Owon CMS101 3 in 1 1000A AC/DC Clamp Meter+ Oscilloscope Bluetooth Recorder Function 18650 Lithium Battery NCV Digital Multimeter User Manual

Model: CMS101 | Brand: OWON

1. Safety Information

(Be sure to read the safety information before using this product.)

General Safety Requirements

  • Before any operations, please read the following safety precautions to avoid any possible bodily injury and prevent this product or any other products connected from damage. In order to avoid any contingent danger, this product is only used within the range specified.
  • Limit operation to the specified measurement category, voltage, or amperage ratings.
  • Do not use the digital clamp meter if it is damaged. Before you use the digital clamp meter, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
  • Do not use the test leads provided for other products. Use only the certified test leads specified for this product.
  • Inspect the test leads for damaged insulation or exposed metal.
  • Before use, verify the digital clamp meter's operation by measuring a known voltage.
  • Only the qualified technicians can implement the maintenance.
  • Always use the specified battery type. The power for the digital clamp meter is supplied with a battery. Observe the correct polarity markings before you insert the batteries to ensure proper insertion of the batteries in the digital clamp meter.
  • Check all Terminal Ratings. To avoid fire or shock hazard, check all ratings and markers of this product. Refer to the user's manual for more information about ratings before connecting to the digital clamp meter.
  • Do not operate the digital clamp meter with the cover or portions of the cover removed or loosened.
  • Do not operate if in any doubt. If you suspect damage occurs to the digital clamp meter, have it inspected by qualified service personnel before further operations.
  • Do not operate this product in wet or damp conditions.
  • Do not operate in an explosive atmosphere.
  • Keep product surfaces clean and dry.
  • Do not apply more than the rated voltage (as marked on the digital clamp meter) between terminals, or between terminal and earth ground.
  • When measuring current, turn off the circuit power before connecting the digital clamp meter in the circuit. Remember to place the digital clamp meter in series with the circuit.
  • When servicing the digital clamp meter, use only the specified replacement parts.
  • Use caution when working above 60 V DC, 30 V AC RMS, or 42.4 V peak. Such voltages pose a shock hazard.
  • When using the test leads, keep your fingers behind the finger guards on the test leads.
  • Remove the test leads from the digital clamp meter before you open the battery cover.
  • To avoid false readings, which may lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears and flashes.
  • Disconnect circuit power and discharge all high-voltage capacitors before testing resistance, continuity, diodes, or capacitance.
  • Use the proper terminals, function, and range for your measurements. When the range of the value to be measured is unknown, set the rotary switch position as the highest range, or choose the auto ranging mode. To avoid damages to the digital clamp meter, do not exceed the maximum limits of the input values shown in the technical specification tables.
  • Connect the common test lead before you connect the live test lead. When you disconnect the leads, disconnect the live test lead first.
  • Before changing functions, disconnect the test leads from the circuit under test.

Measurement Category

The digital clamp meter has a safety rating of 1000 V, CAT III and 600V, CAT IV.

Measurement category definition

  • Measurement CAT I applies to measurements performed on circuits not directly connected to the AC mains. Examples are measurements on circuits not derived from the AC mains and specially protected (internal) mains-derived circuits.
  • Measurement CAT II applies to protect against transients from energy-consuming equipment supplied from the fixed installation, such as TVs, PCs, portable tools, and other household circuits.
  • Measurement CAT III applies to protect against transients in equipment in fixed equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.
  • Measurement CAT IV applies to measurements performed at the source of the low-voltage installation. Examples are electricity meters and measurements on primary over current protection devices and ripple control units.

Safety Terms and Symbols

Terms in this Manual

  • Warning: Warning indicates the conditions or practices that could result in personal injury or death.
  • Caution: Caution indicates the conditions or practices that could result in damage to this product or other property.

Terms on the Product

  • Danger: It indicates an injury or hazard may immediately happen.
  • Warning: It indicates an injury or hazard may be accessible potentially.
  • Caution: It indicates a potential damage to the instrument or other property might occur.

