Contents

COXO CP-1 Dental Air Polisher User Manual

Model: CP-1 | Brand: COXO

1. Introduction

The COXO CP-1 Dental Air Polisher is a versatile 2-in-1 device designed for professional dental use, offering both supragingival and subgingival modes for comprehensive periodontal treatment. It provides comfortable and clean operation with an effective sandblasting effect for tooth cleaning, polishing, and whitening.

COXO CP-1 Dental Air Polisher, 2 in 1 design
Figure 1: COXO CP-1 Dental Air Polisher, featuring a 2-in-1 design for versatile use.
CP-1 Dental Air Polisher features: comfortable, clean, better sandblasting effect, supragingival & subgingival mode Two-in-one
Figure 2: Key features of the CP-1 Dental Air Polisher, highlighting comfort, cleanliness, improved sandblasting, and dual-mode functionality.

2. Safety Precautions

⚠ Warning: Read the following information carefully before using this product.

  • This item may be intended for use by professional health care providers only. The seller may ask you to provide evidence that you are a qualified health care provider and are entitled to buy and use certain items in your country.
  • The instrument must only be used in hospital environments, clinics, or dental offices, by qualified practitioners.
  • In case of failure, please contact professionals for maintenance.
  • Please stop using the product in case of obvious aging or abnormal phenomenon.
  • Please ensure regular testing and maintenance. Do not use if testing and maintenance are not performed.
  • A connector that matches the connection part of the sandblasting scaler must be used; otherwise, it will cause failure of the sandblasting scaler and the connection port.
  • Confirm that the sandblasting scaler is connected to the handpiece tube before use. If connected incorrectly, failure to fix will cause the connection to burst due to the pressure of compressed air, resulting in accidents.
  • Please do not screw the nuts on the connections while the sandblasting scaler is in operation, as these may cause the connections to burst.
  • Adequate use of cooling water is necessary to cool the nozzle and tooth during tooth cleaning, and insufficient supply of water can cause damage to the tooth surface.
  • Always be careful to prevent scattered powder from entering the patient's eyes and nose.
  • Shake before use to eliminate powder agglomeration.
  • Please clean the sandblasted tooth powder left in the equipment powder box after use.
  • Tighten the bottle cap after use to prevent powder deliquescence; it is not suitable to store in high temperature and high humidity environment.
  • The subgingival nozzle is provided in non-sterile state for single use and shall be cleaned by alcohol prior to use.
  • Use evacuating suction equipment to absorb the air or sand powder mixture ejected from the treatment during treatment.

⚠ Caution:

  • Do not use the subgingival handpiece with the supragingival sand, which can cause the sandblasting pipe to be blocked.
  • Do not spray the powder directly into the soft tissues of the mouth, gums, or under the tongue.
  • In order to avoid infection, wear a mask, gloves, and protective glasses during treatment.
  • To prevent the patient from swallowing the powder, rinse the patient frequently if necessary.
  • In order to prevent the backflow of water into the powder box, please first vent and then drain the water during operation.

3. Intended Use & Contraindications

Intended Use:

Used for removing tartar, lentigo, or polishing tooth surface adhered on tooth surface.

Contraindications:

  • Do not use this device on patients with hemophilia; do not use this device on patients who are allergic to sandblasted tooth cleaning powder.
  • Do not use this device on patients with oral mucosal inflammation and gingival bleeding.
  • Use with caution for pregnant women and children.
  • Do not use this device on patients suffering from asthma, chronic bronchitis, and other respiratory diseases.

4. Product Overview

The COXO CP-1 Dental Air Polisher consists of a main handpiece, a powder bottle, and interchangeable nozzles for supragingival and subgingival applications.

Hand holding the COXO CP-1 Dental Air Polisher, demonstrating ergonomic design.
Figure 3: The ergonomic design of the COXO CP-1 allows for comfortable and easy operation.
Diagram showing supragingival and subgingival modes with Prophy and Perio tips and sandblasting examples.
Figure 4: The two-in-one design allows for both supragingival (Prophy) and subgingival (Perio) air polishing, enabling comprehensive periodontal treatment.
Close-up of the COXO CP-1 tip showing 360 degree rotation for easy cleaning and maintenance.
Figure 5: The tip features 360° rotation, making it removable for easy cleaning and maintenance.
Exploded view of the COXO CP-1 showing sectional design for easy loading, unloading, cleaning, and maintenance, and highlighting high temperature & pressure disinfection.
Figure 6: Sectional design allows for simple loading, unloading, and easy cleaning. The device is designed for high temperature and pressure disinfection and sterilization for enhanced safety and cleanliness.
Close-up of the small nozzle design with a 0.7mm caliber for concentrated powder output and accurate scaling.
Figure 7: The small nozzle design features a 0.7mm caliber for more concentrated powder output and accurate scaling.

5. Setup and Assembly

5.1. Attaching the Unit to the Handpiece Hose

  1. Align the hose connector of the sandblasting scaler with the handpiece hose interface of the comprehensive treatment table.
  2. Insert and tighten the nut.

⚠ Caution:

Fully dry the hose interface of the sandblasting scaler and the hose interface of the handpiece before connecting, so as not to block the nozzle due to damp sand powder during use.

5.2. Disassembling and Assembling the Handpiece

  1. Connection: Insert the handpiece alignment into the body unit until it clicks.
  2. Disconnecting: Pull it straight out.
Diagram showing how to connect and disconnect the handpiece.
Figure 8: Illustration of handpiece assembly and disassembly.

⚠ Caution:

Thoroughly dry the interior of the handpiece and connection site prior to connection, so as not to block the nozzle due to damp sand powder during use.

5.3. Subgingival Nozzle Disassembly and Assembly

  1. Installation: When installing the subgingival nozzle, the head needs to be oriented to the nozzle D-shaped head.
  2. Disassembly: Use the disassembling wrench to hook up the nozzle and lock it down, then force it outward to pry it out.
Diagram showing subgingival nozzle disassembly using a wrench.
Figure 9: Steps for subgingival nozzle disassembly.

5.4. Powder Loading

  1. Unscrew and remove the upper cover of the powder bottle and add powder to it.
  2. The maximum amount of powder loaded shall not exceed the MAX scale line. Clean up any powder at the thread position of the powder bottle cover, and then tighten the upper cover.
Diagram showing the powder bottle with MAX fill line.
Figure 10: Powder bottle with MAX fill line indicator.

6. Operation

6.1. Preparation for Use

  1. Adjust the corresponding pressure regulating valve of the dental unit to the appropriate working pressure: 250kPa - 450kPa.
  2. Adjust the water to a suitable atomization state.

⚠ Caution:

Ensure that the dental air polisher is properly installed.

6.2. Sandblasting

6.2.1. Supragingival Sandblasting

  1. Use the supragingival handpiece to spray in the extraoral container for 1-3 seconds in advance to ensure that the air and water can be sprayed evenly before treating the patient.
  2. Usually hold the head in a pen holding position.
  3. When the dental air polisher is used for normal sandblasting and tooth cleaning, it is recommended to maintain a distance of 3-5 mm between the air outlet of the nozzle and the tooth surface, and it is recommended that the sandblasting direction is at an included angle of 30-60° with the tooth surface.
Diagram illustrating the recommended 3-5mm distance and 30-60 degree angle for supragingival sandblasting.
Figure 11: Recommended distance (3-5mm) and angle (30-60°) for effective supragingival sandblasting.

6.2.2. Subgingival Blasting

  1. For nozzle installation, refer to "5.3. Subgingival Nozzle Disassembly and Assembly".
  2. Use the subgingival handpiece to spray in the extraoral container for 1-3 seconds in advance, and ensure that the air and water are evenly sprayed before treating the patient.
  3. Usually hold the head in a pen holding position.
  4. When normal subgingival sandblasting is performed using the subgingival handpiece, it is recommended to use a nozzle for plaque removal from the subgingival 4-9 mm periodontal pockets and lift up and down during use.
  5. It is recommended that each periodontal pocket be sandblasted for no more than 5 seconds.
Diagram illustrating vertical movements for subgingival blasting in 4-9mm periodontal pockets.
Figure 12: Vertical movements for subgingival blasting in 4-9mm periodontal pockets.

7. Maintenance

7.1. Cleaning the Powder

  1. Open the powder cavity cover and clean up the remaining powder cavity.
  2. Use air to blow off the powder remaining in the powder box.
  3. Use alcohol cotton or a dry towel to wipe off the powder residue on the surface of the sandblasting scaler.

