NJTY 58A

Contents

NJTY 58A Digital Multimeter User Manual

High-Precision Digital Display Smart Burn-Proof Household Electrician Capacitance Meter

1. Introduction

The NJTY 58A (also referred to as T58A) is a handheld digital multimeter designed for high-precision measurements in electrical applications. It features True RMS measurement, a large 9999-count LCD digital display with backlight, overload protection, battery undervoltage indication, and input light indication. This instrument is ideal for professionals, factories, schools, enthusiasts, and home use, designed and produced according to the international electrical safety standard IEC-61010.

NJTY 58A Multimeter with key features
Figure 1: NJTY 58A Multimeter highlighting True RMS, diode measurement, capacitance up to 100000uF, gear socket indicator light, and Celsius/Fahrenheit measurement.

2. Safety Information

Always read and understand all safety information and operating instructions before using this meter. Failure to do so may result in electric shock, personal injury, or damage to the meter or equipment under test.

2.1 General Safety Instructions

  • Comply with all standard safety regulations regarding protection against electric shocks and prevention of instrument misuse.
  • Use only the test leads provided with the meter. Inspect them for damage before each use.
  • Do not use the meter if its appearance or the test leads are broken.
  • Incorrect use may compromise the meter's safety functions.
  • Exercise caution when working near bare conductors or busbars.
  • Do not use this instrument in environments with explosive gas, vapor, or dust.
  • Ensure the correct input function and range are selected for the measurement.
  • Input values must not exceed the specified limits for each range to prevent damage.
  • Do not touch unused input terminals when the meter is connected to a circuit.
  • Be cautious of electric shock when measuring voltages exceeding 60V DC or 36V AC.
  • When using test leads, keep your fingers behind the guard rings.
  • Before changing the measurement range, ensure the test leads are disconnected from the circuit under test.
  • For DC measurements, always confirm the absence of AC voltage first, then select an appropriate DC voltage range.
  • Before measuring resistance, diodes, capacitance, or performing continuity tests, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.
  • Do not measure resistance or perform buzzer tests on live circuits.
  • Before current measurement, check the meter's fuse.
  • The meter uses three AAA 1.5V batteries. Replace them immediately when the low voltage symbol (⚠) appears on the display to avoid incorrect readings and potential hazards.
  • Do not measure Category II voltages exceeding 1000V or Category III voltages exceeding 600V.
  • Do not operate the instrument with its housing or parts of the housing removed.

2.2 Safety Symbols

The following symbols are used on the meter and in this manual:

SymbolDescription
⚠Warning: Important safety signs. Refer to the instruction manual before use.
∿AC (Alternating Current)
—DC (Direct Current)
≈AC or DC
⟂Ground
□Double insulation protection
⚊Fuse
CEMeets EU standards
⚡High voltage warning
CAT. IICategory II: 1000V overvoltage protection
CAT. IIICategory III: 600V overvoltage protection

3. Setup and Accessories

3.1 Included Accessories

Your NJTY 58A Multimeter package should contain the following items:

  • 1 x NJTY 58A Digital Multimeter
  • 1 x Instruction Manual (this document)
  • 1 x Pair of Test Leads (red and black)
  • 1 x Thermocouple (for temperature measurement)
  • 3 x 1.5V AAA Batteries (may be pre-installed or separate)
  • 1 x Canvas Bag (for storage and transport)
NJTY 58A Multimeter with accessories
Figure 2: NJTY 58A Multimeter shown with its instruction manual, test leads, and temperature probe.

3.2 Battery and Fuse Replacement

To avoid electric shock or personal injury, always turn off the instrument and disconnect all test leads from the input sockets before replacing batteries or fuses. Use only the specified fuse types.

When the battery low voltage symbol (⚠) appears on the display, replace the batteries immediately. If a fuse blows, replace it with the correct type.

