1. Introduction
This manual provides comprehensive instructions for the OWON DGE1000 Series Single-Channel Arbitrary Waveform Generator. It covers essential safety information, quick start procedures, detailed panel operations, communication with a PC, troubleshooting, specifications, and general care. Please read this manual thoroughly before operating the device to ensure safe and efficient use.

2. General Safety Requirement
Before any operations, please read the following safety precautions to avoid any possible bodily injury and prevent this product or any other products connected from damage. This product is only used within the range specified.
2.1. Safety Terms and Symbols
Terms in this Manual:
- Warning: Indicates conditions or practices that could result in injury or loss of life.
- Caution: Indicates conditions or practices that could result in damage to this product or other property.
Terms on the Product:
- Danger: Indicates an injury or hazard may immediately happen.
- Warning: Indicates an injury or hazard may be accessible potentially.
- Caution: Indicates a potential damage to the instrument or other property might occur.
Safety Symbols on the Product:
- ⚠ Hazardous Voltage
- ℹ Refer to Manual
- ⏚ Protective Earth Terminal
- ⏚ Chassis Ground
- ⏚ Test Ground
2.2. General Precautions
- Use Proper Power Cord: Use only the power cord supplied with the product and certified for use in your country.
- Product Grounded: This instrument is grounded through the power cord grounding conductor. To avoid electric shock, the grounding conductor must be grounded. The product must be grounded properly before any connection with its input or output terminal.
- Check all Terminal Ratings: To avoid fire or shock hazard, check all ratings and markers on the instrument. Refer to the user's manual for more information about ratings before connecting the instrument. Do not exceed any defined ratings.
- Do not operate without covers: Do not operate the instrument with covers or panels removed.
- Use Proper Fuse: Use only the specified type and rating fuse for this instrument.
- Avoid exposed circuit: Do not touch exposed junctions and components when the instrument is powered.
- Do not operate if in any doubt: If you suspect damage, have it inspected by qualified service personnel.
- Use in a well-ventilated area: Inadequate ventilation may cause overheating. Keep the instrument well ventilated and inspect air outlets and fans regularly.
- Do not operate in wet conditions.
- Do not operate in an explosive atmosphere.
- Keep instrument surfaces clean and dry.
- BNC output ports not allowed to input any signal: Do not input voltage, current, or other electrical signals to BNC output ports, as this will damage the device.
3. Setup
3.1. General Inspection
Upon receiving your new generator, perform the following checks:
- Check for Transportation Damage: If the packaging carton or foam protection cushion is seriously damaged, do not discard it until the device and accessories have passed electrical and mechanical property tests.
- Check Accessories: Refer to Appendix A: Accessories in this manual to verify all supplied accessories are present and undamaged. Contact your distributor or local office if any items are missing or damaged.
- Check the Complete Instrument: If the instrument's appearance is damaged, it does not work normally, or fails performance tests, contact your distributor or local office. Keep the original packaging if transportation damage is suspected.
3.2. Power On
- Connect the instrument to an AC power source using the provided DC power adapter.
- Press the power button on the front panel. The backlight of the button will illuminate, and a buzzer sound will confirm power-on.
Warning: To prevent electric shock, ensure the instrument is properly grounded.
4. Operating Instructions
4.1. Front Panel Overview

The front panel features a 3.6-inch TFT LCD for user interface display, a function setting area with a rotary knob and directional keys, an ON/OFF button, a waveform switching area, a channel output control area, and a numeric keypad for parameter input.
4.2. Rear Panel Overview

The rear panel includes air vents for cooling, a DC power input connector, a foot stool for tilting the device, and a USB Device interface for PC remote control via SCPI commands.
4.3. User Interface
The 3.6-inch TFT color LCD displays channel name, load, current modulation mode, and current waveform. Menu settings are fully displayed, and a convenient shortcut key improves man-machine interaction.
4.4. Channel Setting
Press the On/Off button on the front panel to toggle the corresponding channel output. The button's backlight indicates when the channel is active.
