1. Introduction
The NJTY T21L is a multi-purpose digital multimeter designed for high measurement accuracy, fast response speed, and a high safety level. It features a special chip with up to 6000 counts, composed of high-precision AD and a high-speed digital processor. This ensures accurate measurements, high resolution, and fast calculation speeds, including high-speed True RMS calculation for AC 1KHz signals. The device is software calibrated, maintaining accuracy over long-term use.
Its ergonomic design makes it suitable for various industrial applications. The T21L offers numerous measurement functions and a user-friendly interface, catering to professional engineers and maintenance personnel. It can measure DC and AC voltage, DC and AC current, resistance, capacitance, inductance, temperature, diode, and perform circuit continuity testing.
For convenience in low-light conditions, the instrument is equipped with a backlight and an illumination lamp.
2. Safety Information
When using this meter, always comply with standard safety regulations to prevent electric shocks and misuse of the instrument. To ensure personal safety, always use the provided test leads, inspect them for damage before use, and ensure they are intact.
Safety Signs
| Symbol | Description |
|---|---|
![]() | Warning |
| ~ | AC (alternative current) |
| --- | DC (direct current) |
![]() | Ground |
| Π | Double insulation |
| CE | European union standard |
![]() | High voltage warning |
| CAT II | II category overvoltage protection |
Precautions for Use
- Avoid using the meter near devices with large electromagnetic disturbances, as this can cause unstable readings or large errors.
- Do not use the meter if its appearance or the test leads are broken.
- Incorrect use may compromise the meter's safety functions.
- Exercise caution when working around bare conductors or busbars.
- Do not use this instrument near explosive gas, vapor, or dust.
- Always select the correct input function and range for the measurement.
- Do not switch ranges while measuring voltage or current.
- Ensure the input value does not exceed the specified limit for each range to prevent damage.
- Do not touch unused input terminals when the instrument is connected to a circuit under test.
- When measuring voltages exceeding 60 VDC or 30 VAC, use caution to prevent electric shock.
- When using a test lead, keep your finger behind the guard ring.
- Before changing the range, ensure the test leads have been disconnected from the circuit under test.
- Before measuring resistance, diode, capacitance, or performing a continuity test, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.
- Do not measure resistance on a live circuit or perform a buzzer test on a live circuit.
- Before current measurement, check the meter's fuse.
- Before connecting the meter to a circuit for current measurement, turn off the power of the circuit under test.
- When repairing TVs or measuring power conversion circuits, be cautious of high-amplitude voltage pulses to avoid damaging the meter.
3. Product Overview
Panel Diagram
- Illumination (Flashlight)
- NCV sensing area
- LED display
- Buzzer
- Backlight/illumination button
- "FUNC/Hz" selection button
- Gear rotary switch
- VΩmAL°C input socket
- COM input socket
- 10A input socket
- Lock keys (HOLD button)
- NCV indicator
Monitor Symbols Explanation
| Symbol | Description |
|---|---|
![]() | Battery low voltage display |
![]() | Auto power off |
| - | Negative polarity input indication |
| AC | Alternative input indication |
| DC | Direct input indication |
| ))) | In continuity test mode |
![]() | In diode test mode |
| L | Inductance measurement |
| AUTO | Auto ranging mode |
| HOLD | Data hold mode |
| °C, °F | Temperature |
| % | Duty cycle |
| NCV | Non-contact voltage detection mode |
| V, mV | Volt, unit of Voltage |
| A, mA, µA | Ampere, unit of current |
| Ω, kΩ, MΩ | Ohm, unit of resistor |
| Hz, KHz, MHz | Hz, hertz, unit of frequency |
| mF, µF, nF, pF | Farah, unit of capacitance |
| H, mH, µH | Inductance unit |
4. Setup
Battery Installation
The instrument uses three AAA 1.5V batteries as its power supply. Ensure batteries are installed correctly in the battery compartment. When the low voltage symbol (
) appears on the display, replace the batteries immediately. Insufficient battery power can lead to incorrect readings, posing a risk of electric shock or personal injury.
To replace batteries:
- Turn off the instrument.
- Remove all test leads from the input sockets.
- Use a screwdriver to loosen the screws securing the battery cover.
- Remove the battery cover.
- Remove the old batteries.
- Replace with new 3x1.5V AAA batteries.