  1. Turn off the power of the instrument.
  2. Pull all test leads out of the input sockets.
  3. Use a screwdriver to loosen the screws fixing the battery cover.
  4. Remove the battery cover.
  5. Remove the old batteries or the damaged fuse.
  6. Replace with new 3 x 1.5V AAA batteries, observing polarity. If replacing a fuse, use a 600mA/250V fuse for mA range and a 10A/250V fast-blow fuse for 10A range.
  7. Install the battery cover and securely close the screws.

4. Operating Instructions

4.1 Meter Overview and Controls

The NJTY 58A Multimeter features a clear LCD display, a rotary gear switch for function selection, and several function buttons. Models A, B, and C may have slight variations in features (e.g., Model A has Infrared remote control detection and Transistor measurement, Model C has Inductance measurement).

Key Components:

  • LED Display: Shows measurement readings, units, and status indicators.
  • NCV Sensing Area: For non-contact voltage detection.
  • NCV and Buzzer Indicator Light: Visual and audible alerts for NCV.
  • Gear Rotary Switch: Main selector for measurement functions (Voltage, Current, Resistance, Capacitance, Diode, Continuity, Frequency, Temperature, NCV, etc.).
  • Function Buttons:
    • FUNC: Short press to cycle through sub-functions within a gear (e.g., AC/DC, °C/°F, Diode/Continuity).
    • AREL HOLD: Long press for Relative Value measurement; short press for Data Hold.
    • Backlight/Lighting Button: Short press to turn backlight on/off; long press for simultaneous lighting and backlight.
    • RANGE (Model B/C): For manual range selection.
    • MAX MIN: Toggles between maximum and minimum value display.
    • PEAK (Model B/C): For Peak-to-Peak measurement.
  • Input Sockets:
    • VΩHz°C: Positive input for voltage, resistance, frequency, temperature, diode, capacitance, continuity. (Connect red test lead).
    • COM: Common input for all measurements. (Connect black test lead).
    • mA: Input for milliampere current measurements.
    • 10A: Input for high current (up to 10A) measurements.
    • µA mA L (Model C): Input for microampere, milliampere, and inductance measurements.

4.2 Auto Power Off

The meter will automatically power off after approximately 15 minutes of inactivity to conserve battery life. An audible prompt will sound before shutdown. Press any button to restart the meter from hibernation mode.

4.3 Voltage Measurement (ACV & DCV)

Do not measure voltages greater than 1000V DC or 750V AC RMS. Do not apply more than 1000V DC or 750V AC RMS between the common and earth terminals.

  1. Rotate the gear switch to the 'V' (DC Voltage) or '~V' (AC Voltage) position. The corresponding input socket indicator light will flash.
  2. Connect the black test lead to the COM jack and the red test lead to the VΩHz°C jack.
  3. Connect the test leads across the circuit or component to measure the voltage.
  4. Read the voltage value on the LED display. For DC, polarity will be indicated.
Multimeter measuring AC voltage
Figure 3: Multimeter measuring AC voltage, displaying 220.5V at 50Hz.

4.4 Resistance Measurement

Before measuring resistance, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.

  1. Rotate the gear switch to the 'Ω' position.
  2. Connect the black test lead to COM and the red test lead to VΩHz°C.
  3. Connect the test leads across the component to measure resistance.
  4. Read the resistance value on the display. If 'OL' is displayed, the resistance is out of range.

4.5 Diode Measurement & Continuity Test

Before performing diode or continuity tests, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.

  1. Rotate the gear switch to the '→|▶' (Diode/Continuity) position. Press FUNC to switch between diode test and continuity test.
  2. Connect the black test lead to COM and the red test lead to VΩHz°C.
  3. Diode Test: Connect the red lead to the anode and the black lead to the cathode of the diode. The display shows the forward voltage drop. Reverse the leads to check for 'OL' (open circuit).
  4. Continuity Test: Connect the test leads across the circuit or component. A buzzer will sound and the indicator light will turn on if resistance is below approximately 50 ohms.

4.6 Capacitance Measurement

Before measuring capacitance, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.

  1. Rotate the gear switch to the '——F' (Capacitance) position.
  2. Connect the black test lead to COM and the red test lead to VΩHz°C.
  3. Connect the test leads across the capacitor.
  4. Read the capacitance value on the display. Large capacitors may take some time to measure.