4.5. Waveform Setting
Sine, square, ramp, pulse, noise, or arbitrary waves can be set and output. Press the respective waveform selection button on the front panel to enter the corresponding setting interface. Parameters vary by waveform type.
4.5.1. Output Sine Wave
Press the Sine waveform button. The Sine setting menu on the right allows adjustment of Frequency/Period, Amplitude/High Level, Offset/Low Level, and Phase. Use the knob or numeric keypad to set values.
4.5.2. Output Square Wave
Press the Square waveform button. The Square waveform parameters can be set via the Square setting menu, including Frequency/Period, Amplitude/High Level, Offset/Low Level, and Phase.
4.5.3. Output Ramp Wave
Press the Ramp waveform button. The Ramp waveform parameters can be set via the Ramp setting menu, including Frequency/Period, Amplitude/High Level, Offset/Low Level, Phase, and Symmetry.
4.5.4. Output Pulse Wave
Press the Pulse waveform button. The Pulse waveform parameters can be set via the Pulse setting menu, including Frequency/Period, Amplitude/High Level, Offset/Low Level, Phase, Width/Duty, Rise, and Fall.
4.5.5. Output Noise Wave
The generator outputs white noise. Press the Noise waveform button. The Noise waveform parameters can be set via the Noise setting menu, including Amplitude/High Level and Offset/Low Level. The noise wave has no frequency or periodic parameters, and is Gaussian noise with a bandwidth of 20MHz.
4.5.6. Output Arbitrary Wave
Press the Arbitrary waveform button. The Arbitrary waveform parameters can be set via the Arbitrary setting menu, including Frequency/Period, Amplitude/High Level, Offset/Low Level, Phase, Built-in, and Store. The Arbitrary signal consists of system built-in waveforms and user-definable waveforms.
There are 160 types of built-in waveforms (8192 points, up to 10MHz). To select a built-in waveform:
- Press the Arbitrary waveform softkey, then NextPage to enter the next page menu.
- Press Built-in to enter the built-in wave menu.
- Select wave type (Common, Medical Treatment, Standard, Maths, Trigonometric, Window Function, Engineering, Seg Mod, Fan test) using softkeys and NextPage.
- Turn the knob to select the desired waveform (e.g., DC) and press Ok to output.
For a detailed list of built-in waveforms, refer to the Specifications section.
4.5.7. Store and Recall Waveforms
Users can load customized waveforms to the device via PC software. Install the PC software first (refer to Section 5, "Communication with PC").
To read a waveform:
- Visit the official website to obtain and decompress the Waveform Editor software.
- Run the "Waveform Editor" software.
- Select the desired waveform on the instrument.
- In the Waveform Editor software, click the "Read Waveform Icon" button. The waveform will be read and displayed on the screen.
To write and recall a waveform:
Use the Line Draw, Hand Draw, and Point Edit modes in the Waveform Editor to create or edit waveforms. Save and display them on the instrument by writing.
- In the Waveform Editor software, click the "Write waveform Icon" button.
- After successful writing, a "File transfer completed" prompt will appear. Click "OK".
- On the instrument, the screen will show "Any wave has been updated to USERX(X is 0-15)".
- Press the Arbitrary wave button, then NextPage to enter the next page menu.
- Press the Store softkey to enter the file system, then Enter. Select the file name "USERX" that was just written.
- Press the Call out softkey. The screen will display "File read successfully". Press the Arbitrary wave key to view the waveform on the instrument.
Note: The file size is displayed on the right. If 0B is displayed, the file is empty.
4.6. Generate Sweep (Sweep)
In sweep mode, the generator varies its output frequency from a start to a stop frequency within a specified sweep time. Sweep can be generated by Sine, Square, Ramp, or Arbitrary waves (except DC).
To set sweep parameters:
- When the output signal is Sine, Square, Ramp, or Arbitrary wave (except DC), press the front panel Mode function key, then press Sweep to enter sweep mode.
- Press the waveform selection button (Sine, Square, Ramp, Arbitrary) to display the sweep waveform and parameters. Refer to "Output Sine Wave" for parameter details. Press the waveform button again to return to sweep mode, or Mode to return to modulation selection.
