1. Introduction
Thank you for choosing a Voldent AirJet product. This device is a handheld dental tool powered by a dental unit, designed to jet sandblasting powder onto the surface of a patient's teeth for cleaning and polishing.
Key Features:
- Performs supragingival and periodontal sandblasting for all-round plaque removal and periodontal treatment.
- Anti-suction design prevents powder and moisture from being sucked back, reducing cross-infection risk and protecting the dental unit cable.
- Adjustable air flow to meet diverse clinical requirements.
- Compact, lightweight, and ergonomically designed for comfortable handling and reduced fatigue.

Figure 1: AirJet Dental Air Polisher in use, demonstrating supragingival and subgingival modes.
Video: Overview of the AirJet Powder Jet Handpiece, showcasing its features and different models (AirJet 1, AirJet 1S, AirJet 2) and their components.
2. Safety Warnings and Precautions
READ ME before use
- ⚠️ Only the handpiece can be autoclaved. The main body cannot be autoclaved.
- ⚠️ Only the handpiece can be submerged in a disinfectant bath. The main body cannot be submerged in a disinfectant bath.
- ⚠️ Damage caused by incorrect cleaning or disinfection is not covered by the warranty.
- ⚠️ Do not use this device on patients with gum tissue damage or if empty items are clogged by air entering the soft tissue.
- ⚠️ Ensure that the input air pressure is within the rated pressure range (0.35 to 0.45MPa). Otherwise, water leakage may occur or the quick connector may come out.
- ⚠️ Before using, ensure the quick connector or the hose of the dental unit, handpiece, powder chamber, powder chamber cap, air inlet, and outlet pipe are completely dry.
- ⚠️ Please ensure that the air source used is dry, clean, and oil-free. Otherwise, it may cause congestion.
- ⚠️ Please do not disassemble this product by yourself, otherwise your product will no longer enjoy the warranty.
General Precautions:
- This product must be used only by trained and qualified personnel.
- Do not use this device on patients with respiratory illness, mucous membrane injury (greater than 3mm pocket depth), grave ulcers, mucosal inflammation, pulmonary ventilation disorder, liver dysfunction, heart dysfunction, allergies, or those wearing contact lenses.
- Patients needing to limit sodium intake, with hypernatremia, poisoned during pregnancy, kidney deficiency, chronic respiratory disease, or chronic diarrhea should not be treated with this device.
- During operation, the operator and patient should wear protective glasses and masks.
- It is recommended to use a suction device to remove excess powder.
- If powder enters the eyes, rinse immediately with plenty of water and consult an ophthalmologist.
- Do not aim the nozzle directly at fillings, crowns, or dentures, as this may cause damage.
3. Product Components
The AirJet Dental Air Polisher consists of a main body, a sandblasting handpiece, a tail connector, and a powder chamber. Different models (AirJet 1, AirJet 1S, AirJet 2) come with specific configurations.
AirJet 2 Structure and Accessories:
- Subgingival sandblasting handpiece
- Main body
- Powder chamber cap
- Short cleaning needle
- Long cleaning needle
- Cleaning spray nozzle
- Small seal ring
- Powder chamber seal ring
- Cleaning brush
- Periodontal sandblasting handpiece
- Periodontal nozzle cap
- Periodontal nozzle
- Periodontal nozzle wrench
- Power gear adjustment knob

Figure 2: AirJet 2 Structure and Accessories (VD-J3 Supragingival scaling handpiece, VD-J4 Periodontal scaling handpiece, large capacity visible powder chamber, MULTIflex quick coupling).
AirJet 1 and AirJet 1S Specifics:
- AirJet 1: Equipped with VD-J1 supragingival sandblasting handpiece and a 4-holes connector.
- AirJet 1S: Equipped with VD-J1 supragingival sandblasting handpiece and a KAVO MULTIflex quick connector.
- AirJet 2: Equipped with VD-J2 (supragingival) and VD-J1 (periodontal) sandblasting handpieces, and a KAVO MULTIflex quick connector. It features 2 sandblasting modes (G & P).

Figure 3: AirJet 1 with VD-J1 supragingival sandblasting handpiece and 4-holes connector.

Figure 4: AirJet 1S with VD-J1 supragingival sandblasting handpiece and KAVO MULTIflex quick connector.

Figure 5: AirJet 2 with VD-J2 (supragingival) and VD-J1 (periodontal) handpieces and KAVO MULTIflex quick connector.