- Install the battery cover and tighten the screws.
Accessories
The package includes:
- 1x Instruction Manual
- 1x Pair of Test Leads
- 1x Thermocouple
- 3x 1.5V AAA Batteries (Note: Some listings may state 4.3x1.5V AAA AA batteries, but the device uses 3xAAA)
5. Operation
On/Off Operation
Turn the gear rotary switch to any measurement position to automatically power on the device. If there is no operation for approximately 15 minutes after powering on, the instrument will emit a beep sound, indicating it has entered sleep mode. To restart, press any button or rotate the gear switch.
Button Function Description
- "HOLD" Key (Data Hold)
- During measurement, short press the HOLD key to freeze the current display value (voltage, current, resistance, capacitance, inductance). Short press again or rotate the gear switch to exit this mode.
- Press and hold the HOLD button to power on the meter. Three "di-di-di" beeps will indicate that the automatic shutdown function is canceled. The device will restart and restore normal operation.
- "FUNC/Hz" Key (Function Selection)
- When measuring AC/DC voltage or AC/DC current, short press to switch between AC and DC modes. When in AC voltage or AC current measurement, long press "FUNC/Hz" to measure the frequency of 0-1KHz signals with voltage below 250V.
- In the
/
/
range, short press "FUNC/Hz" to cycle through resistance, capacitance, continuity, and diode measurement modes.
- Backlight and Lighting Functions
- Short press the
button to turn on the backlight. Short press again to turn it off. If there is no operation for 30 seconds, the backlight will turn off automatically. - Long press the
button to turn on the flashlight, and the backlight will also light up. Short press again to turn off both. If there is no operation for 30 seconds, both will turn off automatically. - If the backlight is on, long pressing the button for 2 seconds will also turn on the flashlight. If there is no operation for 30 seconds, both will turn off.
The multimeter with its integrated flashlight activated. - Short press the
ACV and DCV Measurement
⚠ Do not measure any voltage greater than AC600V or DC600V to prevent electric shock or damage to the instrument.
- Turn the rotary switch to the "mV" or "V" position, and press "FUNC/Hz" to select DC or AC.
- Connect the black test lead to the COM input socket and the red test lead to the V input socket.
- Use the other ends of the test leads to measure the voltage of the circuit under test (connected in parallel).
- The display will show the measured voltage value. When measuring DC voltage, if the red test lead touches the negative pole, a "-" polarity symbol will be displayed.
Notice:
- In the DC or AC 600mV and 6V ranges, if no test lead is input or connected, the meter may show some readings. To zero the display, short-circuit the "V-Ω" and "COM" terminals.
- Under AC voltage function, long press "FUNC/Hz" to measure the frequency of the AC voltage source.
- The AC voltage value measured by this meter is the True RMS value. This is accurate for sine waves and other waveforms (without DC offset) like square waves, triangle waves, and step waves.
Resistance Measurement
⚠ To avoid damage to the meter or device under test, all power to the circuit under test should be cut off before measuring resistance, and all high-voltage capacitors should be fully discharged.
- Turn the rotary switch to the
position and press "FUNC/Hz" to select the "Ω" range. - Connect the black test lead to the COM input socket and the red test lead to the "V/Ω" input socket.
- Use the other ends of the test leads to measure the resistance of the circuit under test.
- The display will show the measured resistance value.
Notice:
- The resistance value measured on a circuit may differ from the rated value of the resistor due to parallel resistance and capacitance.
- When measuring low resistance, short-circuit the two test leads first to read their resistance value. Subtract this value from the measured resistance of the circuit under test for accuracy.
- If the display shows "OL", the measured value exceeds the range. Switch to a higher range.
- If the meter is open circuit, the display will show "OL".
Diode Measurement
⚠ To avoid damage to the meter or device under test, all power to the circuit under test should be cut off before measuring diodes, and all high-voltage capacitors should be fully discharged.
- Turn the rotary switch to the
position and press "FUNC/Hz" to select the "
" range. - Connect the black test lead to the COM input socket and the red test lead to the V/Ω input socket.
- Connect the black test lead to the cathode and the red test lead to the anode of the diode under test.
- The meter will display the forward conduction voltage value. If the polarity is reversed, the meter will display "OL".
