to affect the measurement result.
Titanium electrode probe for accurate measurements
Figure 4: Titanium Electrode Probe for accurate measurements
Detachable probe design for extended service life
Figure 5: Detachable Probe Design
Waterproof body design for dustproof and splashproof protection
Figure 6: Waterproof Body Design

6. Calibration

Instructions for Calibration:

  • The instrument includes 3 calibration points: 0µS/cm (distilled water), 1413µS/cm, and 12.88mS/cm.
  • For calibration, use a solution with known conductivity values. If the conductivity value of the solution is different from the default value, the calibration value can be adjusted by the "Up key" and "Down key".
  • Under a certain calibration point interface, long press the "Power key" to skip the calibration point and enter the next calibration point.
  • The measurement temperature should be as constant as possible, and there should be no bubbles on the probe.
  • The probe should be suspended in the solution to be stable, and not close to the wall and the bottom of the container.

Calibration Steps:

  1. Long press the "Up key" in the measured state to enter calibration mode. The screen will prompt "CAL", the conductivity unit will be "µS", the measurement value display area will show the current calibration point and measurement value, and the temperature display area will show the current temperature and "CAL".
  2. Before calibration, clean the probe with distilled water and gently wipe with medical cotton. After cleaning, immerse the probe into the solution of the corresponding concentration according to the value indicated by the screen calibration point. Press the "Power key" to stabilize the reading, and the instrument will jump to the next calibration point or the end of the calibration.

Note:

  • If the instrument calibration is successful, "PASS" is displayed.
  • If the instrument calibration fails, "Err" is displayed, indicating that the calibration point is abnormal and needs to be recalibrated.

7. Coefficient Regulation (Temperature Compensation)

In the measurement state, simultaneously press the "Up" and "Down" keys to enter the temperature compensation coefficient adjustment mode. The measurement value display area shows "tC0", and the temperature display area shows the temperature supplement value (default value is 2.1%/°C). Press the "Power key" again in the temperature supplement coefficient adjustment interface to automatically save the adjusted coefficient value and return to the main measurement interface.

Note:

  • In the coefficient adjustment state, short pressing "Up key" or "Down key" can increase or decrease the value.
  • Long pressing can quickly increase or decrease the value.

8. Restore Factory Settings

In the shutdown state, long press the "Power key" and "Down key" at the same time. The instrument will automatically restore the factory calibration parameters, and then enter the normal operating state. "CLr" is displayed if the restoration is successful, and "Err" is displayed if it failed.

9. Backlight

In the measurement state, short press "Down key" to turn the screen backlight on or off.

10. Unit Conversion

Conductivity Unit:

When the concentration of the measurement solution exceeds 1999µS, the unit automatically switches to mS.

Temperature Unit:

Under the measured state, long press "Down key" to switch between °C and °F.

11. Data Lock

During measurement, short press "Up key". The screen shows "HOLD", and the instrument locks the current measurement value. To cancel, just press "Up key" again.

12. Automatic Shutdown

Each time the meter boots, the automatic shutdown function is enabled by default. In the boot state, simultaneously press "Power key" and "Down key". The screen will prompt "APO On" to enable the automatic shutdown function, and "APO OFF" to close it. After about 5 minutes without any key operation, the instrument will automatically shut down.

13. Technical Parameters

Parameter Value
Type SW-762
Measuring Range Conductance: 0-1999µS, 2.00-19.99mS
Temperature: 0-60°C
Resolution Ratio Conductance: 1µS (0-1999µS), 0.01mS (2.00-19.99mS)
Temperature: 0.1°C
Accuracy Conductance: ±2%F.S
Temperature: ±1.0°C
Operational Environment 0-60°C; <85% RH (not condensed)
Storage Environment 0-60°C; <85% RH (not condensed)
Unit Conductance: µS, mS; Temperature: °C, °F
Power Source 2 * 1.5V AAA alkaline battery
Protection Level IP66
Dimensions 180 * 44 * 44mm
Material ABS
Weight About 92g (excluding battery)
Warranty Period 12 months

14. Troubleshooting

  • No Display / Power Issues: Ensure batteries are correctly installed with the correct polarity and are not depleted. Replace batteries if necessary.
  • Inaccurate Readings:
    • Clean the probe thoroughly with distilled water and wipe with medical cotton before each use.
    • Perform calibration as per Section 6. Calibration.
    • Ensure the probe is fully immersed in the solution, but not beyond the water level line, and is not touching the container walls or bottom.
    • Check for air bubbles on the probe during measurement; gently shake the meter to dislodge them.
    • Verify the temperature compensation coefficient is correctly set (Section 7).
  • "Err" Displayed During Calibration: This indicates a calibration failure. Re-clean the probe, ensure the calibration solution is correct, and repeat the calibration process.
  • Slow Response: Ensure the probe is clean and free from residue.

15. User Tips

  • Always rinse the probe with distilled water before and after each use to maintain accuracy and prolong its lifespan.
  • Store the meter with the probe cover on to protect the electrode.
  • For best results, allow the meter to stabilize in the solution for a few seconds before taking a reading.
  • Keep calibration solutions fresh and at a consistent temperature during calibration.
  • The detachable probe design allows for easier cleaning and potential replacement, contributing to the meter's longevity.

16. Warranty and Support

This product comes with a 12-month warranty period from the date of purchase.

For technical support or warranty claims, please contact:

Guangzhou Suwei Electronic Technology Co., Ltd
Brand: Suwei SWEVY
Contact Number: 020-38104332
Website: www.sw1718.com
Address: Room 1314, Building H6, Luogang Aoyuan Square, No.1936, Kaichuang Avenue, Huangpu District, Guangzhou

A user manual in PDF format is also available for download: User manual (PDF).

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.