1. Introduction
The ZOYI ZT-702S is a handheld oscilloscope multimeter that combines the functionalities of a digital oscilloscope and a multimeter into a single, compact device. It features double-injection molding technology for durability, a 2.8-inch IPS full-view color screen for clear readings, and a 9999-count multimeter for precise measurements. Designed for ease of use, its intuitive function button menu interface makes it suitable for various measurement scenarios.
1.1. Safety Instructions
To prevent electric shock, fire, or personal injury, always adhere to the following safety precautions:
- Read this user manual thoroughly before operation.
- Use the product only for its intended purpose. Misuse may impair the protection provided.
- Inspect the device for cracks or plastic defects before each use. Ensure insulators near input ports are intact.
- Always use the correct input port and gear setting as specified in this manual.
- Do not use the product in environments with explosive gases, vapors, or high humidity.
- Keep your fingers behind the guard of the test lead probe during measurements.
- When connected to a circuit, avoid touching unused input ports.
- Disconnect test leads and the circuit before changing measurement modes.
- Be cautious when measuring DC voltages above 36V or AC voltages above 25V, as these can cause serious injury.
- Select the appropriate test gear and range to prevent instrument damage or personal injury.
- Do not operate the device with the front or back cover open.
- Charge the device promptly when the battery voltage is low to maintain measurement accuracy.
- The probe's ground wire is connected to GND. Do not connect it to a high-voltage power supply, especially when charging via USB, to avoid damage or injury.
- When using an oscilloscope probe to measure voltages higher than AC 25V or DC 36V, ensure the USB protective cover is tightly closed to prevent contact with exposed metal parts.
2. Setup and Initial Operation
2.1. Powering On/Off
Press and hold the Power button (red button with power symbol) for 2 seconds to switch the device on or off.
2.2. Input/Output Ports
Familiarize yourself with the input and output ports:
- 10A Port: Current input for measurements up to 10A.
- mA Port: Current input for measurements up to 200mA.
- COM Port: Common (return) port for all measurements.
- VΩHz-|| Port: Input for voltage, resistance, capacitance, diode, continuity, frequency, and temperature measurements.
- CH (BNC) Port: Oscilloscope probe input terminal. Maximum input voltage: 1X < 150V, 10X < 300V (DC+AC peak value).
- Square Wave Signal Output: Metal port for 3V/1KHz constant output. Avoid touching during measurement.
- USB Port: Type-C port for communication with a computer and battery charging.
2.3. Probe Connection and Compensation
When using the oscilloscope probe:
- Safety: Always keep your fingers behind the safety ring on the probe body. Do not touch the metal tip when the probe is connected to a high-voltage power supply. Ensure the measured voltage does not exceed the probe's specifications.
- Manual Probe Offset (Compensation): Uncompensated or improperly compensated probes can lead to measurement errors. To adjust:
- Power on the device and connect the probe to the CH input terminal.
- Input a 4V/1KHz square wave signal.
- Press the AUTO button on the panel to check the waveform display status.
- Adjust the capacitance on the probe using the provided adjustment rod until the waveform displays correctly (neither over-compensated nor under-compensated).
- Probe Attenuation Setting: The probe has an X1/X10 attenuation switch. Ensure this setting matches the probe attenuation option in the oscilloscope's system settings.
- If the probe switch is set to X1, set the oscilloscope to X1.
- If the probe switch is set to X10, set the oscilloscope to X10.
- Note: Using X1 with a 6M2/X1 probe may limit bandwidth to 6MHz. For full bandwidth, use X10 or a higher specification probe.
3. Operating Instructions
3.1. Panel Keys Overview
The device features several function keys:
- F1, F2, F3, F4: Function keys corresponding to on-screen menus.
- Power (REL): Press and hold to power on/off. In multimeter mode, short press to enter relative value (REL) measurement.
- AUTO/RANGE: In oscilloscope mode, short press for automatic waveform acquisition. In multimeter mode, short press to switch to manual range.
- HOLD/SAVE: In oscilloscope mode, short press for STOP/RUN function; long press to save waveform data. In multimeter mode, short press for data hold/cancel.
- MODE: Switches between oscilloscope and multimeter modes.
- Direction Keys (Up, Down, Left, Right): Adjust setting parameters, move cursor, and navigate menus.
- MENU: Opens the system function expansion menu (3 pages).
3.2. Oscilloscope Mode
The oscilloscope mode allows for waveform visualization and analysis.

Figure 2: ZOYI ZT-702S displaying Oscilloscope and Multimeter modes.
