1. Introduction
This handheld digital anemometer is a high-precision air velocity measuring instrument designed for various applications including HVAC inspection, ventilation testing, environmental monitoring, and industrial airflow diagnostics. Depending on the model, it supports measurements of wind speed, air temperature, air volume, and humidity. Advanced models offer features such as data logging, audible and visual alarms, and USB communication for PC data analysis. Both integrated and split probe designs are available to suit different measurement scenarios.

2. Safety Information
Precautions
- Do not use the instrument in explosive or flammable environments.
- Avoid exposing the sensor to strong impacts or high-speed debris.
- Do not immerse the device in water or use it in heavy condensation.
- Ensure correct battery polarity during installation.
- Remove batteries if the instrument will not be used for an extended period.
- Do not exceed the specified measurement range to prevent sensor damage.
Warning Information (EU Market Reference)
WARNING!
- This product contains rotating sensor components. Avoid contact during operation.
- Use only the specified battery types and power supply.
- Do not disassemble or modify the instrument.
- Keep out of reach of children.
- Electronic waste disposal must follow the WEEE Directive and local EU regulations.
Note: The above warning information is for reference only. Final warning language must comply with the official language requirements of the destination EU country.
3. Product Overview
The anemometer features a robust design with an 8-blade fan for accurate and responsive measurements. The display is a high-definition LCD, providing clear readings.
Key Components and Controls


Model TA640A


Model TA641B

Models TA642A/B

Model TA642C (Split Type)

4. Specifications
The following tables detail the technical parameters for each model:
General Specifications
| Specification | TA640A | TA641A | TA641B | TA642A | TA642B | TA642C |
|---|---|---|---|---|---|---|
| Wind Speed Measurement Range | 0.1-30.0 m/s | 0.06-30.0 m/s | 0.06-45.0 m/s | |||
| Wind Speed Accuracy | ±(2% + 0.2) | ±0.2 + (±2% of reading) | ±(2% + 0.2) | |||
| Wind Speed Resolution | 0.1 m/s | 0.01 / 0.1 / 1 | ||||
| Air Temperature Range | 0-45℃ | 0-45℃ (32.0-113.0℉) | ||||
| Air Temperature Accuracy | ±2℃ | ±2℃ (±3.6℉) | ||||
| Air Temperature Resolution | — | 0.1 / 1 | ||||
| Humidity Measurement Range | — | 0~99% RH | — | |||
| Humidity Accuracy | — | ±8% RH | — | |||
| Wind Speed Display | 4-digit display, up to 9999 | |||||
| Sensor Type | Infrared laser beam cutting method | |||||
| Sampling Rate | 2 times/second | |||||
| Repeatability | — | ±2% | ||||
| Data Storage | — | 1000 records | ||||
| USB PC Communication | — | ✔ | ||||
| Auto Power-Off | 15 minutes (manually switchable) | 10 minutes (manually switchable) | 15 minutes (manual shutdown available) | |||
| Operating Environment | -10~50℃, Max 80% RH, indoor altitude < 2000 m | |||||
| Storage Environment | -10~50℃, Max 70% RH (remove battery) | |||||
| Product Dimensions | 168×59×29 mm | 152×55×33 mm | 239×70×35 mm | 175×70×35 mm | ||
| Display Size | 34×31 mm | 37×41 mm | 53×44 mm | |||
| Weight | Approx. 120 g (excluding battery) | Approx. 135 g (excluding battery) | Approx. 210 g (excluding battery) | |||
| Power Supply | 2 × 1.5V AAA (R03/LR03) | 3 × 1.5V AAA (R03/LR03) | 3 × 1.5V AA (LR06) batteries / USB micro port 5V | |||
Wind Speed Range Details
| Unit | Range (TA640A) | Range (TA641A/B) | Range (TA642A/B/C) | Resolution | Accuracy |
|---|---|---|---|---|---|
| m/s | 0.1 ~ 30.0 | 0.06 ~ 30.0 | 0.06 ~ 45.0 | 0.1 / 0.01 / 0.001 | ±(2% + 0.2) |
| km/h | 1.0 ~ 108.0 | 1.0 ~ 108.0 | 1.0 ~ 162.0 | 0.1 / 0.01 / 0.001 | ±(2% + 0.7) |
| ft/m | 10 ~ 5905 | 10 ~ 5905 | 10 ~ 8858 | 1 / 0.1 / 0.001 | ±(2% + 20) |
| mile/h | 0.1 ~ 67.00 | 0.1 ~ 67.00 | 0.1 ~ 100.6 | 0.01 / 0.001 | ±(2% + 0.2) |
| knots | 0.1 ~ 58.32 | 0.1 ~ 58.32 | 0.1 ~ 84.48 | 0.01 / 0.001 | ±(2% + 0.2) |
| Bft | 0 ~ 12 | — | — | ||
Conversion factors: 1m/s = 3.6km/h = 196.85ft/min = 3.280ft/s = 2.237mile/h = 1.944knots. 1knot = 0.7234710361 (Bft).
Wind Temperature Range Details
| Unit | Range | Resolution | Accuracy |
|---|---|---|---|
| ℃ | 0 ~ 45℃ | 0.1℃ | ±2℃ |
| ℉ | 32.0 ~ 113.0℉ | 0.1℉ | ±3.6℉ |
Air Volume Range Details (TA642A/B/C)
| Unit | Range | Resolution | Accuracy |
|---|---|---|---|
| CFM | 0 ~ 9999ft³/min | 0.001 ~ 100 | — |
| CMM | 0 ~ 9999m³/min | 0.001 ~ 100 | — |
| CMS | 0 ~ 9999m³/s | 0.001 ~ 100 | — |


