1. Safety Information
WARNING: Pay special attention to the use of this instrument. Improper use may cause electric shock or damage to the instrument. During use, observe usual safety regulations and safety measures regulated in this manual. To ensure safe operation and full use of functions, please carefully read and observe the methods in this manual.
This instrument complies with safety requirements for electronic measuring instruments: EN-61010-1, EN-61010-2-030, and EN-61010-2-032, level II pollution, and over-voltage standard CAT III 1000V, CAT IV 600V.
1.1 Preparation
- When using this instrument, users must observe standard safety rules:
- General electric shock prevention.
- Prevention of misuse of the instrument.
- After receiving the instrument, check whether it is damaged during transportation.
- After storing and shipping under adverse conditions, check whether the instrument has been damaged.
- Test leads must be in good condition. Before use, check whether the insulation of the leads is damaged and whether metal wire is exposed.
1.2 Symbols
| Symbol | Description |
|---|---|
| Note (refer to use manual for important safety information) | |
| 4 | Able to be used on dangerous electrified conductors. |
| Dual-insulation protection (Category II) | |
| CAT III CAT IV |
Over-voltage according to IEC-61010-1 standard (installation), level III, IV pollution degree 2 refers to protection level of pulse withstand voltage provided. |
| CE | Comply with EU standard. |
| Grounded |
2. Product Overview
The PEAKMETER PM2128 is a multifunctional digital clamp meter designed for high-precision electrical measurements. It features a robust design with a large jaw opening and a clear LCD display.
Image: The PEAKMETER PM2128 Clamp Meter, showcasing its capabilities for True RMS measurement, AC/DC current up to 1000A, AC/DC voltage up to 1000V, and VFD frequency measurement.
2.1 Part Name
Image: Diagram illustrating the main components of the PM2128 Clamp Meter.
- Current clamp: Used in current measurement
- Rotary switch
- Function key
- Liquid crystal display (LCD)
- Input socket
- Function key
- Trigger
- Non-contact voltage detecting & inducing area
2.2 Rotary Switch Description
| Symbol | Function |
|---|---|
| A≈ | AC & DC current measurement |
| TEMP | Temperature measurement |
| Capacitor, on-off and resistance measurement | |
| Hz VFD V | AC/DC voltage measurement |
| LoZ V | Low-impedance voltage measurement |
| uA | DC µA current measurement |
| OFF | Instrument OFF |
2.3 LCD Display Symbols
| Symbol | Description |
|---|---|
| AC & DC | |
| Connection/Disconnection indication | |
| AUTO | Automatic range mode |
| Automatic shutdown indication | |
| Low battery | |
| H | Reading hold state |
| V, A, µA | Volt (voltage), ampere (current) |
| Ω, kΩ, MΩ | ohm, kilohm and megohm (resistance) |
| Hz, kHz | hertz, kilohertz |
| VFD | Variable frequency measurement |
| uF | Microfarad |
| °C°F | Centigrade and Fahrenheit |
| LowZ | Low-impedance measurement mode |
| % | Duty ration measurement |
| REL | Relative measurement |
3. Specifications
The instrument specifies one year as a cycle and shall be re-calibrated under 18°C ~ 28°C, with relative humidity less than 75%.
3.1 Overview
- Select measurement function and range automatically.
- Overload protection throughout the range.
- Max. voltage between measurement terminal and ground: 1000V DC or 1000V AC.
- Operating height: Max. 2000m.
- Display: LCD.
- Max. display value: 6000 digit.
- Polarity indication: Automatic indication, and '-' indicates negative.
- Over range display: 'OL' or '-OL'.
- Sampling time: About 3 times/second, bargraph 10 times/second.
- Unit display: Function and electricity quantity unit display.
- Automatic shutdown time: 10 minutes.
- Power supply: 1.5V AAA battery × 3.
- Battery under-voltage indication: LCD display symbol
. - Temperature coefficient: Less than 0.1 × accuracy/°C.
- Operating temperature: 18°C ~ 28°C.
- Storage temperature: -10°C ~ 50°C.
3.2 Technical Indexes
3.2.1 AC Current
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 0.01A | ± (2.5% reading + 8 digits) |
| 600A | 0.1A | |
| 1000A | 1A |
- Min. input current: 0.1A AC current.
- Max. input current: 1000A AC current.
- Frequency range: 45 ~ 65Hz.
3.2.2 DC Current
| Range | Resolution | Accuracy |
|---|---|---|
| 60A | 0.01A | ± (3% reading + 10 digits) |
| 600A | 0.1A | |
| 1000A | 1A |
- Min. input current: 0.01A DC current.
- Max. input current: 1000A DC current.