Buzzer Test (Continuity)
⚠ To avoid damage to the meter or device under test, all power to the circuit under test should be cut off before measuring continuity, and all high-voltage capacitors should be fully discharged.
- Turn the rotary switch to the
position and press "FUNC/Hz" to select the "
" range. - Connect the black test lead to the COM input socket and the red test lead to the V/Ω input socket.
- Use the other ends of the test leads to measure the resistance of the circuit under test. If the resistance is less than 50Ω, the indicator light will illuminate, the buzzer will sound continuously, and the display will show the measured resistance value.
Capacitance Measurement
⚠ To avoid damage to the meter or device under test, all power to the circuit under test should be cut off before measuring capacitance. All high-voltage capacitors should be fully discharged.
- Turn the rotary switch to the
position and press "FUNC/Hz" to select the "F" range. - Connect the black test lead to the COM input socket and the red test lead to the "F" input socket.
- Touch the two terminal pins of the capacitor with the test leads. The display will show the measured capacitance value.
Notice:
- When measuring large capacitance, it may take some time for the reading to stabilize.
- When measuring polarized capacitors, pay attention to the correct polarity to avoid damaging the instrument and ensure measurement accuracy.
Frequency Measurement
⚠ Do not test any voltage higher than 250V to prevent electric shock or damage to the instrument.
- Turn the rotary switch to the "Hz%" position.
- Connect the black test lead to the COM input socket and the red test lead to the "Hz" input socket.
- Use the other ends of the test leads to measure the frequency of the circuit under test.
- Press the "FUNC/Hz" key to display frequency and duty cycle respectively.
Current Measurement
⚠ Never attempt to make current measurements on a circuit when the open circuit voltage to ground exceeds 250V. If the fuse is blown during measurement, the meter may be damaged or you may be injured. To avoid damage to the meter or the equipment under test, please check the fuse of the meter before measuring current.
When measuring, ensure the correct input socket, function range, and range are used. When the test leads are plugged into the current input socket (µA, mA, A), do not connect the two test leads in parallel to any circuit, as this risks burning out the instrument or causing electric shock.
- Turn the rotary switch to the "µA", "mA", or "A" position, and press "FUNC/Hz" to select AC or DC.
- Connect the black test lead to the COM input socket. If the measured current is less than 600mA, connect the red test lead to the VΩMA input socket. If the measured current is between 600mA and 10A, connect the red test lead to the 10A input socket.
- Disconnect the circuit to be tested. Connect the black test lead to the lower voltage end of the disconnected circuit and the red test lead to the higher voltage end.
- Connect power to the circuit and read the display. If the display shows "OL", the input exceeds the selected range. After powering off, turn the rotary switch to a higher range.
- In the ACA measurement state, long press the "FUNC/Hz" key to display the frequency of the AC signal 0-1KHz.
NCV Test (Non-Contact Voltage Detection)
- Rotate the rotary switch to the "NCV" position.
- Place the top of the meter (NCV sensing area) close to the conductor.
- If AC voltage is detected, the indicator light will illuminate based on signal strength, and the buzzer will emit different frequencies of alarm sounds.
Notice:
- Voltage may still be present even when not indicated. Do not rely solely on non-contact voltage detectors to determine voltage presence, as detection can be affected by factors like socket design, insulation thickness, and type.
- If voltage is input to the meter's input terminals, the voltage sensor indicator light may also light up due to induced voltage.
- External interference (e.g., flashlights, motors) may mistakenly trigger non-contact voltage detection.
Temperature Measurement
- Turn the rotary switch to the "°C (°F)" position. The instrument will display the ambient temperature.
- Remove the test leads and insert the thermocouple into the "COM" and "VΩMA" input terminals, observing correct polarity.
- The meter will then display the approximate temperature value transmitted from the thermocouple.
Inductance Measurement
- Turn the rotary switch to the "L" position.
- Connect the black test lead to the "COM" input socket and the red test lead to the "VΩMAL" input terminal.
- Use the test leads to touch both ends of the inductance device. The display will show the measured inductance value.
6. Maintenance
General Maintenance
To avoid electric shock or damage to the meter, do not allow the inside of the meter to get wet. Before opening the case or battery cover, turn off the power and ensure the test leads are disconnected from the circuit under test.
Clean the instrument case regularly with a damp cloth and a small amount of detergent. Do not use abrasives or chemical solvents. Dirty or damp input sockets can cause oxidation of the metal surface, which may affect readings.
