1. Introduction
The ASAIR AMS3 series gas mass flow meter is a high-performance, stable, and reliable device based on the thermal measurement principle. It incorporates a MEMS flow sensor chip developed by Ausone to accurately measure the mass flow rate of gases. This product offers excellent accuracy, stability, linearity, and fast response, with resistance to high and low temperatures.
Equipped with an internal alarm function and a TFT display, the AMS3 series provides clear status indication. It supports a wide range of outputs including RS485, NPN open collector, 1-5V linear voltage, and 4-20mA linear current, catering to diverse user requirements. The series includes various models to cover a broad spectrum of flow ranges and applications.
1.1 Key Features
- Integrated Mass Flow and Temperature Measurement
- Good repeatability
- Supports multiple gas measurements
- Standard Modbus RTU communication
- 2 NPN open collector outputs (for upper and lower limit alarms)
- TFT screen display with adjustable display direction (up, down, left, right)
- 9-28V DC power supply
1.2 Scope of Application
The AMS3 series is suitable for mass flow monitoring of air, nitrogen, and other dry, clean, non-corrosive gases (excluding flammable and explosive gases). Its applications span university research, fire fighting, environmental monitoring, tobacco industry, intelligent agriculture, food processing, medicine, and other industrial sectors.
Figure 1: ASAIR AMS3 Series Gas Mass Flow Meter with key features and specifications.
2. Safety Warnings
WARNING: Personal Injury Hazard
Do not use this product in safety protection devices or emergency stop equipment, or in any other applications where product failure could lead to personal injury, unless specifically authorized and used for its intended purpose. Always refer to the product data sheet and instructions before installing, handling, using, or maintaining the product. Failure to follow these instructions may result in death or serious personal injury. The manufacturer is not liable for personal injury or death resulting from misuse and disclaims all related claims.
3. Setup and Installation
3.1 Appearance Structure and Interface
The AMS3 series features a sensor duct and a working interface. The working interface includes an LCD display and operation keys (DISP, MODE, Up, Down). The LCD shows the Modbus communication address, cumulative flow rate, and instantaneous flow rate. The sensor duct, located below the working interface, contains the air inlet and outlet. For a detailed schematic diagram, please refer to the full user manual (PDF).
3.2 DB9 Male Connector Definition
The AMS3 series uses a standard DB9 male connector for its interface. The pin definitions are as follows:
| Pin | Definition |
|---|---|
| 1 | External Output 1 |
| 2 | Analog Output |
| 3 | RS485B |
| 4 | RS485A |
| 5 | Power cable |
| 6 | GND |
| 7 | GND |
| 8 | GND |
| 9 | External Output 2 |
Figure 2: DB9 connector on the AMS3 Series Gas Mass Flow Meter.
3.3 AMS3 Series Lead Wire Definition
The AMS3 series lead wire is a DB9 female lead wire, approximately 430mm in length, made of UL2464 material. It consists of 10 wires with the following definitions:
| Lead Wire Color | Definition |
|---|---|
| Orange | External Output 1 |
| Brown | Analog Output |
| Green | RS485B |
| Yellow | RS485A |
| Red | Power cable |
| Black | GND |
| Blue | GND |
| Purple | GND |
| Gray | External Output 2 |
| Transparent Heat Shrink Sleeve | Shielded wire |
3.4 Main Dimensions and Piping Method
The AMS3 series flow meters come in various models with specific dimensions. For detailed dimensional drawings of models like AMS3001H00, AMS3002H00, AMS3003H00, and AMS3006H00, please refer to the comprehensive PDF user manual.
When installing the piping, observe the following guidelines:
- Inlet Connection: Ensure a straight pipe section connected to the AMS3 series air inlet connector that is at least 10 times the length of the inlet connector's inner diameter. Failure to do so may shift accuracy by 2% F.S. or more.
- Outlet Connection: The air tube connected to the air outlet port should have a straight section equal to or greater than 5 times the inner diameter of the air outlet port. Otherwise, accuracy may be shifted by 2% F.S. or more.
- Sensor Duct Mounting: Sensor ducts are recommended to be mounted parallel or perpendicular to the horizontal. If mounted horizontally, the sensor interface must be above the sensor duct and parallel to the horizontal. For other mounting angles, consult customer service.
4. Operation
4.1 Main Screen Interface and Menu Settings
The AMS3 series display shows various icons and information:
- Lock screen icon: Indicates the screen is locked.
- 485 communication identification: Shows RS485 communication status.
- Traffic alarm status: Indicates if an alarm is active.
- Instantaneous flow rate: Current flow measurement.
- Gas Type: Displays the type of gas being measured (e.g., AIR).
- Cumulative flow: Total accumulated flow.
- 485 communication address: The device's Modbus address.
If there is no operation for 30 seconds, the system automatically enters a lock screen state. To unlock and access settings, long-press the MODE key for 5 seconds. Use the Up and Down keys to navigate and select parameters.
The menu system is structured in four levels, allowing configuration of various parameters including fluid selection (Air/N2), alarm switch (on/off), upper and lower flow limits, and accumulated traffic reset. For a complete breakdown of menu functions, refer to the 'Function Introduction' table in the full PDF manual.
4.2 NPN Collector Open Circuit Output
The AMS3 series features two NPN open collector output ports: External Output 1 for the upper limit alarm and External Output 2 for the lower limit alarm. If using the AMS3 series lead wires, NPN1 corresponds to the orange wire, NPN2 to the gray wire, and GND to the black, blue, or purple wires. The load connected to these outputs (e.g., a warning light) should have a resistance of 100Ω or more.
By default, the alarm function is off. To activate, set the upper and lower flow limits. When the instantaneous flow rate exceeds the upper limit or falls below the lower limit, the corresponding NPN output port will close, allowing current to flow through the load and trigger the alarm device.
