HoldPeak HP-770HC Digital Multimeter and HP-605A Clamp Adapter User Manual

Model: HP-770HC (Digital Multimeter) & HP-605A (AC/DC Clamp Adapter)

Brand: HoldPeak

1. Introduction

The HoldPeak HP-770HC Digital Multimeter is a slim, portable, and stable instrument designed for precise electrical measurements. Featuring a 6000-count LCD display with 25mm high characters, it provides clear readings. Its robust circuitry, including large-scale IC A/D converters and overload protection, ensures excellent performance and durability.

This multimeter, when used in conjunction with the HP-605A AC/DC Clamp Adapter, can measure DC & AC voltage, DC & AC current (up to 600A), resistance, capacitance, frequency, duty cycle, temperature, transistor hFE, Non-Contact AC Voltage (NCV) detection, positive diode voltage fall, and audible continuity.

1.1 Safety Information

This meter is designed according to IEC-1010 standards for electronic measuring instruments, with an over-voltage category of 1000V (CAT III) and pollution degree 2. Always follow all safety and operating instructions to ensure safe and proper use.

  • Important Safety Information: Refer to the operating manual for critical safety details.
  • Dangerous Voltage: Be aware that dangerous voltage may be present during operation.
  • Double Insulation: The device features double insulation (Protection Class II).

1.2 Special Cautions for Operation

  • Always use the supplied test leads. If replacement is necessary, use only leads of the same model and electrical specifications.
  • To avoid electric shock, do not use the meters before the cover is in place.
  • Ensure the range switch is in the correct position for the test being performed.
  • Do not exceed the specified input signal limits to prevent electric shock and instrument damage.
  • When measuring TV sets or switched power supplies, be cautious of possible pulses that could damage the circuit.
  • Avoid changing the range switch randomly during measurement.
  • Exercise caution when measuring voltages higher than DC 60V or AC 30V to prevent shock.
  • Replace protection fuses only with the same type and specification.
  • After use, set the function switch to 'OFF' to conserve battery power.
  • If the meter will not be used for an extended period, remove the battery to prevent leakage damage.

2. Components and Key Functions

2.1 HP-770HC Digital Multimeter Panel Layout

HP-770HC Digital Multimeter Key Function Diagram
Diagram of the HP-770HC Digital Multimeter with labeled key functions.
  1. SELECT key: Used to choose between resistance, diode, continuity, capacitance test, DC/AC voltage/current, or °C/°F test. Press and hold to disable auto power off.
  2. HOLD key: Locks the displayed value. Press again to exit.
  3. RANGE key: Enters manual range mode. Press for more than 2 seconds to return to auto mode.
  4. REL▲ key: Enters relative measuring mode, displaying 'REL' and the present reading as reference. Press again to exit.
  5. Hz/Duty key: On ACV/ACA or Hz range, selects Frequency or Duty Cycle measurement.
  6. MAX/MIN key: Locks MAX, MIN, or MAX-MIN values. Press for over 2 seconds to exit.
  7. Rotary Switch: Selects functions and ranges.
  8. VQHz Input Jack: For Voltage, Frequency measurements.
  9. 20A Input Jack: For 20A current measurements.
  10. µAmA T+ Input Jack: For microampere/milliampere current and Temperature measurements.
  11. COM T- Input Jack: Common terminal and Temperature negative input.

Additional features include:

  • NCV Detection Area: For Non-Contact AC Voltage detection.
  • CDS Sensor: Automatically controls LCD backlight based on ambient brightness.
  • NCV Indicator Lights: Red and green lights indicate NCV detection.
  • LCD Display: 6000 counts digit, with full function symbol display.
  • Auto LCD Backlight: Automatically adjusts for visibility.
  • Auto Power Off: The meter switches to standby mode after 15 minutes of inactivity. Press any key to exit.
  • Durable Rubber Holster: Prevents electric shock and protects against drops.
HP-770HC Digital Multimeter with Durable Rubber Holster
The HP-770HC multimeter features a durable rubber holster for protection.

2.2 HP-605A AC/DC Clamp Adapter

HoldPeak HP-605A AC/DC Clamp Adapter
The HoldPeak HP-605A AC/DC Clamp Adapter.

