Victor 9808+ Digital Multimeter Instruction Manual

Model: 9808+

Brand: RuoShui

1. Introduction

The Victor 9808+ is a stable and high-performance digital multimeter powered by batteries. It features an LCD with a 20mm-high display for clear readings and convenient operation. This instrument utilizes a dual-integral A/D converter and offers a wide range of functions including DCV, ACV, DCA, ACA, resistance, capacitance, frequency, inductance, diode, triode, continuity testing, temperature measurement, electromagnetic induction, and live wire judgment. It also includes features such as power on/off, backlight, and a torch, making it an excellent tool for various electrical measurements.

Safety Notes

  • Input of limit values beyond the specified range is prohibited.
  • Voltages less than 36V are considered safety voltages. When measuring voltages higher than DC 36V or AC 25V, always check the connection and insulation of test leads to prevent electric shock.
  • Ensure test leads are disconnected from the testing point when changing functions or ranges.
  • Select the correct function and range to avoid incorrect operation.
  • Do not use the meter for testing if the battery is not installed or the back cover is not tightened.
  • When measuring resistance, capacitance, inductance, diode, or performing on-off tests, do not input voltage signals.
  • Before replacing the battery or fuse, remove test leads from the test point and turn off the power switch.
  • Always comply with local and national safety regulations. Wear personal protective equipment (approved rubber gloves, masks, and flame retardant clothing) to prevent electric shock and arc injury.
  • Use only the correct measurement standard class (CAT), voltage, and current rated probe, test conductor, and adapter.
  • The symbol "A" indicates high voltage. The symbol "GND" refers to ground. The symbol "A" indicates that the manual must be consulted. The battery symbol "" indicates low battery.

2. Package Contents

Upon opening the box, verify the presence of the following items. If any items are missing or damaged, please contact your supplier.

  • Multimeter: 1 pc
  • Color box: 1 pc
  • Test pen: 1 pc
  • Product manual: 1 pc
  • Certificate of conformity: 1 pc
  • Temperature probe: 1 pc
  • K-type probe (-20℃~250℃): 1 pc
  • 1.5V AAA Battery: 4 pcs
  • Instruction Manual: 1 pc
  • Test leads: 1 pair
  • Crocodile clip: 1 pair
  • Cloth bag: 1 pc
Contents of the Victor 9808+ Digital Multimeter package
Contents of the Victor 9808+ Digital Multimeter package: Multimeter, color box, test pen, product manual, certificate of conformity, and temperature probe.

3. Parts and Controls

Labeled diagram of the Victor 9808+ Digital Multimeter
Labeled diagram of the Victor 9808+ Digital Multimeter showing LCD display, data retention/backlight, power button/function switching, range knob, current sockets, alarm indicator light, max/min, relative value/flashlight, triode jack, voltage/resistance/capacitance jack, public jack, flashlight, pen slot, and battery compartment/bracket.
  1. LCD display screen
  2. Data retention/backlight
  3. Power button/Function switching
  4. Range knob
  5. 20A current socket
  6. µA/mA current socket
  7. Alarm indicator light
  8. Max/min
  9. Relative value/flashlight lighting
  10. Triode Jack
  11. Voltage/resistance/capacitance jack
  12. Public Jack (COM)
  13. Flashlight
  14. Pen slot
  15. Battery compartment/bracket
Close-up of the Victor 9808+ Digital Multimeter's backlit large LCD display
Close-up of the Victor 9808+ Digital Multimeter's backlit large LCD display, showing a maximum display of 1999 (3 1/2 Clear) numerical values.

4. Specifications

Accuracy: ± (a% x reading + digit), at (23±5)°C, relative humidity <75%. One year guarantee since production date.

Performance (Note: "A" indicates that the meter has this range; "*" means the meter does not have this range.)

