NJTY T58C

Contents

NJTY T58C Multifunctional Digital LCR Multimeter User Manual

Model: T58C

1. Introduction

1.1 Overview

The NJTY T58C is a handheld digital multimeter featuring True RMS measurement capabilities. It is equipped with a large 9999-count LCD digital display and a backlight that automatically adjusts brightness for optimal readability. The device also includes an LED light for illuminating dark areas, overload protection, battery under-voltage indication, and input light indication. Designed according to the international electrical safety standard IEC-61010, this multimeter is suitable for professionals, factories, schools, enthusiasts, and home use.

NJTY T58C Multimeter held in hand, displaying a reading and highlighting key features like True RMS, 10000uF capacitance, inductance, NCV, and auto shutdown.

Figure 1: NJTY T58C Multimeter highlighting key features.

1.2 Safety Information

When using this meter, adhere to all standard safety regulations to prevent electric shocks and misuse of the instrument. Failure to follow these instructions may result in injury or damage to the meter.

  • Always inspect the provided test pens for any damage before use.
  • Avoid using the meter near devices with large electromagnetic disturbances, as this can cause unstable readings or large errors.
  • Do not use the meter if its housing or test pens are broken.
  • Ensure the correct input function and range are selected for each measurement.
  • Do not exceed the specified input value limits for each range to prevent damage.
  • Avoid touching unused input terminals when the meter is connected to a circuit.
  • Exercise caution when measuring voltages exceeding 60V DC or 36V AC to prevent electric shock.
  • When using test pens, keep your fingers behind the guard ring.
  • Always disconnect test pens from the circuit before changing the measurement range.
  • Before measuring resistance, diodes, capacitance, or performing continuity tests, ensure the circuit is powered off and all high-voltage capacitors are fully discharged.
  • Do not measure resistance or perform buzzer tests on live circuits.
  • Before current measurement, check the meter's fuse and turn off the power to the circuit under test.
  • The meter is powered by three AAA 1.5V batteries. Replace them immediately when the low voltage symbol appears on the display to avoid incorrect readings.
  • Do not exceed 1000V for Category II voltages or 600V for Category III voltages.

1.3 Safety Symbols

The following symbols are used on the meter and in this manual:

Symbol Description
AWarning: Important safety signs. Refer to the instruction manual before use. Improper use can result in damage to the device or its parts.
~AC (Alternating Current)
---DC (Direct Current)
~---AC or DC
⏚Ground
☐Double insulation protection
FUSEFuse
CEMeets EU standards
4High voltage warning
CAT. IICategory II. 1000V overvoltage protection
CAT. IIICategory III. 600V overvoltage protection

1.4 Input Protection Measures

  • The limit voltage for measurement is 1000 DCV or 750 ACV.
  • The meter can withstand AC voltages not exceeding 250V when measuring frequency, resistance, continuity, and diodes.
  • For µA and mA current measurements, protection is provided by a 600mA/250V fuse.
  • For 10A large current measurements, protection is provided by a 10A/250V fuse.

2. Setup

2.1 Battery and Fuse Replacement

To avoid electric shock or erroneous readings, replace batteries and fuses as soon as the low battery symbol (+-) appears on the display. Use only the specified fuses (1A/250V, 10A/250V fast-blow fuses).

  1. Turn off the instrument.
  2. Pull all test pens out of the input sockets.
  3. Use a screwdriver to loosen the screws fixing the battery compartment cover.
  4. Remove the battery cover.
  5. Remove the old batteries or the damaged fuse.
  6. Replace with new 3 x 1.5V AAA batteries or a new fuse (a spare fuse is typically located under the battery cover).
  7. Install the battery cover and tighten the screws.
Contents of the instrument kit including the T58C multimeter, instruction manual, test leads, temperature probe, and instrument kit bag.

Figure 2: NJTY T58C Multimeter and included accessories.

2.2 Initial Checks

Before each use, visually inspect the meter and test leads for any signs of damage. Ensure the test leads are intact and properly insulated. If any damage is found, do not use the instrument and arrange for repair.

