Aideepen TS457

Rate this product if you’ve used it
Contents

Aideepen TS457 High-Precision Lithium Battery Internal Resistance Meter Tester User Manual

Model: TS457 | Brand: Aideepen

1. Introduction

The Aideepen TS457 is a professional entry-level battery voltage internal resistance tester designed for high-precision measurements. It utilizes an AC four-wire method principle for accurate testing of lithium batteries, including 18650 and 21700 types, and other DC5V power sources. This manual provides detailed instructions for setup, operation, and maintenance of your TS457 device.

Play video
Watch video online

Video: Product Introduction and Unboxing of the TS457 Tester.

2. Safety Precautions

  • DO NOT test AC power.
  • DO NOT test voltages above 45V.
  • DO NOT supply power to the device with overvoltage (ensure power supply is DC5V).
  • Ensure proper insulation for any unused wires, especially the yellow charging monitoring cable if not in use.

3. Package Contents

The standard package includes:

  • TS457 Voltage Internal Resistance Tester Unit
  • XH-2.54-3P Power Supply Cable (red, black, yellow wires, 150mm length)
  • XH-2.54-4P Signal Acquisition Cable (400mm length)

Optional accessories (purchased separately):

  • XH2.54-4P Kelvin Clamp
  • XH2.54-4P Battery Test Stand
  • 18650 Battery + Power Supply Adapter Board
TS457 Tester with included cables

Image: TS457 Tester with its standard power supply and signal acquisition cables.

4. Specifications

FeatureSpecification
Product NameTS457 Voltage Internal Resistance Tester
Model NumberFour-line TS457
TypeHousehold Battery Tester
DIY SuppliesElectrical
Brand NameAideepen
OriginMainland China
CertificationCE
Working VoltageDC5V (Do not supply power with overvoltage)
Test PrincipleAC 4-wire method
Voltage Testing Range0.001V ~ 45.00V
Voltage Resolution0.001V (Maximum)
Internal Resistance Range0.1mΩ ~ 6.999Ω
Internal Resistance Resolution0.1mΩ (Maximum)
Data RecordingRecords 5 sets of internal resistance data automatically
Voltage Display Fine-tuningSupported
Internal Resistance Display Fine AdjustmentSupported
Automatic ShutdownAdjustable 1-9 minute standby time
Backlight BrightnessAdjustable 1-9 levels
Product Size79.5 × 42.1 × 27.8mm
Installation Opening77 × 39.3mm
Product Weight45g
Battery IncludedNo

5. Setup

5.1 Power Supply Methods

The TS457 can be powered using a DC5V power supply or an 18650/21700 battery.

5.1.1 DC5V Two-Wire Power Supply

Use the XH-2.54-3P power supply cable (150mm) included in the package.

  • Connect the red wire to the positive electrode of your DC5V power source.
  • Connect the black wire to the negative electrode of your DC5V power source.
  • The yellow wire is a charging monitoring cable. If not used, ensure it is properly insulated.
TS457 DC5V two-wire power supply method

Image: Diagram showing DC5V two-wire power supply connection. The fine red line connects to the positive electrode, the fine black line to the negative electrode. The yellow line is not connected for basic power.

5.1.2 18650 Battery Power Supply (with Power Transfer Board)

If using an 18650 battery, a power transfer board (purchased separately) is required.

TS457 connected to 18650 battery via power transfer board

Image: TS457 connected to an 18650 battery using a power transfer board.

5.2 Test Probes/Braces

To perform measurements, you will need a test brace. This can be a self-made test fixture using the XH-2.54-4P signal acquisition cable or purchased accessories like the Kelvin clamp or battery test stand.

5.2.1 Testing with Kelvin Clamp

Solder the XH-2.54-4P signal acquisition cable to a Kelvin clip or battery test probe.

Kelvin clip test leads

Image: XH2.54-4P cable connected to Kelvin clamps.

Plug the XH2.54-4P connector into the 4P terminal on the back of the meter. Use the Kelvin clamps on both ends of the component to be tested.

Kelvin clamps connected to TS457 and a component

Image: Kelvin clamps connected to the 4P terminal on the back of the meter and a tested component.

5.2.2 Testing with Battery Test Stand

Connect the XH2.54-4P Battery Conversion Test Stand (purchased separately) to the 4P terminal on the back of the meter.

XH2.54-4P Battery Conversion Test Stand

Image: XH2.54-4P Battery Conversion Test Stand.

Insert the battery into the test stand for measurement.

