NJTY T21L

Contents

NJTY T21L Digital LCR Multimeter User Manual

Model: T21L

1. Introduction

This meter is a multi-purpose instrument with high measurement accuracy, fast response speed, and high safety level. It is embedded with a special chip with up to 6000 counts. This chip is composed of high-precision AD and high-speed digital processor. It has accurate measurement, high resolution, fast calculation speed, and can achieve high-speed true effective value calculation of AC 1KHz. The whole machine is software calibrated, and the accuracy remains unchanged for long-term use.

The appearance design of the entire instrument is aesthetically pleasing and suitable for various industrial applications, and the circuit design is safe and reliable. The entire instrument has many measurement functions and a friendly human-machine interface, which can meet the needs of various application groups such as professional engineers and maintenance engineers.

This meter can be used to measure DC and AC voltage, DC and AC current, resistance, capacitance, inductance, temperature, diode, and circuit continuity testing. This instrument is equipped with a backlight and an illumination lamp to facilitate users to read the measurement display value in a dark place.

NJTY T21L Digital LCR Multimeter
Figure 1: NJTY T21L Digital LCR Multimeter
Video 1: Demonstration of NJTY T21L Multimeter functions including NCV, voltage, resistance, capacitance, diode, temperature, inductance, and current measurements.

2. Safety Information

2.1. Safety Instructions

When using this meter, the user must comply with all the standard safety regulations in the following two aspects:

  1. Protection against electric shocks.
  2. Preventing the misuse of the instrument's safety procedures.

To ensure your personal safety, please use the test pen provided with the meter, check before use, and make sure they are intact.

2.2. Safety Signs

SymbolDescription
▲Warning
~AC (alternative current)
---DC (direct current)
⏚Ground
ΠDouble insulation
CEEuropean union standard
⚡High voltage warning
CATIIII category overvoltage protection

2.3. Precautions for Use

  • The use of meter instruments near devices with large electromagnetic disturbances will be unstable and may even cause large errors.
  • Do not use when the appearance of the meter or the test pen is broken.
  • If the meter is not used correctly, the safety functions provided by the meter may fail.
  • Care must be taken when working around bare conductors or buses.
  • Do not use this instrument near explosive gas vapor or dust.
  • The correct input function must be used to measure the range.
  • Do not switch gears while measuring voltage or current.
  • The input value must not exceed the limit of the input value specified for each range to prevent damage to the instrument.
  • Do not touch the unused input when the instrument is connected to the circuit under test.
  • When the measured voltage exceeds 60 VDC or 30 VAC, use caution to prevent electric shock.
  • When measuring with a test pen, place your finger behind the guard ring of the test.
  • Before converting the range, it must be ensured that the test pens have left the circuit under test.
  • Before carrying out a resistance, diode, capacitance measurement or continuity test, the circuit under test must be powered off and all high-voltage capacitors in the circuit under test should be discharged.
  • Do not measure the resistance on a live circuit or perform buzzer test.
  • Before conducting the current measurement, the fuse of the meter should be checked.
  • Before connecting the meter to the circuit under test, the power of the circuit under test should be turned off.
  • When performing TV repairs or measuring power conversion circuits, care must be taken in the high-amplitude voltage pulses in the circuit under test to avoid damage to the meter.

2.4. Input Protection Measures

  • The limit voltage is 600V when the voltage is measuring.
  • The limit voltage is 250 ACV or the equivalent RMS voltage when the capacitance or the diode is measuring.
  • When measuring µA current and mA current, it is protected by a fuse (600mA/250V).

3. Product Overview

3.1. Panel Diagram

The multimeter features a clear LCD display, a rotary switch for function selection, and various input jacks. Key components include:

  1. Illumination
  2. NCV sensing area
  3. LED display
  4. Buzzer
  5. Backlight/illumination button
  6. "FUNC/Hz" selection button
  7. Gear rotary switch
  8. Hz/VΩMAL/°C input socket
  9. COM input socket
  10. 10A input socket
  11. HOLD button
  12. NCV indicator

3.2. Monitor Symbols Explanation

SymbolDescription
Battery low voltage display
Auto power off
−Negative polarity input indication
ACAlternative input indication
DCDirect input indication
)))In continuity test mode
→|In diode test mode
LInductance measurement
AUTOAuto ranging mode
HOLDData hold mode
°C, °FTemperature
%Duty cycle
NCVNon-contact voltage detection mode
V, mVVolt, unit of Voltage
A, mA, µAAmpere, unit of current
Ω, KΩ, MΩOhm, unit of resistor
Hz, KHz, MHzHz, hertz, unit of frequency
mF, µF, nF, pFFarah, unit of capacitance
H, mH, µHInductance unit

4. Setup

4.1. Battery and Fuse Replacement

To avoid erroneous readings or electric shock, replace the battery immediately when the battery low voltage symbol appears. Only use designated fuses (600mA/250V and 10A/250V fast-blow fuses).