3. Operation

3.1 Panel Diagram

Familiarize yourself with the layout of the multimeter:

Panel diagram for the Manual Ranging Model A of the 9999 Counts Digital Multimeter.

Figure 3: Manual Ranging Model A Panel Diagram

Panel diagram for the Auto Ranging Model B/C of the 9999 Counts Digital Multimeter.

Figure 4: Auto Ranging Model B/C Panel Diagram

  1. Illumination
  2. NCV sensing area
  3. NCV and buzzer indicator light
  4. LED display
  5. Infrared remote control detection (Model A only)
  6. hFE test socket
  7. Functions switch button
  8. Lighting/Backlight
  9. Gear rotary switch
  10. VΩHz°C input socket
  11. COM input socket
  12. 10A input socket
  13. mA input socket
  14. Data hold/relative value measurement
  15. Maximum and minimum switching
  16. Peak to Peak Measurement
  17. Range switching button
  18. µA mA L (inductance) input socket

3.2 Display Symbols

Symbol Description
+-Battery low voltage display
⏼Auto power off
-Negative polarity input indication
ACAlternative input indication
DCDirect input indication
▶))In continuity test mode
▶|In diode test mode
AUTOAuto ranging mode
HData hold mode
ΔIn relative value measurement mode
PEAKPeak-to-peak measurement mode
LInductance measurement
MAXMaximum value
MINMinimum value
°C °FTemperature unit
%Duty cycle
4High voltage danger warning (displayed when DC voltage is 60V or AC voltage is above 36V)

3.3 Button Functions

Button Instructions
FUNCShort press to select function (e.g., °C/°F switch for temperature, AC/DC switching for voltage/current, Hz/% for AC, three gears for On/off/diode/resistance).
AREL HOLDRelative Value Measurement: Press and hold for 2 seconds to enter/cancel. Data Hold: Short press to enter/cancel, holds instantaneous measurement value.
*/⏼Short press to turn backlight on/off. Long press for 2 seconds to turn on light and backlight simultaneously; short press again to turn off. Automatically turns off after 30 seconds if not pressed.
RANGESwitches between auto and manual range for Voltage/Resistance/Current.
MAX MINShort press to switch between maximum and minimum value measurement. Long press for 2 seconds to exit.
PEAKIn AC voltage and current signal state, short press to enter Peak-to-Peak measurement. Long press for 2 seconds to turn off.

3.4 Input Sockets

Input Socket Description
COMCommon input for all measurements, connected to the black test lead.
VΩHz%°C/°FPositive input for diode, capacitor, buzzer, temperature, voltage, resistance, and frequency (connected with the red test lead).
µA mA LInput socket for microampere, milliampere, and inductance measurements.
10APositive input terminal for current up to 10A.

3.5 Measurement Instructions

3.5.1 Auto Power Off

The meter will automatically power off after approximately 15 minutes of inactivity. An audible prompt will sound before entering hibernation. Any key press will restart the meter from this mode.

3.5.2 ACV and DCV Measurement

▲ Do not measure any voltage greater than 1000V DC or 750V AC RMS to prevent electric shock or damage to the instrument.

  1. Turn the rotary switch to the V or mV position. The red indicator light of the corresponding input socket will flash.
  2. Connect the black test lead to the COM jack and the red test lead to the V jack.
  3. Measure the voltage value of the circuit under test.
  4. The reading will be shown on the LED display, along with the polarity for DC measurements.

Notes:

  • For AC and DC 1V or mV ranges, if no input or test pen is connected, the meter may show a digital display. Short-circuit the "V/Ω" and "COM" terminals; if the meter displays zero, it is normal.
  • When measuring AC voltage, the frequency of the AC signal (below 30KHz) will be displayed in the upper right corner.
  • When measuring DC voltage, the ambient temperature will be displayed in the upper right corner.
  • The AC voltage value measured is the True RMS value.
NJTY T58C Multimeter performing AC Voltage Testing with probes connected to a wall outlet, displaying a voltage reading.