Figure 6: KAVO MULTIflex quick coupling.
4. Specifications
| Attribute | Value |
|---|---|
| Brand Name | FR |
| Model Number | AirJET1/1S/2 |
| Material | Plastic |
| Size | 1 piece |
| Origin | Mainland China |
| Water Input Pressure | 0.7bar~2.2bar (70~220kPa) |
| Air Input Pressure | 0.35~0.45MPa (3.5~4.5bar) |
| Flow Rate | 50~80ml/min |
| Operating Temperature | 10℃~40℃ |
| Operating Relative Humidity | 30%~75% |
| Operating Air Pressure | 500hPa~1060hPa |
| Storage & Transport Temperature | -10℃~40℃ |
| Storage & Transport Relative Humidity | 30%~75% |
| Storage & Transport Air Pressure | 500hPa~1060hPa |
| Commodity Quality Certification | 3C |
| Hand Instrument | Yes |
| High-concerned chemical | None |
| Sterilization of whole machine at high temperature | NO |
| Adapter | Dental Chair |
| AirJET1 4 hole | Supragingival |
| AirJET1S with KAVO Quick Coupling | Supragingival |
| AirJET2 | Supragingival/Subgingival |
| Handpiece VD-J3 | Supragingival |
| Handpiece VD-J4 for AirJET2 | Subgingival |
| Powder Model | PT-X1, PT-X3, PT-X5, PT-X6, PT-X7 |
| Supragingival Powder | PT-X1, PT-X3, PT-X5 |
| Periodontal/Interproximal Powder | PT-X6, PT-X7 |
Recommended Prophylaxis Powder:

Figure 7: Prophylaxis Powder details including composition, particle size, pH, and application for supragingival and periodontal use.
5. Setup and Installation
5.1. Setting up External Dental Equipment:
Ensure your dental unit's water supply system provides a pressure of 0.7bar-2.2bar (70-220kPa) and a temperature of up to 40°C.
Adjust the air supply pressure to obtain 3.5bar to 4.5bar (350 to 450kPa) at the external dental equipment.
⚠️ Excessive air pressure can cause cracks or ruptures in the powder chamber and cover, potentially leading to human injury.
5.2. Installation and Connection:
⚠️ Before installation, ensure the quick connector, power jet handpiece, powder chamber, powder chamber cap, tail connector, air inlet pipe, and air outlet pipe are completely dry. Use a 3-way syringe for drying.
Connecting Power Jet Handpiece and Quick Connector:
- Use a 3-way syringe to dry the power jet handpiece tail connector and the quick connector.
- For AirJet 1S and AirJet 2: Directly insert the quick connector into the Power Jet Handpiece tail connector, then tighten the quick connector.
- For AirJet 1: Directly insert the quick connector into the Power Jet Handpiece tail connector, then tighten the quick connector.

Figure 8: Connection method for AirJet 1S and AirJet 2.
Connecting Handpiece to Main Body:
After drying the inside of the handpiece and the connector of the main body, directly insert the handpiece into the interface of the main body.
6. Operation
6.1. Adjusting Water Flow:
Before loading the prophylaxis powder, adjust the water flow so that the power jet handpiece receives an even amount of water.
6.2. Loading Prophylaxis Powder:
⚠️ Use only powder provided or recommended by VOLDENT. Ensure the powder chamber is absolutely dry. Do not exceed the 'MAX' limit. The opening of air inlet and outlet pipe should not be covered by powder. Clean the threads of the powder chamber before screwing on the cap. Do not shake the device as this could cause the powder to clog the tubes.
- After connecting the Power Jet Handpiece, unscrew the powder chamber cover.
- Put the prophylaxis powder into the chamber.
- Clean any powder at the thread position of the chamber cover.
- Tighten the chamber cover.

Figure 9: Prophylaxis powder filling process.
6.3. Air Volume Switch Operation (for AirJet 2 only):
The AirJet 2 power jet handpiece has two gears. Rotate the air volume adjustment switch to select between:
- Supragingival sandblasting (right)
- Periodontal sandblasting (left)
6.4. Supragingival Scaling:
For supragingival scaling, adjust water and air flow to achieve desired cleaning and polishing effects. Wear a face mask and eye protection. Place absorbent cotton rolls under the lips and use a high-speed evacuator. Do not direct the nozzle directly at the tooth surface; maintain a distance of 3 to 5mm and vary the angle between 30 to 60 degrees. After treatment, polish the teeth surfaces by setting the water flow rate to the maximum.

Figure 10: Supragingival sandblasting technique.
6.5. Periodontal Scaling:
⚠️ The air powder jet is powerful. It can cause injury to the gums or an emphysema if air is introduced into soft tissue spaces. Do not direct the nozzle directly at the gum tissue or into the gingival sulcus.
⚠️ After treatment, the keratin and protein layer on the tooth surface are completely removed. Advise patients that for 2 to 3 hours following treatment, they should neither smoke nor consume food or drinks that could strongly color the teeth (e.g., tea, coffee).
Installation of Periodontal Nozzle:
- Take the periodontal nozzle and assemble it to the top of the handpiece according to the positioning direction.
- Use the wrench to lock the locking cap to the nozzle.
- After inserting the nozzle in the correct direction, screw on the locking cap, use a wrench to lock the nozzle locking cap, and remove the wrench after locking.