Symbols on the Product

SymbolDescriptionSymbolDescription
---Direct current (DC)⎕Fuse
~Alternating current (AC)⚠Caution, risk of danger (refer to this manual for specific Warning or Caution information)
~Both direct and alternating currentCAT ICategory I overvoltage protection
⏚Ground terminalCAT IICategory II overvoltage protection
CEConforms to European Union directivesCAT IIICategory III overvoltage protection
⎕Equipment protected throughout by double insulation or reinforced insulationCAT IVCategory IV overvoltage protection

2. Setup

General Inspection

After receiving your new digital clamp meter, perform the following checks:

  1. Check whether there is any damage caused by transportation.
    If the packaging carton or foam protection cushion is seriously damaged, do not discard it until the device and its accessories have passed electrical and mechanical property tests.
  2. Check the Accessories.
    The supplied accessories are listed in the Appendix A. Verify that no accessories are lost or damaged. If any are missing or damaged, contact your distributor or local offices.
  3. Check the Complete Instrument.
    If the instrument's appearance is damaged, it does not work normally, or fails performance tests, contact your distributor or local offices. If damage is due to transportation, retain the packaging and inform the transportation department or distributor for repair or replacement.

Battery Installation

The digital clamp meter is powered by a 3.7V (18650) lithium battery.

Warning: To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the low battery indicator appears. Before replacing the battery, turn off the meter, disconnect test leads and any connectors from any circuit under test, and remove test leads from the input terminals. Use only the specified battery type.

Use the following procedure to install the battery:

  1. Power off, remove test leads and any connectors from the input terminals.
  2. Loosen the screws with a suitable Phillips screwdriver and remove the battery cover.
  3. Observe the battery polarity indicated inside the battery compartment, then insert the battery.
  4. Place the battery cover back in its original position and tighten the screws.
Caution: To avoid instrument damage from battery leakage, remove the batteries and store them separately if the digital clamp meter is not going to be used for a long period.

Power On/Off

The device can be turned on by pressing the POWER button at the bottom left of the host.

It can be turned off in the following ways:

  • Manual shutdown: Tap and hold the POWER button.
  • Automatic shutdown: The device will emit a short beep one minute before shutdown, then a long beep during shutdown.
  • Low power automatic shutdown.

Selecting the Range

  • Auto ranging is set as default when the meter is powered on, and "Auto" is displayed.
  • Under automatic range, press the Up or Down arrow keys to enter the manual range mode.
  • Under manual range, each additional press of the Up arrow key sets the next higher range; each additional press of the Down arrow key sets the next lower range.
  • Under manual range, press the A key to enter the auto range mode.

Note: Manual range is not available when measuring capacitance, only in multimeter measurement mode.

3. Operating Instructions

Instrument Panel Overview

Owon CMS101 3-in-1 Clamp Meter with key features highlighted
Figure 3-1: Owon CMS101 3-in-1 Clamp Meter highlighting Oscillograph, AC/DC 1000A, Bluetooth recorder, 18650 lithium battery, 2.8-inch IPS LCD, 320x240 resolution, and dual theme display. Also shows measurement capabilities for AC/DC current, AC/DC voltage, resistances, capacitors, NCV, diodes, frequency, and duty cycle.
Owon CMS101 Oscilloscope, Multimeter, and Other Functions
Figure 3-2: Detailed features of the Oscilloscope (Analog broadband voltage, real-time sampling, input resistance, automatic trigger, automatic measurement, sampling depth), Multimeter (AC/DC current/voltage/resistance/capacitance/frequency/duty cycle, diode/on-off buzzer, NCV, data hold, screen backlight, true RMS), and Other Functions (Bluetooth communication, power supply, LCD screen).

The front panel and keys of the digital clamp meter are shown in the figures above. Key components include:

  • NCV measurement area (1)
  • Safety level (2)
  • Maximum current allowed to be measured (3)
  • Operation indicator light (4)
  • Reading hold button / Run/Stop button (5)
  • Display area (6)
  • F1-F4 keys (7): Multi-function keys; select corresponding menu items in each mode.
  • Direction keys (8): Used for changing range, voltage/current scales, zooming waveforms, time base changing. The 'A' key restores auto gear or performs automatic setup.
  • Measurement input port (9): Input for the measurement signal.
  • Charging port (10): USB-C port for charging.
  • Switch key (11): Briefly press to turn on. After powering on, briefly press to switch between oscilloscope or multimeter mode. Press and hold to turn off.
  • Tab function switch button (12)
  • Clamp head trigger (13): Press to open clamp heads; release to close.
  • Clamp head (14): Current measurement sensor converts AC or DC current into voltage.

