1. Safety Information
WARNING:
Pay special attention to the use of this instrument. Improper use may cause electric shock or damage to the instrument. During use, observe usual safety regulations and safety measures regulated in this manual.
To make full use of instrument functions and guarantee safe operation, please carefully read and observe the use methods in this manual.
This instrument complies with safety requirements for electronic measuring instruments: EN-61010-1, EN-61010-2-030, and EN-61010-2-032, level II pollution, and over-voltage standard CAT III 1000V, CAT IV 600V.
Please observe the safety operation guide and ensure the instrument is used in a safe manner.
1.1 Preparation
- When using this instrument, users must observe standard safety rules:
- General electric shock prevention.
- Prevention of misuse of the instrument.
- After receiving the instrument, check whether it is damaged during transportation.
- After storing and shipping under adverse conditions, check whether the instrument has been damaged.
- The test leads (pens) of the instrument must be in good condition. Before use, check whether the insulation of the leads is damaged, and whether any metal wire is exposed.
1.2 Symbols
| Symbol | Description |
|---|---|
| ⚠ | Note (refer to user manual for important safety information) |
| ⚡ | Able to be used on dangerous electrified conductors. |
| ☒ | Dual-insulation protection (Category II) |
| CAT III CAT IV |
Over-voltage according to IEC-61010-1 standard (installation), level III, IV pollution degree 2 refers to protection level of pulse withstand voltage provided. |
| CE | Comply with EU standard. |
| ☂ | Grounded |
1.3 General Maintenance Advice
- Do not try to open the bottom case to adjust or repair the instrument. Such operations should only be conducted by technicians who fully understand the instrument and the electric shock danger.
- Before opening the instrument's bottom case or battery cover, remove the test leads from the wire being measured.
- To avoid electric shock possibly caused by error reading, when the battery symbol 🔋 displays on the instrument, replace the battery immediately.
- Use a wet cloth and gentle detergent to clean the instrument. Do not use any abrasive or solvent.
- Power off the instrument when not in use, and rotate the range switch to the OFF position.
- If the instrument is not used for a long time, take out the battery to avoid any damage to the instrument.
2. Product Description
2.1 Part Names
- Current clamp: Used in current measurement
- Rotary switch: Selects measurement functions
- Function key: For additional function selection
- Liquid crystal display (LCD): Shows measurement readings and indicators
- Input socket: For connecting test leads
- Function key: For additional function selection
- Trigger: Opens the current clamp jaws
- Non-contact voltage detecting & inducing area: For NCV detection
2.2 Rotary Switch Description
| Symbol | Function |
|---|---|
| A≈ | AC & DC current measurement |
| TEMP | Temperature measurement |
| Ω ⇲ | Capacitor, on-off and resistance measurement |
| V≈ VFD | AC/DC voltage measurement |
| LoZ V | Low-impedance voltage measurement |
| µA | DC µA current measurement |
| OFF | Instrument OFF |
2.3 LCD Display Indicators
| Indicator | Description |
|---|---|
| ≈ ⎯ | AC & DC |
| •)) | Connection/Disconnection indication (Continuity) |
| AUTO | Automatic range mode |
| ⏸ | Automatic shutdown indication |
| 🔋 | Low battery |
| H | Reading hold state |
| V,A, µA | Volt (voltage), ampere (current) |
| Ω, kΩ, MΩ | ohm, kilohm and megohm (resistance) |
| Hz, kHz | hertz, kilohertz |
| VFD | Variable frequency measurement |
| µF | Microfarad |
| °C °F | Centigrade and Fahrenheit |
| LowZ | Low-impedance measurement mode |
| % | Duty ratio measurement |
| REL | Relative measurement |
3. Specifications
The instrument specifies a one-year calibration cycle and shall be re-calibrated under 18°C ~ 28°C, with relative humidity less than 75%.
3.1 Overview
- Select measurement function and range automatically.
- Overload protection throughout the range.
- Max. voltage between measurement terminal and ground: 1000V DC or 1000V AC.
- Operating height: Max. 2000m.
- Display: LCD.
- Max. display value: 6000 digit.
- Polarity indication: Automatic indication, and '-' indicates negative.
- Over range display: 'OL' or '-OL'.
- Sampling time: About 3 times/second, bargraph 10 times/second.
- Unit display: Function and electricity quantity unit display.
- Automatic shutdown time: 10 minutes.
- Power supply: 1.5V AAA battery × 3.
- Battery under-voltage indication: LCD display symbol 🔋.
- Temperature coefficient: Less than 0.1 × accuracy/°C.
- Operating temperature: 18°C ~ 28°C.
- Storage temperature: -10°C ~ 50°C.
3.2 Technical Indexes
3.2.1 AC Current
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 0.01A | ±(2.5% reading + 8 digits) |
| 600A | 0.1A | |
| 1000A | 1A |
- Min. input current: 0.1A AC current.
- Max. input current: 1000A AC current.
