Introduction
This manual provides detailed instructions for the safe and effective use of your NJTY S471PRO Mini Analog Multimeter. This universal high-precision mechanical anti-burn multimeter is designed for various electrical measurements, including AC and DC voltage, DC current, resistance, capacitance, transistor hFE, diode polarity, LED lamp testing, battery voltage, and infrared remote control emission detection. Please read this guide thoroughly before operation and keep it for future reference.

The NJTY S471PRO Mini Analog Multimeter, showcasing its compact design and multiple testing functions.
Safety Information
Always observe safety precautions when using any electrical testing equipment. Failure to do so may result in injury or damage to the meter or equipment under test.
- Ensure the multimeter is in good working condition before use.
- Do not exceed the maximum input values for any range.
- Always disconnect power to the circuit before measuring resistance or capacitance.
- When measuring high voltage (greater than 100V), operate the two test pens with one hand to prevent accidents.
- Replace batteries immediately when the resistance range cannot be adjusted to zero, indicating low battery power.
- This meter uses multiple protection circuits to prevent overload and burnout. However, always select the correct range for the measurement.
Product Overview
Familiarize yourself with the components of your S471PRO Multimeter:

Labeled diagram of the S471PRO Multimeter's front panel and terminals.
Appearance and Name
- Dial
- Mechanical Zero Adjuster (adjusts the mechanical zero position on the left)
- Transistor hFE Measurement Socket
- Gear Knob (20 gears)
- DC current "DCmA/0.5V" Jack
- Voltage Resistance "V/Ω/BATT" Jack
- Common Terminal "COM" Jack
- Ω Adjust Knob
- Infrared Receiving Tube (adaptation function, refer to actual object)
- Infrared Receiving Indicator Light (adaptation function, refer to actual object)
- Spare Fuse (Φ5×20 0.5A inside the box)
- Working fuse (Φ5×20 0.5A inside the box)
- Spare Fuse (Φ5×20 0.5A inside the box)
- 2*1.5V AAA battery Slot
Specifications
The specific specifications may vary according to different models (without prior notice). Please note when purchasing and choose freely.
| Measuring range | Sensitivity and Voltage drop | Accuracy | Error indication method | Dial Scale |
|---|---|---|---|---|
| DC Current | 0-100µA-2.5mA-25mA-500mA | 0.5V | ±3% | DCA |
| DC Voltage | 0-0.5V-2.5V-10V-50V-250V-1000V | 10KΩ/V | ±3% | DCV |
| AC Voltage | 0-10V-50V-250V-1000V | 4.5KΩ/V | ±5% | ACV |
| DC Resistance | R×1 R×10 R×100 R×1k R×100k (Center value 20) | ±10% | Ω | |
| On/off buzzer | R×3 (reference value) When it is lower than 30Ω, the buzzer will sound. | •))) | ||
| Capacitance measurement | C×0.1 C×10 C×100 C×1k C×10k | C (µF) | ||
| LI Detection | R×1: 150mA, R×10: 15mA, R×100: 1.5mA, R×1k: 150µA | LI(mA) | ||
| LV Detection | R×1-R×1k: 0-3V | LV(V) | ||
| Transistor DC Magnification hFE | R×10hFE gear zero adjustment | 0-1000 | hFE | |
| Infrared Remote Control Emission Detection | Vertical Angle ±15°, Distance 1-30cm | Red LED Indication (flashing) | ||
| Battery Level | 1.2V-1.5V-2V-3V-3.6V | RL=12Ω | BATT | |
| Audio Level dB | -10dB~+22dB | 0dB=1mW/600Ω | dB |
Dimensions: 151×82×40mm
Weight: ≤205g (without battery)

