1. Introduction
This manual provides essential information for the safe and efficient operation of the Moray ME260P Digital Pressure Gauge. Please read these instructions carefully before use to ensure proper handling and to prevent potential hazards.
Overview
The ME260P Digital Pressure Gauge utilizes high-precision pressure-sensitive elements and a 24-bit AD processing chip with an industrial MCU for circuit control. It is designed for real-time monitoring and calibration of field pressure in various applications.
Safety Information
Warning!
Indicates a potential hazard which, if not avoided, could result in serious loss or injury.
Caution!
Indicates a potential hazard which, if not avoided, could result in minor injury, property damage, or environmental hazards.
Information!
Provides useful tips, suggestions, and information for efficient and fault-free operations.
The digital pressure gauge must be used within its specified scope of application and in compliance with relevant accident prevention and general safety regulations. The manufacturer is not liable for damages resulting from improper use, non-compliance with instructions, or unauthorized repairs.
2. Product Outline and Components
The Moray ME260P Digital Pressure Gauge is designed for durability and ease of use. Below is an outline of its main components and features.



3. Technical Specifications
| Feature | Specification |
|---|---|
| Model Number | ME260P |
| Measuring Range | 0 ~ 100MPa (various options available, e.g., 0-1MPa, 0-10MPa, 0-40MPa) |
| Pressure Accuracy | ±1% FS |
| Overpressure | 2XFS |
| Display Type | DIGITAL LCD with backlight, 4-digit display |
| Resolution | 4 bit |
| Compensation Temperature | 0 ℃ ~ +50 ℃ |
| Operating Temperature | -10 ℃ ~ +60 ℃ |
| Medium Temperature | -20 ℃ ~ +80 ℃ |
| Storage Temperature | -20 ℃ ~ +60 ℃ |
| Power Supply | 4.5V DC (3 x AAA batteries, not included) |
| Current Consumption | <4.5mA (backlight not turned on) |
| Shell Material | ABS |
| Pressure Sensitive Film Material | Stainless steel 17-4PH |
| Seal Ring Material | NBR |
| Protection Level | IP65 |
| Dial Diameter | 80MM (2 - 3.9 Inches) |
| Connect Screw Thread | External thread (standard) / Internal (customizable, e.g., G1-4, NPT1-4) |
| Medium Type | Gas, water, oil, pure liquid |
| Pressure Type | Gauge pressure (standard), absolute pressure or sealed gauge pressure |
| Certification | CE |
4. Installation
- Select Proper Sealing: Before installing the digital pressure gauge, ensure you select a proper sealing method.
- Medium Compatibility: Verify that the measuring medium (gas, water, oil, pure liquid) is suitable for the instrument material (stainless steel 17-4PH diaphragm, NBR seal ring).
- Thread Connection: The gauge features an external thread (standard) or internal thread (customizable, e.g., G1-4, NPT1-4) for connection. Ensure a secure fit.

The installation thread can be customized to fit various systems. - Avoid Diaphragm Contact:
Caution!
Do not touch the sensitive element diaphragm in any way during installation.
- Proper Handling:
Caution!
Do not install the instrument by rotating the instrument housing. Always use a wrench on the hexagonal fitting to tighten the connection.
5. Operation
5.1 ON/OFF and Backlight
- Short press: Turn on the backlight.
- Long press: Turn the device ON or OFF.

5.2 Zero Clearing
This function allows you to set the current pressure as the zero point for measurement.
- Long press "ZERO/PEAK" button: Use the current pressure as the measurement zero point.
Information!
- The clearing range is within 5% of the full scale.
- If there is a deviation in the display zero point, use the zero clearing function.
- It is recommended to use the Zero clearing function before each use.

5.3 Peak View (Max/Min Value Holder)
This feature allows you to view the maximum and minimum pressure values recorded.
- Short press "ZERO/PEAK" button: Switch between Maximum value and Minimum value.
- Long press "ZERO/PEAK" button: Clears the maximum or minimum value.
5.4 Unit Switching
The gauge supports multiple pressure units for versatility.
- Short press "UNIT" button: Cycle between pressure units (bar, psi, kPa, MPa, kg/cm²).
Information!
After unit switching, the measured value will display the current equivalent pressure.

6. Maintenance
6.1 Battery Replacement
Caution!
Turn off the power supply of the pressure gauge before replacing the battery.
Battery replacement can be performed by the user. The gauge uses 3 x AAA alkaline batteries (4.5V).
Information!
The battery life is dependent on the battery capacity and the operating mode of the instrument. Refer to the Technical Specifications section for details.

6.2 General Maintenance
The digital pressure gauge is generally maintenance-free under normal operation. Any repairs should only be carried out by the manufacturer.
6.3 Recalibration
The instrument can be recalibrated if needed. Please contact your dealer for recalibration services.
7. Troubleshooting and Fault Codes
If the digital pressure gauge displays a fault code, refer to the following table for handling methods:
| Fault Code | Description | Action |
|---|---|---|
| Er-H | Exceeds the upper measurement range. | Check if the measuring range of the pressure gauge is appropriate for the application. |
| Er-S | Pressure sensitive element failure. | Return the unit to the manufacturer for maintenance. |
| Er-D | Data loss. | Return the unit to the manufacturer or customer for calibration. |
| Er-Z | Zero clearing failure. | Refer to section 5.2 (Zero Clearing) of this manual for instructions. |
8. User Tips
- Stable Readings: Ensure stable mounting for consistent and accurate pressure readings.
- Unit Versatility: Utilize the multi-unit switching feature (bar, psi, kPa, MPa, kg/cm²) to adapt the gauge to various measurement requirements and industry standards.
- Easy Data Recording: Take advantage of the clear digital display for easier and more accurate recording of pressure values compared to traditional mechanical gauges.
- Outdoor Use: The IP65 rating makes this gauge suitable for outdoor work, providing protection against dust and water jets.
9. Disassembly, Repair, and Disposal
9.1 Disassembly
Warning!
The residual medium in the digital pressure gauge can be hazardous to people, the environment, and equipment. Always take adequate precautionary measures. Disconnect the test and calibration only after the system has no pressure.
9.2 Repair
Warning!
When returning the digital pressure gauge to the factory for repair, ensure the product is free of any harmful substances (e.g., acid, alkali, or other solutions).
When returning the digital pressure gauge, use the original packaging materials or select appropriate packaging materials to prevent damage during transit.
9.3 Disposal
Improper disposal can cause environmental pollution. Dispose of instrument components and packaging materials in an environmentally compatible manner, in accordance with national waste disposal regulations.
10. Additional Resources
A detailed user manual in PDF format is available for download: