NJTY 58A

Contents

NJTY 58A Multimeter User Manual

Model: 58A

High-Precision Digital Display Smart Burn-Proof Household Electrician Capacitance Meter

1. Introduction

This instrument is a handheld digital multimeter featuring true RMS measurement. It boasts a large 9999-count LCD digital display with automatic backlight adjustment for easy readability in various lighting conditions. Designed with user safety and convenience in mind, it includes overload protection, battery under-voltage indication, and input light indication. This multimeter is an ideal tool for professionals, factories, schools, enthusiasts, and home use.

The NJTY 58A series of digital multimeters are designed and produced in accordance with the international electrical safety standard IEC-61010 for electronic measuring instruments and handheld digital multimeters, ensuring reliable and safe operation.

NJTY T58A Multimeter showing key features like True RMS, diode measurement, capacitance up to 100000uF, gear socket indicator light, and Celsius/Fahrenheit measurement.
The NJTY T58A Multimeter highlighting its core features.

2. Safety Information

2.1 Safety Instructions

  • When using this meter, always comply with standard safety regulations for protection against electric shocks and to prevent misuse of the instrument's safety features.
  • To ensure personal safety, use the provided test leads, checking them for integrity before each use.
  • The use of meter instruments near devices with large electromagnetic disturbances may cause unstable readings or large errors.
  • Do not use the meter if its appearance or the test leads are broken.
  • Incorrect use may compromise the meter's safety functions.
  • Exercise caution when working near bare conductors or busbars.
  • Do not use this instrument in environments with explosive gas, vapor, or dust.
  • Always select the correct input function and range for measurements.
  • Input values must not exceed the specified limits for each range to prevent damage.
  • Do not touch unused input terminals when the instrument is connected to a circuit under test.
  • Use caution to prevent electric shock when measuring voltages exceeding 60V DC or 36V AC.
  • When measuring with test leads, keep your fingers behind the guard ring.
  • Ensure test leads are disconnected from the circuit before changing ranges.
  • For DC functions, always confirm the absence of AC voltage first, then select a DC voltage range equal to or greater than the AC voltage to avoid electric shock from incorrect readings.
  • Before resistance, diode, capacitance, or continuity tests, ensure the circuit under test is powered off and all high-voltage capacitors are fully discharged.
  • Do not measure resistance or perform buzzer tests on live circuits.
  • Before current measurement, check the meter's fuse. Turn off power to the circuit under test before connecting the meter.
  • When performing TV repairs or measuring power conversion circuits, be careful of high-amplitude voltage pulses.
  • The instrument uses three AAA 1.5V batteries. Ensure they are correctly installed.
  • Replace batteries immediately when the low voltage symbol appears, as insufficient power can lead to incorrect readings and potential electric shock.
  • Do not exceed 1000V for Category II voltages or 600V for Category III voltages.
  • Do not use the instrument if its housing or parts are removed.

2.2 Safety Symbols

Symbols used on the meter and in this manual:

Symbol Description
⚠Warning: Important safety signs. Refer to the instruction manual before use. Improper use can result in damage to the device or its parts.
~AC (Alternating Current)
---DC (Direct Current)
≈AC or DC
⊥Ground
□Double insulation protection
🔋Fuse
CEMeets EU standards
⚡High voltage warning (displayed when DC voltage is 60V or AC voltage is above 36V)
CAT. IIII category. 1000V overvoltage protection
CAT. IIIIII category. 600V overvoltage protection

3. Product Overview

3.1 Panel Diagram

The following diagrams illustrate the components of the multimeter's front panel for both manual and auto-ranging models.

