1. Brief Introduction
The NJTY 328D is a multi-function digital clamp meter designed for high measurement accuracy, fast response, and a high safety level. It features a special chip capable of up to 6000 counts and a high-precision A/D converter with a high-speed digital processor for true RMS calculations up to AC 1KHz. This instrument provides accurate measurements, high resolution, and fast operation, with complete software calibration ensuring long-term accuracy.
This meter can be used to measure AC and DC voltage, AC and DC current, resistance, capacitance, temperature, diodes, and perform circuit continuity tests. It is equipped with a backlight for easy reading in dark environments.
2. Safety Information
Please read and understand all safety information before operating this meter. Failure to follow these instructions could result in electric shock, fire, or serious injury.
- This meter is rated for CAT II 600V and Pollution Grade 2.
- The maximum voltage between the measurement end and ground is 600V.
- Ensure the instrument is turned off and all test leads are disconnected before replacing batteries or fuses.
- When measuring current, keep your fingers under the arc-shaped part of the clamp to ensure safety.
- Observe correct polarity when connecting capacitors to protect the meter.
- Do not attempt to measure voltages or currents exceeding the specified maximum ranges.
- Overload protection is provided for various functions (e.g., 250V DC/AC for Resistance, Diode, DCA, ACA, Frequency, Capacitance).
3. Product Overview
The NJTY 328D clamp meter features a compact design with a large clamp jaw for easy testing. It includes a clear LED display and various function controls.
3.1. Symbols on Panel
Refer to the diagram below for an explanation of the symbols and controls on the meter's panel:
- COM input socket
- LED display
- Clear symbol + frequency + percentage (ZERO+Hz+%) button: Press to clear DCA 60A readings or choose frequency/percentage in ACV/ACA ranges.
- True RMS indicator
- Trigger: Used to open and close the clamp.
- DCA direction symbol
- Clamp
- NCV induction plate
- NCV indicator
- Functions switch (Rotary Dial)
- Data hold (HOLD/*) button: Short press to hold data, long press to turn on/off backlight.
- Function choose (FUNC) button: Selects measurement functions.
- V/Ω/Hz/°C input socket
4. Setup
4.1. Initial Inspection
Upon receiving your NJTY 328D Clamp Meter, please check the package contents against the list below:
- NJTY 328D Clamp Meter
- Instruction Manual (this document and the included PDF)
- Test Leads (one pair)
- Thermocouple (one piece)
- Instrument Case
- AAA Batteries (two pieces, for initial installation)
If any items are missing or damaged, please contact your vendor immediately.
4.2. Battery Installation
The meter requires two AAA batteries for operation. Follow these steps to install them:
- Ensure the instrument is turned OFF.
- Pull all test pens out of the input sockets.
- Use a screwdriver to loosen the screws fixing the battery cover on the back of the meter.
- Remove the battery cover.
- Insert two new AAA batteries, observing the correct polarity (+/-).
- Replace the battery cover and tighten the screws securely.
5. Operation Instructions
5.1. General Operation
- ZERO/Hz Key:
- Clears induced digits on the LED screen when in DCA 60A range.
- Used to choose frequency and percentage in ACV and ACA ranges.
- FUNC Key: Used to select specific measuring functions (e.g., DCV/DCA, ACA modes, resistance, diode, buzzer, capacitance modes).
- Data Hold (HOLD/*) Key:
- Short press to enter/exit data hold mode.
- Long press to turn on/off the backlight. The backlight automatically turns off after 15 seconds of no operation.
- Auto Power Off: The meter will automatically turn off the LED screen after approximately 15 minutes of inactivity to save power. Any key press will restart it.
5.2. ACV and DCV Measurement
- Turn the rotary switch to the V position. Use the FUNC button to choose between AC (~) or DC (=) voltage measurement.
- Connect the black test lead to the COM jack and the red test lead to the V/Ω/Hz/°C jack.
- Measure the voltage of the circuit under test by touching the test leads to the desired points.
Note: The meter may show readings in DCV 600mV and 6V ranges even without input voltage or test leads connected. Short circuit V/Ω and COM to zero the meter. For AC voltage, the meter measures True RMS values.
5.3. Resistance Measurement
- Rotate the rotary switch to the Ω position. Press the FUNC button to select the 'Ω' resistance measurement mode.
- Connect the black test lead to the COM input socket and the red test lead to the V/Ω/Hz/°C input socket.
- Use the test leads to measure the resistance value of the circuit. The value will be shown on the display.
Notes:
- To accurately measure low resistance, short-circuit the test leads to read their inherent resistance, then subtract this value from your measurement.
- At 60 megohms range, the reading will stabilize after a few seconds.
- If the display shows "OL" when not in circuit, it indicates the measurement value is out of range.
5.4. Diode Measurement
- Turn the rotary switch to the diode symbol (►|) position. Press the FUNC button to select the diode test mode.