General Characteristics

Table of general characteristics for the Victor 9808+ Digital Multimeter
Table of general characteristics for the Victor 9808+ Digital Multimeter.
Characteristic Value Characteristic Value
Max Display1999 (3 ½ Bit)Dual display√
On/off buzzing√Progress bar√
Sound and light alarm√Low voltage prompt√
Data retention√Over range promptDisplay "OL"
Backlight display√Sampling rateAbout 3 times/second
Relative value√NCV/LIVE√
Flashlight lighting√Working conditions(0-40)℃ <75% RH
Max/min√Storage conditions(-20~60)℃ <85% RH
Automatic shutdown15 minutesPower SourceFour AAA batteries
Diode/transistor√Size189 * 88 * 56mm
T-RMS√Weight406g

Measurement Ranges and Accuracy

Technical parameters table for DC/AC Voltage and DC/AC Current
Technical parameters table for DC/AC Voltage and DC/AC Current.
Function Include or Not
DCVA
ACVA
DCAA
ACAA
Resistance ΩA
Inductance LA
Diode/Continuity testingA
Capacitance CA
Temperature ℃/℉A
Triode hFEA
Frequency FA
Auto power-off APOA
Backlight/TorchA
NCV/LiveA

DC Voltage

Range Accuracy Resolution Input Impedance Overload Protection
200mV±(0.5%+5)0.1mV250VDC/AC RMS
2V±(0.5%+3)0.001VAbout 10MΩ1000V DC/750V AC
20V0.01VRMS
200V0.1V
1000V±(1.0%+10)1V

AC Voltage TRUE RMS Measurement

Range Accuracy Resolution Input Impedance Overload Protection
200mV±(1.0%+5)0.1mV250VDC/AC RMS
2V±(0.8%+5)0.001VAbout 10MΩ1000V DC/750V AC
20V0.01VRMS
200V0.1V
750V±(1.2%+10)1V

Accurate measurement range: 10%-100% of the range; Frequency response: 40Hz-1KHz. For signals below 50Hz, the measurement accuracy increases by 2% on the original basis. Measurement mode (sine wave): True RMS. Crest factor: CF≤3, adding an additional error of 1% to the reading at CF≥2. When the test voltage value is greater than 20% of the range, the accompanying frequency can be displayed.

DC Current (DCA)

Range Accuracy Resolution Load Voltage Overload Protection
200uA±(1.2%+8)0.1uA0.1mV/uAFUSE 200mA/250V
2000uA1uA0.1mV/uA
20mA0.01mA1.1mV/mA
200mA0.1mA1.1mV/mA
20A±(2.0%+5)0.01A20.3mV/AFUSE 20A/250V

20A (test within 10 seconds); Recovery time is 15 minutes.

AC Current (ACA)

Range Accuracy Resolution Load Voltage Overload Protection
200uA±(1.5%+15)0.1uA0.1mV/uAFUSE 200mA/250V
2000uA1uA0.1mV/uA
20mA0.01mA1.1mV/mA
200mA0.1mA1.1mV/mA
20A±(3.0%+10)0.01A20.3mV/AFUSE 20A/250V

Frequency response: 40Hz-1000Hz; Measurement mode (sine wave): true RMS; Crest factor: CF≈3, adding an additional error of 1% to the reading at CF≥2. 20A (test within 10 seconds); Recovery time is 15 minutes. When the test current value is greater than 40% of the range, the accompanying frequency can be displayed.

Resistance (Ω)

Range Accuracy Resolution Fault Current Open Circuit Voltage
200Ω±(0.8%+5)0.1Ω0.586mA
2KΩ±(0.8%+3)0.001KΩ0.586mAabout 1.9V
20KΩ0.01KΩ0.15mA
200KΩ0.1KΩ17.58uA
2MΩ±(1.0%+25)0.001MΩ1.82uAabout 1.8V
20MΩ0.01MΩ0.21uA
200MΩ±(5.0%+30)1MΩ0.21uAabout 1V

Overload protection 250VDC/AC RMS. The measured error doesn't include lead resistance.

Capacitance (C)

Range Accuracy Resolution Overload Protection
20nF±(3.5%+20)0.01nF250VDC/AC RMS
200nF0.1nF
20uF0.01uF
200uF0.1uF
100mFUnspecified0.1mF

Accurate measurement range: 10%-100%. Large capacitance response time: 1mF About 8s. The measured error doesn't include lead capacitance. Unit: 1mF=1000uF, 1uF=1000nF, 1nF=1000pF.

