1. Introduction
This manual provides detailed instructions for the application, operation, and maintenance of the SNDWAY SW-523 and SW-524 Digital Sound Level Meters. Please read this manual thoroughly to ensure proper function and optimal performance of your device.
1.1. Important User Instructions
- Do not use the device in flammable or explosive environments.
- Used batteries should be disposed of according to national or local laws and regulations, not with normal household waste.
- For any quality issues or operational doubts, please contact your local agent or the manufacturer for assistance.
1.2. Brief Overview
The SNDWAY SW-523 and SW-524 sound level meters are designed for measuring environmental noise in various settings. They are suitable for noise projects, quality control, and health prevention applications. Typical usage areas include industries, offices, traffic roads, homes, and for assessing noise from audio equipment, air conditioners, and refrigerators.
2. Detail Packing List
Please check the contents of your package against the list below. If any items are missing, contact your supplier.
| NO. | Item | Unit | QTY |
|---|---|---|---|
| 1 | Sound Level Meter | PC | 1 |
| 2 | Sponge Ball | PC | 1 |
| 3 | USB Data Cable (for SW-524 only) | PC | 1 |
| 4 | 1.5V AAA Battery | PC | 3 |
| 5 | Manual | PC | 1 |
| 6 | Warranty Card | PC | 1 |
| 7 | Gift Box | PC | 1 |
3. Product Features and Parts
3.1. Key Features
- Accuracy: ±1.5dB (based on 94dB@1KHz standard sound source).
- Measurement Range: 30dB~130dB.
- Frequency Response: 31.5Hz~8.5KHz.
- Frequency Weighting: A-weighting and C-weighting.
- Response Time: FAST and SLOW modes.
- Measurement Modes: Max and Min.
- Backlight: For use in dark conditions.
- Data Hold: To freeze current readings.
- Auto Shut-off: Device powers down after ten minutes of inactivity.
- Analog Signal Output: AC and DC analog signal output.
- Computer Connectivity (SW-524 only): USB data cable for connecting to a computer, downloading data records, real-time sampling, and printing charts/data.
3.2. Names of Parts
- Sponge ball: Prevents wind interference.
- Condenser microphone sensor.
- Battery cover (located on the back).
- DC output: Noise digital signal output (10mV/dB, output impedance approx. 100 Ω).
- AC output: Noise analog signal output (output impedance approx. 600 Ω).
- USB port: For computer connection (SW-524 only).
- Adjustment switch: For calibration.
- LCD display screen.
- Control keys.
3.3. Key Functions
| Key Icon | Name | Function |
|---|---|---|
| Power (BACK) | Short press to turn on, extended press to turn off. | |
| MAX MIN | Max/Min | Short press to test maximum, press again for minimum, press again for normal measurement. |
| A/C | A/C Weighting | Short press to switch between A and C weighting. |
| FAST SLOW | Response Time | Short press to switch between FAST and SLOW response times. |
| HOLD | Data Hold | Short press to hold data, press again to return to normal data. |
| REC | Storage Record | Short press to enter storage mode, press again to exit. |
| SET | Backlight/Setting | Short press to turn backlight on/off, extended press to set warning. |
4. LCD Display
- Noise reading
- dB unit
- Measurement range indicator
- Time weighting FAST indicator
- Time weighting SLOW indicator
- MAX (maximum value) indicator
- MIN (minimum value) indicator
- Measurement dynamic display
- Current time display
- FULL (storage full) indicator
- Low power indicator
- USB connecting computer indicator
- Storage mode indicator
- "OVER" warning: Lights when noise exceeds 130dB
- "UNDER" warning: Lights when noise is below 30dB
- Frequency weighting: A/C indicator lights when A/C weighting is selected.
5. Specifications
| Parameter | Value |
|---|---|
| Accuracy | ±1.5dB (at reference sound pressure level, 94dB@1KHz) |
| Frequency Response | 31.5Hz~8.5KHz |
| Dynamic Range | 50dB |
| Measurement Range | 30~130dBA, 35~130dBC |
| Frequency Weighting Characteristics | A-weight and C-weight |
| Digital Display | 4 digits, resolution 0.1dB, sampling frequency 2 times/second (SW-523), 3 and a half (SW-524) |
| Dynamic Scale Display | 30~130dB (1 scale represents 2dB or 1dB) |
| Overload Indication | "OVER" and "UNDER" Symbolic representation |
| AC Signal Output | Yes |
| DC Signal Output | Yes |
| Response Time | FAST (fast) and SLOW (slow) |
| Microphone | 6mm condenser microphone |
| Size Value Lock | MAX and MIN |
| Battery | 3x1.5V AAA |
| Working Temperature and Humidity | 0°C~40°C, 10%~80%RH |
| Storage Temperature and Humidity | -10°C~60°C, 10%~70%RH |
| Time Display | Yes (SW-524 only) |
| Storage Function | 1 record/second, continuous storage for 17 hours (SW-524 only) |
| Communicate with Computer | Communication via MicroUSB (SW-524 only) |
| Alarm Function | High limit alarm can be set (SW-524 only) |
| Size | 215mm*63.5mm*27mm |
6. Setup
6.1. Battery Installation
To install batteries, open the battery cover located on the back of the meter. Insert three 1.5V AAA batteries, ensuring correct polarity. Close the battery cover securely.
6.2. Computer Connection (SW-524 only)
The SW-524 model supports connection to a computer for advanced functions such as downloading recordings, real-time sampling, and printing charts and data. The necessary software package is typically provided as a 'NoiseMeter.rar' file. After unzipping, refer to the installation and operation instructions within the software package before proceeding.
