NJTY T28B

Contents

NJTY T28B Digital Multimeter User Manual

Model: T28B

1. Introduction

Thank you for your purchase of the NJTY T28B multi-meter. This precision instrument is designed to be a valuable tool for improving your work efficiency. Before using the device, please read this manual carefully. If you have any questions, please contact our customer service.

NJTY T28B Multimeter held in hand, displaying '-00.15' and highlighting features like high-precision, true RMS, 100,000uF capacitor measurement, NCV, frequency/duty cycle, and automatic shutdown.
NJTY T28B Multimeter: High-precision digital display with True RMS measurement, capacitance, NCV, frequency, duty cycle, and automatic shutdown.
NJTY T28B Multimeter with test probes and temperature probe, and a small user manual.
The multimeter comes with test probes and a temperature probe.

2. Safety Information

The meter conforms to IEC61010-1 CAT III 600V, 1000V CAT II IEC61010 and pollution level 2.

2.1 Safety Instructions

When using this meter, the user must comply with all standard safety regulations:

  1. Protection against electric shocks.
  2. Preventing the misuse of the multimeter safety procedures.

To ensure your personal safety, please use the test leads provided with the meter. Always check them for integrity before use.

2.2 To Prevent Electrical Shock, Fire, or Personal Injury:

  1. Do not alter, modify, or tamper with the multimeter. Use only as specified.
  2. Do not use the multimeter near explosive gas, vapor, or in damp, wet, or dusty environments.
  3. Do not operate the multimeter with the covers removed or the case open. Hazardous voltage exposure is possible.
  4. Do not apply more than the rated voltage between the terminals or between each terminal and earth ground.
  5. Do not use test leads if they are damaged. Examine them for damaged insulation, exposed metal, or wear.
  6. Use the correct terminals, function, and range for measurements.
  7. The range switch should be placed in the correct range position.
  8. The input value must not exceed the limit of the value specified for each range.
  9. When measuring televisions or switching power supplies, be aware of pulses that can damage the circuit.
  10. Remove the batteries if the multimeter is not used for an extended period or stored above 50°C.

2.3 Safety Symbols

Symbol Description Symbol Description
⚠️ WARNING. Risk of danger. ∿ AC or DC
⚡ High voltage warning ⎔ Double insulated
⎓ DC (Direct Current) ⏚ Fuse
∼ AC (Alternating Current) ⏚ Ground
CE Conforms to European Union directives.
CAT. II Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation.
CAT. III Measurement Category III is applicable to test circuits connected to the distribution part of the building's low-voltage MAINS installation.

3. Overview

3.1 Product Functions Overview

NJTY T28B Multimeter with kickstand, highlighting True RMS, auto shutdown, low voltage indication, overload protection, and magnetic attraction.
Key features include True RMS, automatic shutdown, low voltage indication, overload protection, and magnetic attraction.
NJTY T28B Multimeter alongside its soft rubber casing.
The multimeter is protected by a soft rubber casing.
NJTY T28B Multimeter with its built-in flashlight illuminated.
Built-in flashlight for improved visibility in dark environments.

The multimeter features a large LED display (6000 counts) and a function knob for selecting measurement modes. It includes dedicated input sockets for mA, 20A, COM, and VΩHz°C.

3.2 Backlight / Flashlight

Push the */ button to toggle the backlight on and off. Push this button for 2 seconds to toggle the flashlight on and off. The backlight/flashlight automatically turns off after 30 seconds.

3.3 Auto Power Off

The multimeter will automatically turn off if it is left idle for about 15 minutes. You can press any button to restart the device.

3.4 Display Hold

In display hold mode, the multimeter freezes the display. Push the FUNC HOLD button for 2 seconds to activate/deactivate the display hold ('H' shows on the display).

3.5 Manual and Auto-Ranging

The multimeter defaults to Auto-range. Push the RANGE button for 2 seconds to enter Auto-range mode ('AUTO' shows on the display). Push the RANGE button to enter Manual range. In Manual Range mode, you override Auto-range and select the range yourself. Push the RANGE button to increase the range; after the highest range, the multimeter wraps to the lowest range.

4. Setup

Before any measurement, ensure the multimeter is ready for use:

  1. Battery Installation: Refer to the 'Install Batteries & Replace Fuses' section for instructions on installing or replacing batteries.
  2. Test Leads Check: Always examine the test leads for any damage to insulation or exposed metal before use. Replace if damaged.
  3. Initial Battery Check: Check the battery first. If the low battery indicator is displayed, the battery needs to be changed.
  4. Input Warning: The warning symbol beside the sockets indicates the maximum input of voltage and current. Do not exceed these limits.
  5. Polarity: The red lead is positive in polarity, and the black lead is negative.

5. Operating Instructions

Always switch to the correct range before measurement.

5.1 DC/AC Voltage Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the 'mV' or 'V' ranges and connect the test leads to the loaded circuit.
  3. DC signal will be displayed, indicating DC voltage test. Press FUNC HOLD button to switch to ACV test ('AC' is shown on the screen).