- Press Linear/Log softkey to switch sweep type (Linear or Logarithmic).
- Press Sweep Time softkey to set the sweep time (1ms to 500s).
- Press Start Freq/Centre Freq softkey to highlight Start Freq (and Stop Freq in Stop Freq/Freq Span) to input desired frequencies. Alternatively, set frequency boundaries using center frequency and frequency span.
- Press Source softkey to select the trigger source (Internal or Manual).
4.7. Generate Burst (Burst)
Burst pulse train can produce various waveform functions. Burst can last for a certain number of waveform cycles (N-Cycle Burst) and applies to Sine, Square, Ramp, Pulse, and Arbitrary waveforms (except DC).
To set burst parameters:
- When the output signal is Sine, Square, Ramp, Pulse, or Arbitrary wave (except DC), press the front panel Mode function key, then press Burst to enter burst mode.
- Press the waveform selection button (Sine, Square, Ramp, Pulse, Arbitrary) to display the burst waveform and parameters. Refer to "Output Sine Wave" for parameter details. Press the waveform button again to return to burst mode, or Mode to return to modulation selection.
- Press Cycles/Infinite softkey to highlight Cycles and input the number of cycles (1 to 60,000). If Infinite is selected, the generator outputs a continuous waveform after receiving a trigger signal.
- Burst trigger source can be Internal, External, or Manual. Press Trigger to select the source.
4.8. Output Modulated Waves
Supported modulation types include AM (Amplitude Modulation), FM (Frequency Modulation), PM (Phase Modulation), and FSK (Frequency Shift Keying). Press the Mode function key to select the modulation type. Press Mode again to turn off modulation.
4.8.1. AM (Amplitude Modulation)
The amplitude of the carrier wave varies with the instantaneous voltage of the modulating wave.
To set AM parameters:
- Press Mode, then AM to enter the AM user interface.
- Select carrier wave shape (Sine, Square, Ramp, or Arbitrary wave, except DC).
- Set carrier wave parameters (refer to "Output Sine Wave").
- Select modulating wave shape (Sine, Square, Ramp, or Noise).
- Set modulating wave frequency (2 mHz – 100 kHz).
- Set modulation depth (0% – 100%).
4.8.2. FM (Frequency Modulation)
The frequency of the carrier wave varies with the instantaneous voltage of the modulating wave.
To set FM parameters:
- Press Mode, then FM to enter the FM user interface.
- Select carrier wave shape (Sine, Square, Ramp, or Arbitrary wave, except DC).
- Set carrier wave parameters (refer to "Output Sine Wave").
- Select modulating wave shape (Sine, Square, Ramp, or Noise).
- Set modulating wave frequency (2 mHz – 100 kHz).
- Set frequency deviation (1 uHz ≤ deviation < min(carrier frequency or carrier maximum frequency minus carrier frequency)).
4.8.3. PM (Phase Modulation)
The phase of the carrier wave varies with the instantaneous voltage of the modulating wave.
To set PM parameters:
- Press Mode, then NextPage, then PM to enter the PM user interface.
- Select carrier wave shape (Sine, Square, Ramp, or Arbitrary wave, except DC).
- Set carrier wave parameters (refer to "Output Sine Wave").
- Select modulating wave shape (Sine, Square, Ramp, or Noise).
- Set modulating wave frequency (2 mHz – 100 kHz for internal source only).
- Set phase deviation (0° to 180°).
4.8.4. FSK (Frequency Shift Keying)
FSK shifts the output signal frequency between two frequencies: the carrier frequency and hop frequency. The shift frequency (FSK rate) is determined by the internal signal level.
To set FSK parameters:
- Press Mode, then NextPage, then FSK to enter the FSK user interface.
- Select carrier wave shape (Sine, Square, Ramp, or Arbitrary wave, except DC).
- Set carrier wave parameters (refer to "Output Sine Wave").
- Press Rate softkey to set the FSK rate (2 mHz – 1 MHz).
- Press HopFreq softkey to set the hop frequency.