5. Setup
Battery Installation
Install the correct type and number of batteries according to your model (refer to the specifications table for details). Ensure correct battery polarity during installation.
- TA640A: 2 × 1.5V AAA batteries
- TA641A / TA641B: 3 × 1.5V AAA batteries
- TA642A / TA642B / TA642C: 3 × 1.5V AA batteries


Power On
Power on the instrument using the power key. Ensure the sensor impeller rotates freely without obstruction before use.
Probe Connection (TA642C Split Type)
For the TA642C split-type model, connect the detachable probe using the 5-pin aviation connector. Extend the telescopic rod as required for measurement, up to 1 meter.


6. Operation
Measurement Modes and Unit Selection
Select the desired measurement mode (wind speed, air temperature, air volume, or humidity) using the appropriate buttons on the device. Choose the required wind speed unit (m/s, km/h, ft/m, mile/h, knots) using the "UNIT" button.




Data Reading and Hold Function
Read real-time values directly from the LCD. Use the data hold function to freeze readings. Supported models allow viewing of maximum, minimum, and average values.

Alarms (TA641A/B, TA642 series)
For models with alarms, upper and lower limit thresholds can be set. When the measured value exceeds or falls below the threshold, a buzzer will sound and the warning light will flash.

Data Logging and USB Communication (TA642B/C)
The TA642 series supports data logging, storing up to 1000 records. These records can be viewed directly on the device. For TA642B/C models, stored data can be transferred to a PC via USB. Upper computer software (compatible with Windows 7, Windows 8, and Windows 10) allows for real-time communication, data download, and the export/printing of xlsx data tables and graphs.


Auto Power-Off
The device automatically shuts off after 10 or 15 minutes of inactivity, depending on the model. This feature can be manually deactivated.
7. Maintenance
Cleaning
Clean the housing and sensor gently with a soft dry cloth. Do not use solvents or abrasive cleaners.
Storage
Store the instrument in a dry, well-ventilated place within the specified storage conditions (-10 to 50°C, maximum relative humidity of 70%). Periodically check battery condition and replace when necessary. Ensure the impeller remains clean and free of dust. Remove batteries during long-term storage.
8. Troubleshooting
If the device is not functioning as expected, please ensure:
- Batteries are correctly installed and have sufficient charge.
- The sensor impeller is free from obstructions and rotates smoothly.
- The device is used within its specified operating environment.
- For split-type models, the probe is securely connected.
If issues persist after checking these points, please refer to the contact information in the Support section.
9. User Tips
- Switching Units: Easily toggle between wind speed units (m/s, km/h, ft/min, mile/h, knots) by pressing the "UNIT" button for quick adaptation to different measurement standards.
- Capturing Peak/Trough Data: Utilize the "MAX/MIN" button during measurements to record and review the highest, lowest, and average values, which is useful for analyzing fluctuating conditions.
- PC Data Analysis (TA642B/C): For detailed analysis, connect your TA642B/C model to a PC via USB. The accompanying software allows you to export data in .xlsx format and visualize trends with real-time graphs, making reporting and long-term monitoring more efficient.
10. Support & Compliance
Compliance Disclaimer
This document is provided for product instruction purposes only. It does not constitute legal or regulatory advice. The seller is responsible for ensuring full compliance with applicable EU directives and local market regulations.
Contact Information
For technical consultations or bulk orders, please contact us:
- WhatsApp: +86 13311678550