1. Introduction
The ELECTESLA ES-UL800 is an intelligent ultrasonic level meter designed for precise, non-contact level measurement in various applications. It utilizes ultrasonic technology to determine the distance to the liquid surface, providing reliable data for monitoring and control. This manual provides detailed instructions for the installation, operation, calibration, and maintenance of your ES-UL800 device.

Figure 1: ELECTESLA ES-UL800 Ultrasonic Level Meter with key features (Range: 0-10m, Output: 4-20mA/RS485, Power Supply: DC24V, IP68).
2. Specifications
The ES-UL800 Ultrasonic Level Meter offers robust performance and versatile features:
| Parameter | Value |
|---|---|
| Model Number | ES-UL800 |
| Measuring Range | 0-10m (customizable up to 25m) |
| Blind Area | 0.25-0.5m |
| Accuracy | 0.25% - 0.5% |
| Resolution | 1mm |
| Pressure | Below 4 atmospheres |
| Display | LCD display (shows level height and spatial distance) |
| Analog Output | 4-20mA, 0-10V, 0-5V |
| Digital Output | RS485 Modbus |
| Power Supply | DC24V; AC/DC220V (Optional, built-in lightning protection) |
| Medium | Water, Fuel, Oil, etc. |
| Working Temperature | -40~70℃ (Standard); -20~+60℃ (High temperature customized) |
| Protection Grade | IP68 |
| Certification | CE, FCC, RoHS |
| OEM Logo | Yes |

Figure 2: Key features of the ES-UL800, including clear LCD display, waterproof connection for reduced interference, and a completely enclosed probe for strong penetration and accuracy.
3. Product Dimensions
The ES-UL800 is designed for compact installation. Below are the typical external dimensions:
- Overall Height: Approximately 260mm
- Top Diameter: Approximately 102mm (98mm for mounting)
- Base Diameter (Probe): Approximately 54mm (92mm for mounting)
- Thread Size: G2 (Length 30mm)
Note: Thread material, size, and length can be customized.

Figure 3: The ES-UL800 supports range customization, with 0-10M range being optional, illustrating its adaptability to various tank depths.
4. Installation and Setup
Proper installation is crucial for accurate and reliable performance of the ultrasonic level meter. Ensure the probe surface is level with the measured liquid surface.
4.1. Installation Methods
In open environments, bracket mounting or direct nut mounting are common. For tanks or pools, cut a hole slightly larger than the probe diameter (60mm) at the installation point. Place the instrument and secure it with the flange or nut from bottom to top.
- Plastic Nut Installation: Use a plastic nut made of the same material as the probe to secure the device.
- Bracket Installation: A mounting bracket can be used for secure positioning. This option is often preferred for stability.
- Installation Reference: Ensure the front of the probe extends out from the mounting surface for optimal performance. Avoid situations where the probe is inside a pipe or obstructed.
4.2. Installation Principle Schematic
The ultrasonic sensor emits sound waves that travel to the liquid surface and reflect back. The device calculates the distance based on the time taken for the echo to return. The