- Connect the black test lead to the COM input socket and the red test lead to the V/Ω/Hz/°C input socket.
- Connect the black test lead to the negative electrode and the red test lead to the positive electrode of the diode under test.
- The meter will display the forward bias value. If polarity is reversed, it will show 'OL'.
5.5. Buzzer Test (Continuity)
- Turn the rotary switch to the buzzer symbol (🔊) position. Press the FUNC button to select the continuity test mode.
- Connect the black test lead to the COM input socket and the red test lead to the V/Ω/Hz/°C input socket.
- Measure the resistance of the circuit under test. If the resistance is less than approximately 50 Ω, the buzzer will sound continuously.
5.6. Capacitance Measurement
- Turn the rotary switch to the capacitance symbol (∔) position. Press the FUNC button to select the capacitance measurement mode.
- Connect the black test lead to the COM input socket and the red test lead to the V/Ω/Hz/°C input socket.
- Measure the capacitance value of the circuit under test. The reading will be shown on the display.
Notes:
- Measuring large capacitors may take some time.
- Ensure correct polarity when connecting capacitors to protect the meter.
5.7. Current Measurement (AC/DC)
- Turn the rotary switch to the 60A or 600A AC/DC current range. Press FUNC to choose between AC (~) or DC (=) current.
- Before testing DCA, press ZERO to clear any induced digits on the screen (only for 60A range). Induced digits in ACA will not affect results.
- Press the trigger to open the clamp jaw and clamp it around a single conductor under test. Read the current value on the LED display.
Notes:
- Do not suddenly loosen the trigger to avoid reading changes.
- Do not clamp two conductors with opposite current directions, as they will cancel each other out.
- Keep fingers under the arc-shaped part of the clamp for safety.
- If the display shows "OL", the input exceeds the selected range. Switch to a higher range.
- In ACA testing mode, press the Hz key to test the frequency of the alternative current source (for currents greater than 20A).
5.8. Frequency Measurement
- Turn the rotary switch to the Hz% position.
- Connect the black test lead to the COM jack and the red test lead to the V/Ω/Hz/°C jack.
- Measure the frequency value of the circuit under test.
- Press FUNC to display frequency and duty cycle (0.1%–99.9%) respectively.
5.9. Temperature Measurement
- Turn the rotary switch to the °C (°F) position. The meter will show the surrounding temperature.
- Disconnect the test leads. Connect the thermocouple to the COM and V/Ω/Hz/°C inputs, observing correct polarity.
- The meter will display the approximate temperature from the thermocouple.
5.10. NCV (Non-Contact Voltage) Detection
The red LED on the upper right area of the panel will flash quickly when an electric field exceeding 20V is detected by the inductive sensor in the jaws, indicating the presence of voltage without physical contact.
- Set the rotary function switch to the "NCV" position. The NCV indicator will flash every 1-2 seconds, signifying the instrument is ready.
- When the clamp jaws detect voltage, the NCV LED will flash rapidly.
6. Maintenance
6.1. Battery Replacement
Refer to section 4.2 for detailed instructions on battery replacement. Always use two new AAA batteries and ensure correct polarity.
6.2. Cleaning
To clean the meter, wipe the case with a damp cloth and a mild detergent. Do not use abrasives or solvents. Ensure the meter is completely dry before use.
7. Troubleshooting
- "OL" Display: If the display shows "OL" (Overload), it indicates that the measured value exceeds the selected range. Switch to a higher range or ensure the circuit is within the meter's capabilities.
- Low Voltage Display: A battery symbol (🔋) displayed on the screen indicates that the working voltage is low. Replace the batteries promptly to ensure accurate measurements.
- No Display/Meter Not Turning On: Check if the batteries are installed correctly and are not depleted. Ensure the rotary switch is not in the "OFF" position.
- Inaccurate Readings: Ensure test leads are properly connected and not damaged. Verify the correct function is selected for the measurement. For resistance, consider short-circuiting leads to check for inherent resistance.
8. Specifications
8.1. Comprehensive Indicators
- Operating Conditions: 600V CAT II, Pollution Grade 2
- Height: Under 2000m
- Working Temperature: 0-40°C (<80%RH, not considered <10°C)
- Storage Temperature: -10-60°C (<70%RH, remove battery)
- Test/Calibrate Surrounding Temperature: 20°C ± 2°C
- Max Voltage (Measurement End to Ground): 600V
- Clamp Open Size: 25mm
- Auto Ranging: Yes
- Conversion Rate: Approx. 3s/second
- Display: 6000 counts LED display
- Overload Indication: 'OL' displayed
- Low Voltage Indication: Battery symbol (🔋) displayed
- Input Polarity Indicator: '-' shown automatically
- Battery: 2 x AAA
- Size: 183mm (L) * 47mm (W) * 25mm (H)
- Weight: Approx. 165g (battery included)
8.2. Accuracy Index
Accuracy: ± (%reading + digit), one year warranty from the manufacture date. Conditions: surrounding temperature from 18°C to 28°C, <80%RH.