Frequency (Hz)

Range Accuracy Resolution Overload Protection
10Hz±(1.0%+10)0.01Hz250VDC/AC RMS
100Hz0.1Hz
1kHz0.001kHz
10kHz0.01kHz
100kHz0.1kHz
1MHz0.001MHz
20MHz0.01MHz

Note: For signals below 3Hz, the reading is zero. Input sensitivity: 3.0V. Duty cycle display: (0.1%-99.9%). Frequency input amplitude requirement: 10Hz-10MHz: 3Vrms < input amplitude < 10Vrms. 10MHz-20MHz: 11Vrms < input amplitude < 20Vrms. Duty cycle: 10%-90% range, suitable for square waves of 10Hz-1kHz; 30%-70% range, suitable for square waves of 1kHz to 10kHz. This frequency range is an automatic range test, which can be measured from 10Hz to 20MHz. Instrument sub-display can synchronously display duty cycle.

Temperature (°C/°F)

Range Displaying Range Resolution Overload Protection
(-20-1000)℃<400℃ ±(1.0%+5), ≥400℃ ±(1.5%+15)1℃250VDC/AC RMS
(-4-1832)℉<752℉ ±(1.0%+5), ≥752℉ ±(1.5%+15)1℉

Sensor: K type thermocouple (Ni-Cr-Ni-Si) banana plug.

Inductance (L)

Range Accuracy Resolution Overload Protection
2mH±(2.5%+30)0.001mH36VDC/AC RMS
20mH0.01mH
200mH0.1mH
2H±(5%+20)0.001H
20H0.01H

Accuracy measurement range: 5%-100% of the range. Do not input voltage at this range.

Transistor hFE Data Test

Range Displaying Range Test Condition
hFE NPN or PNP0~1000Basic current is approx. 3uA. Vce is about 1.2V

5. Operating Modes

Before operation, ensure the battery level is sufficient. If the battery level is at 30%, replace it. Always place the knob in the required position for measurement. Pay attention to the "A" symbol next to the test leads socket, which warns not to exceed the indicated voltage and current.

5.1. Key Functions Overview

  • POWER SELECT: Short press to turn on, long press to turn off. In AC/DC voltage and current ranges, short press to switch between AC/DC. In diode/on/off mode, short press to switch between diode and on/off tests. In electromagnetic induction/live wire detection mode, short press to switch between these tests.
  • Data Hold (HOLD B/L): Short press to hold the current measured value on the LCD. Press again to exit. Long press for more than 2 seconds to turn on/off the backlight. Warning: Do not use HOLD to measure unknown potential to avoid electric shock and fire.
  • Maximum/Minimum Value Measurement (MAX/MIN): Available in voltage/current/resistance/temperature/on/off/diode range. Short press to enter MAX/MIN mode. "MAX" or "MIN" will appear on the LCD. Long press for more than 2 seconds to exit.
  • Relative Value Measurement/Torch (REL/): Available in voltage/current/resistance/capacitance/temperature/inductance range. Short press "REL" to enter relative value measurement mode. Long press "REL/" for more than 2 seconds to turn on/off the flashlight.
  • Automatic Power Off (APO): The meter automatically shuts down after 15 minutes of inactivity. To cancel, press and hold "POWER" and "REL/" simultaneously when turning on the device. The "APO" symbol will disappear.
Icons representing multiple function options of the multimeter
Icons representing multiple function options: Frequency, Data retention, Triode, Over range prompt, Low voltage display, MAX/MIN, Capacitance, Inductance, Backlight function, Resistance, Automatic shutdown, On/off buzzing, Diode, Fire line judgment, and 1999 Digital Display.

5.2. AC/DC Voltage Measurement

  1. Turn the knob to the voltage gear. Short press the POWER SELECT key to switch between AC and DC voltage tests. If the voltage is unknown, start with a larger range and gradually reduce.
  2. Insert the red test lead into the "VΩHz" jack and the black test lead into the "COM" jack.
  3. Connect the test leads to the testing point. The screen will display the measured voltage value. For DCV, the voltage and polarity of the red test lead connection will be shown. For AC voltage, the secondary display area shows the accompanying frequency.

