NJTY S18D/A

Contents

NJTY S18D/A Digital Multimeter User Manual

Model: S18D/A | Brand: NJTY

1. Introduction

This digital multimeter is designed and manufactured in compliance with IEC-61010 safety requirements for electronic measuring instruments and handheld digital multimeters. It meets IEC-61010 requirements pertaining to 600V CAT II and pollution degree 2. This manual provides essential information for the safe and effective operation of your device.

The meter is equipped with a backlight to facilitate reading measurements in dark environments.

For a complete and detailed manual, please refer to the PDF document available at NJTY S18D/A User Manual (PDF).

2. Safety Information

2.1. Safety Instructions

  • Before operating this meter, observe all standard safety procedures, including those against electric shock and unintended use.
  • Use only test leads compatible with the meter and ensure they are flawless before use.

2.2. Safety Notices

  • Readings may be unstable or erroneous near strong electromagnetic interference.
  • Do not operate the meter if its appearance is damaged.
  • The safety function is void if the meter is not properly operated.
  • Exercise caution when working near exposed conductors or bus lines.
  • Do not use the meter near explosive gas, vapor, or dust.
  • Ensure measurements are made with correct input terminals, functions, and within the allowable range.
  • Do not exceed the permitted values for each measuring range to prevent damage.
  • Avoid touching unused input terminals when the meter is connected to a circuit.
  • Be careful to avoid electric shock when measured voltage exceeds DC60V or AC36V RMS.
  • When using test leads, keep fingers behind the protective ring.
  • Remove test leads from the circuit before switching measuring ranges.
  • For DC functions, check for absence of AC voltage using AC functions first to prevent incorrect readings and potential electric shock.
  • Before testing resistance, diode, capacitance, or continuity, disconnect power to the circuit and discharge all high-voltage capacitors.
  • Resistance measurement or continuity tests cannot be performed on live electrical circuits.
  • Before current measurement, inspect the meter's protective tube and power off the circuit.
  • Exercise caution when repairing TV sets or measuring power switching circuits to prevent high amplitude voltage impulses from damaging the meter.
  • The meter uses 3x1.5V AAA batteries. Ensure they are correctly installed.
  • The battery low symbol (a battery icon) indicates insufficient power; replace batteries promptly for accurate measurement.
  • During measurement, the voltage should not exceed 600V.

2.3. Safety Symbols

SymbolDescription
⚠Warning, an important safety symbol.
∼AC (Alternating Current)
—DC (Direct Current)
⟂Ground
□□Double insulation protection
⌫Fuse
CECompliant with European Union Directive
⚡High voltage warning
CAT IIClass II 600V over-voltage protection

3. Product Overview

3.1. Multimeter Diagram and Components

Diagram of the NJTY S18D/A Digital Multimeter with numbered parts.
Figure 1: NJTY S18D/A Digital Multimeter Diagram
  1. Flashlight
  2. NCV detection area
  3. LED display
  4. Buzzer
  5. Backlight/Flashlight button
  6. "FUNC/Hz" function selection button
  7. Rotary switch
  8. HzVΩmA℃ input socket
  9. COM input socket
  10. 10A input socket
  11. Hold button
  12. NCV indicator

3.2. Symbols on the Display Unit

Image showing various symbols that appear on the multimeter's LCD display.
Figure 2: Multimeter Display Symbols
SymbolExplanation
🔋Low battery indicator
APOAuto power off indicator
–Negative input polarity
ACInput voltage AC
DCInput voltage DC
♫Multimeter in beep mode
➕Multimeter in diode test mode
AUTOAutomatic range measurement mode
HOLDMultimeter in data hold mode
℃ / ℉Unit of temperature (Celsius / Fahrenheit)
NCVNon-contact AC voltage detection mode
V, mVUnit of volt
A, mA, µAUnit of current
MHz, kHzUnit of frequency
mF, µF, nFUnit of capacitance

3.3. Functional Keys

KeyDescription of functions
HOLDActivates data hold mode.
🔦Short press: Backlight on/off. Long press (2S): Backlight and illumination light on/off.
FUNC/HzSelects AC/DC in A/mA/V/mV positions. Selects ℃/℉ in temperature position. Selects Hz or % mode in Hz% position.

3.4. Input Sockets

Input SocketDescription
COMCommon input terminal. Connect black test lead here.
VmAHz℃Input for voltage, resistance, capacitance, frequency, duty ratio, diode, continuity, and temperature measurements. Connect red test lead here.
10AInput for high current (up to 10A) measurements. Connect red test lead here.