3.2.3 DC Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| 6V | 0.001V | ± (0.5% reading + 5 digits) |
| 60V | 0.01V | |
| 600V | 0.1V | |
| 1000V | 1V |
- Min. input voltage: 0.001V DC.
- Max. input voltage: 1000V DC.
3.2.4 AC Voltage
| Range | Resolution | Accuracy |
|---|---|---|
| 6V | 0.001V | ± (0.8% reading + 5 digits) |
| 60V | 0.01V | |
| 600V | 0.1V | ± (1.0% reading + 5 digits) |
| 1000V | 1V |
- Min. input voltage: 0.001V AC.
- Max. input voltage: 1000V AC (valid value).
- Frequency range: 45 ~ 1000Hz.
3.2.5 Frequency
3.2.5.1 Clamp frequency measurement (via gear A):
| Range | Resolution | Accuracy |
|---|---|---|
| 60.0Hz | 0.1Hz | ± (1.0% reading + 5 digits) |
| 1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz.
- Input signal range: ≥ 2A AC current (valid value).
3.2.5.2 Via gear V:
| Range | Resolution | Accuracy |
|---|---|---|
| 60.0Hz | 0.1Hz | ± (1.0% reading + 5 digits) |
| 1000Hz | 1Hz |
- Measurement range: 40Hz ~ 1000Hz.
- Input signal range: ≥ 0.8V AC voltage (valid value).
3.2.6 Resistance
| Range | Resolution | Accuracy |
|---|---|---|
| 6kΩ | 0.001kΩ | ± (0.8% reading + 3 digits) |
| 60kΩ | 0.01kΩ | |
| 600kΩ | 0.1kΩ | |
| 6MΩ | 0.001MΩ | |
| 60MΩ | 0.01MΩ |
- Overload protection: 250V DC or AC (valid value).
3.2.7 Line On-off Test
| Range | Resolution | Function |
|---|---|---|
| 1Ω | If resistance of the line being measured is less than 30Ω, the instrument buzzer will make continuous alarming sounds. |
- Overload protection: 250V DC or AC (valid value).
3.2.8 Capacitor
| Range | Resolution | Accuracy |
|---|---|---|
| 60.00nF | 0.01nF | ± (4% reading + 3 digits) |
| 600.0nF | 0.1nF | |
| 6.000µF | 1nF | |
| 60.00µF | 10nF | |
| 600.0µF | 100nF | |
| 6.000mF | 1µF | |
| 60.00mF | 10µF |
- Overload protection: 250V DC or AC (valid value).
3.2.9 Temperature
| Range | Resolution | Accuracy |
|---|---|---|
| -20 ~ 1000°C (-4 ~ 1832°F) |
1°C/2°F | ± (1% reading + 2 digits) |
- Overload protection: 250V DC or AC (valid value).
3.2.10 µA Current
| Range | Resolution | Accuracy |
|---|---|---|
| 200.0µA | 0.1µA | ± (0.8% reading + 3 digits) |
- Overload protection: 250V DC or AC (valid value).
4. Setup and Operation Guide
4.1 Initial Setup
- Battery Installation: The PM2128 requires 3x 1.5V AAA batteries. Refer to section 5.1 for detailed instructions on changing the battery.
- Test Lead Connection: For voltage, resistance, capacitance, and µA current measurements, connect the test leads to the input sockets. Ensure the red lead is connected to the positive terminal and the black lead to the negative terminal of the circuit being measured.
4.2 Reading Hold
During measurement, if you need to hold the reading on the display, press the H button. The value on the display will be locked. Press the H button again to cancel the reading hold.
4.3 Backlight/Light
- If the environment light is too dark, press the H/
key for more than 2 seconds to turn on the backlight or flashlight. It will automatically turn off in about 1 minute. - During this period, pressing the H/
key for more than 2 seconds will turn off the backlight.
Image: The PM2128 Clamp Meter demonstrating its flashlight feature for emergency night lighting.
4.4 Automatic Shutdown
- If there is no operation within 10 minutes after starting, the instrument will enter standby state and shut down to save energy.
- After automatic shutdown, press any key to wake the instrument and resume operation.
4.5 AC & DC Current Measurement
- Turn the rotary switch to the AC current range (A≈).
- Hold the trigger to open the clamp.
- Clip one lead of the line being measured within the clamp. The instrument will display the measured current.
- Press the FUNC key to select AC or DC current measurement.
- Press the RAN key to select the range manually.
- During AC current measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured current.
- Long press the FUNC key to start the VFD measurement function.
Image: Diagram illustrating the setup for AC & DC current measurement using the clamp function.
4.6 AC & DC Voltage Measurement
- Turn the rotary switch to AC & DC voltage (Hz VFD V).
- Press the FUNC key to shift between AC and DC voltage measurement modes.