Physical dimensions of the S471PRO Multimeter.
This meter requires use under an ambient temperature of 0-40°C and a relative humidity of 20%-75%. All technical performance indicators are in line with Q/3201TYD02-2016 Industry standards and relevant clauses of IEC51 international standard.
Setup
Battery Installation
The multimeter requires 2 AAA 1.5V batteries for operation. The battery slot is located at position (14) as shown in the "Appearance and Name" diagram. To install or replace batteries:
- Open the back cover of the multimeter.
- Insert 2 AAA 1.5V batteries into the battery slot, ensuring correct polarity.
- Close the back cover securely.
If the resistance range cannot be adjusted to zero, it indicates that the battery power is exhausted and needs replacement.
Operating Instructions
Before any measurement, ensure the pointer is at the mechanical zero position. If not, adjust the "mechanical zero adjuster" (2) until the pointer aligns with zero. Insert the red test lead into the "V/Ω/BATT" (6) jack and the black test lead into the "COM" (7) jack, unless otherwise specified for a particular measurement.
1. DC Current Measurement (DCA)
For measuring 100µA to 500mA:
- Turn the gear knob (4) to the desired current range (e.g., 500mA, 25mA, 2.5mA, 100µA).
- Insert the red test lead into the "DCmA/0.5V" (5) jack and the black test lead into the "COM" (7) jack.
- Connect the test leads in series with the circuit being measured.
- Read the value on the second scale line (DCA) of the dial.

