1. Introduction
The COXO CP-1 Dental Air Polisher is a versatile 2-in-1 device designed for professional dental use, offering both supragingival and subgingival modes for comprehensive periodontal treatment. It provides comfortable and clean operation with an effective sandblasting effect for tooth cleaning, polishing, and whitening.


2. Safety Precautions
⚠ Warning: Read the following information carefully before using this product.
- This item may be intended for use by professional health care providers only. The seller may ask you to provide evidence that you are a qualified health care provider and are entitled to buy and use certain items in your country.
- The instrument must only be used in hospital environments, clinics, or dental offices, by qualified practitioners.
- In case of failure, please contact professionals for maintenance.
- Please stop using the product in case of obvious aging or abnormal phenomenon.
- Please ensure regular testing and maintenance. Do not use if testing and maintenance are not performed.
- A connector that matches the connection part of the sandblasting scaler must be used; otherwise, it will cause failure of the sandblasting scaler and the connection port.
- Confirm that the sandblasting scaler is connected to the handpiece tube before use. If connected incorrectly, failure to fix will cause the connection to burst due to the pressure of compressed air, resulting in accidents.
- Please do not screw the nuts on the connections while the sandblasting scaler is in operation, as these may cause the connections to burst.
- Adequate use of cooling water is necessary to cool the nozzle and tooth during tooth cleaning, and insufficient supply of water can cause damage to the tooth surface.
- Always be careful to prevent scattered powder from entering the patient's eyes and nose.
- Shake before use to eliminate powder agglomeration.
- Please clean the sandblasted tooth powder left in the equipment powder box after use.
- Tighten the bottle cap after use to prevent powder deliquescence; it is not suitable to store in high temperature and high humidity environment.
- The subgingival nozzle is provided in non-sterile state for single use and shall be cleaned by alcohol prior to use.
- Use evacuating suction equipment to absorb the air or sand powder mixture ejected from the treatment during treatment.
⚠ Caution:
- Do not use the subgingival handpiece with the supragingival sand, which can cause the sandblasting pipe to be blocked.
- Do not spray the powder directly into the soft tissues of the mouth, gums, or under the tongue.
- In order to avoid infection, wear a mask, gloves, and protective glasses during treatment.
- To prevent the patient from swallowing the powder, rinse the patient frequently if necessary.
- In order to prevent the backflow of water into the powder box, please first vent and then drain the water during operation.
3. Intended Use & Contraindications
Intended Use:
Used for removing tartar, lentigo, or polishing tooth surface adhered on tooth surface.
Contraindications:
- Do not use this device on patients with hemophilia; do not use this device on patients who are allergic to sandblasted tooth cleaning powder.
- Do not use this device on patients with oral mucosal inflammation and gingival bleeding.
- Use with caution for pregnant women and children.
- Do not use this device on patients suffering from asthma, chronic bronchitis, and other respiratory diseases.
4. Product Overview
The COXO CP-1 Dental Air Polisher consists of a main handpiece, a powder bottle, and interchangeable nozzles for supragingival and subgingival applications.





5. Setup and Assembly
5.1. Attaching the Unit to the Handpiece Hose
- Align the hose connector of the sandblasting scaler with the handpiece hose interface of the comprehensive treatment table.
- Insert and tighten the nut.
⚠ Caution:
Fully dry the hose interface of the sandblasting scaler and the hose interface of the handpiece before connecting, so as not to block the nozzle due to damp sand powder during use.
5.2. Disassembling and Assembling the Handpiece
- Connection: Insert the handpiece alignment into the body unit until it clicks.
- Disconnecting: Pull it straight out.

⚠ Caution:
Thoroughly dry the interior of the handpiece and connection site prior to connection, so as not to block the nozzle due to damp sand powder during use.
5.3. Subgingival Nozzle Disassembly and Assembly
- Installation: When installing the subgingival nozzle, the head needs to be oriented to the nozzle D-shaped head.
- Disassembly: Use the disassembling wrench to hook up the nozzle and lock it down, then force it outward to pry it out.

5.4. Powder Loading
- Unscrew and remove the upper cover of the powder bottle and add powder to it.
- The maximum amount of powder loaded shall not exceed the MAX scale line. Clean up any powder at the thread position of the powder bottle cover, and then tighten the upper cover.

6. Operation
6.1. Preparation for Use
- Adjust the corresponding pressure regulating valve of the dental unit to the appropriate working pressure: 250kPa - 450kPa.
- Adjust the water to a suitable atomization state.
⚠ Caution:
Ensure that the dental air polisher is properly installed.
6.2. Sandblasting
6.2.1. Supragingival Sandblasting
- Use the supragingival handpiece to spray in the extraoral container for 1-3 seconds in advance to ensure that the air and water can be sprayed evenly before treating the patient.
- Usually hold the head in a pen holding position.
- When the dental air polisher is used for normal sandblasting and tooth cleaning, it is recommended to maintain a distance of 3-5 mm between the air outlet of the nozzle and the tooth surface, and it is recommended that the sandblasting direction is at an included angle of 30-60° with the tooth surface.

