Morayelec ME260P

Contents

Moray ME260P Digital Pressure Gauge User Manual

Model: ME260P | Brand: Morayelec

1. Introduction

This manual provides essential information for the safe and efficient operation of the Moray ME260P Digital Pressure Gauge. Please read these instructions carefully before use to ensure proper handling and to prevent potential hazards.

Overview

The ME260P Digital Pressure Gauge utilizes high-precision pressure-sensitive elements and a 24-bit AD processing chip with an industrial MCU for circuit control. It is designed for real-time monitoring and calibration of field pressure in various applications.

Safety Information

Warning!

Indicates a potential hazard which, if not avoided, could result in serious loss or injury.

Caution!

Indicates a potential hazard which, if not avoided, could result in minor injury, property damage, or environmental hazards.

Information!

Provides useful tips, suggestions, and information for efficient and fault-free operations.

The digital pressure gauge must be used within its specified scope of application and in compliance with relevant accident prevention and general safety regulations. The manufacturer is not liable for damages resulting from improper use, non-compliance with instructions, or unauthorized repairs.

Product demonstration video, highlighting key features and operational aspects of the digital pressure gauge.

2. Product Outline and Components

The Moray ME260P Digital Pressure Gauge is designed for durability and ease of use. Below is an outline of its main components and features.

Dimensional drawing of Moray ME260P Digital Pressure Gauge
Dimensional drawing of the ME260P digital pressure gauge, showing its physical dimensions and connection thread (M20*1.5).
Detailed view of pressure gauge components: pressure conduit hole, SS 304 housing, IP65 enclosure, and battery compartment
Detailed view of the 3mm pressure conduit hole for anti-shocking, SS 304 housing for anti-corrosion/anti-aging, IP65 enclosure for outdoor work, and the 4.5V DC (3*AAA) standard battery compartment for easy changing.
Digital pressure gauge display features: LCD High Definition, Backlight, Range Percentage, Zero Clearing
The gauge features an LCD High Definition display with backlight, showing range percentage and supporting zero clearing.

3. Technical Specifications

FeatureSpecification
Model NumberME260P
Measuring Range0 ~ 100MPa (various options available, e.g., 0-1MPa, 0-10MPa, 0-40MPa)
Pressure Accuracy±1% FS
Overpressure2XFS
Display TypeDIGITAL LCD with backlight, 4-digit display
Resolution4 bit
Compensation Temperature0 ℃ ~ +50 ℃
Operating Temperature-10 ℃ ~ +60 ℃
Medium Temperature-20 ℃ ~ +80 ℃
Storage Temperature-20 ℃ ~ +60 ℃
Power Supply4.5V DC (3 x AAA batteries, not included)
Current Consumption<4.5mA (backlight not turned on)
Shell MaterialABS
Pressure Sensitive Film MaterialStainless steel 17-4PH
Seal Ring MaterialNBR
Protection LevelIP65
Dial Diameter80MM (2 - 3.9 Inches)
Connect Screw ThreadExternal thread (standard) / Internal (customizable, e.g., G1-4, NPT1-4)
Medium TypeGas, water, oil, pure liquid
Pressure TypeGauge pressure (standard), absolute pressure or sealed gauge pressure
CertificationCE

4. Installation

  1. Select Proper Sealing: Before installing the digital pressure gauge, ensure you select a proper sealing method.
  2. Medium Compatibility: Verify that the measuring medium (gas, water, oil, pure liquid) is suitable for the instrument material (stainless steel 17-4PH diaphragm, NBR seal ring).
  3. Thread Connection: The gauge features an external thread (standard) or internal thread (customizable, e.g., G1-4, NPT1-4) for connection. Ensure a secure fit.
    Digital pressure gauge with customizable installation thread
    The installation thread can be customized to fit various systems.
  4. Avoid Diaphragm Contact:

    Caution!

    Do not touch the sensitive element diaphragm in any way during installation.

  5. Proper Handling:

    Caution!

    Do not install the instrument by rotating the instrument housing. Always use a wrench on the hexagonal fitting to tighten the connection.

5. Operation

5.1 ON/OFF and Backlight

  • Short press: Turn on the backlight.
  • Long press: Turn the device ON or OFF.
Digital pressure gauge display with ON/OFF button highlighted
The ON/OFF button also controls the backlight with a short press.

5.2 Zero Clearing

This function allows you to set the current pressure as the zero point for measurement.

  • Long press "ZERO/PEAK" button: Use the current pressure as the measurement zero point.

Information!

  • The clearing range is within 5% of the full scale.
  • If there is a deviation in the display zero point, use the zero clearing function.
  • It is recommended to use the Zero clearing function before each use.
Digital pressure gauge with ZERO/PEAK button highlighted for zero calibration
One-key zero calibration feature for accurate measurements.

5.3 Peak View (Max/Min Value Holder)

This feature allows you to view the maximum and minimum pressure values recorded.

  • Short press "ZERO/PEAK" button: Switch between Maximum value and Minimum value.
  • Long press "ZERO/PEAK" button: Clears the maximum or minimum value.

5.4 Unit Switching

The gauge supports multiple pressure units for versatility.

  • Short press "UNIT" button: Cycle between pressure units (bar, psi, kPa, MPa, kg/cm²).

Information!

After unit switching, the measured value will display the current equivalent pressure.

Digital pressure gauge with UNIT button highlighted for multi-unit switching
Multi-unit switch allows selection between bar, psi, kPa, MPa, and kg/cm².

6. Maintenance

6.1 Battery Replacement

Caution!

Turn off the power supply of the pressure gauge before replacing the battery.

Battery replacement can be performed by the user. The gauge uses 3 x AAA alkaline batteries (4.5V).

Information!

The battery life is dependent on the battery capacity and the operating mode of the instrument. Refer to the Technical Specifications section for details.

Digital pressure gauge showing the battery compartment for 3 AAA batteries
The battery compartment is designed for easy access and replacement of 3 AAA batteries.

6.2 General Maintenance

The digital pressure gauge is generally maintenance-free under normal operation. Any repairs should only be carried out by the manufacturer.

6.3 Recalibration

The instrument can be recalibrated if needed. Please contact your dealer for recalibration services.

7. Troubleshooting and Fault Codes

If the digital pressure gauge displays a fault code, refer to the following table for handling methods:

Fault CodeDescriptionAction
Er-HExceeds the upper measurement range.Check if the measuring range of the pressure gauge is appropriate for the application.
Er-SPressure sensitive element failure.Return the unit to the manufacturer for maintenance.
Er-DData loss.Return the unit to the manufacturer or customer for calibration.
Er-ZZero clearing failure.Refer to section 5.2 (Zero Clearing) of this manual for instructions.

8. User Tips

  • Stable Readings: Ensure stable mounting for consistent and accurate pressure readings.
  • Unit Versatility: Utilize the multi-unit switching feature (bar, psi, kPa, MPa, kg/cm²) to adapt the gauge to various measurement requirements and industry standards.
  • Easy Data Recording: Take advantage of the clear digital display for easier and more accurate recording of pressure values compared to traditional mechanical gauges.
  • Outdoor Use: The IP65 rating makes this gauge suitable for outdoor work, providing protection against dust and water jets.

9. Disassembly, Repair, and Disposal

9.1 Disassembly

Warning!

The residual medium in the digital pressure gauge can be hazardous to people, the environment, and equipment. Always take adequate precautionary measures. Disconnect the test and calibration only after the system has no pressure.

9.2 Repair

Warning!

When returning the digital pressure gauge to the factory for repair, ensure the product is free of any harmful substances (e.g., acid, alkali, or other solutions).

When returning the digital pressure gauge, use the original packaging materials or select appropriate packaging materials to prevent damage during transit.

9.3 Disposal

Improper disposal can cause environmental pollution. Dispose of instrument components and packaging materials in an environmentally compatible manner, in accordance with national waste disposal regulations.

10. Additional Resources

A detailed user manual in PDF format is available for download:

View User Manual (PDF)

© 2024 Morayelec. All rights reserved.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.