Connection for DC Current Measurement (Load current < 500mA).
Note: Current measurement requires disconnecting the circuit, which can be unsafe. For higher currents or safer measurement, consider using a clamp meter.
2. AC and DC Voltage Measurement (DCV, ACV)
For measuring DC voltage (2.5V to 1000V):
- Turn the gear knob (4) to the required DCV voltage range (e.g., 10V, 50V, 250V, 1000V).
- Insert the red test lead into the "V/Ω/BATT" (6) jack and the black test lead into the "COM" (7) jack.
- Connect the test leads across the two ends of the circuit under test.
- Read the measured value on the second scale line (DCV) of the dial.
For measuring AC voltage (10V to 1000V):
- Turn the gear knob (4) to the required ACV voltage range (e.g., 10V, 50V, 250V, 1000V).
- Insert the red test lead into the "V/Ω/BATT" (6) jack and the black test lead into the "COM" (7) jack.
- Connect the test leads across the two ends of the circuit under test.
- Read the measured value on the ACV scale line of the dial. For AC 10V, use the special AC 10V scale.
Note: For DC voltage below 0.5V, insert the red test lead into the "DCmA" jack and rotate the gear knob to the 100µA/0.5V gear.
3. DC Resistance Measurement (Ω)
- Ensure batteries are installed.
- Turn the gear knob (4) to the desired resistance range (R×1, R×10, R×100, R×1k, R×100k).
- Insert the red test lead into the "V/Ω/BATT" (6) jack and the black test lead into the "COM" (7) jack.
- Short-circuit the two ends of the test leads. Adjust the Ω Adjust Knob (8) so that the pointer aligns with the "0" position on the right side of the resistance scale line.
- Separate the test leads and connect them across the component to be measured.
- Read the value on the Ω scale line. Multiply the reading by the range multiplier (e.g., for R×10, a reading of 20 means 20 × 10 = 200Ω).
Note: Always readjust the Ω zero adjuster knob after each gear change. When measuring resistance on a circuit board, first disconnect the power supply and fully discharge any large-capacity capacitors. Avoid touching the ends of the resistor with your body during measurement to prevent additional errors.
4. Buzzer Detection (•)))
- Adjust the meter to zero as in the ohm mode.
- Turn the gear knob (4) to the "•)))" position.
- Short-circuit the red and black test leads. The buzzer should emit a long beep (approx. 2KHz).
- Connect the test leads across the circuit under test. If the resistance is lower than about 30Ω, the buzzer will beep, indicating continuity.
5. Transistor DC Parameter Measurement (hFE)
- Turn the gear knob (4) to the R×100 position.
- Short-circuit the test leads and adjust the meter to zero as in the Ω mode.
- Insert the transistor legs (emitter, base, collector) into the appropriate ebc socket of the Transistor hFE Measurement Socket (3). N-type transistors go into N-type holes, P-type into P-type holes.
- The pointer deflection indicates the transistor DC magnification (hFE).
Note: The scale is calibrated by silicon transistors; results for composite or germanium transistors are for reference only.
6. Capacitance Measurement C(µF)
- Turn the gear knob (4) to a capacitance range approximately matching the expected value (C×0.1, C×10, C×100, C×1k, C×10k).
- Adjust the meter to zero as in the Ω range.
- Connect the capacitor to both ends of the meter leads.
- The maximum needle swing indicates the capacitance. The stop position indicates the quality factor (loss resistance). Read the value on the C(µF) capacitance scale line.
| Capacitance range | C×0.1 | C×10 | C×100 | C×1k | C×10k |
|---|---|---|---|---|---|
| Measuring range | 1000pF-5 µF | 0.1 µF-500 µF | 1 µF-5000 µF | 10 µF-50000 µF | 100 µF-5F |
| Remark | Red test lead (-) | Black test lead (+) | |||
Note: Fully discharge capacitors before measurement. Polarized capacitors must be connected with correct polarity.
7. Load Voltage LV(V) / Load Current LI(mA) Measurement
This function measures voltage drop or reverse voltage drop performance parameters of components under different currents. This is useful for testing diodes, LEDs, and transistors.
The measurement method is similar to the Ω gear. Use the 0-3V scale for R×1-R×1k ranges. The full current for each gear is:
| Switch Position (Ω) | R×1 | R×10 | R×100 | R×1k |
|---|---|---|---|---|
| Full Scale Current I | 150mA | 15mA | 1.5mA | 150µA |
| Measuring Range LV | 0-3V | |||
8. Infrared Remote Control Detection
This function determines if an infrared remote control transmitter is working correctly.
- Turn the gear knob (4) to the Infrared Remote Control Detection position.
- Align the infrared transmitter head vertically with the receiving window at the lower left of the dial (within ±15° deviation).
- Press a button on the remote control. If the infrared receiving indicator light (10) flashes, the transmitter is working.
- Vary the distance (1-30cm) to determine the transmitter's output power.
Note: Ensure the transmitter head is perpendicular to the receiving window. Avoid strong direct light on the receiving window during testing.
9. Audio Level Measurement (dB)
This function measures amplifier gain and line loss in decibels, with AC 10V as the reference scale.
- Turn the gear knob (4) to the corresponding AC voltage gear (e.g., AC10V, AC50V, AC250V, AC500V).
- Connect the test leads to the circuit. If a DC voltage component is present, connect a 0.1µF DC blocking capacitor in series with the "V/Ω" jack.
- Read the value on the dB scale. If the indicated value is greater than +22dB, measure at higher voltage limits (above 50V gear) and correct according to the increase value.
| Range | Increase the value by level scale | Level Measurement Range |
|---|---|---|
| AC10V | -10~22dB | |
| AC50V | 14dB | 4~36dB |
| AC250V | 28dB | 18~50dB |
| AC500V | 34dB | 24~56dB |
10. Battery Power Detection (BATT)
- Insert the red test lead into the "V/Ω/BATT" (6) jack and the black test lead into the "COM" (7) jack.
- Turn the gear knob (4) to the BATT gear.
- Read the value on the BATT scale on the dial. "Green" indicates sufficient power, "?" indicates a bad battery, and "Red" indicates low battery and replacement is needed.
Maintenance
Fuse Replacement
The multimeter is equipped with multiple protection circuits. If a fuse burns out:
- Open the back cover of the multimeter.
- Locate the working fuse (12) and the spare fuses (11, 13) near the battery.
- Replace the burnt fuse with a new Φ5x20mm 0.5A/250V fuse.
- Close the back cover securely.
Battery Replacement
If the resistance range cannot be adjusted to zero, it indicates that the battery power in the meter has been exhausted. Replace the batteries immediately to ensure measurement accuracy and prevent battery electrolyte from overflowing and corroding the parts in the meter. Refer to the "Setup" section for battery installation instructions.
Troubleshooting
- Pointer not at zero: Adjust the mechanical zero adjuster.
- Resistance range cannot be adjusted to zero: Replace the AAA 1.5V batteries.
- Inaccurate resistance measurement: Ensure the power supply is disconnected and capacitors are discharged when measuring on a circuit board. Avoid touching resistor ends with your body.
- Fuse blown: Replace with a Φ5x20mm 0.5A/250V fuse.
- Overload/Burnout: Always select the correct measurement range to prevent damage.
User Tips
- When measuring high voltage (above 100V), always handle test leads with one hand to minimize risk.
- For accurate resistance measurements, it is best to remove the component from the circuit.
- Always ensure polarized capacitors are connected with the correct polarity during capacitance measurement.
Standard Configuration of Product

Contents of the NJTY S471PRO Multimeter package.
- Multimeter (1 unit, with 2 fuses Φ5x20mm 0.5A/250V inside the box)
- Test leads (1 pair)
- Instruction manual (1 unit)