The HP-605A is a clamp-on probe designed for measuring AC/DC currents up to 600 amps. It features a 1mV per ampere output and standard 3/4 inch banana plugs, making it compatible with most multimeters that have 1mV sensitivity. Its maximum jaw caliber size is 1.77 inches (45mm), allowing for accurate current measurements without breaking the circuit.

3. Specifications

3.1 HP-770HC Digital Multimeter Specifications

Measurement Range Resolution Accuracy
Max LCD Display 6000 Counts Digit
DC Voltage 60mV/600mV/6V/60V/600V/1000V 0.01mV to 1V ±(0.5% to 0.8% of rdg + digits)
AC Voltage (True RMS) 60mV/600mV/6V/60V/600V/750V 0.01mV to 1V ±(1.0% to 1.5% of rdg + digits)
DC Current 600µA/6000µA/60mA/600mA/6A/20A 0.1µA to 10mA ±(1.2% to 2.0% of rdg + digits)
AC Current (True RMS) 600µA/6000µA/60mA/600mA/6A/20A 0.1µA to 10mA ±(1.5% to 2.5% of rdg + digits)
Resistance 600Ω/6KΩ/60KΩ/600KΩ/6MΩ/60MΩ 0.1Ω to 10KΩ ±(1.0% to 1.5% of rdg + digits)
Capacitance 9.999nF ~ 99.99mF 1pF to 10µF ±(2.5% to 10.0% of rdg + digits)
Frequency 9.999Hz ~ 9.999MHz 0.001Hz to 1kHz ±(0.1% of rdg + 5 digits)
Temperature -20°C ~ 1000°C / -4°F ~ 1832°F 1°C / 1°F ±(3°C/°F + digits)
Duty Cycle 0.1% ~ 99.9% ±(2.0% of rdg + 2 digits)
Transistor hFE Test Range: 0 ~ 1000; Ib=10µA, Vce=1.8V Approx.
Weight Approx. 350g (including battery)
Dimension 195 x 88 x 40mm

For detailed specifications, please refer to the full specification table in the user manual PDF.

Detailed specifications table for HoldPeak HP-770HC
Detailed specifications for the HP-770HC multimeter, including ranges and accuracies for various measurements.

3.2 HP-605A AC/DC Clamp Adapter Specifications

Parameter Value
DC Current 600 A
AC Current 600 A
Jaw Caliber 1.77 inches (45mm)
Output Conversion 1mV/1A
Conversion Accuracy ±3.0% + 5 digits
Output Impedance 10KΩ
Dimension 7.05 x 3.54 x 1.42 inches

4. Operating Instructions

4.1 General Operation Notes

  • Before operation, check the battery. If the battery voltage drops below the proper operating range, the low battery symbol will appear on the LCD display, indicating the battery needs to be changed.
  • Pay attention to the input jack labels (e.g., 'A') which indicate the maximum input voltage or current.
  • Always position the range switch to the desired range before measurement.

4.2 Measuring DC & AC Voltage

  1. Set the rotary switch to the desired 'mV~' or 'V' range position. For AC voltage, press the 'SELECT' button.
  2. Connect the black test lead to the 'COM T-' jack and the red test lead to the 'VQHz' jack.
  3. Connect the test leads across the source or load under measurement.
  4. Read the measurement from the LCD. The polarity of the red lead connection will be indicated for DC voltage.

Note: Do not input voltage greater than 1000V DC or 750V AC, as this may damage the circuit or cause shock. Be cautious when measuring high voltage.

4.3 Measuring DC & AC Current (Multimeter Only)

  1. Set the rotary switch to the desired 'µA~', 'mA~', or 'A' range position. For AC current, press the 'SELECT' button.
  2. Connect the black test lead to the 'COM T-' jack. Connect the red test lead to the 'µAmA T+' jack for currents up to 600mA, or to the '20A' jack for currents up to 20A.
  3. Connect the test leads in series with the load under measurement.
  4. Read the measurement from the LCD. The polarity of the red lead connection will be indicated for DC current.

Note: If the measured value is unknown, start with the highest range. If 'OL' is displayed, it indicates over-range; select a higher range. The maximum current for the µAmA T+ socket is 800mA, and for the 20A socket is 20A. For the 20A range, limit measuring time to less than 10 seconds to maintain precision.