Figure 5: AC Voltage Testing

3.5.3 Resistance Measurement

▲ To avoid damage, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged before measuring resistance.

  1. Rotate the rotary switch to the "Ω" position. The red indicator light will flash.
  2. Connect the black test lead to the COM input socket and the red test lead to the VΩ input socket.
  3. Use the test pens to measure the resistance value.

Notes:

  • Resistance measured on a circuit may differ from the rated resistance.
  • For accurate low resistance measurements, short-circuit the test pens, read the short-circuit resistance, and subtract it from your measurement.
  • High resistance measurements may take a few seconds to stabilize.
  • If the meter displays "OL" when not in circuit, the measurement is out of range.

3.5.4 Diode Measurement

▲ Power off the circuit and discharge all high-voltage capacitors before measuring diodes.

  1. Turn the rotary switch to the diode symbol (▶|) position and press FUNC if necessary. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to VΩ.
  3. Connect the test leads to the negative and positive electrodes of the diode.
  4. The meter will display the forward bias value. If polarity is reversed, it will show 'OL'.

3.5.5 Buzzer Test (Continuity)

▲ Power off the circuit and discharge all high-voltage capacitors before performing a continuity test.

  1. Turn the rotary switch to the continuity symbol (▶)) position and press FUNC if necessary. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to VΩ.
  3. If the resistance is less than about 50 ohms, the indicator light will turn on and the buzzer will sound continuously. If resistance is between 40 and 500 ohms, it displays the resistance value. If greater than 500 ohms, it shows "OL".

3.5.6 Capacitance Measurement

▲ Power off the circuit and fully discharge all high-voltage capacitors before measuring capacitance.

  1. Turn the rotary switch to the capacitance symbol (µF) position. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to the µA mA L input socket.
  3. Measure the capacitance value.

Notes:

  • Measuring large capacitors may take some time.
  • Always ensure large capacitors are discharged before measurement to prevent damage.

3.5.7 Measuring Transistor (Model A only)

▲ Do not apply voltage exceeding 36V DC or AC RMS to the common terminal and hFE terminal.

  1. Turn the rotary switch to the "hFE" position.
  2. Determine if the transistor is NPN or PNP type.
  3. Insert the emitter (e), base (b), and collector (c) pins into the corresponding holes of the hFE test socket.
  4. Read the hFE measurement value on the LCD.

3.5.8 Frequency Measurement

▲ Do not test any voltage higher than 250V to prevent electric shock or damage to the instrument. If input voltage is greater than 30V, use the AC voltage range to measure frequency.

  1. Turn the rotary switch to the Hz% position. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to the Hz jack.
  3. Measure the frequency value of the circuit under test.
  4. The reading will be shown on the LED display, along with the duty cycle % in the upper right corner.

3.5.9 NCV Test (Non-Contact Voltage)

  1. Rotate the rotary switch to the NCV position.
  2. Place the top of the meter close to the conductor.
  3. If AC voltage is detected, the signal strength indicator (low-yellow, high-red) will light up, and the buzzer will emit different frequency alarms.

Notes:

  • NCV detectors should not be solely relied upon to determine voltage presence, as factors like socket depth or insulation thickness can affect detection.
  • External interference (e.g., flashlights, motors) may trigger false positives.
NJTY T58C Multimeter demonstrating NCV (Non-Contact Voltage) electric field sensing near a wall outlet.

Figure 6: NCV Electric Field Sensing

3.5.10 Infrared Remote Control Detection (Model A only)

  1. Turn the rotary switch to the infrared symbol position.
  2. Align the transmitter head of the infrared remote control vertically with the meter's infrared receiving window (deviation ≤ ±15°).
  3. Press the remote control's launch button. If the red light-emitting tube on the right side flashes, the transmitter is working normally.
  4. Moving the transmitter within 1-10cm can determine its output power status.