Fuse Replacement
Only use designated fuses (600mA/250V, 10A/250V fast-blow fuses). To avoid electric shock or personal injury, turn off the meter and disconnect the test leads from any circuit before opening the battery cover to replace a fuse.
To replace fuses:
- Turn off the instrument.
- Remove all test leads from the input sockets.
- Use a screwdriver to loosen the screws securing the battery cover.
- Remove the battery cover.
- Remove the damaged fuse.
- Replace with a new fuse (spare fuses are typically located under the battery cover).
- Install the battery cover and tighten the screws.
7. Specifications
Comprehensive Indicators
- Operating Conditions: 600V CAT II, Pollution grade: 2, Height: under 2000m
- Working Temperature: 0-40°C (<80%RH)
- Storage Temperature: -10-60°C (<70%RH, remove battery)
- Test/Calibration Temperature: 25°C ±2°C
- Max Voltage (Measurement End to Ground): 600V
- Fuse Protection: "µA/mA" fuse 600mA/250V; "A" fuse 10A/250V
- Conversion Rate: Approx. 3s/second
- Display: 6000 counts LED display
- Overload Indication: 'OL' displayed
- Low Voltage Display:
displayed when working voltage is low - Input Polarity Indicator: "-" shown automatically
- Battery: 3 x 1.5V AAA
- Size: 147mm (L) x 71mm (W) x 45mm (H)
- Weight: Approx. 220g (battery excluded)
- Accuracy Index: ± (%reading + digit), one year warranty from manufacture date. Conditions: surroundings temperature from 18°C to 28°C, <80%RH.
Direct Current Voltage (DCV)
| Range | Resolution | Accuracy |
|---|---|---|
| 60mV | 0.01mV | ± (0.5% Readings + 5 digits) |
| 600mV | 0.1mV | |
| 6V | 0.001V | |
| 60V | 0.01V | |
| 600V | 0.1V |
Input resistance: 10MΩ
Maximum input voltage: 600VDC
Alternative Current Voltage (TRMS)
| Range | Resolution | Accuracy |
|---|---|---|
| 60mV | 0.01mV | ± (1% Readings + 4 digits) |
| 600mV | 0.1mV | |
| 6V | 0.001V | |
| 60V | 0.01V | |
| 600V | 0.1V |
Input resistance: 10MΩ
Maximum input voltage: AC600V
Frequency response: 40Hz to 1KHz True RMS
Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 600Ω | 0.1Ω | ± (0.8% Readings + 5 digits) |
| 6kΩ | 0.001kΩ | |
| 60kΩ | 0.01kΩ | |
| 600kΩ | 0.1kΩ | |
| 6MΩ | 0.001MΩ | |
| 60MΩ | 0.01MΩ |
Overload protection: 250V DC/AC
Open circuit voltage: 0.52V
Diode and Buzzer
| Function | Testing Condition | Open-Circuit Voltage |
|---|---|---|
![]() | Displays approximate value of forward voltage of the diode. | 3.2V |
| ))) | Buzzer sounds and red indicator light lights up when internal resistance is lower than 50Ω. | 1V |
Overload protection: 250V DC/AC
Direct Current
| Range | Resolution | Accuracy |
|---|---|---|
| 600µA | 0.1µA | ± (1.2% Readings + 5 digits) |
| 6000µA | 1µA | |
| 60mA | 0.01mA | |
| 600mA | 0.1mA | ± (3% Readings + 5 digits) |
| 6A | 0.001A | |
| 10A | 0.01A |
Overload protection: µA/mA range fuse (600mA/250V); 10A range fuse (10A/250V). When measurement current exceeds 10A, continuous measurement time should not exceed 10 seconds to prevent solder joint melting. After measurement, allow 1 minute before measuring again.
Alternative Current
| Range | Resolution | Accuracy |
|---|---|---|
| 600µA | 0.1µA | ± (1.2% Readings + 5 digits) |
| 6100µA | 1µA | |
| 60mA | 0.01mA | |
| 600mA | 0.1mA | ± (3% Readings + 5 digits) |
| 6A | 0.001A | |
| 10A | 0.01A |
Overload protection: µA/mA range fuse (600mA/250V); 10A range fuse (10A/250V). Maximum input current: mA level: 600mA AC effective value; 10A level: 10A AC effective value. When measurement current exceeds 10A, continuous measurement time should not exceed 10 seconds to prevent solder joint melting. After measurement, allow 1 minute before measuring again. Frequency response: 40Hz-1KHz, True RMS.