4.9. Utility Function Setting
Press the front panel Utility function key to enter the utility menu. This allows setting display, channel, and system parameters. Press Utility again to exit.
4.9.1. Display Settings
- Brightness Control: Press Utility, then Display, then Backlight. Use the knob or numeric keypad to adjust brightness (0% to 100%).
- Screen Saver: If no operation occurs within the set time, the screen enters protection mode. Press Utility, then Display, then ScrSaver to select On or Off. If On, set the screen saver time (1 to 999 minutes) using the knob or numeric keypad.
- Separator: Set the separator for screen display data (Comma, Space, or Nothing) via Utility > Display > Separator.
4.9.2. Channel Settings (Load Setting)
The generator has a 50Ω fixed serial output impedance. If the actual load does not match, the displayed voltage level may be inaccurate. This function matches the displayed voltage with the expected one.
To set the load value:
- Press Utility, then Channel Set.
- Press Load to select High Z or * ohm (default 50Ω).
- To change the load value, select * ohm, then use the knob or numeric keypad to adjust the value (1Ω to 10kΩ).
Warning: Each output has a fixed 50Ω impedance. If the actual load differs from the specified value, the displayed voltage level will not match the actual level.
4.9.3. System Settings
- Language: Press Utility, then System, then Language to switch the display language.
- Beeper: Press Utility, then System, then Beeper to toggle On or Off.
- USB Device Type: Press Utility > System > USB Dev to toggle between PC (internal communication protocol for Waveform Editor) and USBTMC (USBTMC communication protocol standard).
- Restore to Factory Setting: Press Utility > System > NextPage > Factory Set, then Ok to restore default settings.
4.9.4. Firmware Update
Use the rear-panel USB port to update firmware with a USB memory device.
Caution: Do not power off or remove the USB device during update to prevent damage.
To update firmware:
- Press Utility > System > NextPage > Upgrade. A prompt message will appear.
- Press Ok. The external disk icon will display on the PC.
- Obtain firmware upgrade packages (e.g., "AG.upp" file) from sales/service, copy to the displayed external disk.
- Restart the device. The upgrade status will display.
- After upgrade, the instrument will shut down automatically.
- Press the power button to turn on. Press Utility to check the version.
5. Communication with PC
The signal generator supports communication with a computer via a USB port. The Waveform Editor software allows controlling the output and writing files to the generator. Instrument settings can be saved as files (up to 16 settings).
Note: Obtain the Waveform Editor communication software from the official website and install it.
5.1. Install Communication Software
- Set USB device protocol type: Press Utility > System > USB Dev, switch to PC.
- Connection: Connect the USB Device interface on the rear panel to the computer's USB interface with a USB cable.
- Install the driver: Run Waveform Editor software. Follow instructions to install the driver, located in the USBDRV folder (e.g., "C:\Program Files (x86)\DS_Wave\Waveform Editor\USBDRV").
- Host computer communication port setting: Open Waveform Editor, click "Communications" > "Ports-Settings", and select "USB" as the communication port. A green connection status in the lower right corner indicates success.
6. Troubleshooting
6.1. No Display After Power On
If the screen remains black after pressing the power switch:
- Check if the power is connected correctly.
- Verify the voltage selector is in the correct position.
- Check the fuse below the power connector (pry open with a flat-blade screwdriver) to ensure it's the specified type, rating, and in good condition.
- Restart the instrument after these checks.
If the problem persists, contact customer service.
6.2. Output Signal Amplitude Mismatch
If the measured output signal amplitude does not match the displayed value:
- Check if the actual load value of the signal is consistent with the load value set in the instrument (refer to Section 4.9.2, "Channel Settings (Load Setting)").
If other problems occur, try restarting the instrument. If it still doesn't work properly, contact customer service.
7. Specifications
All technical specifications are guaranteed under specified operating conditions (20°C to 30°C, continuous operation for >30 minutes, within calibration interval, and after self-calibration), unless otherwise stated. Specifications marked "Typical" are not guaranteed.