Detailed panel diagram of the NJTY T58A Multimeter, showing numbered parts for manual and auto-ranging models.
Panel Diagram (Model A - Manual Ranging, Model B/C - Auto Ranging)
  1. Illumination
  2. NCV sensing area
  3. NCV and buzzer indicator light
  4. LED display
  5. Infrared remote control detection (A model)
  6. hFE test socket
  7. Functions switch button
  8. Lighting/Backlight
  9. Gear rotary switch
  10. VΩHz°C input socket
  11. COM input socket
  12. 10A input socket
  13. mA input socket
  14. Data hold/relative value measurement
  15. Maximum and minimum switching
  16. Peak to Peak Measurement
  17. Range switching button
  18. µA mA L (inductance) input socket

3.2 Display Symbols

The display uses various symbols to indicate the current mode and status:

Symbol Description
−+Battery low voltage display
⏻Auto power off
−Negative polarity input indication
ACAlternative input indication
DCDirect input indication
♫In continuity test mode
→|In diode test mode
AUTOAuto ranging mode
HData hold mode
△In relative value measurement mode
PEAKPeak-to-peak measurement mode
LInductance measurement
MAXMaximum value
MINMinimum value
°C °FTemperature unit
%Duty cycle

3.3 Button Functions

The following table describes the functions of the buttons on the multimeter:

Button Instructions
FUNCShort press to select function (e.g., °C/°F switch, AC/DC switching, Hz/% when AC is used, On/off/diode/resistance gear switching).
A REL HOLDRelative Value Measurement: Press and hold for 2 seconds to enter/exit. Data Hold: Short press to enter/exit, holding the instantaneous measurement value.
☀️/💡Short press to turn backlight on/off. Long press for 2 seconds to turn on illumination and backlight simultaneously; short press again to turn off. Automatically turns off after 30s if not pressed.
RANGESwitches to manual range for Voltage/Resistance/Current measurements.
MAX MINShort press to switch between maximum and minimum value measurement. Long press for 2 seconds to exit.
PEAKFor AC voltage and current signals, short press to enter Peak-to-Peak measurement. Long press for 2 seconds to turn off.

3.4 Input Sockets

The multimeter features several input sockets for different measurement types:

Input Socket Description
COMCommon input for all measurements, connected to the black test lead or the common input of a dedicated multi-function test socket.
V→|♫ΩHz%°C/°FPositive input for diode, capacitor, buzzer, temperature, voltage, resistance, and frequency (connected with the red test lead).
µA mA LµA mA L (inductance) input socket.
10APositive input terminal for current up to 10A.

4. Initial Setup

4.1 Battery Installation

The multimeter requires three AAA 1.5V batteries for operation. To install or replace batteries:

  1. Turn off the power of the instrument.
  2. Pull all test leads out of the input sockets.
  3. Use a screwdriver to loosen the screws fixing the battery cover.
  4. Remove the battery cover.
  5. Remove old batteries (if any).
  6. Insert new 3 x 1.5V AAA batteries, ensuring correct polarity.
  7. Install the battery cover and tighten the screws.

Warning: Replace batteries immediately when the low voltage symbol appears on the display to avoid erroneous readings and potential electric shock.

5. Operating Instructions

5.1 Auto Power Off

After approximately 15 minutes of inactivity, the multimeter will emit an audible prompt and automatically enter hibernation mode to conserve battery. Any key press will restart the device.

5.2 ACV and DCV Measurement

Warning: Do not measure any voltage greater than 1000V DC or 750V AC RMS to prevent electric shock or damage to the instrument. Do not apply more than 1000V DC or 750VAC RMS voltage between the common and earth.

Direct Voltage Ranges: 1V, 10V, 100V, 1000V

Alternative Voltage Ranges: 1V, 10V, 100V, 750V

  1. Turn the gear rotary switch to the V or ≈V/≈mV position. The red indicator light of the corresponding input socket will flash.
  2. Connect the black test lead to the COM jack and the red test lead to the V jack.
  3. Measure the voltage of the circuit under test with the test leads.
  4. The reading and polarity (for DC) will be shown on the LED display.

Notes:

  • In AC and DC 1V range or at ≈mV range (B and C model), the meter may show a digital display even without input. Short-circuit the "V/Ω" and "COM" terminals; if the meter displays zero, it is normal.
  • When measuring AC voltage, the frequency of the AC signal (below 30KHz) will be displayed in the upper right corner.
  • When measuring DC voltage, the ambient temperature will be displayed in the upper right corner.
  • The AC voltage value measured is the true effective value (RMS).

