1. Product Overview
The 5800H handheld ultrasonic oxygen detector utilizes the principle of ultrasonic waves to accurately measure the concentration, flow rate, and pressure of oxygen in gas. It is specifically designed for use with PSA oxygen generators and concentrators. The device integrates a high-precision pressure sensor to detect the oxygen outlet pressure.
This 3-in-1 gas meter is a professional tool for simultaneously measuring oxygen concentration, flow rate, and pressure, making it ideal for both medical and industrial oxygen systems.

2. Package Contents
Please check the package contents upon receipt to ensure all items are present and undamaged:
- 5800H Oxygen Analyzer (Main Unit) ×1 (with built-in 2000mAh battery)
- USB Cable (Type-C) ×1
- Hose ×1 (silicone tube, 280mm long, 10mm outer diameter, 8mm inner diameter)
- Packing Box ×1
- Packaging Bag (Storage Bag) ×1
- User Manual ×1
- Warranty Card ×1

3. Setup
3.1 Component Identification

- Flow and Oxygen Content Input Port: Used for connecting the hose to measure oxygen concentration and flow.
- Pressure Test Input Port: Used for connecting the hose to measure pressure.
- LCD Backlit Screen: Displays measurement readings.
- ON/OFF Button: Powers the device on or off.
- HOLD Button: Toggles data retention.
- Backlight Button: Toggles the screen backlight.
3.2 Initial Charging
The device is equipped with a 2000mAh lithium battery. When the battery symbol appears on the display, the battery needs to be charged.
- Connect the supplied Type-C USB charging cable to a DC-5V power adapter (not included).
- Connect the other end of the Type-C USB charging cable to the charging port on the meter.
- The red LED indicator will light up, indicating that the battery is charging.
- The green LED indicator will light up when charging is complete.

3.3 Pre-Measurement Preparation
Important: When using the detector with an oxygen generator, please ensure to drain any water from the oxygen generator's humidification bottle or remove the humidification bottle entirely. This prevents water from entering the detector and causing damage to the instrument.

4. Operating Instructions
4.1 Power On/Off
- Press the ON button to turn on the device. The LCD display will light up and show the initial readings.
- Press the OFF button to shut down the device.
4.2 Measuring Flow Rate and Oxygen Concentration
- Insert the provided hose into the "Conc/flow" input port on the top of the detector.
- Connect the other end of the hose to the oxygen source (e.g., PSA oxygen generator outlet).
- Press the ON button to start measuring. The display will show the oxygen purity and flow rate.
4.3 Measuring Pressure
- Insert the provided hose into the "Pressure" input port on the top of the detector.
- Connect the other end of the hose to the point where pressure needs to be measured.
- Press the ON button to start measuring. The display will show the pressure reading.
4.4 Backlight Function
- Press the backlight button (light bulb icon) to turn on the LCD backlight.
- Press the backlight button again to turn off the backlight.
4.5 Data Hold Function
- Press the HOLD button to freeze the current data on the display. An 'H' symbol will appear on the screen.
- Press the HOLD button again to release the data hold and resume live measurements.
4.6 Automatic Shutdown
In the oxygen measurement state, if there is no operation or data input for 5 minutes, the device will automatically shut down to conserve battery. This also applies to the stress test state.
5. Specifications
| Measurement | Range | Resolution | Accuracy | Unit |
|---|---|---|---|---|
| Oxygen Concentration (O2) | 21% to 95.6% | 0.1% | ±1.5% | % |
| Flow Rate (F) | 0 to 10.0 L/min | 0.1 L/min | ±5% F.S (±0.2 L/min) (whichever is greater) | L/min |
| Pressure (P) | 0-199 KPa | 1 KPa | ±1% F.S | KPa |
General Specifications
- Battery: 3.7V 2000mAh lithium battery (included)
- Charging: Type-C USB
- Display: LCD with Backlight
- Auto Shutdown: After 5 minutes of no operation (in oxygen measurement or stress test state)
- Dimensions: 96mm (W) × 202mm (H) × 47mm (D)
- Weight: 350g
- DIY Supplies: Electrical
- Gas Type: Oxygen
- Smart Device: No
- Origin: Mainland China

6. Maintenance
- Handling: During use, handle the instrument with care and avoid collision with other objects to prevent damage and maintain detection accuracy.
- Repair: Do not attempt to disassemble or repair the instrument without proper authorization. If the instrument fails, contact the manufacturer or authorized service center for assistance.
- Cleaning: Keep the device clean and free from dust. Use a soft, dry cloth for cleaning.
7. Troubleshooting
7.1 Common Issues and Solutions
- Inaccurate Readings: Ensure the hose is properly connected to the correct input port and that there are no leaks. Verify the oxygen source is stable.
- Device Not Powering On: Check the battery level. If the battery symbol is displayed, charge the device using the Type-C USB cable.
- Water Ingress: If water enters the detector (e.g., from a humidification bottle), immediately power off the device and contact customer support. Do not attempt to use it.
7.2 Environmental Considerations
- Fire Sources: Always keep the detector away from open flames or high heat sources.
- Dust: While the device is designed for general use, excessive dust can affect performance. Ensure the environment is reasonably clean.
8. User Tips
- Prevent Water Damage: Always remember to drain the humidification bottle of your oxygen generator or remove it directly before connecting the detector to prevent water from entering and damaging the instrument.
- Welding Applications: This device can be used in applications such as stainless steel welding to measure the percentage of oxygen during the purging process, ensuring optimal welding conditions.
- Battery Life: The 2000mAh battery provides approximately 12 hours of operation. Plan charging accordingly for extended use.
- Data Recording: Utilize the data hold function to easily record stable readings, especially in dynamic environments.
9. Warranty and Support
For any problems or concerns regarding your BTMETER 5800H Portable Ultrasonic Oxygen Detector, please contact our customer support team for prompt assistance.
Email: AFTERSALES1010@HOTMAIL.COM
When contacting support, please provide the following information to ensure a quick resolution:
- Your Order Number
- Platform of Your Purchase
- Full Model Number (5800H)
- Description of the Problem (Attaching videos or photos can help us troubleshoot the problems even faster)
Please refer to your Warranty Card for specific warranty terms and conditions.