MESTEK IR05B Dual Laser Infrared Thermometer User Manual

Model: IR05B | Brand: MESTEK

1. Introduction

The MESTEK IR05B is an industrial-grade dual laser infrared thermometer designed for non-contact temperature measurement. It features a wide temperature range, high precision sensors, and the ability to use a K-type thermocouple probe for contact measurements. This manual provides essential information for the safe and effective operation of your device.

Video: Overview of the MESTEK IR05B Infrared Thermometer's features and functions.

2. Safety Information

Please read and understand all safety warnings before using this product. Failure to follow these instructions may result in injury or damage to the device.

  • Do not point the laser directly at eyes or indirectly at the eyes through reflective surfaces.
  • Do not measure through transparent surfaces such as glass or plastic.
  • If the ambient temperature changes significantly/suddenly, wait for 30 minutes before use to allow the device to stabilize.
  • This device is not a medical thermometer and should not be used for measuring human body temperature.
  • Keep the device away from strong electromagnetic fields.

A user manual in PDF format is available for download for more detailed safety and operational guidelines. Please refer to the official product page for the link.

3. Product Overview

The MESTEK IR05B features a robust design with industrial-grade components for reliable performance.

Diagram of MESTEK IR05B internal components
Figure 3.1: Industrial Grade Components. Includes high precision laser indication, high precision sensors, aluminum alloy sensing housing, and a high-speed intelligent chip.

3.1 Panel Functions

Diagram of MESTEK IR05B display panel functions
Figure 3.2: Display Panel Functions.
  1. Emissivity
  2. Measurement indication
  3. Laser indication
  4. Battery undervoltage indication
  5. Data locking (HOLD)
  6. Irradiation measurement temperature value
  7. Temperature unit (°C/°F)
  8. Maximum lock (MAX)
  9. Minimum lock (MIN)
  10. Average (AVG)
  11. Temperature difference value (DIFF)
  12. Alarm
  13. High alarm (HI)
  14. Low alarm (LOW)
  15. K-type thermocouple temperature

4. Setup

4.1 Battery Installation

  1. Open the battery compartment.
  2. Install 2 No. 5 (AA) batteries, ensuring correct polarity.
Diagram showing battery installation and quick measurement steps
Figure 4.1: Battery installation (Step 1) and quick measurement steps.

5. Operating Instructions

5.1 Basic Temperature Measurement

  1. Pull the trigger to turn on the device and start measurement.
  2. Aim the dual lasers at the target object. The lasers indicate the measurement area.
  3. Release the trigger to hold the measurement on the display.
Image showing one-click temperature measurement in an industrial setting
Figure 5.1: One-click temperature measurement for quick response and safety.

5.2 Temperature Unit Switching (°C/°F)

The device supports both Celsius (°C) and Fahrenheit (°F) units. To switch between them:

  • Press the SET button repeatedly until the temperature unit flashes.
  • Use the UP/DOWN buttons to select the desired unit.
  • Short press the trigger to exit setup mode.
Display showing Celsius and Fahrenheit switching
Figure 5.2: Switching between Celsius and Fahrenheit units.

5.3 Measurement Value Modes (MAX/MIN/AVG/DIFF)

The default maximum value for startup can be toggled between minimum, average, and temperature difference values by lightly pressing the appropriate button.

Display showing Max, Min, Avg, Diff values
Figure 5.3: Displaying Max, Min, Avg, and Diff temperature values.

5.4 High and Low Temperature Alarm

The device can be set to alarm when the measured temperature exceeds a predefined high or low threshold.

  • Setting High Alarm: Long press the SET button. The maximum temperature flashes. Use UP/DOWN buttons to adjust the value.
  • Setting Low Alarm: Gently press the SET button again. The minimum temperature flashes. Use UP/DOWN buttons to adjust the value.
  • When the measured temperature exceeds the set alarm value, an alarm indication will appear on the instrument screen.
Image showing the thermometer measuring fish with high and low temperature alarm
Figure 5.4: High and low temperature alarm in action.