- Frequency range: 45 ~ 65Hz.
3.2.2 DC Current
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 0.01A | ±(3% reading + 10 digits) |
| 600A | 0.1A | |
| 1000A | 1A |
- Min. input current: 0.01A DC current.
- Max. input current: 1000A DC current.
3.2.3 DC Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| 6V | 0.001V | ±(0.5% reading + 5 digits) |
| 60V | 0.01V | |
| 600V | 0.1V | |
| 1000V | 1V |
- Min. input voltage: 0.001V DC.
- Max. input voltage: 1000V DC.
3.2.4 AC Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| 6V | 0.001V | ±(0.8% reading + 5 digits) |
| 60V | 0.01V | |
| 600V | 0.1V | ±(1.0% reading + 5 digits) |
| 1000V | 1V |
- Min. input voltage: 0.001V AC.
- Max. input voltage: 1000V AC (valid value).
- Frequency range: 45 ~ 1000Hz.
3.2.5 Frequency
3.2.5.1 Clamp Frequency Measurement (via gear A)
| Range | Resolution | Accuracy |
|---|---|---|
| 60.0Hz | 0.1Hz | ±(1.0% reading + 5 digits) |
| 1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz.
- Input signal range: ≥ 2A AC current (valid value).
3.2.5.2 Frequency Measurement (Via gear V)
| Range | Resolution | Accuracy |
|---|---|---|
| 60.0Hz | 0.1Hz | ±(1.0% reading + 5 digits) |
| 1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz.
- Input signal range: ≥ 0.8V AC voltage (valid value).
3.2.6 Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 6KΩ | 0.001KΩ | ±(0.8% reading + 3 digits) |
| 60KΩ | 0.01KΩ | |
| 600KΩ | 0.1KΩ | |
| 6MΩ | 0.001MΩ | |
| 60MΩ | 0.01MΩ |
- Overload protection: 250V DC or AC (valid value).
3.2.7 Line On-off Test (Continuity)
| Range | Resolution | Function |
|---|---|---|
| •)) | 1Ω | If the resistance of the line being measured is less than 30Ω, the instrument buzzer will make continuous alarming sounds. |
- Overload protection: 250V DC or AC (valid value).
3.2.8 Capacitor
| Range | Resolution | Accuracy |
|---|---|---|
| 60.00nF | 0.01nF | ±(4% reading + 3 digits) |
| 600.0nF | 0.1nF | |
| 6.000µF | 1nF | |
| 60.00µF | 10nF | |
| 600.0µF | 100nF | |
| 6.000mF | 1µF | |
| 60.00mF | 10µF |
- Overload protection: 250V DC or AC (valid value).
3.2.9 Temperature
| Range | Resolution | Accuracy |
|---|---|---|
| -20~1000°C (-4~1832°F) | 1°C/2°F | ±(1% reading + 2 digits) |
- Overload protection: 250V DC or AC (valid value).
3.2.10 µA Current
| Range | Resolution | Accuracy |
|---|---|---|
| 200.0µA | 0.1µA | ±(0.8% reading + 3 digits) |
- Overload protection: 250V DC or AC (valid value).
4. Operation Guide
4.1 Reading Hold
During measurement, if it is required to hold the reading, press the 'H' button. The value on the display will be locked. Press the 'H' button again to cancel the reading hold.
4.2 Backlight/Light
- During measurement, if the environment light is too dark, making it difficult to read, press the 'H/☀' key for more than 2 seconds to open the backlight or light. It will turn off automatically in about 1 minute.
- During this period, if you press the 'H/☀' key for more than 2 seconds, it will turn off the backlight immediately.
4.3 Automatic Shutdown
- If there is no operation within 10 minutes after starting, the instrument will enter standby state and shut down to save energy.
- After automatic shutdown, press any key to wake the instrument and enter the work state.
4.4 AC & DC Current Measurement
Turn the rotary switch to the AC current range and hold the trigger to open the clamp. Clip one lead of the line being measured within the clamp. The instrument will display the measured current. Press the FUNC key to select AC or DC current measurement, and press the RAN key to select the range manually. During AC current measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured current. Long press the FUNC key to start the VFD measurement function.
4.5 AC & DC Voltage Measurement
Turn the rotary switch to AC & DC voltage. Press the FUNC key to shift between AC and DC voltage measurement modes. Connect the test leads to the signal being measured, with the red lead connected to the positive of the signal and the black lead connected to the negative. During AC voltage measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured voltage. Long press the FUNC key to start the VFD measurement function.
4.6 Low-impedance Measurement (LoZ)
Turn the rotary switch to LoZ. Press the FUNC key to shift between AC and DC voltage measurement modes, and connect the test leads to the signal being measured.
Note:
In low-impedance measurement mode, the longest measurement time shall not exceed 1 minute.