5.3 Resistance Measurement

Warning: To avoid damage, power off the circuit and discharge all high-voltage capacitors before measuring resistance.

The unit of resistance is Ohm (Ω).

Resistance Ranges: 1kΩ, 10kΩ, 100kΩ, 1MΩ, 10MΩ, 100MΩ

  1. Rotate the gear rotary switch to the Ω position (A model) or Ω/♫/→| position and press FUNC to select Ω (B&C model). The red indicator light will flash.
  2. Connect the black test lead to the COM input socket and the red test lead to the VΩ input socket.
  3. Use the test leads to measure the resistance value of the circuit. The value will be shown on the display.

Notes:

  • Resistance measured on a circuit may differ from the rated resistance.
  • For accurate low resistance measurement, short-circuit the test leads to read their inherent resistance, then subtract this value from your measurement.
  • High resistance measurements may take a few seconds to stabilize.
  • If the meter is not connected to a circuit, it will display "OL" (Overload), indicating the measurement is out of range.

5.4 Diode Measurement

Warning: To avoid damage, power off the circuit and discharge all high-voltage capacitors before measuring diodes.

To test a diode outside the circuit:

  1. Turn the gear rotary switch to the →| position (A model) or Ω/♫/→| position and press FUNC to select →| (B&C model). The red indicator light will flash.
  2. Connect the black test lead to the COM input socket and the red test lead to the VΩ input socket.
  3. Connect the black test lead to the negative electrode and the red test lead to the positive electrode of the diode under test.
  4. The meter will display the forward bias value. If polarity is reversed, it will show 'OL'.

5.5 Buzzer Test (Continuity)

Warning: To avoid damage, power off the circuit and discharge all high-voltage capacitors before performing a continuity test.

  1. Turn the gear rotary switch to the ♫ position (A model) or Ω/♫/→| position and press FUNC to select ♫ (B&C model). The red indicator light will flash.
  2. Connect the black test lead to the COM input socket and the red test lead to the VΩ input socket.
  3. If the resistance of the circuit under test is not greater than about 50 ohms, the indicator light will turn on and the buzzer will sound continuously. If the resistance is between 40 and 500 ohms, it will display the resistance value. If greater than 500 ohms, it will display "OL".

5.6 Capacitance Measurement

Warning: To avoid damage, power off the circuit and discharge all high-voltage capacitors before measuring capacitance.

Capacitance Ranges: 1nF, 10nF, 100nF, 1µF, 10µF, 100µF, 1mF, 10mF, 100mF

  1. Turn the gear rotary switch to the 100mF position (A model) or ※ position (B&C model). The red indicator light will flash.
  2. Connect the black test lead to the COM input socket and the red test lead to the capacitance input socket.
  3. Measure the capacitance value of the circuit under test. The reading will be shown on the LED display.

Notes:

  • Measuring large capacitors may take some time.
  • Always discharge large capacitors before measurement to avoid damaging the instrument.

5.7 Transistor Measurement (Model A only)

Warning: Do not apply voltage exceeding 36V DC or AC RMS to the common terminal and hFE terminal to prevent electric shock or damage.

  1. Turn the gear rotary switch to the "hFE" position.
  2. Determine if the transistor is NPN or PNP type.
  3. Insert the three pins (e.b.c) of the transistor into the corresponding holes of the hFE test socket.
  4. Read the hFE measurement value on the LCD.

5.8 Frequency Measurement

Warning: Do not test any voltage higher than 250V to prevent electric shock or damage. If input voltage is greater than 30V, use the AC voltage range to measure frequency.

  1. Turn the gear rotary switch to the Hz% position. The red indicator light will flash.
  2. Connect the black test lead to the COM jack and the red test lead to the Hz jack.
  3. Measure the frequency value of the circuit under test.
  4. The reading and duty cycle (%) will be shown on the LED display.