- Connect the test leads to the signal being measured: red lead to the positive, black lead to the negative.
- During AC voltage measurement, press the Hz% key to shift the display between frequency and duty ratio of the measured voltage.
- Long press the FUNC key to start the VFD measurement function.
Image: Diagram illustrating the connection for AC & DC voltage measurement using test leads.
4.7 Low-Impedance Measurement
- Turn the rotary switch to LowZ.
- Press the FUNC key to shift between AC and DC voltage measurement modes.
- Connect the test leads to the signal being measured.
Note:
- In low-impedance measurement mode, the longest measurement time shall not be larger than 1 minute.
4.8 On-off/Resistance/Capacitor Measurement
- Turn the rotary switch to the resistance and capacitor position (
). - Press the FUNC key to shift between on-off, resistance, and capacitor functions.
- When test leads are connected to the resistance being measured, the instrument will display the measured resistance. If the measured value is less than 30Ω, the instrument buzzer will alarm.
Image: Diagram illustrating the connection for On-off, Resistance, and Capacitor measurement.
4.9 Temperature Measurement
- Turn the rotary switch to TEMP.
- Insert the thermocouple probe into the socket, with the positive of the probe connected to the red input terminal.
- The primary display will show the Centigrade of the measured temperature.
- Press the FUNC key to display Fahrenheit of the measured temperature.
4.10 µA Current Measurement
- Turn the rotary switch to µA.
- Connect the test leads to the signal being measured in series. The primary display will show the measured current.
4.11 NCV (Non-contact Voltage Detection)
- Turn the rotary switch to NCV. The instrument will display 'EF'.
- Close the NCV detector to the lead being measured. The instrument can detect whether the lead being measured has >90V AC voltage.
- When the instrument detects AC voltage, the instrument buzzer will alarm, and the NCV alarm light will flash.
Note:
Even if there is no alarm indication, voltage may still exist. Do not depend solely on the non-contact voltage detector to judge whether there is voltage in the lead. Detection operation may be impacted by factors like different socket designs and insulation thickness types, etc.
5. Maintenance
5.1 General Maintenance
- Do not try to open the bottom case to adjust or repair the instrument. Such operations should only be conducted by technicians who fully understand the instrument and the electric shock danger.
- Before opening the instrument bottom case or battery cover, remove the test leads from the wire being measured.
- To avoid electric shock possibly caused by error reading, when the low battery symbol (
) displays on the instrument, replace the battery immediately. - Use a wet cloth and gentle detergent to clean the instrument; do not use any abrasive or solvent.
- Power off the instrument when not in use, and rotate the range switch to the OFF position.
- If the instrument is not used for a long time, take out the battery to avoid any damage to the instrument.
5.2 Change Battery
WARNING: Before opening the battery cover of the instrument, remove the test leads from the circuit being measured to avoid electric shock.
- If the low battery symbol (
) appears, it indicates to change the battery. - Unfasten the bolts on the battery cover of the instrument and remove the cover.
- Change the old battery with new 1.5V AAA batteries (3 pieces).
- Place the battery cover back and fasten the bolts.
Image: Diagram showing the battery compartment and replacement process.
5.3 Change Test Leads
WARNING: When changing the test leads, it is required to replace them with identical leads or leads of the same level. The leads must be in good condition, and the level of the leads is: 1000V 10A.
If the insulation layer of the test lead is damaged, or the metal wire of the lead is exposed, it is required to change the lead immediately to ensure safety and accurate measurements.
6. Troubleshooting
This manual does not provide specific troubleshooting steps for common issues. If you encounter problems not covered by the operational or maintenance sections, please refer to the full user manual PDF or contact customer support.
7. Warranty and Support
The instrument specifies a one-year cycle for re-calibration, implying a standard warranty period for manufacturing defects. For detailed warranty terms and conditions, please refer to the official product documentation.
For further assistance or detailed information, please consult the complete user manual available in PDF format: View User Manual (PDF).
8. User Tips
- Check Package Contents: Upon receiving your PEAKMETER PM2128, ensure all accessories are present. The package typically includes the clamp meter, a user manual, a temperature probe, test leads, and a carrying case.
- Battery Life: Always replace batteries promptly when the low battery indicator appears to ensure accurate readings and prevent potential damage to the instrument.
- Test Lead Condition: Regularly inspect your test leads for any signs of damage to the insulation or exposed metal wires. Damaged leads can pose a safety risk and affect measurement accuracy. Replace them with appropriate 1000V 10A rated leads if necessary.
- Non-Contact Voltage Detection: While NCV is a convenient feature, always exercise caution and do not rely solely on it to confirm the absence of voltage. Use direct contact measurement with test leads for definitive voltage verification.