1. Introduction
Thank you for purchasing the NJTY T28B Multimeter. This precision instrument is designed to be a reliable and efficient tool for various electrical measurements. To ensure safe and effective operation, please read this manual thoroughly before using the device. If you have any questions, please contact our customer service.
Figure 1: NJTY T28B Multimeter
2. Safety Information
2.1. General Safety Instructions
The NJTY T28B Multimeter conforms to IEC61010-1 CAT III 600V, 1000V CAT II IEC61010 and pollution level 2. When using this meter, always comply with standard safety regulations to prevent electrical shock, fire, or personal injury.
- Do not alter, modify, or tamper with the multimeter. Use only as specified.
- Do not use the multimeter near explosive gas, vapor, or in damp, wet, or dusty environments.
- Do not operate the multimeter with the covers removed or the case open.
- Do not apply more than the rated voltage between terminals or between each terminal and earth ground.
- Always use the provided test leads. Examine them for damage (insulation, exposed metal) before each use.
- Use the correct terminals, function, and range for measurements.
- Ensure the range switch is in the correct position.
- The input value must not exceed the specified limit for each range.
- When measuring televisions or switching power supplies, be aware of pulses that can damage the circuit.
- Remove batteries if the multimeter is not used for an extended period or stored above 50°C.
2.2. Safety Symbols
| Symbol | Description | Symbol | Description |
|---|---|---|---|
| ⚠️ | WARNING. Risk of danger. | ⎓ | AC or DC |
| ⚡ | High voltage warning | ⎚ | Double insulated |
| ⎓ | DC (Direct Current) | ⏚ | Fuse |
| ∼ | AC (Alternating Current) | ⏚ | Ground |
| CE | Conforms to European Union directives. | ||
| CAT. II | Measurement Category II is applicable to test and measuring circuits connected directly to utilization points (socket outlets and similar points) of the low-voltage MAINS installation. | ||
| CAT. III | Measurement Category III is applicable to test circuits connected to the distribution part of the building's low-voltage MAINS installation. | ||
3. Product Overview
3.1. Key Features
Figure 2: Key Features of the T28B Multimeter
- True RMS Measurement: Provides accurate readings for non-sinusoidal waveforms.
- Auto Range: Automatically selects the appropriate measurement range.
- NCV Electric Field Induction: Non-Contact Voltage detection for safety.
- Capacitance Measurement: Up to 100,000µF.
- Frequency/Duty Cycle Measurement: For analyzing signal characteristics.
- Automatic Shutdown: Conserves battery life after 15 minutes of inactivity.
- Low Voltage Indication: Alerts when battery power is low.
- Overload Protection: Safeguards the device from excessive input.
- Magnetic Attraction Feature: Allows for hands-free operation.
- Soft Rubber Casing: Provides durability and protection.
Figure 3: Multimeter with Soft Rubber Casing
3.2. Components and Controls
The NJTY T28B Multimeter features a clear LED display, a function knob, and various input sockets. Refer to Figure 1 for a visual representation of the device.
- LED Display: Shows measurement readings, units, and indicators (e.g., AUTO, H, OL).
- Function Knob: Used to select the desired measurement function (e.g., V, A, Ω, Hz, NCV, °C).
- FUNC/HOLD Button: Toggles between sub-functions (e.g., DC/AC) or activates/deactivates display hold.
- Hz/%/Δ Button: For frequency/duty cycle measurements or relative measurements.
- RANGE Button: Toggles between auto-ranging and manual ranging.
- */💡 Button: Toggles backlight and flashlight.
- Input Sockets:
- COM: Common terminal for the black test lead.
- VΩHz°C: Input for Voltage, Resistance, Frequency, and Temperature (red test lead).
- mA: Input for milliampere current measurements (red test lead, fused up to 600mA).
- 20A: Input for high current measurements (red test lead, fused up to 20A).
3.3. Backlight / Flashlight
Press the */💡 button to toggle the display backlight on and off. Press and hold the */💡 button for 2 seconds to toggle the flashlight on and off. The backlight/flashlight automatically turns off after 30 seconds to conserve battery.
Figure 4: Multimeter Flashlight in Use
3.4. Auto Power Off
The multimeter will automatically turn off if left idle for approximately 15 minutes. To restart the device, press any button.
3.5. Display Hold
In display hold mode, the multimeter freezes the current reading on the display. Press and hold the FUNC HOLD button for 2 seconds to activate or deactivate this feature. 'H' will be shown on the display when active.
3.6. Manual and Auto-Ranging
The multimeter defaults to Auto-range mode ('AUTO' shown on display). In this mode, the multimeter automatically selects the range with the best resolution.
- To enter Auto-range mode, press the RANGE button for 2 seconds.
- To enter Manual range mode, press the RANGE button. Each press will increase the range. After the highest range, it wraps back to the lowest.
4. Setup
4.1. Installing Batteries & Replacing Fuses
- Turn off the multimeter.
