1. Introduction
The Yieryi PH-W218EU is an advanced 8-in-1 WiFi online water quality tester designed for continuous monitoring of various parameters in aquariums, pools, hydroponic systems, and other water environments. This smart device integrates with the Tuya Smart APP, allowing for real-time data monitoring, automatic data storage, alarm settings, and data export capabilities.
The tester measures pH, EC (Electrical Conductivity), Salinity, ORP (Oxidation-Reduction Potential), TDS (Total Dissolved Solids), CF (Conductivity Factor), S.G. (Specific Gravity), and Temperature.

2. Specifications
| Parameter | Range | Resolution | Accuracy |
|---|---|---|---|
| pH | 0.00 - 14.00pH | 0.01PH | ±0.03pH |
| ORP | -2000mv - +2000mv | 1mv | ±5mv |
| EC | 0 - 19000us/cm, 0 - 199.0 ms/cm | 1us/cm (<1999us/cm), 10us/cm (>1999us/cm), 0.1ms/cm | ±2%F.S |
| CF | 0 - 1999.0CF | 1CF | ±2%F.S |
| TDS | 0 - 19990ppm, 0 - 199.0ppt | 1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt | ±2%F.S |
| SALT | 0 - 199.9ppt, 0% - 20% | 1ppm (<1999ppm), 10ppm (>1999ppm), 0.1ppt, 0.01% | ±2%F.S |
| S.G. | 0.990 - 1.400 | 0.001 | ±2%F.S |
| Temperature | 0 - 50°C (32 - 122°F) | 0.1°C (0.2°F) | ±1.0°C |
General Specifications:
- Power Supply: DC6V
- Temperature Compensation: 0°C - 50°C
- Operating Temperature: 0°C - 50°C
- Dimensions: 152mm × 80mm × 26mm
- Weight: 245g
- Display Type: DIGITAL
- Backlight: Yes
- Smart Device: Yes (WiFi enabled)
- Battery Included: No
- Origin: Mainland China
- Certification: CE
Package Dimensions:
- Length: 25 cm
- Width: 18 cm
- Height: 7 cm
- Weight: 0.78 kg
3. Setup Guide
3.1 Initial Hardware Setup
- Remove the protective caps from the pH, ORP, and temperature electrodes.
- Connect the pH electrode, ORP electrode, and temperature electrode to their respective ports on the main unit.
- Clean all electrodes with distilled water and dry them well before initial use.
- Connect the power adapter to the main unit and plug it into a suitable power outlet.
- For wall-mounted installation, use the provided accessories to secure the main unit above your water body. Ensure the unit is not exposed to water or damp conditions as it is not waterproof.

3.2 Mobile App Connection (Tuya Smart)
- Download the App: Search for "Tuya Smart" in your mobile APP store (available for Apple iOS and Android systems) and download it.
- Register/Log In: After the download is complete, open the app. If you are a new user, register for an account. Otherwise, log in to your existing Tuya Smart account.
- Add Device: On the app's homepage, click "Add Device" or the '+' icon. The app will search for nearby devices. Ensure your phone's WiFi and Bluetooth are enabled.
- Connect to WiFi: Ensure both your phone and the PH-W218EU device are connected to the same 2.4GHz WiFi network. Enter your WiFi password in the app.
- Initiate Pairing: On the PH-W218EU main unit, long press the "MODE EC/ORP-" button to enter pairing mode. The app should then automatically search for and add the device.
- Confirm Connection: Once the device is found and added, a confirmation message will appear in the app. Click "Finish" to complete the setup.

4. Operation
4.1 Basic Measurement
- Ensure the device is powered on and electrodes are connected.
- Immerse the pH electrode, temperature electrode, and TDS electrode into the testing solution.
- Stir the electrodes lightly in the solution and wait a few seconds for the readings to stabilize on the display.
- Press the "TEMP CAL" key to switch between °C and °F temperature units.
- Press the "MODE EC/ORP-" key to cycle through different measurement values such as TDS, EC, SALT, S.G., and CF.