How to Use the Multimeter

Instrument Interface

The digital clamp meter uses four 2 safety banana plug input ends: COM and VΩ.

Multimeter interface elements:

  • Measurement type indication: A,DC (DC current), A,AC (AC current), V,DC/mV,DC (DC voltage), V,AC/mV,AC (AC voltage).
  • Range indication: "Manu" for manual range, "Auto" for automatic range.
  • Current measurement range.
  • Automatic shutdown sign: Displays when enabled, hides when closed.
  • Bluetooth sign: Displays when enabled, hides when closed.
  • Battery power and external power supply indication.
  • Inrush mode (Only in ACA mode).
  • Reading hold mode.
  • Current measurement menu.
  • Current measurement unit.
  • DC/AC/On-Off/Diode/Capacitance mode.
  • Range simulation strip.
  • Operation menu.

Measurement Units

SignDescriptionValue
MMega1E+06 (1,000,000)
kkilo1E+03 (1,000)
mmilli1E-03 (0.001)
μmicro1E-06 (0.000001)
nnano1E-09 (0.000000001)
SignDescriptionMeasurement type
AAmpereCurrent
ΩOhmResistance
HzHertzFrequency
%PercentDuty cycle
FFaradCapacitance

Input Terminals

The terminal connections for different measurement functions are:

Warning: Before starting any measurement, observe the rotary switch position of the digital clamp meter, and then connect the test leads to the correct terminals.
Caution: To avoid damaging the digital clamp meter, do not exceed the rated input limit.
Measurement FunctionInput TerminalsOverload Protection
V, mV, Ω, ⏚, ⏚, NCVCOM, VΩ1000 VAC/1000 VDC
A, ACCOM, VΩ250 VAC/300 VDC

Making Measurements

Measuring DC or AC Current
Warning: Do not attempt to measure current if the open-circuit voltage between the live and ground exceeds 250V, as this may damage the digital clamp meter and pose a risk of electric shock or personal injury.
Caution: To avoid damaging the digital clamp meter or the measured device, check the digital clamp meter before measuring current. Ensure the correct input terminals, function settings, and range are used for the measurement.
  • The current measurement function must be operated between 0°C and 40°C. Avoid suddenly releasing the trigger; the Hall effect sensor is sensitive not only to magnetic fields but also to heat and mechanical stress. Impact can temporarily alter readings.
  • For accurate measurement, ensure the conductor is positioned at the center of the clamp heads. Positioning outside the center can introduce an additional error of ±1.0% of the reading.
  1. Press F1 to switch to DC Current measurement mode (A,DC will be displayed). Press F1 again to switch to AC Current measurement mode (A,AC will be displayed).
  2. Hold the trigger to open the clamp heads. Place the conductor to be measured in the clamp and then slowly release the trigger until the clamp heads are fully closed. Ensure the conductor is centered in the clamp. The clamp meter can measure only one current conductor at a time; measuring two or more conductors simultaneously will result in incorrect readings.
  3. Read the display. Press Up or Down arrow keys to enable and cycle through the manual ranges. If "OL" is displayed, it indicates the input exceeds the range.
  4. In AC current mode, press Tab to switch to the second page of the menu, then press F3 to enable the surge current measurement mode. Start the electrical device to measure its inrush current. Press F3 again to exit this mode.
  5. In AC current mode, press Tab to switch to the second page of the menu. Press F4 to enable the AC current frequency measurement mode. Read the frequency of the AC current from the display. Press F4 once more to exit this mode.

Note: After testing DC current, especially with large currents, the open-circuit baseline might be high. Perform an AC current test to eliminate residual magnetic signals from the clamp heads. If there is residual magnetism, press F3 to select "Zero." This will temporarily subtract the current residual magnetism from the readings. The residual magnetism value will return after restarting the instrument.

Measuring AC or DC Voltage
Warning: Do not measure any voltage of over 1000 Vdc or 1000 Vac rms to avoid instrument damage or electric shock. Do not apply more than 1000 Vdc or 1000 Vac rms between the common terminal and the earth ground to avoid instrument damage or electric shock.

This digital clamp meter displays DC voltage values as well as their polarity. Negative DC voltages will display a negative sign on the left of the screen.