3.2.1. LCD Display Elements (Oscilloscope Mode)
The display shows various parameters:
- Running Status: RUN (auto acquisition), WAIT (normal trigger, waiting), T'D (captured data), STOP (locked waveform).
- Time Base: Current time base position.
- Time Base Scale: Horizontal time base scale value (e.g., 250us).
- Voltage Scale: Vertical voltage scale value (e.g., 1.0V).
- Trigger Level: Current trigger voltage value.
- Battery Power Status: Battery level and charging indicator.
- Horizontal Cursor: Horizontal time base position of the trigger.
- Waveform: Display of the collected signal.
- FreQ: Frequency of the collected signal.
- Vertical Cursor: Triggered vertical voltage position.
- VPP: Peak-to-peak voltage of the collected signal.
- Vmax: Maximum voltage of the collected signal.
3.2.2. Menu Navigation (Oscilloscope Mode)
Press MENU to access extended settings:
- Coupling: AC/DC.
- TR-MODE (Trigger Mode): Auto, Normal, Single.
- TRIG (Trigger Edge): Rising, Falling.
- PROBE: X1, X10.
- LANGUAGE: Simplified Chinese, English.
- AUTO OFF: Cancel, 15Min, 30Min, 60Min, 120Min.
- BK LIGHT (Backlight): 30%, 50%, 80%, 100%.
- MORE MEAS (More Measurements): ON/OFF, Vmin, RMS, Prd.
- CALIBRATE: ON/OFF (for baseline calibration).
- RESET: ON/OFF (to factory defaults).
- STORAGE: ENTER (to view/acquire saved waveforms).
- VERSION: Displays firmware version.
3.2.3. Auto Settings
Press the AUTO button to automatically identify waveform type (sine/square) and adjust control mode for optimal display.
3.2.4. Vertical System
Controls waveform voltage amplitude, scale, and position:
- Voltage/Scale Setting: Press F1 (VOL/TIME menu). Use Up/Down keys to increase/decrease voltage setting. Range: X1 (20mV/div - 10V/div), X10 (200mV/div - 100V/div).
- Vertical Position: Press F2 (MOVE menu). Use Up/Down keys to move waveform position up/down.
3.2.5. Horizontal System
Controls time base and horizontal position:
- Horizontal Scale (Time Base): Press F1 (VOL/TIME menu). Use Left/Right keys to change horizontal scale (zoom in/out).
- Horizontal Position: Select F2 (MOVE menu). Use Left/Right arrow keys to move waveform left/right. Press and hold MENU to return horizontal cursor to center.
- Rolling Mode: Automatically activated when horizontal time base is slower than 100ms/div, suitable for low-speed signals.
3.2.6. Triggering System
Sets conditions for waveform acquisition:
- Trigger Cursor Setting: Press F3 (TRIGGER menu). Use Up/Down keys to adjust trigger cursor position.
- Trigger Mode: Press MENU, then F2. Options:
- Auto: Real-time waveform collection and refresh.
- Normal: Locks and displays waveform when signal amplitude reaches trigger level. Continues collecting.
- Single: Locks and displays waveform, then stops acquisition (STOP state). Press HOLD to re-trigger.
- Trigger Edge: Press MENU, then F3. Select Rising or Falling edge.
3.2.7. Measurement
- Automatic Measurement: Press AUTO to automatically identify and display waveforms.
- Manual Measurement: Manually set parameters (voltage, time base, cursor, trigger, coupling, probe attenuation) and connect probe.
- Measured Values: FreQ, VPP, Vmax are always displayed. Vmin, RMS, Prd can be toggled ON/OFF via MENU -> Right Direction Key -> F4.
3.2.8. Saving and Viewing Waveforms
- Save: Press and hold HOLD/SAVE for 2 seconds. Waveform data is saved as a picture in flash memory.
- View/Acquire: Press MENU, navigate to the third page using Right Direction Key, then press F3 to save data to Flash. Connect Type-C cable to PC, access USB disk, and open the 'pic' folder.
3.3. Multimeter Mode
The multimeter mode provides various electrical measurement functions.
3.3.1. LCD Display Elements (Multimeter Mode)
The display shows:
- HOLD: Indicates data hold is active.
- Battery Power Status: Battery level and charging indicator.
- Analog Pointer: Changes with main display data, indicates scale position.
- Measurement Symbol: AC, DC, resistance, capacitance, diode, buzzer symbols.
- Negative: Negative sign prompt.
- Main Display: Measured value (max 9999 counts).
- Unit Symbol: Unit of measured data.