- Pull all test leads out of the input sockets.
- Use a screwdriver to loosen the screws holding the battery cover.
- Remove the battery cover.
- Remove old batteries (4 * 1.5V AAA) or damaged fuses (F 600mA/250V, F 20A/250V).
- Replace with new batteries or fuses and tighten the screws.
Note: Only use the specified fuses to avoid electric shock or personal injury.
4.2. Connecting Test Leads
Always ensure test leads are correctly connected to the appropriate input sockets for the desired measurement. The red lead is positive, and the black lead is negative.
Figure 5: Multimeter with Test Probes and Temperature Probe
5. Operating Instructions
5.1. General Measurement Guidelines
- Check the battery first; if the battery symbol is displayed, replace the batteries.
- The warning symbol beside the sockets indicates maximum input limits.
- Switch to the correct range before measurement.
5.2. DC/AC Voltage Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to the 'mV' or 'V' ranges.
- Connect the test leads to the loaded circuit.
- DC signal will be displayed. Press FUNC HOLD to switch to ACV test ('AC' will be shown).
Note: Do not measure voltages above DC1000V or AC750V to prevent damage. In low voltage AC/DC ranges, the multimeter may display numbers without input; short-circuiting the 'VΩHz°C' and 'COM' ports should return the display to zero, indicating normal operation.
Figure 6: AC Voltage Measurement
5.3. Resistance Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to the 'Ω' range. ('Ω' will be shown on the display).
- Connect the test leads to the resistor to be tested.
Note: 'OL' indicates overload or open circuit. For resistors > 60MΩ, it may take a few seconds to read. Ensure the circuit is powered off and capacitors are discharged before testing. For accurate low resistance measurements, short-circuit test leads and subtract their resistance from the reading.
5.4. Diode Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to the diode symbol and press FUNC HOLD twice. (Diode symbol will be shown).
- Touch the diode anode with the red probe and the cathode with the black probe.
Note: 'OL' indicates an open circuit or reversed polarity.
5.5. Continuity Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to the continuity symbol and press FUNC HOLD. (Continuity symbol will be shown).
- Contact the probes to the measured circuit or resistance.
Indication: The multimeter will buzz and the indicator light will be yellow if resistance is < 40Ω. If resistance is between 40Ω and 60Ω, the buzzer will not sound, and the indicator light will be red. 'OL' indicates an open circuit. Ensure the circuit is powered off and capacitors are discharged before testing.
5.6. Capacitance Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to the capacitance symbol.
- Connect the test leads to the circuit to be tested.
Note: Ensure the circuit is powered off and capacitors are completely discharged before testing. Larger capacitances may take longer to stabilize the reading.
5.7. Frequency/Duty Cycle Measurement
- Insert the black lead into the 'COM' port and the red lead into the 'VΩHz°C' port.
- Switch the function knob to 'Hz%' and press the Hz/%/Δ button to select Frequency or Duty Ratio.
- Connect the test leads to the circuit to be tested.
Note: Do not test voltages higher than 250V to prevent electric shock or damage.
5.8. DC/AC Current Measurement
- Insert the black lead into the 'COM' port.
- For current under 600mA, insert the red lead into the 'mA' port. For current between 600mA and 20A, insert the red lead into the '20A' port.
- Switch the function knob to 'µA', 'mA', or 'A' ranges.
- 'DC' signal will be displayed. Press FUNC HOLD to switch to AC current test ('AC' will be shown).
- Connect the test leads to the loaded circuit in series.
Note: 'OL' indicates overload. Overload protection is provided by fuses (F600mA/250V for mA range, F20A/250V for 20A range). For currents over 5A, limit test time to less than 10 minutes, with a 1-minute cool-down period after each measurement.
5.9. NCV Test (Non-Contact Voltage)
- Turn the function knob to 'NCV'.
- Gradually approach the detected point with the NCV probe.
Indication: Weak AC signals will cause the yellow indicator to light up with a slow beeping sound. Strong AC signals will cause the red indicator to light up with a fast beeping sound.
Note: Do not rely solely on NCV for voltage detection; always verify with direct contact measurement. External interference (flashlights, motors) may trigger false positives.
5.10. Temperature Measurement
- Turn the function knob to '°C'. The multimeter will display the ambient temperature.
- Press FUNC HOLD to switch between '°C' and '°F'.
- Insert the thermocouple: positive (red) into 'VΩHz°C' and negative (black) into 'COM'.
- Contact the measured object with the thermocouple probe.
Note: When measuring temperature, ensure the thermocouple probe does not touch any charged objects to prevent electric shock or damage.
