NJTY DT-6013

Contents

NJTY DT-6013 Digital Capacitance Meter User Manual

High Precision Digital Capacitance Meter for Measuring Capacitor Capacity

1. Product Overview

The NJTY DT-6013 is a high-precision digital capacitance meter designed for accurate measurement of capacitor capacity. It features a dedicated chip with up to 6000 counts, composed of high-precision AD and a high-speed digital processor. With 10 ranges from 600pF to 100mF, it offers a wide measurement range and superior instrumentation performance.

NJTY DT-6013 Digital Capacitance Meter front view
Front view of the DT-6013 Capacitance Meter, displaying '0.000 uF'. Key features include high-precision, up to 100,000 uF capacitor measurement, 0.1pF minimum resolution, and automatic shutdown.

General Characteristics

  • Display: 6000 LCD display.
  • Measurement Method: Double integral A/D conversion.
  • Low Battery Indication: LCD display.
  • Backlight: Long press the "HOLD/*" key to turn on/off. Automatically turns off after approximately 30 seconds of no operation.
  • Data Hold: Press the "HOLD/*" key during measurement to hold the last reading. Press again or change measurement function to exit.
  • ZERO Function: Clear button to clear sensor numbers on the LCD when the machine is turned on.
  • Fuse: 0.2A fuse.
  • Power Supply: 2 x 1.5V AA batteries.
NJTY DT-6013 Digital Capacitance Meter features
The DT-6013 offers quick one-step measurement, 6000 counts, LCD screen with backlight, measurement up to 100,000 µF, range adjustment, minimum measurement of 1pF, and sampling for 0-5 seconds.

2. Safety Information

Please read the following safety guidelines carefully before using the DT-6013 meter. The design complies with IEC 10 10 clause safety standards.

  • Do not apply voltage to the testing ends to avoid damaging the meter.
  • When measuring components within a circuit, ensure the circuit is powered off and all capacitors are fully discharged before connecting the test probes.
  • Always disconnect the testing leads from the capacitor before shifting the measurement range.
  • Select the appropriate measurement range to prevent misoperation.
  • Never forcibly short-circuit the input terminals. Prolonged short-circuiting can quickly drain the battery and potentially damage the internal circuit.

3. Setup

Battery Installation

  1. Ensure the meter is turned off.
  2. Locate the battery compartment on the back of the meter.
  3. Open the battery cover.
  4. Insert two 1.5V AA batteries, observing the correct polarity (+/-) as indicated inside the compartment.
  5. Close the battery cover securely.
Battery compartment of DT-6013
The DT-6013 requires two AA batteries for operation (batteries not included).

Protective Case

The meter comes with a soft rubber protective case. It is recommended to use this case to protect the device from accidental drops and impacts.

DT-6013 with protective case
The DT-6013 is supplied with a soft rubber protective case for enhanced durability.

4. Operating Instructions

General Operation

  • Power On/Off: Use the rotary switch to select a measurement range to turn on the meter. Turn to 'OFF' position to power off.
  • Backlight: Long press the "HOLD/*" button to toggle the display backlight. The backlight will automatically turn off after approximately 30 seconds of inactivity.
  • Data Hold: During a measurement, press the "HOLD/*" button to freeze the current reading on the display. Press it again or change the measurement function to release the hold.
  • ZERO Function: If the display shows a non-zero value when no component is connected (especially for ranges below 1.6nF), press the "ZERO" button to clear the display.

Capacitance Testing

  1. Discharge Capacitor: Before connecting, ensure the capacitor to be measured is fully discharged to prevent electric shock and damage to the meter.
  2. Select Range: Turn the rotary switch to the appropriate capacitance range (e.g., 600pF, 6nF, 60nF, 600nF, 6uF, 60uF, 600uF, 6mF, 60mF, 100mF).
  3. Connect Test Leads: Insert the red test lead (with alligator clip) into one of the input jacks marked 'CX MAX 30V' and the black test lead into the middle input jack. Connect the alligator clips to the capacitor pins, observing correct polarity if applicable.
  4. Read Measurement: The capacitance value will be displayed on the LCD.
DT-6013 measuring a capacitor
Example of the DT-6013 measuring a 1000 µF capacitor, displaying '0.914 mF'.

5. Troubleshooting

  • "OL" Display: If the display shows "OL", it indicates that the measurement range has been exceeded. Select a higher capacitance range.
  • Leading Zeros: If the display shows one or more zeros before the measured value, consider selecting a lower range to improve measurement resolution and accuracy.
  • Non-Zero Display (below 1.6nF): If the display is not zero when no component is connected in ranges below 1.6nF, press the "ZERO" button to clear it.
  • Stray Capacitance: When using test pens, they can introduce stray capacitance. The meter is designed to subtract this during measurement. For highly accurate measurements, use particularly short wires to minimize stray capacitance.
  • Unstable Readings for Large Capacitors: An open-circuit capacitance will display as "0" in all ranges. If a large capacitor with serious leakage or breakdown shows an unstable value, it suggests the capacitor is leaking. Confirm with other measuring tools.

6. Maintenance

  • Ensure the back cover is securely fixed before operating the meter.
  • Always disconnect the circuit power before opening the back cover for any reason.
  • Clean the meter only with a damp cloth and a small amount of mild detergent. Do not use any chemical solvents.
  • If the meter exhibits any abnormal behavior, discontinue use immediately and seek professional repair.

7. Specifications

Environmental Conditions

  • Operation Temperature: 0°C to 40°C (32°F to 104°F), relative humidity < 85%
  • Storage Temperature: -10°C to 50°C (14°F to 122°F), relative humidity < 85%
  • Accuracy Ensured Temperature: 23°C ±5°C (73.4°F ±9°F), relative humidity < 75%

Measurement Ranges and Accuracy

RangeResolutionAccuracy
600pF0.1pF±(2%+10)
6nF0.001nF±(0.6%+10)
60nF0.01nF±(0.6%+10)
600nF0.1nF±(0.6%+10)
6uF0.001uF±(0.6%+10)
60uF0.01uF±(0.6%+10)
600uF0.1uF±(0.6%+10)
6mF0.001mF±(2%+10)
60mF0.01mF±(4%+10)
100mF0.1mF±(5%+10)

8. User Tips

  • Precision and Value: Users have noted the meter's precision and good value for its cost, performing comparably to more expensive instruments.
  • Ease of Use: The meter is generally considered easy to use and functions well for its intended purpose.
  • Initial Testing: Upon receiving the meter, it's recommended to perform initial tests to familiarize yourself with its operation and verify functionality.

9. Warranty and Support

For detailed information regarding warranty, service, or further technical support, please refer to the official user manual PDF provided by the manufacturer. You can access the PDF manual here: NJTY DT-6013 User Manual (PDF).

For any issues or inquiries not covered in this manual, please contact the seller or manufacturer's customer support.