4.7 On-off/Resistance/Capacitor Measurement
Turn the rotary switch to the resistance and capacitor position. Press the FUNC key to shift between on-off (continuity), resistance, and capacitor functions. When test leads are connected to the resistance being measured, the instrument will display the measured resistance. If the measured value is less than 30Ω, the instrument buzzer will alarm.
4.8 Temperature Measurement
Turn the rotary switch to TEMP. Insert the probe of the thermocouple into the socket, with the positive of the probe connected to the red input terminal. The primary display will show the measured temperature in Centigrade. Press the FUNC key to display the measured temperature in Fahrenheit.
4.9 µA Current Measurement
Turn the rotary switch to µA, and connect the test leads to the signal being measured in series. The primary display will show the measured current.
4.10 NCV (Non-contact Voltage Detection)
Turn the rotary switch to NCV. The instrument will display 'EF'. Bring the NCV detector close to the lead being measured. The instrument can detect whether the lead being measured has >90V AC voltage. When the instrument detects AC voltage, the instrument buzzer will alarm, and the NCV alarm light will flash.
Note:
Even if there is no alarm indication, voltage may still exist. Do not solely depend on the non-contact voltage detector to judge whether there is voltage in the lead. Detection operation may be impacted by factors like different socket designs and insulation thickness types, etc.
5. Maintenance
5.1 Change Battery
WARNING:
Before opening the battery cover of the instrument, remove the test leads from the circuit being measured to avoid electric shock.
- If the battery symbol 🔋 appears, it indicates that the battery needs to be changed.
- Unfasten the bolts on the battery cover of the instrument and remove the cover.
- Change the old battery with new 1.5V AAA batteries (3 required).
- Place the battery cover back and fasten the bolts.
5.2 Change Test Leads (Pens)
WARNING:
When changing the test leads, it is required to replace them with identical leads or leads of the same level. The leads must be in good condition, and the level of the leads should be 1000V 10A.
If the insulation layer of the test lead is damaged, or the metal wire of the lead is exposed, it is required to change the lead immediately.
6. Accessories
The PEAKMETER PM2128 Clamp Meter package typically includes the following items:
- Test Leads (Pens): Level: 1000V 10A (1 set)
- User Manual (1 copy)
- 1.5V AAA Battery (3 pieces)
- Cloth Bag (1 piece)
- K-Type Thermocouple (1 piece)
For a more detailed version of the user manual, you may refer to the official PDF document available at: PEAKMETER PM2128 User Manual (PDF).
7. Troubleshooting
If you encounter issues with your PEAKMETER PM2128 Clamp Meter, consider the following general troubleshooting steps:
- No Display/Power On: Check if the batteries are correctly installed and have sufficient charge. Replace batteries if the low battery indicator is shown or if the device does not power on.
- Incorrect Readings: Ensure the correct measurement function is selected on the rotary switch. Verify that test leads are properly connected and in good condition. For current measurements, ensure the clamp jaws are fully closed around a single conductor.
- Intermittent Readings: Check for loose connections of test leads. Ensure the measurement environment is stable and free from strong electromagnetic interference.
- Buzzer Alarms Continuously: For continuity tests, this indicates a low resistance connection. For NCV, it indicates detected AC voltage. Refer to the specific function's operation guide.
- 'OL' or '-OL' Display: This indicates an over-range condition. The measured value exceeds the selected range. Switch to a higher range if available, or ensure the input is within the instrument's maximum capabilities.
If the problem persists after attempting these steps, please contact customer support or a qualified technician.
8. User Tips
Based on general user feedback for similar devices, here are some tips to enhance your experience with the PM2128 Clamp Meter:
- Always start with the highest range when measuring an unknown value to prevent overloading the meter.
- Utilize the 'Hold' function (H button) to freeze readings in hard-to-see locations.
- The backlight feature is very useful in dimly lit environments; remember to activate it with a long press of the 'H/☀' button.
- Keep the meter and its accessories clean and stored in the provided carrying case to protect it from dust and damage.
- Regularly check the condition of your test leads for any signs of wear or damage to ensure safety and accurate measurements.
9. Frequently Asked Questions (FAQ)
This section addresses common questions about the PEAKMETER PM2128 Clamp Meter.
- Q: Does this meter measure both AC and DC current?
A: Yes, the PM2128 is capable of measuring both AC and DC current up to 1000A. - Q: Are batteries included with the product?
A: Yes, 3 AAA batteries are typically included. - Q: How do I know when to replace the batteries?
A: A low battery symbol (🔋) will appear on the LCD display. - Q: Can I measure temperature with this device?
A: Yes, it supports temperature measurement using the included K-Type thermocouple. - Q: What is NCV and how does it work?
A: NCV stands for Non-Contact Voltage detection. It allows you to detect the presence of AC voltage (>90V) without direct contact, indicated by an alarm and flashing light.
10. Warranty and Support
For warranty information, technical support, or service inquiries, please refer to the manufacturer's official website or contact the seller directly. Keep your purchase receipt as proof of purchase for any warranty claims.