4.2 Calibration Procedures
Regular calibration ensures accurate readings. Always use fresh, appropriate standard buffer solutions for calibration.
4.2.1 pH Calibration (3-point correction)
- Prepare three standard buffer solutions: pH 6.86, pH 4.00, and pH 9.18 (at 25°C). For accurate calibration, pour each solution into two separate clean beakers: one for cleaning the electrode and one for calibration.
- Plug in and turn on the device.
- Calibrate pH 6.86: Immerse the electrode into the pH 6.86 standard buffer solution. Stir gently until the value stabilizes. Long press the "TEMP CAL" button for 5 seconds. The screen will display "06.86" and enter the pH 6.86 automatic correction mode. The meter automatically identifies standard buffer solutions.
- Calibrate pH 4.00: Clean the electrode with distilled water. Immerse it into the pH 4.00 standard buffer solution. Long press the "TEMP CAL" button for 5 seconds when the display value is stable. The screen will display "04.00" and enter the pH 4.01 automatic calibration mode.
- Calibrate pH 9.18: Clean the electrode with distilled water. Immerse it into the pH 9.18 standard buffer solution. Long press the "TEMP CAL" button for 5 seconds when the display value is stable. The screen will display "09.18" and enter the pH 9.18 automatic calibration mode.
- Alternative Calibration Points: Users can choose calibration solutions of 6.86, 4.00, 9.18 or 7.00, 4.00, 10.00. The factory default calibration is 6.86, 4.00, 9.19.
- Switching Calibration Points: If the value 6.86 is flashing, immediately press "TEMP CAL" to switch to 7.00 calibration. Be careful not to use the wrong calibration solution.
4.2.2 EC Calibration
- Turn on the meter. Press the "MODE EC/ORP-" key to select the EC-ms display.
- Immerse the electrode into distilled water for about five minutes to activate it.
- Immerse the electrode into EC calibration solution 12.88ms/cm (at 25°C) and stir gently.
- When the reading is stable, long press the "CAL EC/ORP+" key for about 5 seconds. The screen will display "12.80" or similar.
- Press the "MODE EC/ORP-" or "CAL EC/ORP+" button until the display shows a value between "12.70" and "12.90".
- Clean the electrode with distilled water and dry it with filter paper.
- For 1.413ms/cm calibration, immerse the electrode into the 1.413ms/cm solution (at 25°C) and stir gently. The display value should match or be close to the calibration solution value.
- Clean the electrode with distilled water and dry it with filter paper after calibration.
4.2.3 ORP Calibration
- Plug in the power to turn on the device.
- Immerse the ORP electrode into distilled water for five minutes.
- Immerse the ORP electrode in ORP calibration solution of 265mv (25°C) and stir gently.
- When the reading is stable, press and hold the "ORP CAL" button for about 5 seconds. The screen will display "265" or similar.
- Press the "MODE EC/ORP-" or "CAL EC/ORP+" button to adjust the value until it shows "265".
- Clean the electrode with distilled water and dry it with filter paper.
- For 100mv calibration, immerse the electrode into the 100mv solution (at 25°C) and stir gently. The display value should match or be close to the calibration solution value.
- Clean the electrode with distilled water and dry it with filter paper after calibration.
4.3 Tuya Smart App Functions
The Tuya Smart app provides comprehensive control and monitoring features:
- Real-time Monitoring: View all 8 parameters (pH, EC, Salinity, ORP, TDS, CF, S.G., Temperature) in real-time on your smartphone.
- Data Logging: The device automatically stores data for up to 30 days.
- Data Export: Export historical data to Excel format and send it to your email for further analysis. This can be done hourly, daily, weekly, or monthly.
- Alarm Settings: Set high and low alarm thresholds for parameters like pH. The app will send push notifications to your phone when values exceed or fall below these thresholds.
- Language Conversion: The app automatically converts language based on your phone's language and region settings.