4.4 Measuring DC & AC 600A Current (with HP-605A Clamp Adapter)

HP-605A DC and AC 600A Current Measurement Instructions
Step-by-step instructions for measuring DC and AC 600A current using the HP-605A clamp adapter.
  1. Set the multimeter's selector switch to the desired '600A~' range position. For AC current, press the 'SELECT' button.
  2. Connect the black banana plug of the HP-605A AC/DC Current Clamp Adapter to the 'COM T-' socket and the red banana plug to the 'µAmA T+' socket on the multimeter.
  3. Set the HP-605A AC/DC Current Clamp Adapter to '1mV/A' range.
  4. When performing DC current measurement, always rotate or push the DCA zero adjuster on the Clamp Adapter until the multimeter reads zero.
  5. Clamp the jaws around the single conductor to be measured. Center the conductor within the jaw using the centering marks as guides.
  6. Read the result from the multimeter's LCD panel. The arrow on the jaw indicates the DC current direction of positive current flow (positive to negative).
Multimeter and clamp adapter measuring current in an electrical panel
Demonstration of the HP-770HC multimeter and HP-605A clamp adapter measuring current in an electrical panel.

4.5 Measuring Resistance

  1. Set the rotary switch to the desired 'Ω→(-00)→H' range position.
  2. Connect the black test lead to the 'COM T-' jack and the red test lead to the 'VQHz' jack.
  3. Connect test leads across the resistance under measurement.
  4. Read the measurement from the LCD.

Note: Max input overload is 500V rms for less than 10 seconds. For resistances above 1MΩ, the meter may take a few seconds to stabilize. If 'OL' is displayed, it indicates an open circuit or over-range. When checking in-circuit resistance, ensure the circuit is de-energized and all capacitors are discharged.

4.6 Measuring Capacitance

  1. Set the rotary switch to the 'Ω→(-00)→H' range position, then press 'SELECT' to choose capacitance measurement.
  2. Connect the black test lead to the 'COM T-' jack and the red test lead to the 'VQHz' jack.
  3. Connect test leads across the capacitance under measurement.
  4. Read the measurement from the LCD.

Note: Capacitors should be discharged before testing. Large capacitances (600µF-60mF range) may take longer to stabilize (about 10 seconds). For small capacitances (≤1µF), press 'REL▲' before measuring for better accuracy.

4.7 Measuring Frequency & Duty Cycle

  1. Set the rotary switch to the desired 'Hz' range position.
  2. Connect the black test lead to the 'COM T-' jack and the red test lead to the 'VQHz' jack.
  3. Press the 'Hz/Duty' key to select Frequency or Duty Cycle test.
  4. Connect the probe across the source or load under measurement.
  5. Read the measurement from the LCD.

4.8 Measuring Temperature

  1. Set the rotary switch to the desired '°C/°F' range position, then press 'SELECT' to choose °C or °F measurement.
  2. Connect the black banana plug of the sensor to the 'COM T-' jack and the red banana plug to the 'µAmA T+' jack.
  3. Place the sensor probe into the temperature field under measurement.
  4. Read the measurement from the LCD.

Note: The included K-type thermocouple has a limit of 250°C (300°C for short periods). For higher temperatures, use a special probe. Do not change the thermocouple arbitrarily, and do not input voltage in temperature function.

4.9 Diode & Audible Continuity Testing

  1. Set the rotary switch to the 'Ω→(-00)→H' range position, then press 'SELECT' to choose Diode or Audible continuity measurement.
  2. Connect the black test lead to the 'COM T-' jack and the red test lead to the 'VQHz' jack.
  3. For diode testing, connect the test leads across the diode. The display will show the approximate forward voltage.
  4. For audible continuity, connect the test leads to two points of the circuit. If resistance is less than approximately 500Ω, the buzzer will sound.

Note: Ensure power is cut off and all capacitors are discharged before performing these measurements.

4.10 Transistor hFE Test

  1. Set the rotary switch to the 'hFE' range position.
  2. Determine if the transistor is NPN or PNP and identify the Emitter, Base, and Collector leads.
  3. Insert the leads into the proper holes in the socket on the front panel.
  4. Read the approximate hFE value from the LCD.