- Measurement Mode: AUTO (automatic range) or MANU (manual range).
- Max/Min: Maximum/minimum readings.
- AVG/Hz/Temp: Average value (DC voltage, resistance, capacitance), AC frequency, Fahrenheit temperature.
3.3.2. Measurement Functions
- AC/DC Voltage:
- Insert black lead into COM, red into VΩHz-||.
- For < 99.99mV, press F4 (mV range) twice for AC mV.
- For > 99.99mV, press F1 (DC voltage) twice for AC voltage.
- Contact test points and read. Do not exceed max rated voltage.
- AC/DC Current:
- Insert black lead into COM, red into 10A or mA port (select based on expected current).
- Press F3 for current mode. F4 corresponds to mA.
- Press corresponding menu key again to switch between DC and AC.
- Disconnect circuit, connect leads in series, power on, read. Do not exceed max rated current.
- Resistance:
- Insert black lead into COM, red into VΩHz-||.
- Press F2 to enter resistance gear.
- Touch test points and read. Ensure circuit power is off and capacitors are discharged. Do not input voltage.
- Continuity:
- Insert black lead into COM, red into VΩHz-||.
- In resistance mode, press F2 to enter continuity mode.
- Connect probes. A buzzer indicates a short circuit.
- Diode:
- Press F2 in on-off gear to enter diode gear.
- Connect red lead to diode positive, black to negative. Read forward voltage. 'OL' indicates reverse polarity or damaged diode. Do not input voltage. Disconnect power and discharge capacitors.
- Capacitance:
- Insert black lead into COM, red into VΩHz-||.
- Press F2 in diode gear to enter electronic gear.
- Connect red lead to capacitor positive, black to negative. Read after stable.
- Temperature:
- Connect black lead to COM, red to VΩHz-||.
- Push F4 three times to enter Temperature Mode. Primary display shows Celsius, secondary shows Fahrenheit.
- Touch probes to desired test points and read. Do not input voltage.
4. Maintenance
4.1. General Maintenance
Only qualified personnel should attempt to repair or service the product beyond replacing batteries and fuses.
4.2. Cleaning the Product
Wipe the product with a damp cloth and mild detergent. Do not use abrasive cleaners or solvents. Ensure input signals are removed before cleaning. Dirt or moisture in terminals can affect readings.
4.3. Charging the Battery
When the battery symbol appears in the upper right corner of the screen, charge the device:
- Connect the Type-C data cable to a DC 5V output adapter or a computer's USB port.
- During charging, the screen will display a charging symbol.
- When fully charged, the screen will display a full battery symbol.
- The red light on the power button will be on during charging. It will flash if no battery is installed.
4.4. Battery Storage
If the instrument is not used for more than 6 months, charge the battery to 50%-70% and store it in a cool, dry environment. If the lithium battery shows signs of rust, leakage, or bulging, remove and discard it immediately.
4.5. Replacing the Battery
The device uses a rechargeable 18650 lithium battery. If standby time significantly shortens, replace it with the same type of 18650 battery. Ensure correct positive and negative polarity during installation.
4.6. Replacing the Fuses
If a fuse is blown or malfunctions:
- Remove test leads and turn off the device.
- Loosen the four screws on the back cover and the screw on the battery door.
- Remove the battery door and back cover.
- Replace the fuse with a new fuse of the same type.
- Reassemble the back cover and battery door, then fasten the screws.
5. Troubleshooting
- Device freezes or crashes: If the device freezes or crashes during use, please restart it.