7.1. Waveforms
| Category | Description |
|---|---|
| Bandwidth (DGE1030) | 30 MHz |
| Bandwidth (DGE1060) | 60 MHz |
| Sample Rate (DGE1030) | 125 MSa/s |
| Sample Rate (DGE1060) | 300 MSa/s |
| Vertical Resolution | 14 bits |
| Channel | 1 |
| Standard Waveforms | Sine wave, square wave, ramp wave, pulse wave, noise |
| Arbitrary Waveforms | Sinc, exponential rise, exponential decline, electrocardiogram, Gaussian, semi-positive, Lorentz, dual audio, DC voltage totaling more than 160 kinds |
7.2. Frequency Characteristics
| Waveform | Frequency Range |
|---|---|
| Sine (DGE1030) | 1 μHz ~ 30MHz |
| Sine (DGE1060) | 1 μHz ~ 60MHz |
| Square (DGE1030) | 1 μHz ~ 15MHz |
| Square (DGE1060) | 1 μHz ~ 20MHz |
| Pulse (DGE1030) | 1 μHz ~ 15MHz |
| Pulse (DGE1060) | 1 μHz ~ 20MHz |
| Ramp (DGE1030) | 1 μHz ~ 1MHz |
| Ramp (DGE1060) | 1 μHz ~ 2MHz |
| Noise wave (-3 dB) | 20 MHz BW(AWGN) |
| Arbitrary wave | 1 μHz ~ 10 MHz |
| Frequency resolution | 1 μHz or 7 significant figures |
| Frequency stability | ±30 ppm at 0±40°C |
| Frequency aging rate | ±30 ppm per year |
7.3. Amplitude Characteristics
| Parameter | Specification |
|---|---|
| Output amplitude (DGE1030, High Z) | 2mVpp ~ 20Vpp (≤ 10MHz), 2mVpp ~ 8Vpp (≤ 30MHz) |
| Output amplitude (DGE1030, 50Ω) | 1mVpp ~ 10Vpp (≤ 10MHz), 1mVpp ~ 4Vpp (≤ 30MHz) |
| Output amplitude (DGE1060, High Z) | 2mVpp ~ 20Vpp (≤ 10MHz), 2mVpp ~ 8Vpp (≤ 60MHz) |
| Output amplitude (DGE1060, 50Ω) | 1mVpp ~ 10Vpp (≤ 10MHz), 1mVpp ~ 4Vpp (≤ 60MHz) |
| Amplitude accuracy | ± (1% of setting + 1 mVpp) (Typical 1kHz sine, 0V offset) |
| Amplitude resolution | 1mVpp or 4 digits |
| DC offset range (DGE1030, High Z) | ±(10 Vpk - Amplitude Vpp/2) (≤ 10MHz), ±(4Vpk - Amplitude Vpp/2) (≤ 30MHz) |
| DC offset range (DGE1030, 50Ω) | ±(5 Vpk - Amplitude Vpp/2) (≤ 10MHz), ±(2 Vpk - Amplitude Vpp/2) (≤ 30MHz) |
| DC offset range (DGE1060, High Z) | ±(10Vpk - Amplitude Vpp/2) (≤ 10MHz), ±(4Vpk-Amplitude Vpp/2) (≤ 60MHz) |
| DC offset range (DGE1060, 50Ω) | ±(5 Vpk - Amplitude Vpp/2) (≤ 10MHz), ±(2Vpk-Amplitude Vpp/2) (≤ 60MHz) |
| DC offset accuracy | ± (1% of setting| + 1 mV + amplitude Vpp * 0.5%) |
| Offset resolution | 1 mVpp or 4 digits |
| Output Impedance | 50Ω (Typical) |
7.4. Signal Characteristics
| Parameter | Specification |
|---|---|
| Sine Bandwidth flatness (≤10MHz) | ±0.3dB |
| Sine Bandwidth flatness (<35MHz) | ±0.5dB |
| Sine Bandwidth flatness (≤60MHz) | ±1dB |
| Sine Harmonic distortion (DC to 1MHz) | <-65dBc |
| Sine Harmonic distortion (1MHz to 35MHz) | <-60dBc |
| Sine Harmonic distortion (35MHz to 60MHz) | <-50dBc |
| Sine Total harmonic distortion | < 0.2%, 10 Hz to 20 kHz, 1 Vpp |
| Sine Non-harmonic distortion (≤10MHz) | <-70dBc |
| Sine Non-harmonic distortion (>10MHz) | <-70dBc + 6dB/ sound interval |
| Square Rise/fall time | < 20ns |
| Square Jitter (rms), typical (1Vpp, 50Ω) | 200ps + 30ppm |
| Square Overshoot | < 5% |
| Ramp Linearity | < 1% of peak output (typical 1 kHz, 1 Vpp, symmetry 50%) |
| Ramp Symmetry | 0% to 100% |
| Pulse Period (DGE1030) | 67 ns to 1 Ms |
| Pulse Period (DGE1060) | 50 ns to 1 Ms |
| Pulse Width | ≥ 24ns |
| Pulse Rise and fall time | ≥ 15ns |
| Pulse Overshoot | < 5% |
| Pulse Jitter (rms), typical (1Vpp, 50Ω) | 200ps + 30ppm |
| Noise Types | Gaussian white noise |
| Noise Bandwidth (-3dB) | 20 M |