5.9 NCV Test (Non-Contact Voltage)

  1. Rotate the gear rotary switch to the NCV position.
  2. Place the top of the meter close to the conductor.
  3. If AC voltage is detected, the signal strength indicator (low-yellow, high-red) will light up, and the buzzer will emit different frequency alarms.

Notes:

  • Do not rely solely on NCV detectors to determine if voltage is present, as factors like socket depth, insulation thickness, and type can affect detection.
  • External interference (flashlights, motors) may accidentally trigger NCV detection.
NJTY T58A Multimeter performing an NCV non-contact voltage test, showing a reading of '13' and a dashed line indicating detection.
NCV non-contact voltage test in progress.

5.10 Infrared Remote Control Detection (Model A only)

  1. Turn the gear rotary switch to the ※ position.
  2. Align the transmitter head of the infrared remote control vertically with the infrared receiving window (deviation ≤ ±15°).
  3. Press the remote's launch button. If the red light-emitting tube on the right side flashes, the transmitter is working normally.
  4. Moving the transmitter within 1-10cm can determine its output power status.

Notes:

  • Keep the multimeter's receiving window away from strong light sources when testing.
  • The remote control's transmitter head must be perpendicular to the receiving window (max deviation ±15°).
  • Radio remote controls cannot be detected.

5.11 Temperature Measurement

  1. Turn the gear rotary switch to the °C/°F position. The instrument will display the ambient temperature. The red indicator light of the socket will flash.
  2. Remove the test leads and connect the thermocouple with correct polarity to the COM and V/Ω inputs.
  3. Place the thermocouple on the surface of the object or in the environment to be measured.
  4. The display will show the temperature measurement value.

5.12 Inductance Measurement (Model C only)

  1. Turn the gear rotary switch to the "L" position. The red indicator light of the corresponding input socket will flash.
  2. Connect the black test lead to the COM input terminal and the red test lead to the µA mA L input terminal.
  3. Touch both ends of the inductance device with the test leads. The display will show the measured inductance value.

5.13 Low Impedance Measurement (Model A only)

Automatic switching measurement of AC and DC voltage:

  • DC threshold: 3V
  • AC threshold: 15V
  1. Turn the gear rotary switch to the "LoZ" position. The red indicator light of the corresponding input socket will flash.
  2. Connect the black test lead to the COM input terminal and the red test lead to the V/Ω input terminal.
  3. Connect the test leads in parallel to the power supply or load to be measured. The display will show the measured voltage.

5.14 Current Measurement

Warning: When measuring current, the meter must be connected in series to the circuit under test. Do not attempt to measure circuits in parallel or mains power in parallel. If the fuse blows, you may damage the meter or injure yourself.

To avoid damage to the instrument or equipment, pre-judge the current size, select the appropriate current range, and check the instrument's fuse before measurement.

  1. Turn the gear rotary switch to the appropriate current position (mA or 10A). The red indicator light of the corresponding input socket will flash.
  2. Connect the black test lead to the COM input socket.
  3. If the measured current is less than 1A, connect the red test lead to the "mA" input socket.
  4. If the measured current is between 1A and 10A, connect the red test lead to the "10A" input socket.
  5. Connect the test leads in series with the circuit under test, then turn on the power. The display will show the current measurement value.
  6. If the display shows "OL", the input exceeds the selected range. Turn off power, set the rotary switch to a higher range, and measure again (Model A only).
Diagram showing how to connect the multimeter for current measurement in series with a load, indicating AC and DC connections.
Current measurement setup diagram.

6. Maintenance

6.1 General Maintenance

  • To avoid electric shock or damage, do not allow the inside of the meter to get wet.
  • Before opening the case or battery cover, turn off the power and ensure test leads are disconnected from the circuit.
  • Clean the instrument case regularly with a damp cloth and a small amount of detergent. Do not use abrasives or chemical solvents, as they can oxidize metal surfaces and affect readings.
  • To clean input sockets: Turn off the meter, unplug all test leads, remove dirt, and clean each socket with a new cotton ball moistened with detergent or lubricant to prevent moisture-related contamination.
  • When opening the instrument housing, be aware of residual capacitance. Dangerous voltages may persist even after the instrument is turned off.
  • If any abnormality is observed, stop using the instrument immediately and send it for repair. Ensure it is inspected and qualified before reuse.
  • If not in use for a long time, remove the battery and store the meter away from high temperatures and humidity.