  1. Press F2 to switch DC voltage measurement mode (*,DC will be displayed). Then press F2 to switch into AC Voltage measurement mode (*,AC will be displayed). Then press F2 to switch into mV/V measurement function, press Up or Down arrow keys to adjust measurement scales.
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Probe the test points and read the display.
  4. In AC voltage mode, press Tab to switch to the second page of the menu. Press F4 again to enable the AC voltage frequency measurement mode. Read the frequency of the AC voltage from the display. Press F4 once more to exit this mode.

Note: When the measured voltage exceeds 42V DC, the instrument's LCD will display a high voltage warning symbol "⚡" and the screen will show a red alert.

Measuring Resistance
Caution: To avoid possible damage to your digital clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before measuring resistance.
  1. Press F3 to enter resistance testing mode ("Res" will be displayed).
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Probe the test points to measure the resistance in the circuit.
Measuring Continuity
Caution: To avoid possible damage to your digital clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing for continuity.
  1. Press F3 to enter continuity testing mode ("Cont" will be displayed).
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Probe the test points to measure the resistance in the circuit. If the reading is below 50 Ω, the digital clamp meter will beep continuously.
Measuring Diodes
Caution: To avoid possible damage to your digital clamp meter or to the equipment under test, disconnect the circuit power and discharge all high-voltage capacitors before testing diodes.
  1. Press F3 to enter diode testing mode ("Diode" will be displayed).
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Connect the red test lead to the positive terminal of the diode and the black test lead to the negative terminal.
  4. Read the diode forward bias. If the test lead connection is reversed, the digital clamp meter will display "OL".
Measuring Capacitance
Caution: To avoid possible damage to the digital clamp meter or to the equipment under test, disconnect circuit power and discharge all high-voltage capacitors before measuring capacitance. Use the DC voltage function to confirm that the capacitor is fully discharged.
  1. Press F4 to enter capacitance testing mode ("Cap" will be displayed).
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Probe the test points to measure the capacitance in the circuit.
Measuring Frequency
  1. Press F4 to enter frequency measurement mode ("Freq" will be displayed).
  2. Connect the black test lead to the COM terminal and the red test lead to the VΩ terminal.
  3. Probe the test points to measure the frequency in the circuit.
Measuring NCV (Non-Contact Voltage)
  1. Press F4 to enter NCV testing mode ("NCV" will be displayed). Bring the front end of the clamp head close to the measured conductor for AC voltage detection.
  2. To detect whether there is AC voltage or an electromagnetic field in the area, the LCD will display a dash ("-"). The display will indicate four levels based on the voltage detected: "-", "--", "---", and "-----". Additionally, the buzzer will emit a continuous intermittent sound, and the NCV LED will light up. If no voltage is detected, the LCD will display "EF".

Multimeter Features

Setting Scale
  1. Press A to enter automatic gear mode.
  2. In the automatic gear mode, press Up or Down arrow keys to enter manual gear mode.
Setting Range
  1. Press F1 to switch into DC current measurement and press Up or Down arrow keys to adjust measurement scales.
  2. Press F1 to switch into AC current measurement, press Tab to access the second page menu, press F3 or F4 to select Inrush, Hz/A function, and then press Up or Down arrow keys to adjust measurement scales.
Making Max/Min Measurements

Press Tab to switch into the second page menu and press F1 to enter Max/Min measurement. When the button is pressed, it will automatically enter the manual gear mode. Press the button again to exit the Max/Min mode and resume the automatic gear mode.

Making Relative Measurements

When making relative measurements, the reading is the difference between a stored reference value and the input signal.

  1. Press Tab to switch into the second page menu and press F2 to enter the relative mode. A "Δ" symbol will be shown on the display. The measurement value when pressing F2 is stored as the reference value. In this mode, Δ (current reading) = input value - reference value.
  2. Press F2 again to exit the mode.

In relative measurement, the manual range mode will be activated automatically (this function is only available under the manual range mode).

Value Hold Mode

The value hold mode keeps the current reading on the display.

  1. Press HOLD on the right; the current reading will be maintained, and "HOLD" will be displayed.
  2. Press HOLD again to exit the mode.

How to Use the Oscilloscope

Play video
Watch video online
Video: Owon CMS101 Oscilloscope Function Overview. This video demonstrates how to switch between multimeter and oscilloscope functions, adjust voltage/current ranges, set trigger levels, and measure AC current and voltage waveforms.