| Arbitrary wave Bandwidth | 10M |
| Arbitrary wave Waveform length | 2 to 8K points |
| Arbitrary wave Sampling rate (DGE1030) | 125 MSa/s |
| Arbitrary wave Sampling rate (DGE1060) | 300 MSa/s |
| Arbitrary wave Amplitude accuracy | 14 bits |
7.5. Modulation Characteristics
| Modulation Type | Parameters |
|---|---|
| AM, FM, PM, FSK | |
| AM Carrier | Sine wave, square wave, ramp wave, arbitrary wave (except DC) |
| AM Modulated signal source | Internal |
| AM Internal modulation waveform | Sine wave, square wave, ramp wave, white noise |
| AM Internal amplitude modulation frequency | 2 mHz to 100 kHz |
| AM Depth | 0.0% to 100.0% |
| FM Carrier | Sine wave, square wave, ramp wave, arbitrary wave (except DC) |
| FM Modulated signal source | Internal |
| FM Internal modulation waveform | Sine wave, square wave, ramp wave, white noise |
| FM Internal modulation frequency | 2 mHz to 100 kHz |
| FM Frequency offset | 1 µHz ≤ offset < carrier frequency |
| PM Carrier | Sine wave, square wave, ramp wave, arbitrary wave (except DC) |
| PM Modulated signal source | Internal |
| PM Internal modulation waveform | Sine, square, ramp, white noise |
| PM Internal phase modulation frequency | 2 mHz to 100 kHz |
| PM Phase deviation range | 0° to 180° |
| FSK Carrier | Sine wave, square wave, ramp wave, arbitrary wave (Except DC) |
| FSK Modulated signal source | Internal |
| FSK Internal modulation waveform | 50% square wave |
| FSK rate | 2 mHz to 1MHz |
| FSK hopfreq | 2 mHz ≤ offset < maximum frequency of corresponding carrier |
7.6. Sweep Characteristics
| Parameter | Specification |
|---|---|
| Carrier | Sine wave, square wave, ramp wave, arbitrary wave (Except DC) |
| Minimum/maximum starting frequency | 1 µHz(minimum)/ maximum frequency of corresponding carrier |
| Minimum/maximum termination frequency | 1 μHz(minimum)/ maximum frequency of corresponding carrier |
| Types | Linear, logarithmic |
| Sweep time | 1 ms to 500 s ± 0.1% |
| Trigger source | Internal, manual |
7.7. Burst Characteristics
| Parameter | Specification |
|---|---|
| Waveform | Sine wave, square wave, ramp wave, pulse wave and arbitrary wave (Except DC) |
| Types | N-cycle |
| N-cycle trigger source | Internal, manual |
| Carrier frequency | 1 µHz ≤ Offset ≤ Maximum frequency of corresponding carrier /2 |
| N-cycle trigger cycle (DGE1030) | 34 ns – 1 Ms (Min = Cycles * Period) |
| N-cycle trigger cycle (DGE1060) | 17 ns – 1 Ms (Min = Cycles * Period) |
| Periodicity | 1 ~ 60000 (Max = Burst Period / Period) / infinite |
7.8. General Specifications
| Parameter | Specification |
|---|---|
| Display type | 3.6-inch color LCD display |
| Display resolution | 480 Horizontal x 272 Vertical pixels |
| Display color | 65536 colors, 16 bits, TFT |
| Voltage | DC 5V / 2A |
| Power consumption | Less than 10W |
| Working temperature | 0°C to 40°C |
| Storage temperature | -20°C to 60°C |
| Relative humidity (<35°C) | ≤ 90% relative humidity |
| Relative humidity (35°C to 40°C) | ≤ 60% relative humidity |
| Operating height | 3,000 meters |
| Non-operation height | 12,000 meters |
| Dimension (L x H x W) | 200mm x 68.5mm x 73.6mm |
| Weight | Approx. 0.5 kg |
| Recommended calibration interval | One year |
8. Maintenance & Care
8.1. General Maintenance
- Do not store or leave the instrument where the liquid crystal display will be exposed to direct sunlight for long periods.