Battery inserted into test stand connected to TS457

Image: Battery inserted into the test stand, connected to the TS457 device.

6. Operating Instructions

6.1 Power On/Off

  • When powered on for the first time, the device will automatically turn on.
  • To manually shut down, press and hold the R key for 3 seconds.
  • To turn on the device when it is off, click the R key.

6.2 Interface Display

6.2.1 Main Interface (Not Tested)

When no test is being performed, the voltage will display as zero, and the internal resistance will show as infinite (OL), indicating it exceeds the range.

TS457 display showing 0.0000V and OL for resistance

Image: TS457 display in standby mode, showing 0.0000V and 'OL' for resistance.

6.2.2 Main Interface (Measuring)

When measuring, the display will show the voltage and internal resistance. You can manually record the measured internal resistance value.

TS457 display showing voltage and resistance readings

Image: TS457 display showing active voltage and resistance readings.

6.3 Data Recording

After measuring the voltage and internal resistance value, click the M key to record the measured internal resistance value. The device can automatically record the last 5 groups of internal resistance measurement data.

To reset the recorded internal resistance measurement value, long press the ▼ button for 3 seconds.

6.4 Menu Navigation and Settings

Long press the M key for 3 seconds to enter the menu.

TS457 menu display

Image: TS457 menu interface showing options like Language, VOL-REF, RES-REF, STANDBY, BRIGHTNESS, RESET, EXIT.

  • Switch Menu Options: Click the M key to switch between menu options.
  • Adjust Settings: Use the ▲ or ▼ buttons to adjust the selected setting.
  • Language: Set the display language (e.g., ENGLISH).
  • Voltage Display Fine-tuning (VOL-REF): Adjusts the displayed voltage value. This only affects the display and has no effect on the actual voltage measurement.
  • Internal Resistance Display Fine Adjustment (RES-REF): Adjusts the displayed internal resistance value. This only affects the display and has no effect on the actual resistance measurement.
  • Standby Time Adjustment (STANDBY): Adjusts the automatic shutdown standby time from 1 to 9 minutes. Default is 2 minutes.
  • Backlight Brightness Adjustment (BRIGHTNESS): Adjusts the screen backlight brightness from 1 to 9 levels.
  • Restore Factory Settings (RESET): Click ▲ or ▼ to restore factory settings. "OK" will be displayed after restoration.
  • Exit Menu (EXIT): Click ▲ or ▼ in the "Exit" menu option to exit the system menu.

6.5 Screen Display Flipping

Long press the ▲ button for 3 seconds to flip the display orientation.

6.6 Automatic Shutdown

If no resistance or voltage is detected, the device will automatically turn off according to the "Energy saving Settings" time configured in the system menu. The default is 2 minutes.

TS457 display showing auto shutdown countdown

Image: TS457 display showing an auto shutdown countdown (e.g., 00:10).

7. Maintenance

  • Keep the device clean and dry. Avoid exposure to dust, moisture, and extreme temperatures.
  • Do not use abrasive cleaners or solvents on the device. Wipe with a soft, dry cloth.
  • Store the device in a safe place when not in use to prevent physical damage.
  • Regularly check cables and connectors for any signs of wear or damage. Replace if necessary.

8. Troubleshooting

  • Device does not power on: Ensure the DC5V power supply is correctly connected and providing power, or that the 18650 battery is charged and properly installed with the adapter board. Check the power supply cable for damage.
  • No readings or 'OL' displayed during test: Verify that the test probes (Kelvin clamps or battery test stand) are securely connected to both the device and the component being tested. Ensure the component is within the specified voltage and resistance ranges.
  • Inaccurate readings: Check for loose connections. If the issue persists, consider performing a factory reset via the menu settings. Remember that voltage and resistance fine-tuning only adjust the display, not the actual measurement.
  • Screen is dim: Adjust the backlight brightness in the menu settings (BRIGHTNESS option).
  • Device shuts off unexpectedly: Check the 'STANDBY' setting in the menu. If it's set to a short duration, the device will automatically power off after inactivity.

9. User Tips

  • For consistent and accurate measurements, always ensure a stable power supply.
  • When testing different battery types, ensure the test probes are appropriate for the battery's terminals to get a good contact.
  • Utilize the data recording feature to track changes in battery performance over time, which can be useful for identifying degrading batteries.
  • Adjust the backlight brightness to suit your working environment for optimal visibility and reduced eye strain.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact your retailer or the manufacturer directly. Keep your purchase receipt as proof of purchase.

Documents and downloads

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.

Was this page helpful?