6.2 Fuse Replacement

Warning: Use only the specified fuses (mA range: F 600mA/250V; A range: F 10A/250V fast-blow fuses).

To replace a fuse:

  1. Turn off the power of the instrument.
  2. Pull all test leads out of the input sockets.
  3. Use a screwdriver to loosen the screws fixing the battery cover.
  4. Remove the battery cover.
  5. Remove the damaged fuse.
  6. Replace with a new, specified fuse (a spare fuse may be under the battery cover).
  7. Install the battery cover and tighten the screws.

7. Technical Specifications

7.1 Comprehensive Indicators

  • Operating Conditions: 600V CAT III and 1000V CAT II
  • Pollution Grade: 2
  • Height: Under 2000m
  • Working Temperature: 0-40°C (<80%RH, not considered <10°C)
  • Storage Temperature: -10-60°C (<80%RH, remove battery)
  • Test or Calibration Ambient Temperature: 20°C ±2°C
  • Maximum Voltage between Measurement End and Ground: 1000V DC or 750V AC
  • Fuse Protection: mA range: F 600mA/250V; A range: F 10A/250V
  • Conversion Rate: About 3s/second
  • Display: 9999 counts LED display
  • Overload Display: 'OL' displayed
  • Low Voltage Display of Battery: −+ displayed when working voltage is low
  • Input Polarity Indicator: '-' shown automatically
  • Battery: 3 x 1.5V AAA
  • Size: 187mm (L) * 95mm (W) * 55mm (H)
  • Weight: About 331g (battery excluded)

7.2 Accuracy Index

7.2.1 Direct Current Voltage

Range Resolution Accuracy
1V/≈mV0.1mV (A model)
0.01mV (B/C model)
±(0.5% readings + 5 digits)
10V/≈V0.001V
100V/≈V0.01V
1000V/≈V0.1V±(0.8% readings + 5 digits)

Input resistance: 10M Ω

Maximum input voltage: 1000VDC or 750VAC RMS

7.2.2 Alternative Current Voltage

Range Resolution Accuracy
1V/≈mV0.1mV (A model)
0.01mV (B/C model)
±(1% readings + 4 digits)
10V/≈V0.001V
100V/≈V0.01V
750V/≈V0.1V±(1.2% readings + 4 digits)

Input resistance: 10M Ω

Maximum input voltage: 1000VDC or 750VAC RMS

Frequency response: 10Hz to 1KHz true RMS

7.2.3 Frequency

Range Resolution Accuracy
10Hz0.01Hz
100Hz0.1Hz
1KHz0.001KHz±(1.5% readings + 5 digits)
10KHz0.01KHz
100KHz0.1KHz
1MHz0.001MHz
10MHz0.01MHz

Input voltage range: 200mV-10V AC RMS

Overload protection: 250V DC/AC

7.2.4 Resistance

Range Resolution Accuracy
1KΩ0.1Ω
10KΩ0.001KΩ
100KΩ0.01KΩ±(0.8% readings + 5 digits)
1MΩ0.1KΩ
10MΩ0.001MΩ±(1.2% readings + 5 digits)
100MΩ0.01MΩ

Overload protection: 250V DC/AC

Open circuit voltage: 1.4V

7.2.5 Diode

Function Range Resolution Condition
Diode0-4V0.001VForward direct current: about 1mA; Open circuit voltage: about 3.2V. Displays approximate value of forward voltage drop.