The software allows for data storage every minute, continuous storage for up to 17 hours, and continuous coexistence of data up to 61200 groups.
To monitor data in real-time, click the relevant icon in the software (e.g., 'real time sampling'). The software acquires data from the instrument every second.
The software also provides an interface for exporting and downloading recorded data.
7. Operation
7.1. Noise Measurement
Press the Power key to turn on the device. After approximately two seconds, the meter will enter measurement mode, and the noise reading will be displayed on the screen.
- If the reading is below 30dB, the "UNDER" indicator will light up.
- If the reading is over 130dB, the "OVER" indicator will light up.
7.2. Warning Setting (SW-524 only)
To set a warning value, perform an extended press on the SET key. The screen will display "AL" flashing along with the current warning value. Use the HOLD or REC keys to adjust the warning value. An extended press on either of these keys will increase the adjustment range. Short press the SET key to store the new warning value, or short press the Power (BACK) key to cancel the setting mode.
Note: The warning value will reset to 130dB every time the device is turned on to prevent misoperation.
When the noise being tested exceeds the set warning value, the meter will emit a "beep beep beep" sound and stop testing, with the tested value flashing on the screen. Testing will resume once the warning sound stops. If the noise remains above the warning value, the warning sound will continue.
7.3. Frequency Weighting
The meter offers two types of frequency weighting:
- A-weighting: Simulates the frequency characteristics of low-intensity noise (below 55dB), primarily used for civil applications like environmental noise testing in residential areas.
- C-weighting: Simulates the frequency characteristics of high-intensity noise, mainly for industrial measurements such as noise in aircraft industries or engine noise.
Upon turning on, the meter defaults to A-weighting. Short press the A/C key to switch to C-weighting.
7.4. Response Time
The meter has two response modes:
- FAST mode: Suitable for circumstances where noise changes rapidly, closely mimicking the reaction of human ears. This is the default mode upon turning on.
- SLOW mode: Ideal for steady noise conditions.
Short press the FAST/SLOW key to switch between FAST and SLOW modes.
7.5. Measurement Mode (MAX/MIN)
After turning on, the meter begins normal noise testing. Short press the MAX/MIN key to enter maximum measurement mode. Press it again to enter minimum measurement mode. Pressing it a third time will return to general measurement mode.
7.6. Data Hold
Short press the HOLD key to freeze the current reading on the display. Short press it again to release the data hold and resume live measurement.
7.7. Data Storage (SW-524 only)
Short press the REC key to enter storage mode. The device will store a set of data every second, and the REC indicator will flash. The meter can continuously store data for up to 17 hours. If the storage memory becomes full, the FULL indicator will flash. To clear the storage, short press the Power (BACK) key. While in storage mode, A/C weighting conversion, FAST/SLOW mode changes, MAX/MIN mode, and data hold functions are unavailable.
7.8. Time Setting (SW-524 only)
Perform an extended press on the SET key to enter the warning value setting mode. Short press it again to enter the time setting mode. The display will show the year, month, and day flashing. Use the HOLD or REC keys to adjust the values. Short press the SET key to store the current setting and move to the next parameter (e.g., from year to month). During the setting process, you can short press the Power (BACK) key to exit.
8. Device Adjustment (Calibration)
This device is factory-adjusted. However, it may require adjustment if it has been unused for a long period or if the sensor has been changed. It is recommended to adjust the device annually using a standard 94dB@1KHz sound source.
- Conditions Setting: Set the meter to A-weighting, FAST response time, and the 30~130dB measurement range.
- Microphone Placement: Carefully insert the microphone into the hole of the standard sound source.
- Adjustment: Turn on the sound source and adjust its output to 94dB. Use a small slot-type screwdriver to adjust the calibration knob located inside the hole on the bottom of the device until the screen displays 93.8dB.
9. Maintenance
9.1. Battery Replacement
- If the device will not be used for an extended period, remove the batteries to prevent corrosion from leakage.
- Replace batteries as soon as possible if the low power indicator is shown on the screen.
9.2. Surface Cleaning
Avoid using ethyl alcohol or other diluents as they can have a corrosive effect on the device's body surface. Instead, gently clean the device with a soft, damp cloth.
10. Common Problems
10.1. No Readings on Screen
- No power or damaged batteries: Replace the batteries.
10.2. No Connection with Computer (SW-524 only)
- USB data cable damaged: Replace the USB data cable.
- Computer ports damaged: Try connecting to a different USB port on your computer.
- No Windows system: The software requires Windows (XP, Win7, Win8, Win10). Ensure your computer meets this requirement.
- USB port of the device damaged: Contact factory service for repair.
11. User Tips
Based on general usage and common inquiries:
- Always ensure the sponge ball is properly attached to the microphone to minimize wind noise, especially when measuring outdoors.
- For consistent readings, try to maintain a consistent distance and angle between the meter's microphone and the noise source.
- When comparing noise levels, use the same frequency weighting (A or C) and response time (FAST or SLOW) settings for all measurements.
- If using the SW-524 with computer software, regularly back up your stored data to prevent loss.
12. Warranty and Support
This product comes with a warranty card included in the package. Please refer to the warranty card for specific terms and conditions regarding warranty coverage. For technical support, service, or inquiries, please contact your local distributor or the manufacturer directly.
A PDF version of the user manual is available for download. You can find it at: User manual (PDF).