6. Specifications
6.1. General Specifications
- Maximum Display: 6000 counts
- Low Battery Indication: Displayed on screen
- Operation Temperature: 0°C – 40°C
- Storage Temperature: -10°C – 60°C
- Battery Type: 4 * 1.5V AAA Batteries
- Fuse: F 600mA/250V, F 20A/250V
- Data Hold: 'H' displayed
6.2. Measurement Specifications
| Temperature | ||
|---|---|---|
| Range | Resolution | Accuracy |
| °C | 1°C | ± (1.0% + 3) |
| °F | 1°F | ± (1.0% + 3) |
| Temperature Range: -20°C ~ 1000°C / -4°F ~ 1832°F | ||
| AC Voltage | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 600mV | 0.1mV | ± (1% + 4) |
| 6V | 1mV | |
| 60V | 10mV | |
| 600V | 100mV | ± (0.8% + 5) |
| 750V | 1V | |
| Input Impedance: 10MΩ | ||
| Maximum Input Voltage: 750V AC RMS | ||
| Frequency Response: 10Hz ~ 1KHz; T-RMS | ||
| DC Voltage | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 600mV | 0.1mV | ± (0.5% + 5) |
| 6V | 1mV | |
| 60V | 10mV | |
| 600V | 100mV | ± (0.8% + 5) |
| 1000V | 1V | |
| Input Impedance: 10MΩ | ||
| Maximum Input Voltage: 1000V DC RMS | ||
| Diode / Continuity | ||
|---|---|---|
| Function | Range | Details |
| Diode | 0-3V | Forward direct current: about 1mA. OCV about 3.2V |
| Continuity | 600Ω | Resistor < 40Ω: buzz with yellow light. 40Ω < Resistor < 60Ω: buzz with red light. OCV about 1V |
| AC Current | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 600µA | 0.1µA | ± (1.5% + 5) |
| 60mA | 0.01mA | |
| 600mA | 0.1mA | |
| 20A | 10mA | ± (2.0% + 5) |
| Max Input Current: mA: 600mA RMS, 20A: 20A RMS | ||
| Frequency Response: 10Hz to 1KHz T-RMS | ||
| DC Current | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 600µA | 0.1µA | ± (4.0% + 5) |
| 60mA | 0.01mA | |
| 600mA | 0.1mA | |
| 20A | 10mA | ± (2.0% + 5) |
| Max Input Current: mA: 600mA RMS, 20A: 20A RMS | ||
| Resistance | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 600Ω | 0.1Ω | ± (0.8% + 5) |
| 6kΩ | 1Ω | |
| 60kΩ | 10Ω | |
| 600kΩ | 100Ω | |
| 6MΩ | 1kΩ | ± (1.2% + 5) |
| 60MΩ | 10kΩ | |
| Input Voltage Range: 200mV - 10V AC RMS | ||
| Overload Protection: 250V DC/AC | ||
| Capacitance | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 6nF | 0.001nF | ± (4.0% + 5) |
| 60nF | 0.01nF | |
| 600nF | 0.1nF | |
| 6µF | 1nF | |
| 60µF | 10nF | ± (5.0% + 5) |
| 600µF | 100nF | |
| 6mF | 1µF | |
| 100mF | 10µF | |
| Overload Protection: 250V DC/AC | ||
| Frequency | ||
|---|---|---|
| Range | Resolution | Accuracy |
| 9.999Hz | 0.001Hz | ± (1.5% + 5) |
| 99.99Hz | 0.01Hz | |
| 999.9Hz | 0.1Hz | |
| 9.999KHz | 0.001KHz | |
| 99.99KHz | 0.01KHz | |
| 999.9KHz | 0.1KHz | |
| 9.999MHz | 0.001MHz | |
| Input Voltage Range: 200mV - 10V AC RMS | ||
| Overload Protection: 250V DC/AC | ||
7. Maintenance
7.1. Cleaning
- Turn off the multimeter and remove the test leads.
- Wipe the case with a damp cloth and clean it with mild detergent.
- Ensure no dirt or moisture is present in the terminals, as this can affect readings.
- If any abnormality occurs, stop using the multimeter immediately.
8. Troubleshooting
This section addresses common issues you might encounter with your multimeter.
| Problem | Possible Cause | Solution |
|---|---|---|
| 'OL' displayed | Overload, open circuit, or incorrect range selected. | Switch to a higher range, check for open circuits, or ensure correct lead connection for diode/continuity. |
| Low battery symbol displayed | Batteries are low. | Replace the 4 * 1.5V AAA batteries as per Section 4.1. |
| Incorrect readings | Dirty terminals, incorrect function/range, or damaged test leads. | Clean terminals, verify function/range, inspect and replace test leads if damaged. |
| Multimeter does not turn on | Dead batteries or blown fuse. | Replace batteries or check/replace fuses as per Section 4.1. |
9. User Tips
No specific user tips are available at this time. Always refer to the safety guidelines and operating instructions for optimal performance.
10. Warranty and Support
For any questions, technical support, or warranty inquiries, please contact our customer service via email at: HEMITEK2022@126.COM. Please provide your product model (T28B) and a detailed description of your issue.
11. Additional Resources
A comprehensive user manual in PDF format is available for download. You can access it via the following link: NJTY T28B Multimeter User Manual (PDF).