Overload protection: 250V DC/AC

7.2.6 Buzzer

Function Range Resolution Description Condition
♫ 500Ω0.1ΩBuzzer sounds when resistance is less than 50Ω. Red indicator light turns off if resistance is greater than 50Ω.Open-circuit voltage: about 1.7V

Overload protection: 250V DC/AC

7.2.7 Transistor (A model only)

Function Description Condition
hFEApproximation of monitor readout of hFE value (0-1000)Base current is about 10µA, Vce is about 2.8V

7.2.8 Capacitance

Range Resolution Accuracy
1nF0.001nF
10nF0.01nF
100nF0.1nF±(4.0% readings + 5 digits)
1µF1nF
10µF0.001µF
100µF0.01µF
1mF0.1µF±(5.0% readings + 5 digits)
10mF0.001mF
100mF0.1mF

7.2.9 Inductance L (Model C only)

Range Resolution Accuracy
10mH0.001mH
100mH0.01mH
1H0.1mH±(4.0% readings + 5 digits)
10H0.001H
100H0.01H

7.2.10 Temperature

Range Resolution Accuracy
°C1°C±(1.0% + 3) readings
°F1°F±(1.0% + 3) readings

Overload protection: 250V DC/AC

7.2.11 Direct Current

Range Resolution Accuracy
1mA/µA0.1µA
100mA (Model A)0.01mA±(1.2% readings + 5 digits)
1A/mA0.2mA
10A0.001A±(2% readings + 5 digits)

Overload protection: mA range (1A/250V) and 10A range (10A/250V)

Max input current: mA: 1A RMS, 10A: 10A RMS

When current is more than 5A, test time should be less than 10s, and a one-minute cool-down period should be observed after such measurement.

7.2.12 Alternative Current

Range Resolution Accuracy
1mA/µA0.1µA
100mA (Model A)0.01mA±(1.5% readings + 5 digits)
1A/mA0.2mA
10A0.001A±(2% readings + 5 digits)

Overload protection: mA range (1A/250V) and 10A range (10A/250V)

Max input current: mA: 1A RMS, 10A: 10A RMS

When current is more than 5A, test time should be less than 10s, and a one-minute cool-down period should be observed after such measurement.

Frequency response: 40Hz to 1KHz true RMS

7.2.13 LoZ (Low Impedance)

Range Resolution Identification threshold Accuracy
750(ACV)1V15V±(2% readings + 5 digits)
1000(DCV)3V

8. Troubleshooting

This section provides general guidance for common issues. For detailed troubleshooting, refer to the specific operation instructions for each function.

  • No Display / Faint Display: Check battery level. If the low voltage symbol is displayed, replace batteries immediately. Ensure batteries are installed with correct polarity.
  • "OL" Displayed: This indicates an overload or that the measurement is out of the selected range. Switch to a higher range or ensure the circuit is within the meter's capabilities. For resistance, it can also mean an open circuit.
  • Inaccurate Readings: Ensure test leads are fully inserted and in good condition. Check for proper range selection. Ensure the circuit under test is powered off and capacitors discharged for resistance, diode, and capacitance measurements. Replace batteries if low.
  • No Buzzer Sound in Continuity Test: Check if the resistance is above 50 ohms.
  • NCV Not Detecting Voltage: Ensure the NCV sensing area is close to the conductor. Be aware that insulation thickness and external interference can affect NCV detection.
  • Meter Not Turning On: Check battery installation and charge.

9. User Tips

  • Check Included Accessories: The multimeter typically comes with test leads, a temperature probe, and an instruction manual. Some packages may also include a canvas bag for storage. Verify all components upon receipt.
  • Battery Life: To maximize battery life, always turn off the multimeter when not in use, or rely on the auto power-off feature. Remove batteries if storing the device for extended periods.
  • Probe Care: Keep your test leads clean and inspect them regularly for any signs of wear or damage. Damaged leads can lead to inaccurate readings or safety hazards.
NJTY T58A Multimeter displayed with its accessories: an instruction manual, red and black test leads, and a temperature probe.
Multimeter with standard accessories.

10. Warranty and Support

For detailed warranty information and customer support, please refer to the official documentation provided with your product. A comprehensive user manual in PDF format is available for download or viewing:

View User Manual (PDF)

This PDF contains additional diagrams, safety warnings, and detailed operational procedures that may not be fully covered in this summary.

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.