Caution: To avoid damage, do not expose the instrument or probe to sprays, liquids, or solvents.
8.2. Cleaning
Inspect the instrument and probes as often as operating conditions require. To clean the exterior:
- Wipe dust from the instrument and probe surface with a soft cloth. Do not scuff the transparent LCD protection screen.
- Disconnect power before cleaning. Clean with a wet soft cloth (not dripping water). Use soft detergent or fresh water. Avoid corrosive chemical cleaning agents.
Warning: Before powering on again, ensure the instrument is completely dry to avoid electrical short circuits or injury from moisture.
9. Accessories
The following accessories are included with your OWON DGE1000 Series Arbitrary Waveform Generator:
- 1 x BNC to alligator cable
- 1 x USB cable
- 1 x USB to DC cable
- 1 x Power adapter
- 1 x Quick Guide
- 1 x Packaging

10. Warranty & Support
10.1. General Warranty
The product is warranted to be free from defects in materials and workmanship for a period of 2 years from the date of purchase by the original purchaser. Accessories (probes, battery) are warranted for 12 months. This warranty is non-transferable.
If the product is defective during the warranty period, we will repair or replace it without charge. Replaced parts, modules, and products become our property.
To obtain service, notify our company before the warranty expires. You are responsible for packaging and shipping the defective product to the designated service center, along with proof of purchase.
This warranty does not cover defects, failures, or damage caused by improper use, inadequate maintenance, or care. We are not obligated to provide service for damage resulting from unauthorized repairs, improper connection to incompatible equipment, use of non-approved supplies, or modifications that increase servicing difficulty.
Please contact the nearest Sales and Service Offices for services.
Except for the after-sales services provided, we offer no other guarantees, including implied guarantees for marketability and special-purpose acceptability. We are not responsible for any indirect, special, or consequential damages.
10.2. Product Support
For product support, visit: www.owon.com.hk/download
Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
No. 19, Heming Road
Lantian Industrial Zone, Zhangzhou 363005 P.R. China
Tel: +86-596-2130430
Fax: +86-596-2109272
Web: www.owon.com E-mail: info@owon.com.cn
11. User Tips
- Grounding is Crucial: Always ensure the device is properly grounded before connecting any inputs or outputs to prevent electric shock and ensure accurate measurements.
- Software for Custom Waveforms: Utilize the Waveform Editor PC software for creating and managing custom arbitrary waveforms. This expands the device's versatility significantly.
- Check Load Settings: If your measured amplitude differs from the displayed value, verify that the load impedance setting in the device matches your actual circuit load.
- Ventilation: Keep the device in a well-ventilated area to prevent overheating, especially during prolonged use. Regularly check air vents for obstructions.
- Firmware Updates: Periodically check the manufacturer's website for firmware updates to ensure optimal performance and access to new features. Follow the update instructions carefully.