4. Setup

4.1. Battery Installation

The meter uses three 1.5V AAA batteries. To install or replace batteries:

  1. Turn off the multimeter.
  2. Unplug all test leads from the sockets.
  3. Unscrew the screw on the back cover of the multimeter.
  4. Open the back cover.
  5. Remove the old batteries.
  6. Replace with new 1.5V AAA batteries, ensuring correct polarity.
  7. Close the back cover and tighten the screws.

4.2. Initial Inspection

Before first use, inspect the meter for any physical damage. Ensure test leads are in good condition. If any abnormality is observed, do not use the instrument and send it for repair.

5. Operating Guidelines

5.1. Normal Operation

Hold Mode

In hold mode, the displayed reading is maintained. To enter/exit:

  1. Press the "HOLD" button. The reading will be held, and "H" will appear on the LCD.
  2. Press "HOLD" again to exit and resume normal measurement.

Backlight & Illumination

The meter features both backlight and a flashlight for visibility in dark areas.

  • Backlight: Short press the backlight/flashlight button (🔦) to turn the backlight on/off. It automatically turns off after 20 seconds of inactivity.
  • Illumination (Flashlight): Long press the backlight/flashlight button (🔦) for 2 seconds to turn on both the backlight and the flashlight. Short press again to turn off the flashlight. It automatically turns off after 30 seconds of inactivity.
Multimeter demonstrating flashlight lighting for use in dark environments.
Figure 3: Flashlight Function

Auto Power Off (APO)

If no operations are performed for 15 minutes, the meter will beep and automatically power off to conserve battery life.

5.2. Measurement Guidelines

Always ensure the meter is set to the correct function and range before making any measurement.

AC/DC Voltage Measurement

Do not attempt to measure voltages exceeding 600V (DC/AC) to prevent electric shock or damage to the meter.

  • Ranges: DC Voltage: 600mV, 6V, 60V, 600V. AC Voltage: 600mV, 6V, 60V, 600V.
  • Steps:
    1. Turn the rotary switch to the "mV" or "V" position. Press "FUNC/Hz" to select DC or AC voltage.
    2. Connect the black test lead to the COM input socket and the red test lead to the V input socket.
    3. Connect the other ends of the test leads in parallel to the circuit being measured.
    4. Read the voltage value on the LCD. For DC voltage, the display will show the polarity.
  • Notes:
    • If the meter displays information within 600mV DC or 6V AC without input, short circuit "V - Ω" and "COM" terminals to reset to zero.
    • In AC voltage function, press "FUNC/Hz" to measure AC voltage frequency.
    • The AC voltage measured is True RMS (Root Mean Square), accurate for sine, square, triangular, and step waves (without DC offset).

Electrical Resistance Measurement

Ensure all power sources are cut off and high-voltage capacitors are discharged before measuring resistance.

  • Ranges: 600Ω, 6kΩ, 60kΩ, 600kΩ, 6MΩ, 60MΩ.
  • Steps:
    1. Turn the rotary switch to the Ω position. Press "FUNC/Hz" to select the Ω range.
    2. Connect the black test lead to COM and the red test lead to V/Ω.
    3. Read the resistance value on the LCD.
  • Notes:
    • Measured resistance may differ slightly from rated values due to parallel resistance.
    • For low resistance, short circuit test leads, record reading, then subtract from measured resistance for accuracy.
    • At 60MΩ, allow a few seconds for the reading to stabilize.
    • "OL" indicates the measured value is over range or an open circuit.

Diode Test

Ensure all power sources are cut off and high-voltage capacitors are discharged before testing diodes.

  • Steps:
    1. Turn the rotary switch to the diode test position (➕). Press "FUNC/Hz" to select the diode function.
    2. Connect the black test lead to COM and the red test lead to V/Ω.
    3. Connect the black test lead to the negative pole and the red test lead to the positive pole of the diode.
    4. The meter displays the forward bias value. If polarity is reversed, "OL" will be displayed.
  • Note: A normal diode typically shows a forward voltage drop of 0.5V to 0.8V.

Beep Continuity Test

Ensure all power sources are cut off and high-voltage capacitors are discharged before performing a continuity test.

  • Steps:
    1. Turn the rotary switch to the continuity test position (♫). Press "FUNC/Hz" to select the function.
    2. Connect the black test lead to COM and the red test lead to V/Ω.
    3. Measure the resistance of the circuit. If resistance is ≤ 50Ω, the beeper will sound continuously.

Capacitance Measurement

Ensure all power sources are cut off and high-voltage capacitors are discharged before measuring capacitance. Use the DC voltage position to confirm capacitors are discharged.