Figure 11: Handpiece with 360-degree rotation for easier access.
Periodontal Sandblasting Operation:
- After filling the powder, spray it in an outer container for 1-3 seconds to ensure even air and water spray before treating the patient.
- Hold the powder jet handpiece in a pen-holding position.
- When performing normal periodontal scaling, use the nozzle to remove plaque from the periodontal pocket (1-3mm below the gingival) with up and down lifting and scaling motions.
- Scaling is recommended for less than 5 seconds per periodontal pocket site.
- Ensure the water channel of the handpiece is unobstructed. Adjust water volume and air pressure according to dental plaque or pigment conditions. Avoid staying too long in one area.

Figure 12: Full range of clinic applications for the AirJet.
7. Maintenance
7.1. Daily Maintenance:
- At the end of each day, use long and short needles to clean each pipeline of the equipment.
- Replace the powder chamber cap seal ring, seal ring 1, and seal ring 2 every 6 months.
- Perform ultrasonic cleaning of the handpiece with hot water soak once a week.
7.2. Cleaning, Disinfecting, and Sterilizing:
Powder Cleaning:
- Uncover the chamber cap and empty the powder chamber.
- Keep the unit operating after shutting off the water supply. Use compressed air to blow out remaining water and powder in the pipe.
- Blow out remaining powder in the chamber and connector with dry compressed air.
Handpiece and Body Cleaning:
- Remove the handpiece.
- Insert the cleaning nozzle into the handpiece until a "click" sound is heard, ensuring it's securely fastened.
- Use a 3-way syringe to blow air inside the handpiece through the cleaning nozzle for 3-5 seconds to thoroughly clean the handpiece tubes.
- Wash the powder chamber cap, cap ring, cap dome, and handpiece with distilled or deionized water.
- Clean the threads of the powder chamber, body surface, and handpiece surface with alcohol (ethanol, isopropanol).
- If the nozzle is clogged, use a thru-needle to unblock.
Sterilization:
Enclose the handpiece in a sterilization bag and seal the bag.
Sterilize the handpiece at up to 135°C in the autoclave (134°C for 20 minutes or 132°C for 15 minutes).
⚠️ Only the handpiece can be autoclaved under high temperature and pressure.
⚠️ Ensure no chemical liquid is on the handpiece surface and it is completely dry before sterilization.
⚠️ Do not clean or scour this device with high acidity cleanser or disinfectant.
⚠️ Submerge only the handpiece in a disinfectant bath. Do not submerge the unit in a disinfectant bath.
⚠️ It must be sterilized before first use and after each patient use.
⚠️ Clean up within 30 minutes after treatment; high temperature and high pressure sterilization treatment should be taken place within 2 hours after cleaning.
8. Troubleshooting
| Type of Problem | Solutions |
|---|---|
| No powder/air jet coming from the unit |
|
| Water comes into the powder chamber | Check the connector of dental unit hose and quick connector. Replace the seal ring if necessary. |
| Water leakage from the connection between main body and handpiece | Replace the seal ring 1 and 2. |
| Air or powder leakage from the chamber cap | Check the seal and the cleanliness of the thread on the powder chamber and on the cap. Replace the seal ring if necessary. |
| Weak air and powder jet, low cleaning efficiency | Refill or change the powder. Clean the handpiece and all the relevant tubes. |
In case of any fault that could not be solved by the above methods, please contact your dealer or the nearest service center.
9. User Tips
- Adapter Compatibility: Some users have noted that the included connector might be Midwest, while their dental terminal is Borden. Be prepared to acquire an appropriate adapter if needed for your specific dental unit.
- AirJet 1 Tip Usage: The AirJet 1 model is designed for supragingival treatment. Subgingival tips are not compatible with AirJet 1 and should only be used with models designed for periodontal treatment.
- Powder Availability: If you require additional prophylaxis powder, please contact customer service for details on purchasing compatible powders.
10. Warranty and Support
After-Sale Service:
If this product fails to function normally due to quality problems after purchase, please contact your local dealer or the company with the warranty card for replacement or repair. Damages resulting from non-adherence to operation instructions or wear and tear of parts are excluded from warranty.
⚠️ The warranty of your product will be canceled if you attempt to open it.
Manufacturer Information:
Voldent Medical Technology Co., Ltd.
Address: 1st Floor, Building 4, Plot D-04-2a, Information Industrial Park, Qixing District, Guilin, 541004, Guangxi, P.R.China
Tel: +86-773-6751115
E-mail: voldent.china@hotmail.com
Website: www.voldent-medical.com
11. Environmental Protection
The device does not contain harmful ingredients and can be disposed of and destroyed in accordance with relevant local regulations.
12. Symbols
| Symbol | Description |
|---|---|
| i | Read Instructions |
| ⚠️ | Caution! |
| SN | Serial Number |
| Σ | Valid Period |
| 134°C | Sterilizable in a steam sterilizer (autoclave) at the temperature specified |
| Water-disinfector for thermal disinfection | |
| Not autoclavable. Thermal disinfection not allowed | |
| CE | CE Mark |
| % | Humidity Limitation |
| Atmospheric Pressure Limitation | |
| Temperature Limitation | |
| MD | Medical Device |