Note:

  • Do not measure voltages above DC1000V or AC750V to prevent damage to the multimeter.
  • In AC/DC 600mV and AC 6V ranges, if the multimeter displays numbers with no input, short-circuit the 'VΩHz°C' port and 'COM' port. If the display returns to zero, the multimeter is normal.
NJTY T28B Multimeter performing AC voltage measurement, with test leads connected to a power socket and a warning about incorrect insertion.
Example of AC voltage measurement. Ensure correct insertion of test leads.

5.2 Resistance Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the 'Ω' range ('Ω' shows on the display).
  3. Connect the test leads to the resistor to be tested.

Note:

  • 'OL' means overload. Switch to a higher range. 'OL' also displays if open-circuited.
  • If the tested resistor is > 60M Ω, it may take a few seconds to read.
  • Before testing a resistor in a circuit, ensure the circuit is powered off and high-voltage capacitors are discharged.
  • To measure low resistance accurately, short-circuit the two test leads and read the resistance value. Subtract this value from the measured resistance.

5.3 Diode Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the diode symbol and press FUNC HOLD button 2 times (diode symbol shows on the display).
  3. Touch the diode anode with the red probe and the diode cathode with the black probe.

Note:

  • 'OL' will be displayed if open-circuited or if test leads are connected reversely.

5.4 Continuity Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the continuity symbol and press FUNC HOLD button (buzzer symbol shows on the display).
  3. Contact the probes to the measured circuit or resistance.
  4. The multimeter will buzz and the indicator light is yellow if resistance is < 40Ω. If resistance is > 40Ω and < 60Ω, the buzzer does not buzz, and the indicator light is red.

Note:

  • 'OL' will be displayed if open-circuited.
  • Before continuity testing, ensure the circuit is powered off and high-voltage capacitors are discharged.

5.5 Capacitance Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the capacitance symbol and connect the test leads to the circuit to be tested.

Note:

  • Before testing a capacitor in a circuit, ensure the circuit is powered off and high-voltage capacitors are discharged.
  • For larger capacitance, it may take longer for the reading to stabilize.

5.6 Frequency/Duty Measurement

  1. Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
  2. Switch the knob to the 'Hz%' range and use the Hz/%/Δ button to select Frequency or Duty Ratio.
  3. Connect the test leads to the circuit to be tested.

Note:

  • Do not test any voltage higher than 250V to prevent electric shock or damage.

5.7 DC/AC Current Measurement

  1. Insert the black lead into the 'COM' port. For current under 600mA, insert the red lead into the 'mA' port. For current between 600mA - 20A, insert the red lead into the '20A' port.
  2. Switch the knob to the 'μA', 'mA', or 'A' ranges.
  3. 'DC' signal will be displayed, indicating DC current test. Press FUNC HOLD button to switch to AC current test ('AC' shows on the display).
  4. Connect the test leads to the loaded circuit.

Note:

  • 'OL' means overload. Switch to a higher range.
  • Do not measure voltages above 250V.
  • Pay special attention to safety when measuring high voltage.
  • Test known current with the multimeter before use.
  • When current is more than 5A, test time should be less than 10 minutes, and 1 minute should be given to stop testing after such measurement.

5.8 NCV Test (Non-Contact Voltage)

  1. Turn the knob to the 'NCV' position.
  2. Gradually approach the detected point with the NCV probe.
  3. When weak AC signals are sensed, the yellow indicator lights up with a slow beeping sound. For strong AC signals, the red indicator lights up with a fast beeping sound.

Note:

  • Do not rely solely on NCV for voltage presence; socket depth, insulation thickness, etc., can affect detection.
  • Induced voltage may cause the indicator to light up even without direct contact.
  • External interference (flashlights, motors) may trigger NCV detection.

5.9 Temperature Measurement

  1. Turn the knob to the '°C' position (the multimeter will display the ambient temperature).
  2. Switch to '°C' or '°F' by pressing the FUNC HOLD button.
  3. Insert the positive (red) thermocouple into the 'VΩHz°C' input and the negative (black) into the 'COM' input.
  4. Contact the measured object with the thermocouple probe.

Note:

  • When measuring temperature with thermocouples, the probe should not touch a charged object to prevent damage or personal injury.

6. Maintenance

6.1 Cleaning

  1. Turn off the multimeter and remove the test leads.
  2. Wipe the case with a damp cloth and clean it with mild detergent.
  3. Dirt or moisture in the terminals can affect readings.
  4. If there is any abnormality, stop using the multimeter immediately.

6.2 Install Batteries & Replace Fuses

Check the battery first. If the low battery indicator is displayed, the battery needs to be changed.

  1. Turn off the power of the multi-meter.
  2. Pull all the test leads out of the input sockets.
  3. Use a screwdriver to loosen the screws that hold the battery cover.
  4. Remove the battery cover.
  5. Remove the old battery or the damaged fuse.
  6. Replace with new batteries (4 * 1.5V AAA Batteries) or new fuses (F 600mA/250V, F 20A/250V) and tighten the screw.

Note:

  • Only use the specified fuse (600mA /250V, 20A /250V fast melt fuse) to avoid electric shock or personal injury.
  • Turn off and check that the test has been disconnected from the measured circuit before opening the battery cover and when replacing it with new batteries.

