meterk MESR-100 / M6013

Contents

MESR-100/M6013 ESR Capacitance Ohm Meter

User Instruction Manual

1. Introduction

The Meterk MESR-100/M6013 is an auto-ranging capacitor ESR (Equivalent Series Resistance) and Low Ohm Meter designed for professional capacitance and resistance testing. It supports in-circuit testing, making it a versatile tool for electronics repair and diagnostics. This manual provides detailed instructions for the safe and effective use of your device.

Key Features:

  • Auto-ranging ESR and Low Ohm Meter.
  • Measuring range from 0.001 to 100.0 Ohms.
  • Supports in-circuit testing.
  • Utilizes a true 100 KHz sine wave for accurate ESR measurement.
  • High resolution up to 0.001 Ohm.
  • 128x64 dot matrix LCD for clear display.
  • Embedded 25V capacitor table for quick reference.
  • Ergonomic design with a built-in stand.
  • Powered by 2x AA 1.5V batteries or external 5V micro-USB.
  • Dual terminal for fast and easy inspection.

What is ESR?

ESR refers to the Equivalent Series Resistance within a capacitor. By using a 100kHz test frequency, the capacitive reactance (impedance) becomes negligible, allowing the meter to measure the true series resistance. A high ESR value typically indicates a faulty or degraded electrolytic capacitor, which can lead to circuit malfunctions such as excessive ripple and poor filtering.

2. Setup and Preparation

2.1 Package Contents:

  • MESR-100/M6013 ESR Capacitance Ohm Meter
  • Test Leads with Alligator Clips
  • User Manual (this document)

2.2 Battery Installation:

The device requires two (2) AA 1.5V batteries (not included). Open the battery compartment cover located on the back of the device. Insert the batteries, ensuring correct polarity (+/-), then close the cover securely.

2.3 External Power:

The meter can also be powered via a standard 5V micro-USB cable (not included). Connect the micro-USB cable to the port on the side of the device and to a suitable USB power source.

2.4 Connecting Test Leads:

Insert the banana plugs of the test leads into the corresponding red (+) and black (-) jacks on the front of the meter.

MESR-100 with test leads connected
Figure 1: MESR-100 with test leads connected.

2.5 Important Safety Precaution: Discharge Capacitors!

ALWAYS DISCHARGE THE CAPACITOR BEFORE TESTING.

Failure to discharge capacitors can result in damage to the meter and potential injury. To discharge a capacitor, you can:

  • Use a screwdriver to short circuit the capacitor terminals (for small capacitors).
  • For larger capacitors, use a 10 Ohm resistor in series with a screwdriver to short circuit the terminals for 5 to 10 seconds.

Although the meter has internal diodes to protect against high voltage, a large surge current and voltage from an undischarged capacitor can still damage the device.

3. Operating Instructions

3.1 Power On/Off:

  • To Power On: Press and hold the orange ON/OFF button (circular button) for 2 to 3 seconds.
  • To Power Off: Press and release the ON/OFF button.

3.2 Auto/Manual Mode:

  • Auto Range: Press and release the