Introduction to the User Interface of the Oscilloscope

The oscilloscope interface includes:

  • Time base display: In the horizontal direction, each grid represents a specific time interval.
  • Scale: In the vertical direction, each grid represents a specific voltage or current value.
  • Trigger status: Indicates "Trig" (trigger detected), "Ready" (pre-trigger data obtained), "Scan" (continuously collecting and displaying waveform data), or "Stop" (stop collecting waveform data).
  • Automatic shutdown sign: Displays when enabled, hides when closed.
  • Bluetooth sign: Displays when enabled, hides when closed.
  • Battery power and external power supply indication.
  • RMS value.
  • Trigger horizontal displacement.
  • Trigger level position.
  • Measurement value.
  • Operation menu.
  • Waveform display area.
  • Channel waveforms.
  • Frequency value.

Measuring System

Automatic Measurement

In oscilloscope mode, the screen displays six automatic measurement types: Freq, Rms, Max, Min, Vpp, and Avg.

The automatic measurement of voltage parameters

The oscilloscopes provide automatic voltage measurements including Avg, Vpp, Rms, Max, Min and Vamp.

  • Avg: The arithmetic mean over the entire waveform.
  • Vpp: Peak-to-Peak Voltage.
  • Rms: The true Root Mean Square voltage over the entire waveform.
  • Max: The most positive peak voltage measured over the entire waveform.
  • Min: The most negative peak voltage measured over the entire waveform.
Measurement
  1. Press POWER button to switch into oscilloscope mode.
  2. Press F1 to switch into voltage measurement and then press Up or Down arrow keys to adjust measurement gear. Press Left or Right arrow keys to adjust time base gear.
  3. Press F1 to switch into current measurement and then press Up or Down arrow keys to adjust measurement gear. Press Left or Right arrow keys to adjust time base gear.
  4. Press F1 to switch into current measurement. If there is residual magnetism, press the F3 key to select "Zero." This will temporarily subtract the current residual magnetism from the readings.
  5. Press F4 or HOLD. While in pause mode, press Up or Down arrow keys to zoom in or out the waveform.
  6. Press Tab to enter the second page menu. Press F1 to set the trigger as rise or fall. Press F2 or F3 to adjust the trigger level position.

System Settings

Press Tab to enter the third page menu. Press F1 ~ F4 to select the desired function menu.

  1. Brightness setting: Press F1; the brightness varies between Low, Mid, and High.
  2. Automatic shutdown setting: Press F2 to open or close the function. When the function is turned on, it will automatically shut down after a specified period of time. When the function is turned off, manual shutdown is required.
  3. Theme setting: Press F3 to change the display theme.
Owon CMS101 2.8-inch IPS LCD Dual Theme Display
Figure 3-3: The 2.8-inch IPS LCD features a dual theme display. The brightness of the screen backlight can be adjusted for different light environments to ensure a clearer display.

How to Use Bluetooth

Owon CMS101 Bluetooth Logger Features and App Interface
Figure 3-4: Bluetooth Logger Features. The user's cell phone connects to the meter via APP, setting up automatic recording of measured values, displaying them in chart mode, and storing data in CSV format for further analysis.
  1. On your mobile device, scan the QR code below to install the free multimeter app.
  2. Open the installed application on your mobile device.
  3. Turn on the multimeter, press and hold Tab until the Bluetooth symbol appears on the display.
  4. Click on "Device List" in the bottom navigation bar of the app.
  5. Click the "+" icon in the upper left corner to begin searching for devices and list out the multimeters found.
  6. If Bluetooth is not enabled on your mobile device, a prompt box will appear indicating "Bluetooth is not turned on". Manually open Bluetooth on your mobile device before connecting.
  7. Activate "Filter device" to hide incompatible multimeters.
  8. After "BDM" appears in the list of available devices, click and select it to connect to the mobile device.

Note: When Bluetooth is enabled, the auto power-off function is disabled. After Bluetooth is turned off, the auto power-off function will be restored.

4. Maintenance

General Maintenance

  • Do not store or place the instrument in a location where the LCD screen will be exposed to direct sunlight for a long time.
  • Caution: Do not let spray, liquid or solvent touch the instrument or probe to prevent damage to the instrument or probe.

