NJTY T6005, T6010

Contents

NJTY T6005 / T6010 Analog Insulation Resistance Tester User Manual

USB Rechargeable 500V / 1000V Megaohmmeter

1. Introduction and Overview

This manual provides essential instructions and safety guidelines for the safe operation and maintenance of your NJTY T6005 or T6010 Analog Insulation Resistance Tester. Please read this document thoroughly before using the instrument.

The symbol "▲" on the instrument indicates that users must follow relevant safety instructions detailed in this manual. Pay special attention to safety rules and warning notes to prevent electric shock and damage to the instrument.

The T60 series pointer-type insulation testers, including models T6015, T6010, and T6005, are designed for modern electrical testing. They feature a compact size, portability, high accuracy, and ease of operation. The T6005 model supports 500V insulation testing, while the T6010 supports 1000V insulation testing. These testers are ideal for measuring the insulation resistance of various electrical equipment and insulating materials such as transformers, motors, cables, switches, and appliances. They are suitable for routine maintenance, testing, and verification of electrical equipment.

Front view of the NJTY T6010 Analog Insulation Resistance Tester
Figure 1: Front view of the NJTY T6010 Analog Insulation Resistance Tester.

2. Safety Rules and Precautions

  1. Read this manual carefully before use.
  2. Do not use the instrument if the battery cover is not securely closed, as this poses an electric shock hazard.
  3. Before use, ensure the test leads' insulation is intact and undamaged.
  4. To avoid damaging the instrument, perform insulation measurements only on disconnected circuits.
  5. Do not press the "Test Button" when measuring voltage.
  6. To avoid electric shock, do not touch the test leads or the circuit being tested during measurement.
  7. Avoid touching the tested circuit immediately after testing, as residual charge may cause electric shock.

3. Product Features

  • Wide measurement range, up to 2000 MΩ.
  • AC voltage measurement below 500V (T6010 and T6005 models). T6015 model can measure AC and DC voltage below 500V.
  • Low resistance buzzer alarm: alerts the user when measured resistance is below the alarm setting value.
  • Working indicator lights: include battery undervoltage status (red light), automatic shutdown after 15 minutes (single green light), and continuous working indication (double green lights).
  • Uses a 3.7V-18650 rechargeable battery and Type-C charging interface for convenience.
  • Wider dial design for a broader indication range and more accurate readings.
  • Automatic discharge function: automatically discharges residual charge after shutdown to prevent electric shock.
  • Strong output load capacity and large output short-circuit current.
  • Transistor withstand voltage measurement scale: directly reads the withstand voltage value (load voltage) of the material being tested.
NJTY T6010 tester attached to an electrical cabinet via its strong magnet
Figure 2: The tester features a strong magnet on the back, allowing it to be attached to iron objects for hands-free operation.
Rear view of the NJTY T6010 tester showing magnet and 18650 battery compartment
Figure 3: Rear view of the tester, highlighting the strong magnet and the 18650 lithium battery compartment.

4. Panel Diagram and Components

Familiarize yourself with the components of your insulation resistance tester:

Front panel diagram of T6005/T6010 and T6015 models
Figure 4: Front panel layout of the T6005/T6010 models (top) and T6015 model (bottom).
  1. Charging indicator light
  2. Work indicator light
  3. Gear position rotary switch (T6015 model only)
  4. EARTH input terminal (connected to the ground terminal of the circuit under test)
  5. LINE input terminal (connected to the circuit under test)
  6. Test button
  7. Pointer zero adjustment
  8. Insulation meter reading dial

5. Setup and Initial Checks

5.1 Mechanical Zero Adjustment

Before starting any insulation measurement, verify that the pointer aligns with the "∞" (infinity/zero position on the left) mark on the insulation meter scale. If it does not, use a small screwdriver to rotate the "pointer zero adjustment" screw on the meter to correct the zero position.

5.2 Test Lead Connection

Insert the red and black test leads into the L (LINE) and E (EARTH) ports of the meter, respectively.

6. Operating Instructions

6.1 Insulation Resistance Measurement

  1. **Start and Shut Down:** Press and hold the "test button" for 1 second to turn on the meter. The green "working" indicator light on the upper right corner of the dial will illuminate. The meter will automatically shut down after 15 minutes of inactivity.
  2. **Continuous Working Mode:** To switch to continuous working state, press the "test button" three times within 5 seconds after powering on. Both green indicators in the upper right corner will light up, and the automatic shutdown function will be canceled.
  3. **Manual Shut Down:** To manually shut down the meter, press and hold the "test button" for 1 second after 5 seconds of powering on.
  4. **Undervoltage Indication:** If the working indicator light is red, it signifies an undervoltage state, and the meter will automatically shut down after 15 minutes.
  5. **Reading Measurement:** After starting the insulation measurement, touch the test leads to the insulator being measured and read the value indicated by the pointer scale.
Insulation resistance testing in progress with the NJTY T6010 tester
Figure 5: Performing insulation resistance testing on an electric motor.

6.2 AC and DC Voltage Measurement

  1. Ensure no insulation measurement is active (working indicator light is off).
  2. For the T6015 model, switch to the corresponding voltage range. T6010 and T6005 models can measure directly.
  3. Use the test leads to touch the two ends of the voltage to be measured.
  4. Read the measured value according to the pointer scale.
  5. **Note:** When T6015 measures DC voltage, insert the red test lead into "+" and the black test lead into "-".