Note: If "OL" is displayed, the value is beyond the current range. Select a higher range. When measuring high voltage (above 36V), wear personal protective equipment. Input voltage must not exceed DC1000V or AC750V. Disconnect test leads and circuit after measurement.

5.3. AC/DC Current Measurement

  1. Turn the knob to the current gear. Short press the SELECT key to switch between AC and DC current tests.
  2. Insert the red test lead into the "µA/mA" jack (Max. 200mA) or "20A" jack (Max. 20A), and the black test lead into the "COM" jack.
  3. Connect the test leads in series with the power supply or circuit under test. The measured current value and polarity will be displayed.

Note: Turn off circuit power before connecting. If the range is uncertain, start with the max range. Max input current is 200mA or 20A; larger currents will blow the fuses. Test for less than 10 seconds when testing 20A current. When measuring large current (above 10A), wear personal protective equipment. Do not input more than DC 36V or AC 25V to the current jack and "COM" jack.

5.4. Resistance Measurement

  1. Set the knob to a proper resistance range.
  2. Input the red test lead to "Ω" terminal and the black one to "COM" terminal.
  3. Connect the test leads across the resistance to be tested.

Note: "OL" indicates an open circuit or over range. Adjust the knob to a proper range. For values over 1MΩ, readings may take a few seconds to stabilize. For low resistance, short-circuit probes and press "REL" to measure relative value. When measuring online resistance, ensure power is off and capacitors are discharged. Do not input any voltage at resistance range.

5.5. Capacitance Measurement

  1. Switch to a proper capacitance range.
  2. Input the red test lead to "F" terminal and the black test lead to "COM" terminal.
  3. Connect the test leads to the two sides of the tested capacitor (red test lead is "+").

Note: "OL" indicates over range. For capacitance below 20nF, residual reading may appear due to distributed capacitance; use "REL" key for relative value measurement. For serious leakage or breakdown, values may be unstable. For large capacitance, readings may take a few seconds to stabilize. Discharge the capacitor sufficiently before testing.

5.6. Transistor hFE Measurement

  1. Set the knob to "hFE" range.
  2. Verify the transistor type (NPN or PNP) and insert the emitter, base, and collector into the proper jack.

5.7. Diode and Continuity Test

  1. Input the black test lead to "COM" terminal and the red one to "" terminal (red lead is "+").
  2. Turn the knob to "" range. It defaults to on/off range. Press SELECT to convert to diode testing. If the buzzer sounds and the continuity alarm light is on, resistance is less than (50 ± 20) Ω.
  3. For diode range, connect test leads to the diode. Readings of 500mV to 800mV are normal for silicon PN junctions. "OL" indicates an open circuit or incorrect polarity.

Note: Do not input voltage at this "" function.

5.8. Frequency Measurement

  1. Set the knob to frequency range.
  2. Input the red test lead to "Hz" terminal and the black test lead to "COM" terminal.
  3. Connect test leads or a shielded cable to the signal source or load.

Note: Frequency input amplitude requirements: 10Hz-10MHz: 3Vrms < input amplitude < 10Vrms. 10MHz-20MHz: 11Vrms < input amplitude < 20Vrms. Strictly follow amplitude requirements. Using a shielded cable is better for small signals in noisy environments. Do not input voltage exceeding 36V DC or AC peak. This range is automatic (10Hz to 20MHz). Sub-display shows duty cycle.

5.9. Inductance Measurement

  1. Input the black test lead to "COM" terminal and the red test lead to "mA" terminal.
  2. Turn the knob to the corresponding inductance range, then connect the test leads to the inductance to be tested.

Note: "OL" indicates the measured inductance exceeds the range. For 2mH or 20mH range, short circuit test leads and press "REL" for relative value measurement. Do not input voltage at this range.

5.10. Temperature Measurement

  1. Insert the negative pole of the thermocouple sensor into the "COM" socket and the positive pole into the "℃/℉" socket.
  2. Place the thermocouple working end (measuring temperature end) above or inside the object to be measured. The temperature value will be displayed directly in Celsius.