Battery Type: 3 x 1.5V AAA batteries.

Steps to replace:

  1. Turn off the power of the instrument.
  2. Pull all the test pens out of the input socket.
  3. Use a screwdriver to loosen the screws fixing the battery cover.
  4. Remove the battery cover.
  5. Remove the old battery or the damaged fuse.
  6. Replace with new 3 x 1.5V AAA batteries or a new fuse (the spare fuse is under the battery cover).
  7. Install the battery cover and close the screws.
NJTY T21L Multimeter back with magnetic adsorption
Figure 2: Back of the multimeter showing magnetic adsorption feature and battery compartment.

The video demonstrates opening the back shell, revealing a large plastic chip and two fuses (0.6A and 10A alloy resistor with 0.5% precision).

5. Operating Instructions

5.1. On/Off Operation

Turn the gear switch to any function position to turn on the multimeter. If there is no operation for about 15 minutes after powering on, the instrument will make a beep sound to indicate that it has entered sleep mode. Press any button or rotate the gear switch to restart.

5.2. Button Functions

  • "HOLD" key: Data Hold
    During measurement, short press the HOLD key to keep the current measurement value displayed. Short press again or rotate the gear switch to cancel. Press and hold the HOLD button to turn on the meter, and three beeps will cancel the automatic shutdown function.
  • "FUNC/Hz" key: Function Selection
    Used to switch between AC/DC voltage and current measurements. Long press to switch to frequency measurement (0-1KHz signals with voltage less than 250V). In the resistance/capacitance/diode/continuity range, short press to cycle through these modes.
  • Backlight and Lighting Functions
    Short press the backlight button to turn on/off the backlight. It turns off automatically after 30 seconds of no operation. Long press the button to turn on the lighting (flashlight) and backlight simultaneously. Short press again to turn off both.
NJTY T21L Multimeter with flashlight on
Figure 3: Multimeter showing the flashlight feature for use in dark environments.

5.3. ACV and DCV Measurement

Warning: Do not measure any voltage greater than AC600V or DC600V to prevent electric shock or damage to the instrument.

  1. Turn the rotary switch to the "mV" or "V" position, and press the "FUNC/Hz" key to select DC or AC.
  2. Connect the black test lead to the COM input socket and the red test lead to the V input socket.
  3. Use the other two ends of the test pen to measure the voltage value of the circuit under test (connected in parallel).
  4. The display shows the measured voltage value. For DC voltage, a "-" symbol indicates negative polarity.

Notice: In 600mV and 6V ranges, short-circuit the "V-Ω" and "COM" terminals to zero the display if no test pen is input. The AC voltage value measured is the true effective value (RMS).

5.4. Resistance Measurement

Warning: To avoid damage, cut off all power and discharge high-voltage capacitors before measuring resistance.

  1. Turn the rotary switch to the "Ω" position and press "FUNC/Hz" to select the "Ω" range.
  2. Connect the black test pen to COM and the red test pen to V/Ω input socket.
  3. Measure the resistance of the circuit under test.
  4. The display shows the measured resistance value.

Notice: The measured resistance may differ from the rated value due to parallel components. For accurate low resistance measurements, short-circuit the test leads first, read the resistance, and subtract this value from the final measurement.

5.5. Diode Measurement

Warning: To avoid damage, cut off all power and discharge high-voltage capacitors before measuring diodes.

  1. Turn the rotary switch to the "→|" position and press "FUNC/Hz" to select the "→|" range.
  2. Connect the black test lead to COM and the red test lead to V/Ω input socket.
  3. Connect the black test pen to the cathode and the red test pen to the anode of the diode.
  4. The meter displays the forward conduction voltage. If polarity is reversed, "OL" is displayed.

5.6. Buzzer Test (Continuity)

Warning: To avoid damage, cut off all power and discharge high-voltage capacitors before performing continuity test.