Cleaning

Check the instrument and probe frequently. Clean the external surface as follows:

  1. Wipe floating dust outside the instrument and probe with a soft cloth. When cleaning the LCD, be careful not to scratch the transparent LCD protection screen.
  2. Wipe the instrument with a damp but non-dripping soft cloth. Disconnect the power supply. It can be scrubbed with soft detergent or water. Do not use any abrasive chemical cleaning agent to avoid damaging the instrument or probe.
Warning: Please make sure the instrument is dry before re-energizing to avoid electrical short circuit or personal injury caused by moisture.

Charging and Replacement of Battery

During long-term storage, the battery may be too low due to self-discharge. Use the attached adapter to pre-charge the device for 0.5 to 1 hour (depending on storage time) before turning it on. If the device is not used for a long time, it is recommended to charge it at regular intervals to avoid over-discharge.

Battery Charging

The lithium battery may not be fully charged when delivered. This device has an approximate charging time of 8 hours and a peak discharge battery life of about 18.5 hours. The power supply and battery indicator symbols in the upper right corner of the screen are explained as follows:

  • ■ symbol indicates the power-on charging status;
  • ■ symbol indicates battery power supply;
  • ■ symbol indicates that there is only about five minutes of use time left.

Please charge as soon as possible according to the relevant tips to avoid damage to the battery.

Charging Method: Connect the digital clamp meter to a computer or other equipment through a USB data cable (Type-C) for charging (pay attention to the load capacity of the power supply equipment to avoid abnormal operation).

Note: To avoid overheating of the battery during charging, the ambient temperature must not exceed the allowable value given in the technical specifications.

Charging Indicator: The power button turns red when charging and turns off when charging is complete.

Replacement of Lithium Battery

Generally, the battery does not need to be replaced. However, when necessary, it can only be replaced by qualified personnel, and only lithium batteries of the same specification can be used.

5. Troubleshooting

If you encounter issues with your Owon CMS101, consider the following:

  • Device not powering on: Ensure the 18650 lithium battery is correctly installed with the right polarity and sufficiently charged. If stored for a long time, pre-charge for 0.5 to 1 hour.
  • Inaccurate readings: Verify that the correct measurement function and range are selected. For current measurements, ensure the conductor is centered in the clamp heads.
  • Bluetooth connection issues: Ensure Bluetooth is enabled on both the device and your mobile phone. Follow the app's instructions for pairing.
  • Display issues: Check battery level. Adjust screen brightness in system settings if the display is hard to read.

6. Technical Specifications

Standard conditions: The environment temperature is 18°C to 28°C, the relative humidity is less than 80%.

Note:

  • When measuring AC voltage, AC current, accuracy guarantee range is 10% to 100% of the range.
  • When measuring DC voltage, DC current or capacitance, accuracy guarantee range is 5% to 100% of the range.

Multimeter Specification

FunctionRangeResolutionAccuracy
DC Voltage (V)20.000mV0.001mV± (0.1%+20dig)
200.00mV0.01mV± (0.1%+6dig)
2.0000V / 20.000V / 200.00V0.0001V / 0.001V / 0.01V± (0.1%+5dig)
1000.0V0.1V± (0.15%+5dig)
AC Voltage (V)200.00mV0.01 mVVRMS, Freq range: 40Hz-1000Hz
± (0.6%+10dig) for 2.0000V-200.00V
± (0.8%+10dig) for 1000.0V
2.0000V0.0001V
20.000V / 200.00V0.001V / 0.01mV
1000.0V0.1V
DC Current (A)20.00A0.01A± (2.0%+10dig)
200.0A0.1A± (2.0%+5dig)
1000A1A
AC Current (A)20.00A0.01AVRMS, Freq range: 40Hz-1000Hz
± (3.0%+10dig)
200.0A0.1A± (2.5%+5dig)
1000A1A
Inrush Current (A)20.00A0.01AVRMS, Freq range: 40Hz-1000Hz
± (10.0%+10dig)
200.0A0.1A
1000A1A
NCVSupport
Resistance (Ω)200.00Ω0.01Ω± (0.8%+10dig)
2.0000KΩ0.0001KΩ± (0.3%+10dig)
20.000KΩ0.001KΩ
200.00KΩ0.01KΩ
Capacitance (F)2.000nF0.001nF± (5.0%+10dig)
20.00nF / 200.0nF0.01nF / 0.1nF± (3.0%+10dig)
2.000μF / 20.00μF0.001μF / 0.01μF
200.0μF0.1μF
2.000mF / 20.00mF0.001mF / 0.01mF
Frequency (Hz)200.00Hz0.01Hz± (0.1%+5dig)
2.0000kHz0.0001kHz
20.000kHz / 200.00kHz0.001kHz / 0.01kHz
2.0000MHz0.0001MHz
20.000MHz0.001MHz
Duty Cycle (%)0.1%-99.9% (Typical: Vrms=1V, f=100Hz)0.10%± (1.2%+3dig)
0.1%-99.9% (≥1 kHz)± (2.5%+10dig)
Diode3.0000V0.0001VOpen circuit voltage 3.2V
On-Off0~200.00Ω0.1ΩBuzzer limit 50Ω; Measurement value displayed from 00 to 200.00, "OL" if value exceeds.

