1. Overview
The SZBJ BM500A is a digital insulation resistance tester designed for measuring the insulation resistance of various electrical materials and equipment such as transformers, motors, cables, switches, and other electrical apparatuses. It features switchable output test voltages (250V/500V/1000V) and a measurement range up to 1999MΩ. Its compact structure and portability make it an ideal tool for electrical and electronic testing, maintenance, and inspection.
2. Safety Information
Warning: Important safety information. The user must read this manual carefully before operation.
Danger: High-voltage electric shock is present!
- Always read the operating instructions carefully before using the meter.
- Do not use the meter if its rear cover is not in place, as this may cause electric shock.
- Ensure the insulating layer of the test cables is in good condition without any breakage.
- To avoid electric shock, do not touch the test cables or the circuit under test during a test.
- Ensure test cables are securely inserted into the terminals.
- During testing, do not exceed the specified maximum input value for any range.
- Do not operate the switch knob to change ranges during a test, as this may damage the meter.
- Before performing an insulation test, ensure the range selection switch is set to an appropriate voltage range.
- Do not perform insulation resistance tests in combustible environments, as sparks may cause an explosion.
- Stop using the meter if its case or test cables are broken and metal is exposed.
- When opening the rear cover to change batteries, ensure test cables are removed and the range switch is in the OFF position.
- Remove batteries if the meter will not be used for a long time.
- When the battery symbol () is displayed, replace the battery promptly to maintain measurement accuracy.
- The meter features double-insulated protection (☐).
3. Name and Function of Parts
- LCD: Liquid Crystal Display.
- Data-hold Switch (HOLD): Used to freeze the displayed measurement.
- Rotary Switch: Used to select functions and measurement ranges (250V, 500V, 1000V, OFF).
- L (LINE) Input Terminal: Connects to the line terminal of the object being tested.
- G Input Terminal: Shielded terminal for insulation resistance measurements.
- E (EARTH) Input Terminal: Connects to the EARTH terminal of the object being tested.
- TEST/STOP Key: Used to initiate and stop insulation resistance measurement.
- Indicator Light of High Voltage: Illuminates when high voltage is output.
4. Setup
4.1. Connecting Test Terminals
- Insert the red test pen into the L terminal socket of the meter.
- Insert one end of the test cable with a red test clamp into the E terminal socket of the meter.
- Insert one end of the test cable with a black test clamp into the G terminal socket of the meter.
- Ensure all connections are secure.
The cable connected to the E terminal is the earthing cable. The cable connected to the L terminal is the circuit cable. The cable connected to the G terminal is the shielding cable, used to connect to the surface of the object under test to prevent surface leakage and improve impedance test accuracy.
4.2. Rated Voltage Selection
Before starting an insulation resistance test, turn the rotary switch knob to select the desired rated voltage (250V, 500V, or 1000V) corresponding to your test requirements.
5. Operating Instructions: Insulation Resistance Measurement
Special Attention in Operation:
- Before testing, ensure the circuit to be tested is NOT electrified. Do not measure any electrified equipment or circuit.
- During the test, the meter outputs dangerous voltage. Operate with extreme care.
- Ensure the object under test is securely clamped and keep your hands away from the test clamps before pressing the TEST key to output high voltage.
- When using 500V for resistance less than 1.5MΩ, or 1000V for resistance less than 5MΩ, the measurement time must not exceed 10 seconds.
5.1. Performing the Test
- Connect the test pen to the other terminal of the object under test.
- Press the TEST/STOP key. The red indicator light will turn on, indicating that high voltage output is active. The actual high voltage value will be displayed on the meter.
- Once the test begins, the numerical resistance value will be displayed on the LCD.
Attention: Do not short-circuit the two test pens when high voltage is output, or attempt to measure insulation resistance immediately after high voltage output. Improper operation can generate sparks, cause fire, and damage the meter.
5.2. Power Off
- After the test, release the TEST/STOP key. The red indicator light will turn off, indicating that high voltage output has been disconnected.
- Rotate the range switch to the OFF position. The LCD screen will turn off.
- For capacitive loads, first test the target for residual charge to ensure it's discharged, then remove the test lines.
6. Maintenance
Warning! Always turn off the power and disconnect test pens before opening the meter case or battery cover to prevent electric shock.
- Battery Replacement: When the battery symbol () is displayed, the battery needs to be replaced. Open the battery cover and replace the old battery with a new one of the same type (9V battery, 6F22 or NEDA 1604) to ensure normal operation.
- Cleaning: Keep the meter and test pens clean, dry, and undamaged. Use a clean cloth or an appropriate eradicator to clean the meter's case. Do not use abrasive or organic solvents.
- Storage: Avoid mechanical damage, shock, and impact. Store away from high temperatures and strong magnetic fields.
7. Troubleshooting
- Meter not powering on or inaccurate readings: Check the battery level. If the battery symbol () is displayed, replace the 9V battery (6F22 or NEDA 1604).
- No display or erratic readings: Ensure test cables are securely connected and undamaged. Verify the rotary switch is set to the correct range.
- Overload indication ("1" displayed): This indicates the measurement exceeds the meter's range. Select a higher voltage range if applicable, or ensure the circuit is within the meter's capabilities.
- Meter damage: If the case or test cables are broken, discontinue use immediately to prevent electric shock.
8. Technical Characteristics and Operating Description
Accuracy is specified for a period of one year after calibration, at 18°C to 28°C (64°F to 82°F) with relative humidity up to 75%.
| Parameter | Specification |
|---|---|
| Rated Voltage | 250V/500V/1000V |
| Open Circuit Voltage | 90%~110% of the rated voltage |
| Measurement Range | 0.001~1999MΩ |
| Minimum Resolution | 0.001 MΩ |
| Accuracy | ±(5%RDG+5) |
| Measurement Resistance Value (Lower Limit) |
|
| Short Circuit Current | Not less than 1.5mA |
General Characteristics:
- Overload Indication: "1" will be displayed for overload.
- Display Mode: Liquid crystal display; maximum display: 1999.
- Sampling Rate: 2.5 times per second.
- Voltage Indication: Displays actual insulation test voltage. LED light indicates high voltage output status.
- Operating Environment: 0°C-40°C, less than 75%RH.
- Storage Environment: -10°C-60°C, less than 80%RH.
- Insufficient Battery Indication: Battery symbol () is displayed.
- Power Supply: 9V battery, 6F22 or NEDA 1604.
- Dimensions: 69mm x 138mm x 31mm (with holster).
- Weight: Approx. 170g (including battery).
9. User Tips
- For optimal results, ensure the meter is fully charged or has fresh batteries before critical measurements.
- If you are new to insulation resistance testing, consider watching online video tutorials to familiarize yourself with the process.
- Always double-check your connections and selected voltage range before initiating a test to prevent damage to the meter or inaccurate readings.
10. Warranty and Support
The product includes a conformity certificate and warranty card. Please refer to these documents for specific warranty terms and conditions.
10.1. Included Accessories
- Test cable: one set
- Operating instructions: one copy (this manual)
- Conformity certificate/Warranty card: one piece
For further details or to download the original user manual, you may refer to the official PDF document: SZBJ BM500A User Manual (PDF).