7.2. Cleaning the Nozzle and Handle

  1. Remove the handpiece, unscrew the nozzle, and pull out the nozzle.
  2. Use an air gun to aim at the nozzle and blow off the powder in the nozzle.
  3. Use an air gun to blow the front and rear ends of the handpiece to clean the residual moisture on the handpiece.
  4. If the nozzle is blocked, use a needle to unclog it.

7.3. Daily Maintenance

Pay attention to daily inspection of the handpiece for cracking, deformation, or falling off and other defects to ensure that the handpiece can be used normally. Pay attention to the inspection and cleaning of the sealing ring to ensure that the sealing ring is complete and clean; otherwise, it will affect the product sealing and the normal operation.

  • Keep the sandblasting scaler dry. Rain, moisture, and various liquids may dampen the powder and affect normal use.
  • Do not throw, beat, or vibrate the sandblasting scaler. Rough use of the sandblasting scaler will damage the strength and sealing of the equipment.
  • Do not use paint to smear the instrument; smearing will leave debris in the detachable parts, which will affect the normal operation.
  • Do not use a needle to poke the connector end of the tail wire, which may affect the normal use of the device.

7.4. Disinfection and Sterilization

This item is a hand instrument. Sterilization is recommended using high temperature and high pressure sterilization.

Sterilization Information
Sterilizable ComponentsHandpiece
Sterilization MethodRecommended use of high temperature and high pressure sterilization
Sterilization Conditions134°C, 5Min

Please use the sterilization bag which meets the national regulations for packaging before sterilization.

⚠ Caution:

Except for the handpiece, the other components cannot be sterilized under high temperature and high pressure.

8. Troubleshooting

Troubleshooting Guide
Fault PhenomenonFailure CauseResolution
No air or powder from nozzleClogged NozzleIn the absence of sandblasted powder in the sand powder bottle, remove the handpiece and clean the nozzle with a needle.
Clogged powder thru holeClean powder thru hole in sand powder bottle with needle.
Too little powder and air ejectionFeed water of sand powder bottleReplace the O-ring on the upper cover of sand powder bottle.
Air path in handpiece is blockedUnloading the handpiece without sand powder bottle and cleaning the air passage of the handpiece with a needle.
Water leakage at handpieceDamaged the O-ring connectionReplace the O-ring.

9. Specifications

9.1. Technical Specifications

COXO CP-1 Technical Data
ParameterValue
Input Air Pressure250kPa - 450kPa
Input Water Pressure70kPa - 220kPa
Spray Water Flow>30ml/min
Gas Flow<40NL/min
Noise<68dB
Product Weight0.145kg
Nozzle Caliber (Powder)0.7mm
Nozzle Caliber (Water)1.4mm
MaterialMetal
Commodity Quality CertificationCE, 3C

9.2. Operating Environment and Storage Conditions

Environmental Conditions
ConditionOperating TemperatureOperating HumidityAtmospheric Pressure
Operating+5°C ~ +40°C20% RH ~ 80% RH86kPa ~ 106kPa
ConditionStorage TemperatureStorage HumidityAtmospheric Pressure
Storage-10°C ~ +55°C<93%RH50kPa ~ 106kPa

10. User Tips

  • Replacement Tips: For replacement Perio Flow tips or other parts, it is recommended to contact the manufacturer or authorized seller to ensure compatibility and quality.
  • Connection Type: This device is an air-driven polisher and connects to a dental unit's air and water lines, not an electrical cable.
  • Polishing Powder: Always use specialized dental sandblasting powder (prophy powder) designed for dental use. Do not use household products like baking soda, as this can damage the device and be ineffective or harmful for dental treatment.

11. Warranty and Support

11.1. Guarantee

The Dental Scaler and Air Polisher has a 12-month warranty.

This product (such as through hole, replacement of sealing device, etc.) can be repaired by professional operators or nursing staff. If damage to structural components is found, please contact the after-sales service department of the company for repair.

11.2. Manufacturer Contact

Foshan COXO Medical Instrument Co., Ltd
BLDG 4, District A, Guangdong New Light Source Industrial Base, South of Luocun Avenue Nanhai District Foshan 528226 Guangdong China
Web: www.coxotec.com

For further details, you may refer to the official user manual PDF: View User Manual (PDF)

12. Recycling and Disposal

Dispose of waste equipment in accordance with national regulations and standards to ensure that all parts are not contaminated during the disposal process.

© 2022 COXO. All rights reserved.