6.2.2. Subgingival Blasting
- For nozzle installation, refer to "5.3. Subgingival Nozzle Disassembly and Assembly".
- Use the subgingival handpiece to spray in the extraoral container for 1-3 seconds in advance, and ensure that the air and water are evenly sprayed before treating the patient.
- Usually hold the head in a pen holding position.
- When normal subgingival sandblasting is performed using the subgingival handpiece, it is recommended to use a nozzle for plaque removal from the subgingival 4-9 mm periodontal pockets and lift up and down during use.
- It is recommended that each periodontal pocket be sandblasted for no more than 5 seconds.

7. Maintenance
7.1. Cleaning the Powder
- Open the powder cavity cover and clean up the remaining powder cavity.
- Use air to blow off the powder remaining in the powder box.
- Use alcohol cotton or a dry towel to wipe off the powder residue on the surface of the sandblasting scaler.
7.2. Cleaning the Nozzle and Handle
- Remove the handpiece, unscrew the nozzle, and pull out the nozzle.
- Use an air gun to aim at the nozzle and blow off the powder in the nozzle.
- Use an air gun to blow the front and rear ends of the handpiece to clean the residual moisture on the handpiece.
- If the nozzle is blocked, use a needle to unclog it.
7.3. Daily Maintenance
Pay attention to daily inspection of the handpiece for cracking, deformation, or falling off and other defects to ensure that the handpiece can be used normally. Pay attention to the inspection and cleaning of the sealing ring to ensure that the sealing ring is complete and clean; otherwise, it will affect the product sealing and the normal operation.
- Keep the sandblasting scaler dry. Rain, moisture, and various liquids may dampen the powder and affect normal use.
- Do not throw, beat, or vibrate the sandblasting scaler. Rough use of the sandblasting scaler will damage the strength and sealing of the equipment.
- Do not use paint to smear the instrument; smearing will leave debris in the detachable parts, which will affect the normal operation.
- Do not use a needle to poke the connector end of the tail wire, which may affect the normal use of the device.
7.4. Disinfection and Sterilization
This item is a hand instrument. Sterilization is recommended using high temperature and high pressure sterilization.
| Sterilizable Components | Handpiece |
|---|---|
| Sterilization Method | Recommended use of high temperature and high pressure sterilization |
| Sterilization Conditions | 134°C, 5Min |
Please use the sterilization bag which meets the national regulations for packaging before sterilization.
⚠ Caution:
Except for the handpiece, the other components cannot be sterilized under high temperature and high pressure.
8. Troubleshooting
| Fault Phenomenon | Failure Cause | Resolution |
|---|---|---|
| No air or powder from nozzle | Clogged Nozzle | In the absence of sandblasted powder in the sand powder bottle, remove the handpiece and clean the nozzle with a needle. |
| Clogged powder thru hole | Clean powder thru hole in sand powder bottle with needle. | |
| Too little powder and air ejection | Feed water of sand powder bottle | Replace the O-ring on the upper cover of sand powder bottle. |
| Air path in handpiece is blocked | Unloading the handpiece without sand powder bottle and cleaning the air passage of the handpiece with a needle. | |
| Water leakage at handpiece | Damaged the O-ring connection | Replace the O-ring. |
9. Specifications
9.1. Technical Specifications
| Parameter | Value |
|---|---|
| Input Air Pressure | 250kPa - 450kPa |
| Input Water Pressure | 70kPa - 220kPa |
| Spray Water Flow | >30ml/min |
| Gas Flow | <40NL/min |
| Noise | <68dB |
| Product Weight | 0.145kg |
| Nozzle Caliber (Powder) | 0.7mm |
| Nozzle Caliber (Water) | 1.4mm |
| Material | Metal |
| Commodity Quality Certification | CE, 3C |
9.2. Operating Environment and Storage Conditions
| Condition | Operating Temperature | Operating Humidity | Atmospheric Pressure |
|---|---|---|---|
| Operating | +5°C ~ +40°C | 20% RH ~ 80% RH | 86kPa ~ 106kPa |
| Condition | Storage Temperature | Storage Humidity | Atmospheric Pressure |
|---|---|---|---|
| Storage | -10°C ~ +55°C | <93%RH | 50kPa ~ 106kPa |
10. User Tips
- Replacement Tips: For replacement Perio Flow tips or other parts, it is recommended to contact the manufacturer or authorized seller to ensure compatibility and quality.
- Connection Type: This device is an air-driven polisher and connects to a dental unit's air and water lines, not an electrical cable.
- Polishing Powder: Always use specialized dental sandblasting powder (prophy powder) designed for dental use. Do not use household products like baking soda, as this can damage the device and be ineffective or harmful for dental treatment.
11. Warranty and Support
11.1. Guarantee
The Dental Scaler and Air Polisher has a 12-month warranty.
This product (such as through hole, replacement of sealing device, etc.) can be repaired by professional operators or nursing staff. If damage to structural components is found, please contact the after-sales service department of the company for repair.
11.2. Manufacturer Contact
Foshan COXO Medical Instrument Co., Ltd
BLDG 4, District A, Guangdong New Light Source Industrial Base, South of Luocun Avenue Nanhai District Foshan 528226 Guangdong China
Web: www.coxotec.com
For further details, you may refer to the official user manual PDF: View User Manual (PDF)
12. Recycling and Disposal
Dispose of waste equipment in accordance with national regulations and standards to ensure that all parts are not contaminated during the disposal process.