8.3. Measurement Ranges and Accuracy
| Range | Resolution | Accuracy |
|---|---|---|
| 600mV | 0.1mV | ± (0.5% readings + 5 digits) |
| 6V | 1mV | |
| 60V | 10mV | |
| 600V | 100mV | ± (0.8% readings + 5 digits) |
| Input resistance: 10MΩ, Max input voltage: DC600V | ||
| Range | Resolution | Accuracy |
|---|---|---|
| 6V | 1mV | ± (1% readings + 4 digits) |
| 60V | 10mV | |
| 600V | 100mV | ± (1.2% readings + 10 digits) |
| Input resistance: 10MΩ, Max input voltage: 600V, Frequency response: 40Hz to 1KHz true RMS | ||
| Range | Resolution | Scope | Accuracy |
|---|---|---|---|
| °C | 1°C | -20°C --- 1000°C | ± (1.0 + 3) readings |
| °F | 1°F | -4°F --- 1832°F | ± (1.0 + 3) readings |
| Overload protection: 250V DC/AC, Open circuit voltage: 2.4V | |||
| Range | Resolution | Accuracy |
|---|---|---|
| 600Ω | 0.1Ω | |
| 6KΩ | 1Ω | ± (0.8% readings + 5 digits) |
| 60KΩ | 10Ω | |
| 600KΩ | 100Ω | |
| 6MΩ | 1KΩ | ± (1.2% readings + 5 digits) |
| 60MΩ | 10KΩ | |
| Overload protection: 250V DC/AC | ||
| Function | Test Conditions |
|---|---|
| Diode (►|) | Displays approximate value of diode forward voltage drop. |
| Buzzer (🔊) | Sounds when circuit resistance is lower than 50 Ω. |
| Overload protection: 250V DC/AC | |
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 10mA | ± (3% readings + 3 digits) |
| 600A | 100mA | |
| Overload protection: 250V DC/AC | ||
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 10mA | ± (2.5% readings + 3 digits) |
| 600A | 100mA | |
| Overload protection: 250V DC/AC | ||
| Range | Resolution | Accuracy |
|---|---|---|
| 60Hz | 0.01Hz | |
| 600Hz | 0.1Hz | ± (1.5% readings + 5 digits) |
| 6KHz | 0.001KHz | |
| 60KHz | 0.01KHz | |
| 600KHz | 0.1KHz | |
| 10MHz | 0.001MHz | |
| Overload protection: 250V DC/AC, Input Voltage Range: 200mV --- 10V DC | ||
| Range | Resolution | Accuracy |
|---|---|---|
| 60nF | 0.01nF | |
| 600nF | 0.1nF | ± (4.0% readings + 5 digits) |
| 6uF | 1nF | |
| 60uF | 10nF | |
| 600uF | 100nF | |
| 6mF | 0.1uF | ± (5.0% readings + 5 digits) |
| 100mF | 0.001mF | |
| Overload protection: 250V DC/AC | ||
9. User Tips
- Always start with the highest range when measuring an unknown value to prevent overloading the meter.
- For current measurements, ensure the clamp jaw completely encloses only one conductor. Measuring multiple conductors carrying current in different directions will result in an inaccurate reading.
- Utilize the backlight feature in dimly lit areas by long-pressing the HOLD/* button.
- Remember to turn off the meter when not in use to conserve battery life, although it has an auto-power-off function.
- Keep the test leads and meter clean and free from dust or moisture for optimal performance.
10. Frequently Asked Questions (FAQ)
- Q: Why does my meter show "OL"?
- A: "OL" stands for Overload. It means the measured value is higher than the selected range. Try switching to a higher range setting.
- Q: How do I know if the battery is low?
- A: A battery symbol (🔋) will appear on the display when the battery voltage is low. Replace the batteries as soon as possible.
- Q: Can I measure AC and DC current simultaneously?
- A: No, you must select either AC or DC current mode using the FUNC button after setting the rotary switch to the current measurement range.
- Q: What does True RMS mean?
- A: True RMS (Root Mean Square) means the meter can accurately measure AC voltage and current for both pure sine waves and non-sine wave forms (like square waves, triangular waves, etc.), providing a more accurate reading than average-responding meters.
11. Warranty and Support
The NJTY 328D Digital Clamp Meter comes with a one-year warranty from the date of manufacture, covering defects in materials and workmanship under normal use. This warranty does not cover damage caused by misuse, accident, neglect, alteration, contamination, or abnormal operating conditions.
For detailed warranty terms, troubleshooting assistance, or technical support, please refer to the complete user manual PDF available at: NJTY 328D User Manual (PDF). You may also contact your point of purchase for further assistance.