6.3 Transistor Withstand Voltage Measurement

  1. Connect the transistor to be tested to the meter, observing polarity.
  2. Press the test button to directly read the BV(V) voltage scale. This value represents the withstand voltage of the transistor (the DC load voltage value L(V) at both ends of the material being tested).
  3. The maximum withstand voltage measurement for T6015 and T6010 insulation meters is 1000V. The maximum for the T6005 insulation meter is 500V.

7. Charging and Replacing Batteries

  1. **Safety First:** To prevent electric shock, always remove the test leads from the instrument before charging or replacing batteries.
  2. **Battery Status:** When the instrument is operating normally, a green working indicator light signifies sufficient battery power. A red working indicator light indicates low battery (undervoltage), requiring charging or replacement.
  3. **Charging:** Use a 5V USB power supply (e.g., power bank, car charger) with the Type-C charging interface on the instrument.
  4. **Charging Indicator:** The charging indicator light will be red during charging and turn green when fully charged.
  5. **Important:** Do not initiate any tests while the instrument is charging.
  6. **Battery Replacement:** If the battery fails, replace it with a battery of the same specification (18650 rechargeable lithium battery).

8. Maintenance

This instrument is a precision electronic device and requires careful maintenance to ensure longevity and accuracy.

  1. **Long-term Storage:** If the instrument will not be used for an extended period, recharge the battery every two months to maintain its health.
  2. **Storage Environment:** Store the instrument in a dry place, away from high humidity.
  3. **Physical Handling:** Avoid strong vibrations or impacts, as these can damage the meter's internal components or circuitry.
  4. **Internal Modifications:** Do not attempt to modify the internal circuitry of the instrument. If a malfunction occurs, contact customer support for assistance.

9. Specifications

General Specifications:

  • **Power Supply:** 1 x 3.7V 18650 rechargeable lithium battery
  • **Dimensions:** 161 × 96 × 45 mm
  • **Weight:** Approximately 534 g (including accessories)
  • **Operating Temperature:** 0°C - 40°C, Relative Humidity <75% (non-condensing)
  • **Storage Temperature:** -10°C - 50°C, Relative Humidity <85%
  • **Calibration Temperature:** 23°C ± 5°C, Relative Humidity <70%

Measurement Specifications:

Model T6015 T6005 T6010
Rated Test Voltage 1000V 500V 100V 500V 1000V
Low-Resistance Alarm <2MΩ <1MΩ <200kΩ <2MΩ <2MΩ
Transistor Withstand Voltage Test 1000V 500V 100V 500V 1000V
Output Short-Circuit Current Max 0.75 mA Max 0.65 mA Max 1.3 mA
Measurement Range 0-2000MΩ 0-1000MΩ 0-2000MΩ
Measurement Accuracy 1M-2000MΩ ±10% 1M-1000MΩ ±10% 1M-2000MΩ ±10%
Voltage Measurement Range DCV500V, 50V; ACV500V (50HZ) ACV500V (50HZ)
Insulation Test Output Voltage 90%-120% of Rated Voltage
Discharge Resistance 10MΩ
Voltage Measurement Error ±5% (Full Scale)
Battery Detection 3.3V Low Voltage Indication
Insulation Between the movement circuit and the case: >50MΩ

10. Package Contents

The following items are included in your product package:

Contents of the T6005 Insulation Resistance Tester package
Figure 6: T6005 Insulation Resistance Tester Configuration Diagram, showing all included accessories.
  • T60 Series Insulation Resistance Tester x 1 (T6005 or T6010 model)
  • Portable Tool Bag x 1
  • Insulation Resistance Tester Test Leads x 1 pair (includes crocodile clips)
  • User Manual x 1 (PDF version available here)
  • 18650 Rechargeable Lithium Battery x 1
  • USB Type-C Charging Cable x 1

11. Troubleshooting

This section provides guidance for common issues you might encounter. For problems not listed here, please refer to the maintenance section or contact customer support.

  • **Instrument does not power on:** Check if the battery is properly inserted and charged. Ensure the battery cover is securely closed.
  • **Working indicator light is red:** This indicates low battery. Recharge the battery using the provided Type-C cable and a 5V USB power supply.
  • **Inaccurate readings:** Perform mechanical zero adjustment as described in Section 5.1. Ensure test leads are properly connected and undamaged.
  • **No measurement when test button is pressed:** Ensure the test leads are firmly connected to the circuit under test and that the circuit is disconnected from power sources for insulation tests.
  • **Automatic shutdown occurs too quickly:** Check battery level. If in continuous working mode, ensure the test button was pressed three times within 5 seconds after power-on.

12. User Tips

  • Users have reported that the device works well and is reliable for its intended purpose.
  • The magnetic back is a convenient feature for hands-free operation in electrical panels.
  • Always double-check the mechanical zero adjustment before critical measurements for optimal accuracy.

13. Warranty and Support

For warranty information, technical support, or service inquiries, please refer to the contact information provided by your retailer or the manufacturer's official website. Always provide your product model (T6005 or T6010) and serial number when seeking support.