5.11. Non-Contact Voltage (NCV) / Live Wire Recognition

  1. Turn the knob to the "NCV/Live" position (defaults to NCV function).
  2. For NCV, place the upper part of the instrument close to the tested live AC power line. When AC voltage is sensed, the red indicator light will flash and the buzzer will sound. The closer to the AC power line, the stronger the induced AC voltage and faster the flashing/buzzing.
  3. For Live wire recognition, input the red test lead to "VΩHz" terminal. Short press SELECT to switch to Live function. Use the red test lead to contact the tested position or socket. An audible and visual alarm indicates a live wire. No change indicates a zero wire or no power.
Diagram illustrating terminal block connections for current, voltage, resistance, continuity, diode, and capacitance measurements, with examples for voltage, current, and inductance.
Diagram illustrating terminal block connections for current, voltage, and inductance measurements.
Examples of connecting the multimeter for Capacitor, On/Off/Diode, Resistance, Temperature, Transistor (hFE), and Frequency measurements.
Examples of connecting the multimeter for Capacitor, On/Off/Diode, Resistance, Temperature, Transistor (hFE), and Frequency measurements.

6. Maintenance

6.1. General Care

  • Do not attempt to modify the electric circuit.
  • Keep the meter away from water, dust, and shock.
  • Do not store or operate the meter under conditions of high temperature, high humidity, combustible, explosive, or strong magnetic fields.
  • Wipe the case with a damp cloth and detergent; do not use abrasives or alcohol.
  • If the meter will not be used for a long time, remove the batteries to avoid leakage.

6.2. Replacing Battery or Fuse

  1. Remove the test leads from the circuit under test and from the input jacks. Long press the POWER button to turn off the meter.
  2. Remove the screws on the battery door with a screwdriver and remove the battery door and bracket.
  3. Remove the old battery or blown fuse and replace it with a new 1.5V battery or new fuse.
  4. Cover the battery door and install the screws.
  5. Battery specification: Four 1.5V AAA batteries.
  6. Fuse Specification: mA input fuse "FS1": 5x20mm 200mA 250V. 20A input fuse "FS2": 5x20mm 20A 250V.

Note: When the battery symbol "" is displayed, the battery level is 30%. Replace the battery immediately to maintain measurement accuracy.

Opening the battery compartment of the multimeter
Opening the battery compartment to replace batteries or fuses.
Internal view of the multimeter circuit board, highlighting the 200mA and 20A glass explosion-proof fuses
Internal view of the multimeter circuit board, highlighting the 200mA and 20A glass explosion-proof fuses for intelligent burn prevention.

7. Troubleshooting

If your meter is not working properly, the following methods can help you quickly solve the problem. If the fault persists, please contact the repair center or dealer.

Fault Solution
No reading on LCDPower off, then turn on. Set a correct mode. Replace battery.
The signal appearsReplace battery.
No current inputReplace fuse.
Big errorReplace battery.
LCD is darkReplace battery.

8. User Tips

  • The multimeter requires 4 AAA batteries for operation.
  • For precise maintenance work, consider using needle-type test leads.
  • When measuring inductance in the 0-2mH or 0-20mH ranges, a 100uH base error may occur. To correct this, short-circuit the probes and press the 'REL' button to zero out the reading for each measurement.
  • Capacitance measurements in the 100mF range may have lower resolution (only one decimal place) and can take several seconds to complete (e.g., 2,200uF in 3 seconds, 10,000uF in 5 seconds).
  • The continuity buzzer has a hold time of approximately 1 second, which might be longer than desired for rapid short-circuit detection.
  • Be aware that the multimeter may shut down if the input voltage exceeds 2V RMS when measuring frequencies.

9. Warranty and Support

This manual is subject to change without notice. The contents are considered correct. If you find any errors or omissions, please contact the manufacturer for assistance.

The company does not bear responsibility for accidents or harm caused by incorrect user operation. The functions described in this manual are not intended as a reason to use the product for a special purpose.

For further support or warranty claims, please refer to the contact information provided at the point of purchase or visit the official RuoShui website.