  1. Turn the rotary switch to the "))) " position and press "FUNC/Hz" to select the "))) " range.
  2. Connect the black test lead to COM and the red test lead to V/Ω input socket.
  3. Measure the resistance of the circuit under test. If resistance is less than 50Ω, the indicator light will light up, and the buzzer will sound continuously.

5.7. Capacitance Measurement

Warning: To avoid damage, cut off all power and discharge high-voltage capacitors before measuring capacitance.

  1. Turn the rotary switch to the "F" position and press "FUNC/Hz" to select the "F" range.
  2. Connect the black test pen to COM and the red test pen to the input socket.
  3. Touch the two terminal pins of the capacitor with a test pen. The display will show the measured capacitance value.

Notice: Measuring large capacitance may take time to stabilize. Pay attention to polarity when measuring polarized capacitors.

5.8. Frequency Measurement

Warning: Do not test any voltage higher than 250V.

  1. Turn the rotary switch to the "Hz%" position.
  2. Connect the black test pen to COM and the red test pen to the Hz input socket.
  3. Measure the frequency value of the circuit under test.
  4. Press the "FUNC/Hz" key to display frequency and duty cycle respectively.

5.9. Current Measurement (AC/DC)

Warning: Never attempt to make current measurements on a circuit when the open circuit voltage to ground exceeds 250V. Check the fuse before measuring current. When measuring, use the correct input socket and function range. Do not connect the two test pens in parallel to any circuit.

  1. Turn the rotary switch to the "A" or "mA" or "µA" position, and press "FUNC/Hz" to select AC or DC.
  2. Connect the black test lead to COM. For currents less than 600mA, connect the red test pen to VΩMA. For currents between 600mA and 10A, connect the red test pen to 10A.
  3. Disconnect the circuit to be tested. Connect the black test lead to the lower voltage end and the red test lead to the higher voltage end of the disconnected circuit.
  4. Connect power to the circuit and read the display. If "OL" is displayed, the input exceeds the selected range. After powering off, switch to a higher range.
  5. In ACA measurement, long press "FUNC/Hz" to display the frequency of the AC signal 0-1KHz.

5.10. NCV Test (Non-Contact Voltage Detection)

Rotate the rotary switch to the "NCV" position and place the top of the meter (NCV sensing area) close to the conductor. If the meter detects AC voltage, the indicator light lights up based on signal strength, and the buzzer emits different frequencies of alarm sound.

Notice: Voltage may still be present even when not indicated. Do not rely solely on non-contact voltage detectors. Interference sources (flashlights, motors, etc.) may trigger false detections.

5.11. Temperature Measurement

  1. Turn the rotary switch to the "°C (°F)" position. The instrument will display the ambient temperature.
  2. Remove the test leads and insert the thermocouple to the "COM" and "VΩMA" input terminals according to the correct polarity.
  3. The meter displays the approximate temperature value transmitted from the thermocouple.

5.12. Inductance Measurement

  1. Turn the rotary switch to the "L" position.
  2. Connect the black test pen to COM and the red test pen to VΩMAL input terminals.
  3. Use the test pen to touch both ends of the inductance device. The display will show the measured inductance value.
NJTY T21L Multimeter measuring inductance
Figure 4: Multimeter performing an inductance measurement on a component.

6. Technical Specifications

6.1. Comprehensive Indicators

  • Operating conditions: 600V CAT II, Pollution grade: 2, Height: under 2000m
  • Working temperature: 0-40°C (<80%RH)
  • Storage temperature: -10-60°C (<70%RH, take off battery)
  • Test or calibrate surrounding temperature: 25°C ±2°C
  • Biggest voltage between measurement end and ground: 600V
  • Fuse protection: "µA/mA" fuse 600mA/250V; "A" fuse 10A/250V
  • Conversion rate: about 3s/second
  • Display: 6000 counts LED display
  • Overload: 'OL' displayed
  • Low voltage display of battery: Displayed when the working voltage is low
  • Input polarity indicator: "-" shown automatically
  • Battery: 3 x 1.5V AAA
  • Size: 147mm (L) x 71mm (W) x 45mm (H)
  • Weight: about 220g (battery excluded)

6.2. Accuracy Index

Accuracy: ± (%reading+digit), one year warranty from the manufacture date. Conditions: surroundings temperature from 18°C to 28°C, <80%RH.

