ZOYI MD1

Contents

ZOYI MD1 Mini Smart Tweezer LCR SMD ESR Tester User Manual

Model: MD1 (ZT-MD1)

1. Overview

The ZOYI MD1 is a mini digital LCR tweezers Bridge Tester, designed for extended operation with its lithium polymer battery. It features an OLED display for a clear interface, intelligently identifying and accurately measuring resistance, capacitance, and inductance across a wide range with high precision. Its compact and elegant design ensures portability. The flexible measuring arms with detachable gold-plated tips provide reliable operation, making it suitable for various SMD or other LCR parameter measurements.

This device also includes a user calibration learning mode, allowing for calibration based on standard resistors. Factory calibration data can be restored by resetting to factory settings.

ZOYI MD1 LCR Bridge SMD Test Clip features

Figure 1: ZOYI MD1 LCR Bridge SMD Test Clip highlighting key features such as Resistance (10mΩ-10MΩ), Diode measurement, Lithium battery power, Component automatic identification, Adjustable highest accuracy (0.5%), Inductance (1uH-60H), Frequency (100Hz-10KHz adjustable), and Capacitance (1pF-20mF).

2. Safety Precautions

To prevent electric shock, fire, or personal injury, please read the following safety precautions before use:

  1. Do not apply input voltage during measurement. Do not measure "live circuits." Disconnect power before online measurement and ensure all capacitors in the circuit are discharged.
  2. Discharge capacitors before measuring capacitance.
  3. Avoid prolonged use or storage in high temperature, high dust, or direct sunlight environments.
  4. Do not use this product in explosive gas, vapor, or humid environments.

3. Setup

3.1. Package Contents

Before using your ZOYI MD1, please ensure all items are present in the package:

  1. LCR tweezers bridge (main unit)
  2. Instructions manual (this document and the included paper manual)
  3. Customized tweezers storage bag
  4. Charging line (USB Type-A to Type-C)
  5. Packaging color box
ZOYI MD1 package contents

Figure 2: Contents of the ZOYI MD1 package, including the LCR tweezers bridge, instructions, customized storage bag, charging line, and packaging box.

3.2. Charging the Device

The ZOYI MD1 is powered by a built-in 300mA lithium battery, providing sustainable operation for up to 12 hours. When the battery icon in the top right corner of the screen shows one red bar after prolonged use, it needs to be charged promptly. If continued use persists until the battery shutdown voltage is reached, the buzzer will emit intermittent beeps and then shut down automatically.

  1. Insert the Type-C data cable to connect the tweezer meter to a DC 5V output adapter or a USB port on a computer to begin charging.
  2. During charging, the charging indicator light will remain solid red. When fully charged, the indicator light will change from red to blue.

Note:

  • Do not exceed the charging voltage range of DC 5V.
  • The charging indicator light (Label 2 on the operation panel) is solid red while charging and turns blue when fully charged.
ZOYI MD1 charging via Type-C interface

Figure 3: ZOYI MD1 being charged directly via its Type-C interface, highlighting the built-in lithium battery for extended use.

3.3. Replacing Tweezer Tips

If the tweezer tips show signs of wear, oxidation, or damage, they can be replaced:

  1. Loosen the 4 screws from each tweezer tip using a 3.5PM2*3 screwdriver. Ensure the tool matches the screw type.
  2. Align the new tweezer head with the slot and insert it.
  3. Tighten each screw securely.
Instructions for replacing the tweezer head on ZOYI MD1

Figure 4: Step-by-step guide for replacing the tweezer head: 1. Loosen screws, align head with slot, insert; 2. Tighten screws.

4. Operating Instructions

4.1. Operation Panel

Label No.NameFunction
1TYPE-C InterfaceUsed to connect the data cable for battery charging and computer communication.
2Charging Indicator LightSolid red while charging. Turns blue when fully charged.
3OLEDDisplays all interfaces and values.
4Power Button / MODE1. Short press to turn on, long press (2s) to turn off.
2. In measurement interface, short press to switch between AUTO, L, C, R, Diode, and Continuity modes.
3. In settings interface, short press to confirm or exit.
5Right Arrow Key / FREQ / MENU1. In measurement interface, short press to switch frequencies.
2. Long press to enter or exit the settings interface.
3. In settings interface, short press to move down the menu and adjust settings.
6Left Arrow Key / HOLD / LEVEL1. In measurement interface, short press to hold current value.
2. Long press to switch measurement voltage.
3. In settings interface, short press to move up the menu and adjust settings.
7Measurement ArmHold the measurement arm with your fingers to operate and measure.
8Tweezer TipsContact both ends of the component being measured with the left and right tweezer tips.