  • Ranges: 6nF, 60nF, 600nF, 6µF, 60µF, 600µF, 6mF, 100mF.
  • Steps:
    1. Turn the rotary switch to the capacitance position.
    2. Connect the black test lead to COM and the red test lead to the capacitance input socket.
    3. Connect the other ends of the test leads to the capacitor and read the value on the LCD.
  • Notes:
    • Large capacitance measurements require a stabilization period.
    • Pay attention to polarity when measuring polarized capacitors to avoid damage.

Frequency Measurement

Do not attempt to measure frequency if the voltage exceeds 250V (DC/AC) to prevent electric shock or damage.

  • Steps:
    1. Turn the rotary switch to the "Hz%" position.
    2. Connect the black test lead to COM and the red test lead to the Hz input socket.
    3. Connect the other ends of the test leads to the circuit and read the frequency on the LCD.
    4. Press "FUNC/Hz" to display frequency and duty ratio.

Current Measurement

Do not attempt to measure current in a circuit if the voltage between open-circuit voltage and ground exceeds 250V. If the fuse blows, you risk damage or injury. Always inspect the protective tube before current measurement.

  • Ranges: DC Current: 60mA, 600mA, 10A. AC Current: 60mA, 600mA, 10A.
  • Steps:
    1. Turn the rotary switch to the appropriate current position. Press "FUNC/Hz" to switch between DC/AC current.
    2. Connect the black test lead to COM. For currents < 600mA, connect the red test lead to the mA input socket. For currents 600mA-10A, connect the red test lead to the 10A input socket.
    3. Connect the black test lead to the negative pole of the load and the red test lead to the positive pole of the load.
    4. Connect power to the circuit and read the displayed value. If "OL" is shown, the input is over range; switch to a higher range.
    5. In AC current function, press "FUNC/Hz" to measure the frequency of the AC current source.

NCV (Non-Contact Voltage) Test

This feature allows detection of AC voltage without direct contact.

  • Steps:
    1. Turn the rotary switch to the NCV position.
    2. Place the top of the meter (NCV detection area) near the conductor.
    3. If AC voltage is detected, signal density indicators (high, medium, low) will light up, and the beeper will sound at different frequencies.
  • Notes:
    • Voltage may be present even without indication. Do not rely solely on NCV for safety checks.
    • Detection can be affected by socket design, insulation thickness, and external interference (e.g., flashlight, motor).
Multimeter demonstrating NCV (Non-Contact Voltage) detection.
Figure 4: NCV Detection in action

Temperature Test

The meter comes with a thermocouple for temperature measurement.

  • Ranges: -20℃ to 1000℃ / -4℉ to 1832℉.
  • Steps:
    1. Turn the rotary switch to the "℃/℉" position. The meter will display ambient temperature.
    2. Insert the red plug of the thermocouple into the VΩmA socket and the black plug into the COM socket.
    3. Read the temperature value on the display after it stabilizes.
  • Note: The maximum measuring temperature for the K-type thermocouple is 300℃.
Multimeter demonstrating temperature test with a thermocouple inserted into a glass of water.
Figure 5: Temperature Measurement

6. Maintenance

6.1. Safe Maintenance Habits

  • Pull out test leads before opening the multimeter case or removing the battery cover.
  • Use appropriate replacement parts for repairs.
  • Disconnect all relevant power before turning on the meter for maintenance.
  • Calibration and maintenance should be performed by professionals.
  • If any abnormality is observed, stop using the instrument immediately and send it for qualified repair.

6.2. Battery & Fuse Replacement

Battery Replacement

Follow the steps outlined in the Setup section (4.1).

Fuse Replacement

The meter uses F600mA/250V for mA range and F10A/250V for 10A range. To replace a fuse:

  1. Turn off the multimeter.
  2. Unplug all test leads from the sockets.
  3. Remove the rubber cover.
  4. Unscrew the rear screws with a screwdriver.
  5. Gently push out the old fuse and replace it with a new one of the correct rating (usually 0.6A for general purpose).
  6. Close the back cover and tighten the screws.

7. Troubleshooting

  • Unstable/Erroneous Readings: May occur near strong electromagnetic interference. Move away from such equipment.
  • "OL" Display: Indicates the measured value is over the selected range, or an open circuit. Select a higher range or check circuit connections.
  • Low Battery Symbol: Appears when battery power is insufficient. Replace batteries promptly for accurate measurements.
  • No Display/Meter Not Turning On: Check battery installation and ensure batteries are not depleted.
  • No Continuity Beep: Ensure the circuit resistance is below 50Ω and the meter is in continuity mode.
  • Inaccurate Capacitance Reading: Ensure polarized capacitors are connected with correct polarity and allow time for large capacitance readings to stabilize.
  • No NCV Detection: Ensure the meter is in NCV mode and the detection area is close to the conductor. External interference can affect detection.