6.3. Measurement Ranges and Accuracy

FunctionRangeResolutionAccuracy
DC Voltage (DCV)60mV0.01mV± (0.5%Readings+5digits)
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V
Input resistance: 10MΩ, Maximum input voltage: 600VDC
AC Voltage (TRMS)60mV0.01mV± (1%Readings+4digits)
600mV0.1mV
6V0.001V
60V0.01V
600V0.1V
Input resistance: 10MΩ, Maximum input voltage: AC600V, Frequency response: 40HZ to 1KHZ true RMS
Resistance600Ω0.1Ω± (0.8%Readings+5digits)
6KΩ0.001KΩ
60KΩ0.01KΩ
600KΩ0.1KΩ
6MΩ0.001MΩ
60MΩ0.01MΩ
Overload protection: 250V DC/AC, Open circuit voltage: 0.52V
Diode & BuzzerDiode TestN/AForward voltage display, Open-circuit voltage: 3.2V
Continuity TestN/ABuzzer sounds if resistance < 50Ω, Red indicator lights up. Open-circuit voltage: 1V
DC Current600µA0.1µA± (1.2%Readings+5digits)
6000µA1µA
60mA0.01mA
600mA0.1mA± (3%Readings+5digits)
6A0.001A
10A0.01A
Overload protection: µA/mA range fuse (600mA/250V); 10A range fuse (10A/250V). Continuous measurement time < 10s for current > 10A.
AC Current600µA0.1µA± (1.2%Readings+5digits)
6100µA1µA
60mA0.01mA
600mA0.1mA± (3%Readings+5digits)
6A0.001A
10A0.01A
Overload protection: µA/mA range fuse (600mA/250V); 10A range fuse (10A/250V). Maximum input current: mA level: 600mA AC; 10A level: 10A AC. Continuous measurement time < 10s for current > 10A. Frequency response: 40Hz-1KHz, true RMS.
Frequency9.999Hz0.001Hz± (1.5%Readings+5digits)
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999KHz0.001KHz
99.99KHz0.01KHz
999.9KHz0.1KHz
9.999MHz0.001MHz
Input voltage range: 200mV-10V AC RMS, Overload protection: 250V DC/AC
Capacitance6000pF1pF± (4.0%Readings+5digits)
60nF0.01nF
600nF0.1nF
6µF0.001µF
60µF0.01µF
600µF0.1µF
6mF0.001mF
60mF0.01mF
100mF0.1mF
Overload protection: 250V DC/AC
Temperature°C: -20°C ~ 1000°C1°C± (1.0%+3) Readings
°F: -4°F ~ 1832°F1°F± (1.0%+3) Readings
Overload protection: 250V DC/AC
Inductance600µH0.1µH± (4.0%Readings+5digits)
6mH0.001mH
60mH0.01mH
600mH0.1mH
6H0.001H
20H0.01H
Overload protection: 250V DC/AC

7. Maintenance

7.1. General Maintenance

To avoid electric shock or damage to the meter, do not get the inside of the meter wet. Before opening the case or battery cover, turn off the power and check that the test pen is disconnected from the circuit under test.

Clean the instrument case regularly with a damp cloth and a small amount of detergent. Do not use abrasives or chemical solvents. Dirty or damp input sockets can cause oxidation of the metal surface, which may affect readings.

7.2. Long-Term Storage

When not in use for a long time, please remove the battery to prevent leakage and avoid storing in high temperature and humidity environments.

8. User Tips

  • The multimeter is excellent for checking capacitors in-circuit and inductance measurement, providing fast and accurate results.
  • The conductivity (buzzer) function works perfectly and is a useful feature.
  • Overall, the equipment is very useful, with accurate voltage and resistance measurements.
  • The magnetic adsorption on the back is convenient for hands-free operation.

9. Troubleshooting

  • "OL" Display: If the display shows "OL" during measurement, it means the measured value exceeds the selected range. Switch to a higher range setting.
  • Low Battery Indication: When the battery low voltage symbol appears, replace the batteries immediately to ensure accurate readings and prevent potential safety hazards.
  • Unstable Readings: If readings are unstable or show large errors, ensure the meter is not being used near devices with large electromagnetic disturbances.
  • No Display/Power: Check battery installation and charge. Ensure the rotary switch is not in the "OFF" position.

10. Warranty and Support

The NJTY T21L Digital LCR Multimeter comes with a one-year warranty from the manufacture date, covering defects in materials and workmanship under normal use. For technical support, service, or warranty claims, please contact your retailer or the manufacturer's customer service department.


Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.