7. Specifications

7.1 General Specifications

  • Maxim Display: 6000 counts
  • Low Battery Indication: Displayed
  • Operation Temperature: 0°C – 40°C
  • Storage Temperature: -10°C – 60°C
  • Battery Type: 4 * 1.5V AAA Batteries
  • Fuse: F 600mA/250V, F 20A/250V
  • Data Hold: 'H' displayed

7.2 Measurement Ranges and Accuracy

Frequency

Range Resolution Accuracy
9.999Hz0.001Hz± (1.5% + 5)
Input voltage range: 200mV - 10V AC RMS
Overload protection: 250V DC/AC
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999KHz0.001KHz
99.99KHz0.01KHz
999.9KHz0.1KHz
9.999MHz0.001MHz

Temperature

Range Resolution Temperature Accuracy
°C1°C-20°C ~ 1000°C± (1.0% + 3)
°F1°F-4°F ~ 1832°F± (1.0% + 3)

AC Voltage

Range Resolution Accuracy
600mV0.1mV± (1% + 4)
6V1mV
60V10mV
600V100mV
750V1V± (0.8% + 5)

Input Impedance: 10MΩ
Maximum input voltage: 750V AC RMS
Frequency Response: 10Hz~1KHz; T-RMS

DC Voltage

Range Resolution Accuracy
600mV0.1mV± (0.5% + 5)
6V1mV
60V10mV
600V100mV
1000V1V± (0.8% + 5)

Input Impedance: 10MΩ
Maximum input voltage: 1000V DC RMS

Diode/Continuity

Function Range Details
Diode0-3VForward direct current: about 1mA. OCV about 3.2V
Continuity600ΩResistor < 40Ω, buzz with yellow light. 40Ω < resistor < 60Ω, buzz with red light. OCV about 1V

AC Current

Range Resolution Accuracy
600μA0.1μA± (1.5% + 5)
60mA0.01mA
600mA0.1mA
20A10mA± (2.0% + 5)

Overload protection: mA range with F600mA/250V and 20A range with F20/250V
Max input current: mA: 600mA RMS, 20A: 20A RMS

DC Current

Range Resolution Accuracy
600μA0.1μA± (4.0% + 5)
60mA0.01mA
600mA0.1mA
20A10mA± (2.0% + 5)

Overload protection: mA range with F600mA/250V and 20A range with F20/250V
Max input current: mA: 600mA RMS, 20A: 20A RMS

Resistance

Range Resolution Accuracy
600Ω0.1Ω± (0.8% + 5)
6kΩ1Ω
60kΩ10Ω
600kΩ100Ω
6MΩ1kΩ± (1.2% + 5)
60MΩ10kΩ

Input voltage range: 200mV-10V AC RMS
Overload protection: 250V DC/AC

Capacitance

Range Resolution Accuracy
6nF0.001nF± (4.0% + 5)
60nF0.01nF
600nF0.1nF
6μF1nF
60μF10nF
600μF100nF
6mF1μF± (5.0% + 5)
100mF10μF

Overload protection: 250V DC/AC

8. Troubleshooting

This section provides general guidance for common issues. For persistent problems, please contact support.

  • 'OL' Displayed: This indicates an overload or an open circuit. Ensure the test leads are properly connected and switch to a higher range if applicable. For resistance, diode, continuity, and capacitance measurements, 'OL' can also mean an open circuit.
  • No Display or Faint Display: Check the batteries. If the low battery indicator is shown, replace the batteries. If the device does not power on, ensure batteries are correctly installed.
  • Inaccurate Readings: Ensure the correct function and range are selected for the measurement. Check test leads for damage or dirt/moisture in the terminals. Calibrate if necessary (refer to professional service).
  • Multimeter Turns Off Unexpectedly: The device has an auto power-off feature after 15 minutes of inactivity. Press any button to restart. If it turns off during active use, check battery levels.
  • NCV False Positives: External interference from flashlights, motors, or induced voltage can trigger NCV. Use NCV as an indicator, not a definitive confirmation of voltage presence.

9. User Tips

  • Always Prioritize Safety: Before any measurement, especially involving high voltage or current, review the safety instructions.
  • Check Test Leads: Make it a habit to inspect your test leads for any signs of wear or damage before each use.
  • Power Off Circuits: For resistance, continuity, and capacitance measurements, always ensure the circuit under test is powered off and any high-voltage capacitors are fully discharged.
  • Use the Right Range: While auto-ranging is convenient, manually selecting the appropriate range can sometimes provide faster or more precise readings.
  • Mind the Current Limits: Be aware of the maximum current limits for the mA and 20A input ports to prevent fuse blowouts or damage to the meter.
  • Keep it Clean: Regularly clean the multimeter's case and ensure terminals are free from dirt and moisture to maintain accuracy.

10. Support

For any questions or support regarding your NJTY T28B Digital Multimeter, please contact our customer service:

Email: HEMITEK2022@126.COM

A detailed user manual in PDF format is also available for download. You can find it at: NJTY T28B User Manual (PDF)

© 2024 NJTY. All rights reserved.