4.2. Setting Operations

  1. Long press the right arrow key (5) to enter the settings interface. Short press the left or right arrow key (6) to select the menu to be adjusted. After selecting the menu, press the power key (4) to confirm. The selected menu will turn yellow. Then, press the left or right arrow key to adjust the parameters. After adjusting, press the power key to confirm and exit.
  2. To exit the settings interface, long press the right arrow key or power off the device.

4.3. Calibration Instructions

When calibration is needed, follow these instructions. Calibration includes 0Ω, 1Ω, 10Ω, 100Ω, 1KΩ, 10KΩ, 100KΩ, 1MΩ, 10MΩ, OPEN, and individual range calibration.

  1. Enter the settings interface, select the calibration settings menu, and short press the power key to confirm. The calibration settings menu will turn yellow, and the screen will prompt "Waiting for calibration."
  2. Short press the right arrow key to select the range to be calibrated. The corresponding range value will appear in the middle column of the right side of the screen.
  3. Fully contact both ends of the standard resistor with the tweezer tips, then press the left arrow key. The screen will prompt "Calibrating."
  4. Wait patiently for 20 seconds. When calibration is complete, the buzzer will beep once, and the screen will display "Calibration successful." If the calibration value is incorrect, the buzzer will beep twice, and the screen will display "Calibration failed."

Notes:

  • Use chip carbon film resistors for calibration, not wire-wound resistors.
  • For 0Ω calibration, short-circuit the tweezer tips completely together.
  • For OPEN calibration, do not connect any resistor, keeping the tweezer tips in an open circuit state.
  • To return to factory calibration data, perform a factory reset.

4.4. Measurement Modes

The ZOYI MD1 can automatically identify components or be set to specific measurement modes. The main and secondary parameters are displayed simultaneously on the OLED screen.

Main and secondary parameter measurement display on ZOYI MD1

Figure 5: The ZOYI MD1 display showing main parameter measurement (up to 9999 characters for L/C/R tests) and secondary parameter measurement (up to 4 decimal places for D/Q tests) simultaneously.

4.4.1. Measuring Resistance

  1. Use the AUTO mode or manually switch to the R mode.
  2. Contact both ends of the resistor to be tested with the tweezer tips.
  3. Read the measurement value displayed on the screen.

Notes:

  • Ensure full contact with the surface when measuring low resistance. Surface oxidation can affect measurement accuracy.
  • If the measurement value exceeds the range, the screen will display "OL."

4.4.2. Measuring Capacitance

  1. Use the AUTO mode or manually switch to the C mode.
  2. Contact both ends of the capacitor to be tested with the tweezer tips.
  3. Read the measurement value displayed on the screen.

Notes:

  • Discharge the capacitor before measurement to avoid damaging the instrument.
  • If the measurement value exceeds the range, the screen will display "OL."
Explanation of the difference between Bridge Measurement of Capacitors and Multimeter Measurement

Figure 6: Comparison illustrating that a standard multimeter only measures capacitance value, while a bridge (like ZOYI MD1) can determine capacitor quality by measuring its D value.

4.4.3. Measuring Inductance

  1. Use the AUTO mode or manually switch to the L mode.
  2. Contact both ends of the inductor to be tested with the tweezer tips.
  3. Read the measurement value displayed on the screen.

4.4.4. Measuring Diodes

Manually switch to the diode mode. Contact the positive and negative terminals of the diode with the tweezer tips, and read the diode's polarity and voltage value.

Notes:

  • Determine the diode's polarity according to the diode symbol direction displayed on the screen and the corresponding measurement object with the tweezer tips.
  • The product's diode forward voltage measurement value is for reference comparison only, with a measurement range of Vf 0.1V-0.6V. Values beyond this range will display "OL." For higher measurement requirements, select a more professional diode measuring instrument.

4.4.5. Continuity Measurement

Switch manually to the continuity mode. Touch the tweezer tips to both ends of the point to be tested. A beep will sound when the circuit is closed, and observe the measurement reading on the screen.

Note:

  • Due to the influence of parallel capacitors during online measurement, continuity measurement results are for reference comparison only and should not be used as a standard for determining continuity.

5. Maintenance

5.1. Cleaning

Do not attempt to open or repair the product, or alter its circuitry. Clean with a damp cloth and mild detergent; do not use corrosive agents or solvents. Dust or moisture in the test ports may affect reading accuracy.

5.2. Firmware Upgrade

  1. With the device powered off, press and hold the left arrow key, then simultaneously press the power key. The screen will display "USB-Boot" prompt.
  2. Connect the TYPE-C data cable to the computer. The computer will recognize the LCR meter as a disk.
  3. Drag the prepared upgrade file into the LCR meter disk. The upgrade will proceed automatically.
  4. The screen will automatically switch to the measurement interface upon successful upgrade.