Notes:

  • Avoid strong light directly illuminating the receiving window.
  • Radio remote controls cannot be detected.

3.5.11 Temperature Measurement

  1. Turn the rotary switch to the °C/°F position. The meter will display the ambient temperature. The red indicator light of the corresponding socket will flash.
  2. Remove the test leads and connect the thermocouple with correct polarity to the COM and V/Ω inputs.
  3. Place the thermocouple on the object or in the environment to be measured.
  4. The display will show the temperature measurement value.

3.5.12 Inductance Measurement (Model C only)

  1. Turn the rotary switch to the "L" position. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to the µA mA L input terminals.
  3. Touch both ends of the inductance device with the test pens.
  4. The display will show the measured inductance value.
NJTY T58C Multimeter performing Inductance Testing with probes connected to a small transformer.

Figure 7: Inductance Testing

3.5.13 Low Impedance Measurement (Model A only)

Automatic switching measurement of AC and DC voltage (DC threshold 3V, AC threshold 15V).

  1. Turn the rotary switch to the "LoZ" position. The red indicator light will flash.
  2. Connect the black test lead to COM and the red test lead to the V/Ω input terminals.
  3. Connect the test leads in parallel to the power supply or load to be measured.
  4. The display will show the measured voltage.

3.5.14 Current Measurement

▲ When measuring current, the meter must be connected in series with the circuit under test. Do not attempt to measure circuits in parallel or mains power in parallel. If the fuse blows, you may damage the meter or injure yourself.

Before current measurement, pre-judge the current size, select the appropriate current gear, and check the instrument's fuse.

  1. Turn the rotary switch to the appropriate current position (µA, mA, or 10A). The red indicator light will flash.
  2. Connect the black test lead to the COM input socket. For currents less than 1A, connect the red test lead to the "mA" input socket. For currents between 1A and 10A, connect the red test lead to the "10A" input socket.
  3. Connect the test pens in series with the circuit under test, then turn on the power.
  4. The display will show the current measurement value. If "OL" is displayed, the input exceeds the selected range. Turn off power, set to a higher range, and measure again (Model A only).
Diagram illustrating the setup for current measurement using the multimeter, showing connection in series with a load.

Figure 8: Current Measurement Setup

4. Maintenance

4.1 General Maintenance

  • Before opening the instrument case or battery cover, always pull out the test pens.
  • Ensure all relevant power is disconnected before opening the meter. Avoid static electricity to prevent damage.
  • Clean the instrument case regularly with a damp cloth and a small amount of mild detergent. Do not use abrasive cleaners or chemical solvents.
  • For long-term storage, remove the batteries and store the meter in a cool, dry place away from high temperatures and humidity.

4.2 Cleaning Input Sockets

  1. Turn off the meter and unplug all test leads from the input sockets.
  2. Remove any dirt or debris from the sockets.
  3. Clean each socket with a new cotton ball moistened with detergent or lubricant. Lubricant can help prevent moisture-related contamination.

5. Troubleshooting

  • "OL" Display: Indicates that the measurement value is out of the selected range, or in diode mode, the polarity is reversed. For resistance, it can also mean very high resistance.
  • Low Battery Symbol: Replace the 3xAAA 1.5V batteries immediately to ensure accurate readings.
  • Unstable Readings: Can be caused by electromagnetic disturbances in the environment or insufficient battery power. Move away from interference sources or replace batteries.
  • Meter Abnormality: If the instrument shows any unusual behavior, stop using it immediately and send it for repair. Do not attempt to use it until it has been inspected and qualified.
  • Frequency Counter Inaccuracy: While functional, the frequency counter may provide rough estimates rather than precise measurements, especially for 60Hz AC wall voltage.