5.5 Adjustable Emissivity

Emissivity is adjustable from 0.1 to 1.0 to achieve higher precision measurements according to different material surfaces.

  • Press the SET button until the emissivity flashes.
  • Use the UP/DOWN buttons to adjust the emissivity value.
  • Short press the trigger to exit setup mode.
Table comparing emissivity values for different substances
Figure 5.5: Emission Rate Comparison Table for various materials. The default emissivity preset is 0.95.

5.6 K-type Thermocouple Measurement

For contact temperature measurement, use the K-type thermocouple probe.

  1. Insert the K-type thermocouple plug into the thermocouple socket of the instrument, observing positive and negative pole markings.
  2. Ensure the thermocouple probe contacts the measured object.
Image showing K-type thermocouple temperature measurement
Figure 5.6: K-type thermocouple in use for contact temperature measurement. Exposure temperature range: -40~2000°C (-40~3632°F). K-type temperature range: -10~500°C (14~932°F).

5.7 Measurement Distance (D:S Ratio)

The device has a Distance-to-Spot (D:S) ratio of 50:1, meaning that at a distance of 50 units, the measurement spot diameter is 1 unit. This ratio is crucial for accurate positioning and measurement.

Diagram illustrating the D:S ratio of 50:1 for accurate measurement
Figure 5.7: Safety measures and D:S ratio of 50:1. The ratio of accurate positioning measurement, measurement distance, and measurement target diameter.

5.8 Specific Measurement Tips (Water/Oil Temperature)

When measuring boiling liquids, ensure that smoke or steam from the boiling water does not come into direct contact with the temperature sensing area. Maintain a measurement angle of approximately 45° to 60° to avoid inaccurate readings caused by steam.

Diagram showing correct and incorrect methods for measuring water/oil temperature
Figure 5.8: Correct measurement method for water/oil temperature to avoid steam interference.

6. Maintenance

  • Cleaning: Use a soft, damp cloth to clean the device. Do not use abrasive cleaners or solvents.
  • Storage: Store the device in a dry, cool place when not in use.
  • Battery Replacement: Replace batteries when the low battery indicator appears on the display.
  • Sleep Mode: The thermometer will automatically enter a sleep state after 35 seconds without any operation to conserve energy.
Image illustrating the sleep mode feature for energy saving
Figure 6.1: Sleep mode activates after 35 seconds of inactivity.

7. Troubleshooting

  • Inaccurate Readings:
    • Check emissivity settings for the material being measured.
    • Ensure the D:S ratio is appropriate for the measurement distance.
    • Avoid steam or smoke interference when measuring liquids.
    • Allow the device to acclimate to ambient temperature if there's a sudden change.
  • No Display/Device Not Turning On:
    • Check battery installation and replace batteries if necessary.
  • Alarm Not Functioning:
    • Verify high and low alarm settings are correctly configured.

8. Specifications

Table of product parameters for MESTEK IR05B
Figure 8.1: Product Parameters Table.
Parameter Value
ModelIR05B
Temperature range-40°C~2000°C (-40°F~3632°F)
Accuracy-40°C~800°C (±1.5%), 800°C~2000°C (±3%)
K-type range-10°C~500°C (14°F~932°F)
K-type accuracy±(1.5%+2°C/4°F)
DisplayVA reverse display color screen
D:S50:1
Spectral response8~14um
Resolving power0.1°C/0.1°F
Emissivity0.10~1.00
Laser<1mW /620~690nm Level 2
Response time<0.5S
Automatic shutdownAbout 35 seconds
Usage environment0~50°C / 10%~90%RH
Storage temperature-20°C~60°C
Size180x120x60mm
Power source2x1.5V AA battery
WeightApproximately 265g (excluding battery)

9. User Tips

  • For highly reflective surfaces, consider applying a matte black paint or tape to the measurement area to get a more accurate reading, then measure the temperature of the painted surface.
  • Always ensure the target object fills the measurement spot indicated by the lasers for the most accurate results.
  • When using the K-type probe, ensure good contact with the surface for stable readings.