Note: Do not connect an external voltage to the measuring terminals.

4.11 Non-Contact AC Voltage (NCV) Detection

HP-770HC NCV Function Demonstration
Demonstration of the NCV function, showing detection of live and null wires.
  1. Set the rotary switch to the 'NCV' range position. The NCV green LED light will illuminate.
  2. Hold the meter so its top is vertically and horizontally centered and contacting the conductor. When live voltage (≥90V AC rms) is detected, the NCV red and green LED lights will alternately illuminate with an audible sound.

Note: Even without LED indication, voltage may still be present. Do not rely solely on the non-contact voltage detector to determine voltage presence. Detection can be affected by socket design, insulation thickness, and other factors. Keep the meter away from electrical noise sources (fluorescent lights, motors, etc.) during tests, as these can trigger the NCV function and invalidate the test.

5. Maintenance

5.1 Battery Replacement

  1. When the low battery symbol appears on the LCD, the battery needs to be changed.
  2. Before replacing the battery, set the selector switch to 'OFF' and remove the test leads from the terminals.
  3. Open the battery cabinet cover using a screwdriver.
  4. Replace the old battery with two new 1.5V 'AA' R6P batteries of the same type.
  5. Close the battery cabinet cover and fasten the screw securely.

5.2 Fuse Replacement

The HP-770HC multimeter is equipped with an 800mA/250V fast fuse for temperature and current circuits up to 600mA, and a 20A/250V fuse for the 20A range.

  1. Ensure the meter is disconnected from all external circuits, set the selector switch to 'OFF', and remove the test leads.
  2. Open the battery cabinet cover using a screwdriver.
  3. Replace the old fuse with a new one of the same type and rating: 5x20mm 800mA/250V fast fuse or 5x20mm 20A/250V fast fuse.
  4. Close the battery cabinet cover and fasten the screw securely.

5.3 General Care

  • Replace test leads if they are exposed or damaged, using only leads with the same specifications.
  • Clean the meter with a moist cloth or a small amount of detergent. Do not use chemical solutions.
  • Ensure the back cover is properly closed and screwed before using the meter.
  • If any abnormality occurs, stop operation immediately and send the meter for maintenance.
  • Remove the battery if the meter will not be used for a long time.

6. User Tips

  • The HP-605A clamp adapter is highly recommended for its practicality and usefulness when paired with a multimeter.
  • The HP-605A clamp adapter reliably measures up to 600A, and potentially slightly more.
  • The HP-605A is designed to work seamlessly with various HoldPeak multimeters, including models like HP-770M, HP-770N, HP-770K, HP-90EPD, HP-90EPC, HP-6688G, and HP-6688H.
  • When performing DC current measurements with the clamp adapter, always press the 'REL' key on the multimeter to zero it out first.
  • For AC current measurements with the clamp adapter, always clamp around the live wire only, or use an AC line splitter.
  • Ensure the black banana plug of the HP-605A is connected to the 'COM' port and the red banana plug to the 'µA mA' port on the multimeter.
  • Press the 'SELECT' button on the multimeter and the AC/DC button on the Clamp Adapter to choose 'DC' or 'AC' measurement.
  • The arrow on the right side of the clamp adapter's jaw indicates the DC current direction (from positive to negative).

7. Troubleshooting

  • 'OL' Displayed: If the meter displays 'OL' (Overload), it indicates an over-range condition. Select a higher measurement range.
  • Low Battery Indicator: If the low battery symbol appears, replace the batteries promptly to ensure accurate readings and proper operation.
  • Inaccurate Readings: Ensure test leads are fully inserted and not damaged. Check that the correct measurement function and range are selected. For resistance measurements, confirm the circuit is de-energized and capacitors are discharged.
  • No Display/Power: Check battery installation and charge. If the issue persists after battery replacement, the fuse might need checking or replacement.
  • NCV Function Not Responding: Ensure the meter is held correctly, and there are no strong electrical noise sources nearby that could interfere with detection.

8. Warranty and Support

All HoldPeak products come with a 1-year warranty from the store of purchase.

For further assistance or technical support, please refer to the official user manual PDF or contact the seller directly.

User Manual (PDF): A comprehensive user manual in PDF format is available for download. Click here to view the PDF.