Oscilloscope Specification

CharacteristicsInstruction
Analog bandwidthVoltage: 1MHz, Current: 1KHz
Sample modeReal-time sample
Real-time sampling rate5.0 MSa/s
Channel1
Input impedance≥10 MΩ
Maximum input voltageMaximum peak voltage 1000V
Maximum sample currentMaximum peak current 1000A
Scan speed2.5 us/div - 10 s/div
Time base accuracy±(0.01 % + 0.1 div)
Sensitivity30 mV/div - 500 V/div
Displacement range±3 grid
Sensitivity accuracy±(5%+0.2div)
Measurement valueRms, Freq, Max, Min, PK-PK, Avg
Trigger modeAuto
Trigger typeRise, Fall
Bluetooth communicationA smart phone can be used to view the measurement data of the multimeter on the mobile phone side, perform remote control, display data charts, and store the measurement data in CSV format.
Automatic shutdownWhen all functions are not used, the meter will automatically shut down in about 10 minutes (default is 10 minutes, can be canceled).
True effective value√
Display modeDMM or OSC
Return zero√

General Specifications

CharacteristicValue
Input protection√
Digital hold√
PowerSingle section 18650 3.7V
Low battery displayWhen power is low, a low power window appears, and the device will shut down automatically after a period.
Backlight function√
LCD Size2.8 inch
WeightApprox. 0.35kg
Dimension248mm (L) x 94.5mm (W) x 37.8mm (D)
Interval Period of AdjustmentOne year is recommended for the calibration interval period.

Note: The signal is within 5dB attenuation with an analog bandwidth of up to 1MHz.

7. User Tips

  • Oscilloscope Current Measurement: Be aware that the oscilloscope function may be inaccurate for currents below 20A. For current measurements, especially below 1A, the visual representation on the oscilloscope graph might be challenging; the device performs better with higher current values (e.g., above 10A).
  • AC/DC Mode Switching: The device requires manual switching between AC and DC modes for certain measurements. It does not automatically detect and switch.
  • Charging Indicator: The power button LED indicates charging status: it turns red when charging and turns off when charging is complete.
  • Charging Cable: The device uses a Type-C cable for charging.

8. Warranty and Support

General Warranty

We warrant that the product will be free from defects in materials and workmanship for a period of 1 year from the date of purchase by the original purchaser. This warranty applies only to the original purchaser and is not transferable.

If the product is defective during the warranty period, we will either repair or replace it without charge for parts and labor. Replaced parts, modules, and products may be new or reconditioned. All replaced items become our property.

To obtain service, the customer must notify us of the defect before the warranty expires. The customer is responsible for packaging and shipping the defective product to the designated service center, along with proof of purchase.

This warranty does not cover defects, failures, or damage caused by:

  • Improper use or inadequate maintenance and care.
  • Damage from attempts to install, repair, or service by unauthorized personnel.
  • Damage from improper use or connection to incompatible equipment.
  • Damage or malfunction caused by the use of non-approved supplies.
  • Servicing a product modified or integrated with other products if such modification increases servicing time or difficulty.

Excepting the after-sales services provided in this summary or the applicable warranty statements, we will not offer any guarantee for maintenance declared or hinted, including but not limited to the implied guarantee for marketability and special-purpose acceptability. We should not take any responsibilities for any indirect, special, or consequent damages.

Product Support

For product support, visit: www.owon.com.hk/download

Fujian LILLIPUT Optoelectronics Technology Co., Ltd.
No. 19, Heming Road
Lantian Industrial Zone, Zhangzhou 363005 P.R. China

Tel: +86-596-2130430