4.7 Current Measurement (AC & DC)

Always connect the meter in series with the circuit under test for current measurement. Do not connect in parallel. Pre-judge the current magnitude to select the correct input jack and range. Check the fuse before measurement.

  1. Rotate the gear switch to the 'mA' or '10A' (AC or DC current) position. Press FUNC to switch between AC and DC.
  2. Connect the black test lead to COM.
  3. For currents up to 1000mA, connect the red test lead to the mA jack. For currents between 1A and 10A, connect the red test lead to the 10A jack.
  4. Open the circuit where current is to be measured and connect the meter in series.
  5. Power on the circuit and read the current value. If 'OL' is displayed, the current exceeds the selected range; turn off power, select a higher range, and re-measure.
Multimeter measuring AC current
Figure 4: Multimeter measuring AC current from a light bulb, displaying 263.3mA at 50Hz.

4.8 NCV (Non-Contact Voltage) Test

  1. Rotate the gear switch to the 'NCV' position.
  2. Place the top of the meter close to the conductor suspected of carrying AC voltage.
  3. If AC voltage is detected, the NCV indicator light will illuminate (low-yellow, high-red) and the buzzer will sound with varying frequencies.
Multimeter performing NCV test
Figure 5: Multimeter performing a non-contact voltage test.

4.9 Frequency Measurement

Do not test voltages higher than 250V to prevent electric shock or damage.

  1. Rotate the gear switch to the 'Hz%' position.
  2. Connect the black test lead to COM and the red test lead to VΩHz°C.
  3. Connect the test leads across the circuit to measure frequency.
  4. Read the frequency value and duty cycle (%) on the display.

4.10 Temperature Measurement

  1. Rotate the gear switch to the '°C/°F' position. The meter will display the ambient temperature.
  2. Remove the test leads and connect the thermocouple to the VΩHz°C and COM inputs, observing polarity.
  3. Place the thermocouple tip on the surface of the object or in the environment to be measured.
  4. Read the temperature value on the display.

4.11 Transistor Measurement (Model A Only)

Do not apply voltage exceeding 36V DC or AC RMS to the common and hFE terminals.

  1. Rotate the gear switch to the 'hFE' position.
  2. Determine if the transistor is NPN or PNP type.
  3. Insert the emitter, base, and collector pins of the transistor into the corresponding holes of the hFE test socket.
  4. Read the hFE (current gain) value on the LCD.

4.12 Inductance Measurement (Model C Only)

  1. Rotate the gear switch to the 'L' position.
  2. Connect the black test lead to COM and the red test lead to the µA mA L input terminal.
  3. Connect the test leads across the inductor.
  4. Read the inductance value on the display.

4.13 Low Impedance Measurement (LoZ - Model A Only)

This function automatically measures AC and DC voltage with a low input impedance to reduce ghost voltages.

  1. Rotate the gear switch to the 'LoZ' position.
  2. Connect the black test lead to COM and the red test lead to VΩHz°C.
  3. Connect the test leads in parallel to the power supply or load to be measured.
  4. Read the voltage value on the display.

5. Maintenance

5.1 General Care

  • To avoid electric shock or damage, do not allow moisture inside the meter.
  • Clean the instrument case regularly with a damp cloth and a small amount of mild detergent. Do not use abrasive cleaners or chemical solvents.
  • Dirty or moist input sockets can cause oxidation and affect readings. To clean, turn off the meter, unplug leads, remove dirt, and clean with a cotton ball moistened with detergent or lubricant.

5.2 Long-Term Storage

  • If the meter will not be used for an extended period, remove the batteries to prevent leakage and damage.
  • Store the meter in a cool, dry place, away from high temperatures and humidity.

5.3 Service and Repair

  • If any abnormality is observed (e.g., incorrect readings, physical damage), stop using the meter immediately and send it for repair by qualified personnel.
  • When opening the instrument case for service, ensure all power is disconnected and take precautions against static electricity.
  • Only specified replacement parts should be used for servicing.

6. Troubleshooting