6. Specifications

6.1 Comprehensive Indicators

  • Operating Conditions: 600V CAT III and 1000V CAT II
  • Height: Under 2000m
  • Pollution Grade: 2
  • Working Temperature: 0 - 40 °C (<80%RH, not considered <10°C)
  • Storage Temperature: -10 - 60 °C (<80%RH, remove battery)
  • Test or Calibration Ambient Temperature: 20 °C ± 2 °C
  • Max Voltage between Measurement End and Ground: 1000V VDC or 750VAC
  • Fuse Protection: mA range: F 600mA/250V; A range: F 10A/250V
  • Conversion Rate: About 3s/second
  • Display: 9999 counts LED display
  • Overload Display: 'OL' displayed
  • Low Voltage Display: +- displayed when voltage is low
  • Input Polarity Indicator: '-' shown automatically
  • Battery: 3 x 1.5V AAA
  • Dimensions: 187mm (L) x 95mm (W) x 55mm (H)
  • Weight: About 331g (battery excluded)

6.2 Accuracy Index

6.2.1 Direct Current Voltage

Range Resolution Accuracy
1V / mV0.1mV (A model), 0.01mV (B/C model)±(0.5% readings + 5 digits)
10V / V0.001V±(0.5% readings + 5 digits)
100V / V0.01V±(0.5% readings + 5 digits)
1000V / V0.1V±(0.8% readings + 5 digits)

Input resistance: 10MΩ; Max input voltage: 1000VDC or 750VAC RMS.

6.2.2 Alternative Current Voltage

Range Resolution Accuracy
1V / mV0.1mV (A model), 0.01mV (B/C model)±(1% readings + 4 digits)
10V / V0.001V±(1% readings + 4 digits)
100V / V0.01V±(1% readings + 4 digits)
750V / V0.1V±(1.2% readings + 4 digits)

Input resistance: 10MΩ; Max input voltage: 1000VDC or 750VAC RMS; Frequency response: 10Hz to 1KHz True RMS.

6.2.3 Frequency

Range Resolution Accuracy
10Hz0.01Hz±(1.5% readings + 5 digits)
100Hz0.1Hz±(1.5% readings + 5 digits)
1KHz0.001KHz±(1.5% readings + 5 digits)
10KHz0.01KHz±(1.5% readings + 5 digits)
100KHz0.1KHz±(1.5% readings + 5 digits)
1MHz0.001MHz±(1.5% readings + 5 digits)
10MHz0.01MHz±(1.5% readings + 5 digits)

Input voltage range: 200mV-10V AC RMS; Overload protection: 250V DC/AC.

6.2.4 Resistance

Range Resolution Accuracy
1KΩ0.1Ω±(0.8% readings + 5 digits)
10KΩ0.001KΩ±(0.8% readings + 5 digits)
100KΩ0.01KΩ±(0.8% readings + 5 digits)
1MΩ0.1KΩ±(0.8% readings + 5 digits)
10MΩ0.001MΩ±(1.2% readings + 5 digits)
100MΩ0.01MΩ±(1.2% readings + 5 digits)

Overload protection: 250V DC/AC; Open circuit voltage: 1.4V.

6.2.5 Diode

Function Range Resolution Condition
Diode0-4V0.001VForward direct current: ~1mA; Open circuit voltage: ~3.2V; Displays approximate forward voltage drop.

Overload protection: 250V DC/AC.

6.2.6 Buzzer (Continuity)

Function Range Resolution Description Condition
Continuity500Ω0.1ΩBuzzer sounds with red indicator when resistance <50Ω. Indicator turns off and beep stops when >50Ω.Open-circuit voltage: ~1.7V

Overload protection: 250V DC/AC.