8. Specifications

8.1. General Characteristics

  • Display: 6000 counts LCD screen, automatically displays unit symbol.
  • Operating Mode: Auto range
  • Operating Temperature: 0 - 40 ℃
  • Operating Humidity: < 80% RH
  • Storage Temperature & Humidity: -10 - 60 ℃ (< 70% RH, remove batteries)
  • Ambient Temperature for Testing/Calibration: 20 ℃ ± 2 ℃
  • Max Allowable Voltage (terminal to ground): AC600V
  • Overrange Indication: "OL" on LCD.
  • Input Polarity Indication: "-" automatically appears.
  • Power: 3 x 1.5V AAA batteries
  • Dimensions: 147mm (L) x 78mm (W) x 45mm (H)
  • Weight: Approximately 220g (excluding batteries)
  • DIY Supplies: Electrical
  • Certification: CE, RoHS
  • Origin: Mainland China

8.2. Precision Indicators

DC Voltage

RangeResolutionAccuracy
600mV0.1mV± (0.5% reading + 5 digits)
6V1mV
60V10mV
600V100mV

Input resistance: 10MΩ; Max input voltage: DC600V

AC Voltage (True RMS)

RangeResolutionAccuracy
600mV0.1mV± (1% reading + 4 digits)
6V1mV
60V10mV
600V100mV

Input resistance: 10MΩ; Max input voltage: AC600V; Frequency response: 40Hz - 1KHz TRMS

Frequency

RangeResolutionAccuracy
9.999Hz0.001Hz± (1.5% reading + 5 digits)
99.99Hz0.01Hz
999.9Hz0.1Hz
9.999kHz0.001kHz
99.99kHz0.01kHz
999.9kHz0.1kHz
9.999MHz0.001MHz

Input voltage range: 200mV-10V AC RMS; Overload protection: 250V DC/AC

Electrical Resistance

RangeResolutionAccuracy
600Ω0.1Ω± (0.8% reading + 5 digits)
6kΩ1Ω
60kΩ10Ω
600kΩ100Ω
6MΩ1kΩ
60MΩ10kΩ

Overload protection: 250V DC/AC; Open-circuit voltage: 2.4V

Diode and Buzzer

FunctionMeasuring RangeResolutionTest Condition
Diode0-3V0.001VForward DC current: approx 1mA; Open-circuit voltage: approx 3.2V. Displays approx value of diode forward voltage drop.
Buzzer≤ 600Ω0.1ΩBuzzer sounds if internal resistance < 40Ω (yellow light on). If 40-60Ω, red light on.

Overload protection: 250V DC/AC

Capacitance

RangeResolutionAccuracy
60nF0.01nF± (4.0% reading + 5 digits)
600nF0.1nF
6µF1nF
60µF10nF
600µF100nF
6mF1µF
100mF10µF

Overload protection: 250V

DC Current

RangeResolutionAccuracy
60mA0.01mA± (1.2% reading + 5 digits)
600mA0.1mA
10A10mA

Overload protection: mA range with F600mA/250V; 10A range fuse with F10A/250V. For currents > 5A, test time < 10 seconds, with 1 min rest between measurements.

AC Current

RangeResolutionAccuracy
60mA0.01mA± (1.5% reading + 5 digits)
600mA0.1mA
10A10mA

Overload protection: mA range with F600mA/250V; 10A range fuse with F10A/250V. Max input current: mA range: 600mA AC TRMS; 10A range: 10A AC TRMS. For currents > 5A, test time < 10 seconds, with 1 min rest between measurements. Frequency response: 10Hz-1KHz TRMS.

Temperature

RangeResolutionAccuracy
-20℃ to 1000℃1℃± (1.0% + 3 digits)
-4℉ to 1932℉1℉

9. User Tips

  • Value for Money: Users often appreciate the balance of features and cost. This multimeter offers a wide range of functions at a competitive price point.
  • Backlight Use: The backlight and flashlight features are very useful in dimly lit environments, making measurements easier to read. Remember the auto-off timers to conserve battery.
  • Auto-Ranging: The auto-ranging feature simplifies operation by automatically selecting the correct measurement range, reducing the chance of errors.

10. Warranty and Support

For warranty information, technical support, or service inquiries, please contact the manufacturer, NJTY, or your point of purchase. Always refer to the official documentation for the most current warranty terms.

A comprehensive user manual in PDF format is available for download: NJTY S18D/A Digital Multimeter User Manual (PDF).

Ask a question about this manual

Ask about setup, troubleshooting, compatibility, parts, safety, or missing instructions. Manuals+ will review the question and use this page’s manual context to help answer it.

After a successful submission, we securely hash the supplied account or form email before sending it to Microsoft Advertising for conversion attribution. Privacy details.