Frequency
| Range | Resolution | Accuracy |
|---|---|---|
| 9.999Hz | 0.001Hz | ± (1.5% Readings + 5 digits) |
| 99.99Hz | 0.01Hz | |
| 999.9Hz | 0.1Hz | |
| 9.999KHz | 0.001KHz | |
| 99.99KHz | 0.01KHz | |
| 999.9KHz | 0.1KHz | |
| 9.999MHz | 0.001MHz |
Input voltage range: 200mV-10V AC RMS
Overload protection: 250V DC/AC
Capacitance
| Range | Resolution | Accuracy |
|---|---|---|
| 6000pF | 1pF | ± (4.0% Readings + 5 digits) |
| 60nF | 0.01nF | |
| 600nF | 0.1nF | |
| 6µF | 0.001µF | |
| 60µF | 0.01µF | |
| 600µF | 0.1µF | |
| 6mF | 0.001mF | |
| 60mF | 0.01mF | |
| 100mF | 0.1mF |
Overload protection: 250V DC/AC
Temperature
| Range | Resolution | Measurement Range | Accuracy |
|---|---|---|---|
| °C | 1°C | -20°C ~ 1000°C | ± (1.0% + 3) Readings |
| °F | 1°F | -4°F ~ 1832°F | ± (1.0% + 3) Readings |
Overload protection: 250V DC/AC
Inductance
| Range | Resolution | Accuracy |
|---|---|---|
| 600µH | 0.1µH | ± (4.0% Readings + 5 digits) |
| 6mH | 0.001mH | |
| 60mH | 0.01mH | |
| 600mH | 0.1mH | |
| 6H | 0.001H | |
| 20H | 0.01H |
Overload protection: 250V DC/AC
8. Troubleshooting
- "OL" Display: If the display shows "OL" during a measurement, it indicates that the measured value exceeds the current range. Switch the rotary dial to a higher range setting. This also appears if the circuit is open during resistance or diode tests.
- Unstable Readings / Large Errors: This can occur if the meter is used near devices with strong electromagnetic interference. Move away from such devices or turn them off if possible.
- Meter Not Powering On: Ensure the batteries are correctly installed and have sufficient charge. Replace batteries if the low voltage indicator is shown or suspected.
- No Continuity Beep: If the buzzer does not sound during a continuity test, ensure the resistance is below 50Ω and the circuit is not live.
- Inaccurate Readings: Check battery level, ensure test leads are intact and properly connected, and verify the correct measurement function and range are selected. For low resistance measurements, perform a lead resistance compensation.
- Fuse Blown During Current Measurement: This indicates an overload. Replace the fuse with the correct type and rating (600mA/250V for µA/mA ranges, 10A/250V for A range). Always ensure the circuit is off before replacing fuses.
9. User Tips
- Packaging: If ordering online, consider requesting extra bubble wrap for shipping to protect the product box, although the meter itself is generally robust.
- Build Quality: Users report good build quality and a solid feel. The integrated magnet on the back is useful for attaching the meter to metal surfaces during work.
- Probes: While functional, some users suggest that the included probes might not be the most sensitive and could be upgraded for better performance.
- Battery Type: The use of AAA batteries is convenient for easy replacement.
- Display: The screen is generally easy to read, and the backlight is effective in low-light conditions.
- Flashlight: The built-in flashlight is a practical feature for working in dark environments.
- Spare Fuses: The inclusion of spare glass fuses is a thoughtful addition, ensuring quick replacements if needed.
10. Frequently Asked Questions (FAQ)
- Q: Does this multimeter measure inductance?
- A: Yes, the NJTY T21L is capable of measuring inductance. Refer to the 'Inductance Measurement' section for instructions.
11. Warranty and Support
The NJTY T21L Digital LCR Multimeter comes with a one-year warranty from the manufacture date, covering defects in materials and workmanship under normal use. For technical support or warranty claims, please contact your retailer or the manufacturer directly.
For a detailed user manual in PDF format, please refer to the official document available at: NJTY T21L User Manual (PDF).