6. Troubleshooting

  • "OL" Display: This indicates that the measured value exceeds the device's measurement range for the current mode. Check the component's expected value or switch to a different range if available.
  • Calibration Failed: Ensure you are using appropriate standard resistors (chip carbon film, not wire-wound) and following the calibration steps precisely. For 0Ω calibration, short-circuit the tips completely. For OPEN calibration, ensure nothing is connected to the tips.
  • Incorrect Component Recognition in AUTO Mode: If the device incorrectly identifies a component (e.g., shows capacitance for a resistor), try switching to the specific manual measurement mode (R, C, L) for that component.
  • Unstable or No Measurement for Very Low Capacitance: The device may have limitations for measuring capacitance below approximately 30pF, and measurements can be unstable at very low values. This is common for many handheld testers.
  • Device Not Turning On: Ensure the battery is charged. Connect the device to a power source using the Type-C cable and check the charging indicator light.

7. Specifications

7.1. Technical Specifications

FunctionRangeAccuracy at 100HzAccuracy at 1KHzAccuracy at 10KHzMeasurement Range
Resistance1MΩ-10MΩ5%±55%±51%±5Automatic Mode: 10mΩ-10MΩ
Resistance Mode: 10mΩ-10MΩ
1KΩ-1MΩ1%±50.5%±50.5%±5
10mΩ-1KΩ1%±50.5%±50.5%±5
Capacitance1mF-20mF5%±55%±52%±5Automatic Mode: 50pF-5mF
Capacitance Mode: 1pF-20mF
1uF-1mF2%±52%±50.5%±5
1nF-1uF2%±50.5%±50.5%±5
1pF-1nF5%±55%±55%±5
Inductance1H-60H2%±55%±52%±5Automatic Mode: 5uH-60H
Inductance Mode: 1uH-60H
1mH-1H2%±52%±50.5%±5
10uH-1mH2%±50.5%±50.5%±5
1uH-10uH5%±55%±5
ParametersL, C, R, D, Q, Rs
ModesAutomatic mode, resistance mode, capacitance mode, inductance mode, diode mode, continuity mode
Frequencies100Hz, 1KHz, 10KHz
Voltages0.3V, 0.6V

7.2. General Technical Specifications

FeatureSpecification
DisplayOLED
RangeAutomatic
Testing Rate1 time/second
Data Hold√
Language SelectionChinese, English
Screen OrientationLeft, Right
Volume Setting0%~100%
Backlight Brightness10%~100%
Auto Power OffOff, 5 Min, 120 Min
Calibration Setting0Ω-10MΩ, OPEN
Restore to Factory Default Settings√
Firmware Upgrade√
Low Battery Indicator√

7.3. Mechanical Technical Specifications

FeatureSpecification
Dimensions146*30*18mm
Weight37g
Battery TypeLithium polymer rechargeable battery 3.7V 400mAh
MaterialABS
Warranty PeriodOne year

7.4. Environmental Specifications

EnvironmentParameterValue
Operating EnvironmentTemperature0~40°C
Humidity<75%
Storage EnvironmentTemperature-20~60°C
Humidity<80%

8. User Tips

  • Frequency and Voltage Settings: While the device supports 100Hz, 1KHz, and 10KHz frequencies and 0.3V/0.6V test voltages, ensure you are in the correct mode to adjust these. Some users might find the device defaults to 1KHz and 0.6V in certain automatic modes.
  • Manual Mode for Specific Components: If the automatic identification feature sometimes misidentifies a component, switch to the specific manual mode (Resistance, Capacitance, or Inductance) for more accurate and consistent readings.
  • Low Capacitance Limitations: Be aware that measurements for very low capacitance values (e.g., below 30pF) might be unstable or outside the device's reliable measurement range.
  • Discharge Capacitors: Always discharge capacitors before testing to prevent damage to the device and ensure accurate readings.
  • Calibration Resistors: For calibration, use chip carbon film resistors rather than wire-wound resistors for best accuracy.
  • Refer to Full Manual: For any further details or specific diagrams, please refer to the original PDF user manual available at ZOYI MD1 User Manual (PDF).

9. Warranty and Support

9.1. Limited Warranty and Scope of Rights and Responsibilities

From the date of purchase, this product is covered by a one-year warranty. This warranty excludes damages caused by accidents, negligence, misuse, modification, contamination, and abnormal operating environments.

This manual is subject to change without prior notice. The contents of this manual are believed to be accurate. If users find any errors or omissions, please contact the manufacturer.

The company is not liable for accidents and damages caused by user error. The functions described in this manual are not intended to justify the use of the product for special purposes.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.