5. Maintenance
- Electrode Cleaning: Always clean the electrodes with distilled water before and after each use, especially before and after calibration. Dry them gently with a clean, soft cloth or filter paper.
- Storage: After use, turn off the meter and replace the protective caps on all electrodes. Store the device in a dry, safe place.
- Re-calibration: The electrodes must be re-calibrated periodically. Re-calibration is particularly important if:
- The device has not been calibrated for a long time.
- The device is in regular and long-term continual use.
- High accuracy is required for your measurements.
- Avoid Water Exposure: The main unit is not waterproof. Ensure it does not get wet or damp during use or storage to prevent malfunction.
6. Troubleshooting
- Inaccurate Readings: If readings appear inaccurate, perform a full calibration for the affected parameters (pH, EC, ORP) using fresh standard buffer solutions. Ensure electrodes are clean.
- Device Not Turning On: Check the power adapter connection and ensure it is plugged into a working power outlet.
- WiFi Connection Issues:
- Ensure your mobile phone and the device are connected to the same 2.4GHz WiFi network.
- Verify the WiFi password entered in the Tuya Smart app is correct.
- Make sure the device is in pairing mode (long press "MODE EC/ORP-" button).
- Restart your router, phone, and the device, then try connecting again.
- App Not Displaying Data: Check the WiFi connection status of the device and your phone. Ensure the device is properly added and online in the Tuya Smart app.
- Device Malfunction after Water Exposure: The tester is not waterproof. If it gets wet, immediately disconnect power and allow it to dry completely. If issues persist, contact customer support.
7. User Tips
- Probe Replacement: While not explicitly detailed, the modular design with separate electrode connections suggests that individual probes (pH, ORP, Temperature, TDS) are replaceable. This is a common feature for such devices, allowing for extended product life.
- Salinity Calibration: Although the manual details pH, EC, and ORP calibration, ensure that if specific salinity calibration solutions are available, they are used to maintain accuracy for salinity measurements.
- Cable Length: Users often inquire about cable length for installation flexibility. While not specified, consider the placement of the main unit relative to the water body to ensure probes can reach the desired depth.
- Power Source: The device uses a DC6V power adapter. Avoid using incompatible power sources, such as USB hubs, unless explicitly stated as compatible by the manufacturer, to prevent damage.
8. Frequently Asked Questions (FAQ)
- Q: How do I connect the device to WiFi?
- A: Download the Tuya Smart app, register/log in, click 'Add Device', ensure your phone and the tester are on the same 2.4GHz WiFi, enter your password, and long press the 'MODE EC/ORP-' button on the tester to initiate pairing.
- Q: Are the pH probes replaceable?
- A: The manual indicates separate connections for pH, ORP, and temperature electrodes, suggesting they are designed to be replaceable. Specific replacement instructions or part numbers are not provided in this manual.
- Q: Where can I get a temperature sensor for this device?
- A: The PH-W218EU comes with a dedicated temperature electrode as part of its standard package, which is used for temperature measurement and compensation.
- Q: How do I calibrate salinity?
- A: The manual provides detailed calibration steps for pH, EC, and ORP. While salinity is measured, specific calibration steps for salinity are not explicitly outlined. It is generally calibrated in conjunction with EC. Refer to the EC calibration section for related procedures.
- Q: What is the length of the electrode cables?
- A: The exact length of the electrode cables is not specified in this manual. Please consider your installation requirements when setting up the device.
9. Warranty and Support
9.1 Warranty Information
All defects in materials and manufacture of this device are guaranteed for one year from the date of purchase. Within this one-year period, please return the parts to the dealer or our office for free repair if the damage is not caused by the user's negligence or improper operation.
9.2 Additional Resources
For a downloadable PDF version of the user manual, please refer to the official product page or contact customer support. A PDF version is available at: User Manual (PDF)