6.2.7 Transistor (A model only)

Function Description Condition
hFEApproximation of monitor readout of hFE value (0-1000)Base current: ~10µA, Vce: ~2.8V

6.2.8 Capacitance

Range Resolution Accuracy
1nF0.001nF±(4.0% readings + 5 digits)
10nF0.01nF±(4.0% readings + 5 digits)
100nF0.1nF±(4.0% readings + 5 digits)
1µF1nF±(4.0% readings + 5 digits)
10µF0.001µF±(4.0% readings + 5 digits)
100µF0.01µF±(4.0% readings + 5 digits)
1mF0.1µF±(5.0% readings + 5 digits)
10mF0.001mF±(5.0% readings + 5 digits)
100mF0.1mF±(5.0% readings + 5 digits)

6.2.9 Inductance L (Model C only)

Range Resolution Accuracy
10mH0.001mH±(4.0% readings + 5 digits)
100mH0.01mH±(4.0% readings + 5 digits)
1H0.1mH±(4.0% readings + 5 digits)
10H0.001H±(4.0% readings + 5 digits)
100H0.01H±(4.0% readings + 5 digits)

6.2.10 Temperature

Range Resolution Measurement Range Accuracy
°C1°C-55°C ~ 1000°C±(1.0% + 3) readings
°F1°F-67°F ~ 1832°F±(1.0% + 3) readings

Overload protection: 250V DC/AC.

6.2.11 Direct Current

Range Resolution Accuracy
1mA / µA0.1µA±(1.2% readings + 5 digits)
100mA (Model A)0.01mA±(1.2% readings + 5 digits)
1A / mA0.2mA±(2% readings + 5 digits)
10A0.001A±(2% readings + 5 digits)

Overload protection: mA range (1A/250V) and 10A range (10A/250V); Max input current: mA: 1A RMS, 10A: 10A RMS. For currents >5A, test time should be <10s, followed by a 1-minute cool-down period.

6.2.12 Alternative Current

Range Resolution Accuracy
1mA / µA0.1µA±(1.5% readings + 5 digits)
100mA (Model A)0.01mA±(1.5% readings + 5 digits)
1A / mA0.2mA±(2% readings + 5 digits)
10A0.001A±(2% readings + 5 digits)

Overload protection: mA range (1A/250V) and 10A range (10A/250V); Max input current: mA: 1A RMS, 10A: 10A RMS. For currents >5A, test time should be <10s, followed by a 1-minute cool-down period. Frequency response: 40Hz to 1KHz True RMS.

6.2.13 LoZ (Low Impedance)

Range Resolution Identification Threshold Accuracy
750(ACV)1V15V±(2% readings + 5 digits)
1000(DCV)1V3V±(2% readings + 5 digits)

7. User Tips

  • Accuracy for Home Use: The multimeter offers good accuracy for general home use and hobbyist projects. DC voltage measurements are highly accurate, and AC voltage measurements are generally good. However, for lab-grade precision, specialized equipment may be required.
  • Frequency Counter: The frequency counter may provide approximate values. For critical applications requiring high-precision frequency measurements, consider its limitations.
  • Probe Insertion Indicators: The flashing probe insertion points are a helpful feature, guiding you to the correct ports for the selected function.
  • Automatic Features: Benefit from automatic scaling and automatic display illumination, which adjusts to ambient light levels for better readability.
  • NCV Sensitivity: The Non-Contact Voltage (NCV) feature may be more effective at detecting AC voltage at wall switches compared to deep wall outlets.
  • Inductance Measurement: If experiencing issues with inductance readings, try turning off the automatic mode if available on your specific model.
  • Quality of Leads: The included test leads are noted for being flexible and of good quality.

8. Frequently Asked Questions (FAQ)

Q: How does the PEAK key work?
A: In AC voltage and current signal states, short pressing the PEAK key activates the Peak-to-Peak measurement function. To turn it off, long press the PEAK key for 2 seconds. Refer to section 3.3 Button Functions for more details.

9. Warranty and Support

This product is designed for reliability and performance. For detailed warranty information and support, please refer to the official product documentation or contact the seller.

A comprehensive user manual in PDF format is available for download. This document contains additional detailed instructions and diagrams:

View User Manual (PDF)

9.1 Accessories

The following accessories are typically included with your multimeter:

  • One instruction manual
  • One pair of test leads
  • One thermocouple
  • 3 x 1.5V AAA batteries (may be pre-installed or included separately)
  • One canvas bag