For example, if the upper limit is 20 L/min and the lower limit is 5 L/min: if the flow rate drops below 5 L/min, External Output 2 closes; if it exceeds 20 L/min, External Output 1 closes.
4.3 Voltage Output
The relationship between the output voltage and instantaneous flow rate is given by the formula:
Instantaneous flow rate = [(Instantaneous output voltage - zero flow voltage) / (Full range voltage - zero flow voltage)] * Full scale flow rate
For instance, for a 2000 L full-scale gas mass flowmeter with a 3V output voltage, 1V zero flow voltage, and 5V full-scale voltage, the instantaneous flow rate would be calculated as: [(3-1) / (5-1)] * 2000 = 1000 L/min.
4.4 RS485 Communication
The AMS3 series gas mass flowmeter uses RS485 communication with a standard Modbus RTU protocol. The device acts as a slave, with a default address of 0x01 (modifiable). Communication parameters include baud rates of 4800, 9600, 19200, 115200 bps, 8-bit data, 1 stop bit, and no parity checking. For detailed Modbus RTU communication register definitions and examples, refer to the full PDF user manual.
5. Maintenance
5.1 Zero Point Calibration
The AMS3 series supports zero point calibration for both air and nitrogen gases. This function can be accessed through the system settings menu on the device. Regular calibration helps maintain measurement accuracy.
5.2 Cleaning
If traffic records decrease or the sensor is suspected to be contaminated, check for foreign objects blocking the inlet/outlet ports or the sensor itself. Clean or replace the sensor as necessary to ensure optimal performance.
6. Troubleshooting
This section provides guidance for common issues encountered with the AMS3 series flow meter.
| Fault Phenomenon | Reason for Malfunction | Troubleshooting/Handling Measures |
|---|---|---|
| No display on the main interface |
|
|
| Instantaneous flow display not zero before measurement |
|
|
| No traffic records |
|
|
| Decreased traffic records |
|
|
| Communication is abnormal |
|
|
7. Specifications
7.1 Model and Range Comparison
| Model | Range | Accuracy | Output Methods | Mechanical Interface |
|---|---|---|---|---|
| AMS3001H00 | 10~1000 L/min | ±3%F.S. (100~1000 L/min) | RS485, NPN open collector, 1~5V linear voltage | 2-RC 1/2 |
| AMS3002H00 | 20~2000 L/min | ±3%F.S. (200~2000 L/min) | RS485, NPN open collector, 1~5V linear voltage | 2-RC 3/4 |
| AMS3003H00 | 30~3000 L/min | ±3%F.S. (300~3000 L/min) | RS485, NPN open collector, 1~5V linear voltage | 2-RC1 |
| AMS3006H00 | 60~6000 L/min | ±3%F.S. (600~6000 L/min) | RS485, NPN open collector, 1~5V linear voltage | 2-RC1 1/2 |
7.2 General Technical Specifications
| Technical Indicator | Description |
|---|---|
| Repeatability | ±1%F.S. |
| Response time | 1s |
| Power supply method | External power supply 9 to 28V DC |
| Display method | TFT display |
| Working Pressure Range | 0.1~1.5MPa |
| Maximum Withstanding Voltage | 2.25MPa |
| Standard calibration gases | Air (25°C, 1 standard atmosphere) |
| Operating temperature | 0~+50°C |
| Power wastage | <0.85W |
| Housing Material | PC |
| Duct Material | 6061 Aluminum Alloy |
| Installation | Horizontal mounting is optimal, Zero calibration is recommended for vertical mounting |
7.3 Product Properties
- High-concerned chemical: None
- Theory: MEMS
- Material: Metal
- Type: Gas Sensor
- Output: Digital Sensor (RS485/NPN/1~5V)
- Brand Name: ASAIR
- Origin: Mainland China
- Use: Gas Mass Flow Meters
- Accuracy: ±3%F.S.
- Repeatability: ±1%F.S.
- Power supply: 9~28VDC
- Maximum working pressure: 2.25MPa
- Operating temperature: -10~+60°C
- Original manufacturer: Guangzhou Aosong
- L/T (Lead Time): 15 Working days
- MOQ (Minimum Order Quantity): 1
- Measuring Medium: Air/N2
8. User Tips
- Calibration Certificate: For applications requiring certified accuracy, inquire about the availability of calibration certificates (e.g., CNAS) from the seller or manufacturer.
- Optimal Mounting: While horizontal mounting is optimal, if vertical mounting is necessary, zero calibration is highly recommended to ensure accurate readings.
- Preventing Blockages: Regularly check the inlet and outlet ports and the sensor for any foreign objects or contamination to prevent decreased traffic records and ensure consistent performance.
9. Warranty and Support
9.1 Quality Assurance
Guangzhou Aosong Electronics Co., Ltd. provides quality assurance to direct buyers of its products, calculated from the date of shipment. If the product is proven to be defective during the warranty period, the company will provide free repair or replacement services, as detailed in the Aosong Electronics product manual with technical specifications.
The company is responsible for products with defects only when applied in situations that meet the technical requirements. It does not guarantee application in non-recommended special scenarios or make commitments regarding reliability when used with non-company supporting products or circuits.
9.2 Warranty Period
| Product Category | Defects Liability Period |
|---|---|
| AMS3 series gas mass flowmeter | 12 months |
| Wire or accessories | 6 months |
This manual is subject to change without prior notice. The final interpretation rights belong to Guangzhou Aosong Electronics Co., Ltd. Copyright ©2024, ASAIR®.
9.3 Further Information
For more detailed diagrams, communication protocols, and comprehensive instructions, please refer to the official PDF user manual available at: AMS3 Series Instruction Manual (PDF).