6. Specifications
6.1. General Specifications
| Characteristic | Value |
|---|---|
| Display | 9999 Counts LCD |
| Ranging | Auto/Manual |
| Material | ABS+TPE |
| Update Rate | 3 Times/Second |
| True RMS | Yes |
| Data Hold | Yes |
| Backlight | Yes |
| Low Battery Indication | Yes |
| Auto Power Off | Yes |
6.2. Mechanical Specifications
| Characteristic | Value |
|---|---|
| Dimension | 177*89*40mm |
| Weight | 345g (NO BATTERY) |
| Battery Type | 18650 Battery * 1 |
| Warranty | One year |
6.3. Environmental Specifications
| Parameter | Operating Range | Storage Range |
|---|---|---|
| Temperature | 0~40°C | -20~60°C |
| Humidity | <75% | <80% |
6.4. Multimeter Specifications
| Function | Range | Resolution | Accuracy |
|---|---|---|---|
| DC Voltage (V) | 999.9mV | 0.1mV | ±(0.5%+3) |
| 9.999V | 0.001V | ||
| 99.99V | 0.01V | ||
| 999.9V | 0.1V | ||
| DC Voltage (mV) | 9.999mV | 0.001mV | |
| 99.99mV | 0.01mV | ||
| AC Voltage (V) | 999.9mV | 0.1mV | ±(1.0%+3) |
| 9.999V | 0.001V | ||
| 99.99V | 0.01V | ||
| 750.0V | 0.1V | ||
| AC Voltage (mV) | 9.999mV | 0.001mV | |
| 99.99mV | 0.01mV | ||
| *Frequency response of ACV: 40Hz-1kHz | |||
| DC Current (A/mA) | 9.999A | 0.001A | ±(1.0%+3) |
| 999.9mA | 0.1mA | ||
| DC Current (µA/mA) | 99.99mA | 0.01mA | ±(0.8%+3) |
| 9999µA | 1µA | ||
| AC Current (A) | 9.999A | 0.001A | ±(1.2%+3) |
| 999.9mA | 0.1mA | ||
| AC Current (µA/mA) | 99.99mA | 0.01mA | ±(1.0%+3) |
| 9999µA | 1µA | ||
| Frequency response of AC Current: 40Hz-1kHz | |||
| Resistance | 99.99Ω | 0.01Ω | ±(1.0%+3) |
| 999.9Ω | 0.1Ω | ±(0.5%+3) | |
| 9.999KΩ | 0.001KΩ | ||
| 99.99KΩ | 0.01KΩ | ||
| 999.9KΩ | 0.1KΩ | ±(1.5%+3) | |
| 9.999MΩ | 0.001MΩ | ±(3.0%+5) | |
| 99.99MΩ | 0.01MΩ | ±(3.0%+5) | |
| Capacitance | 9.999nF | 0.001nF | ±(5.0%+20) |
| 99.99nF | 0.01nF | ±(2.0%+5) | |
| 999.9nF | 0.1nF | ||
| 9.999µF | 0.001µF | ||
| 99.99µF | 0.01µF | ±(5.0%+5) | |
| 999.9µF | 0.1µF | ||
| 9.999mF | 0.001mF | ||
| Frequency (Measures only to 100kHz under ACV setting) | 99.99Hz | 0.01Hz | ±(0.1%+2) |
| 999.9Hz | 0.1Hz | ||
| 9.999kHz | 0.001kHz | ||
| 99.99kHz | 0.01kHz | ||
| 999.9kHz | 0.1kHz | ||
| Temperature | (-20~1000)°C | 1°C | ±(2.5%+5) |
| (-4~1832)°F | 1°F | ||
| Diode | Yes | ||
| Continuity | Yes | ||
6.5. Oscilloscope Specifications
| Characteristic | Instructions |
|---|---|
| Bandwidth | 10MHz |
| Sampling Method | Real-time sampling |
| Real-time Sample Rate | 48MSa/s |
| Channels | 1 |
| Input Coupling | AC/DC |
| Input Resistance | 1MΩ, @16pf |
| Probe Attenuation | X1, X10 |
| Max Input Voltage | X1<150V, X10<300V (DC+AC peak) |
| Sample Rate Range | 1.5Sa/s - 48MSa/s |
| Waveform Interpolation | (sinx)x |
| Sweep Speed Range | 100ns/div - 20s/div |
| Time Base Accuracy | 20ppm |
| Record Length | Max 64Kbyte |
| Sensitivity | 20mV/div - 10V/div |
| Displacement Range | 4 cells (positive and negative) |
| Analog Bandwidth | 10MHz |
| Low Frequency Response | >10Hz |
| Rise Time | <10ns |
| DC Gain Accuracy | ±3% |
| Automatic Measurement | Period, frequency, peak-to-peak value, maximum value, minimum value, effective value |
| Trigger Method | Auto, normal, single |
| Trigger Edge | Rising edge, falling edge |
7. User Tips
- Utilize the automatic measurement mode for ease of use and quick setup in both oscilloscope and multimeter functions.
- The device comes with a 2000mAh 18650 battery. If extended operation time is desired, this battery can be replaced with a higher capacity 18650 battery.
8. Warranty and Support
8.1. Limited Warranty
Customers enjoy a one-year warranty from the date of purchase. This warranty does not cover:
- Fuses and disposable batteries.
- Damage resulting from misuse, accident, neglect, alteration, or contamination.
- Damage due to abnormal operating or handling conditions.
- Failures caused by use outside of the product's specifications.
- Normal wear and tear of mechanical components.
8.2. Support
For further assistance or inquiries not covered in this manual, please refer to the seller's support channels.